ENGLISH Welcome The AirCurve™ 10 ST-A is a bilevel positive airway pressure device. WARNING Read this entire guide before using the device. CAUTION In the US, Federal law restricts this device to sale by or on the order of a physician. Indications for use AirCurve 10 ST-A The AirCurve 10 ST-A device is indicated to provide non-invasive ventilation for patients weighing more than 30 lb (13 kg) or more than 66 lb (30 kg) in iVAPS mode with respiratory insufficiency or obstructive sleep apnea (OSA).
At a glance The AirCurve 10 ST-A includes the following: Device HumidAir™ humidifier (if supplied) Air tubing Power supply unit Travel bag SD card (already inserted).
About your device 1 Air outlet 6 HumidAir humidifier 2 Air filter cover 7 Screen 3 Retention clip 8 Adapter cover 4 Power inlet 9 SD card cover 5 Serial number and device number 10 LED alarm indicator About the control panel Start/Stop button Press to start/stop therapy. Press and hold for three seconds to enter power save mode. Dial Turn to navigate the menu and press to select an option. Turn to adjust a selected option and press to save your change.
Setup CAUTION Do not overfill the humidifier as water may enter the device and air tubing. 1. With the device on a stable level surface, grip the retention clip on the back of the device and pull up to open. Note: The retention clip is shown in the open position. 2. (a) Plug the power connector into the device power inlet then (b) push down the retention clip to secure in place. Connect one end of the power cord into the power supply unit and the other end into the power outlet. 3.
Starting therapy 1. Fit your mask. 2. Press Start/Stop or breathe normally if SmartStart™ is enabled. You will know that therapy is on when the Monitoring screen is displayed. The pressure bar shows the inspiratory and expiratory pressures in green. The green bar will expand and contract as you breathe in and out. The screen will go black automatically after a short period of time. You can press Home or the dial to turn it back on.
Power save mode Your AirCurve 10 ST-A device records your therapy data. In order to allow it to transmit the data to your care provider, you should not unplug the device. However, you can put it into power save mode to save electricity. To enter power save mode: Press and hold Start/Stop for three seconds. The screen goes black. To exit power save mode: Press Start/Stop once. The Home screen is displayed.
Ramp Down Ramp Down is intended to make stopping therapy more comfortable by reducing your pressure over a fixed 15 minute period. This option will only be available to you via your care provider. To enable Ramp Down: 1. In My Options, turn the dial to highlight Ramp Down and then press the dial. 2. Turn the dial to select On and then press the dial to save the change. To start Ramp Down: 1. Press the Start/Stop button.
Mask Fit Mask Fit is designed to help you assess and identify possible air leaks around your mask. To check Mask Fit: 1. Fit the mask as described in the mask user guide. 2. In My Options, turn the dial to highlight Run Mask Fit and then press the dial. The device starts blowing air. 3. Adjust the mask, mask cushion and headgear until you get a Good result. To stop Mask Fit, press the dial or Start/Stop. If you are unable to get a good mask seal, talk to your care provider.
Muting activated alarms To mute the alarm: 1. Press the dial. A list of activated alarms will appear and the flashing alarm mute icon will be displayed in the top right corner of the screen. The alarm will be muted for 2 minutes. 2. To return to the previous screen, highlight OK and press the dial. Once the condition that activated the alarm is corrected, the alarm sound and flashing icon will stop. If the condition that activated the alarm remains after 2 minutes, the alarm will re-occur.
Caring for your device It is important that you regularly clean your AirCurve 10 ST-A device to make sure you receive optimal therapy. The following sections will help you with disassembling, cleaning, checking and reassembling your device. Disassembling 1. Hold the humidifier at the top and bottom, press it gently and pull it away from the device. 2. Open the humidifier and discard any remaining water. 3. Hold the cuff of the air tubing and gently pull it away from the device.
Checking You should regularly check the humidifier, air tubing and the air filter for any damage. 1. Check the humidifier: Replace it if it is leaking or has become cracked, cloudy or pitted. Replace it if the seal is cracked or torn. Remove any white powder deposits using a solution of one part household vinegar to 10 parts water. 2. Check the air tubing and replace it if there are any holes, tears or cracks. 3. Check the air filter and replace it at least every six months.
The data is usually transmitted after therapy has stopped. In order to make sure that your data is transferred, leave your device connected to the mains power at all times and make sure that it is not in Airplane Mode. Notes: Therapy data might not be transmitted if you use it outside of the country or region of purchase. Devices with wireless communication might not be available in all regions.
When using the device on a plane: Make sure the humidifier is completely empty and inserted into your device. The device will not work without the humidifier inserted. Turn on Airplane Mode. To turn on Airplane Mode: 1. In My Options, turn the dial to highlight Airplane Mode and then press the dial. 2. Turn the dial to select On and then press the dial to save the change. The Airplane Mode icon is displayed at the top right of the screen.
Problem/possible cause Solution Air pressure in my mask seems too low (it feels like I am not getting enough air) Ramp may be in progress Wait for air pressure to build up or turn Ramp Time off. . Ramp Down may be in progress . Press Start/Stop to stop therapy then press Start/Stop to restart and continue therapy. Non-vented mask is used. Only use a vented mask. Mask vents might be blocked. Check if you have sufficient venting. Unblock mask vents if necessary.
Problem/possible cause Solution My screen and buttons are flashing but there is no sound or message Software upgrade is in progress. Software upgrade takes approximately 10 minutes to complete. Displays message: Read only card, please remove, unlock and re-insert SD card SD card switch may be in the lock (read-only) position. Move the switch on the SD Card from the lock position to the unlock position and then re-insert it.
Problem/possible cause Solution Displays message: Low Sp02 detected SpO2 has dropped below the pre-set alarm level. Check the attachment of the sensor. If the problem persists contact your care provider. Displays message: No SpO2 data, check your oxi sensor attachment to module/finger Oximeter sensor is not attached properly. Ensure that the oximeter sensor is attached properly to the module and to your finger. Oximeter sensor may be faulty.
Reassembling parts Some parts of your device are designed to easily come off in order to avoid damage to the parts or the device. You can easily reassemble them as described below. To insert the humidifier seal: 1. Place the seal into the lid. 2. Press down along all edges of the seal until it is firmly in place. To reassemble the humidifier lid: 1. Insert one side of the lid into the pivot hole of the base. 2. Slide the other side down the ridge until it clicks into place.
Supplemental oxygen must not be used while smoking or in the presence of an open flame. Always make sure that the device is turned on and airflow generated before the oxygen supply is turned on. Always turn the oxygen supply off before the device is turned off, so that unused oxygen does not accumulate within the device enclosure and create a risk of fire. Do not perform any maintenance tasks while the device is in operation. The device should not be used adjacent to or stacked with other equipment.
Technical specifications 90W power supply unit AC input range: DC output: Typical power consumption: Peak power consumption: Environmental conditions Operating temperature: Operating humidity: Operating altitude: Storage and transport temperature: Storage and transport humidity: 100–240V, 50–60Hz 1.0–1.5A, Class II 115V, 400Hz 1.5A, Class II (nominal for aircraft use) 24V 3.
Physical - device and humidifier Dimensions (H x W x D): Air outlet (complies with ISO 5356-1:2004): Weight (device and cleanable humidifier): Housing construction: Water capacity: Cleanable humidifier - material: 4.57" x 10.04" x 5.91" (116 mm x 255 mm x 150 mm) 22 mm 47.
General The patient is an intended operator. Operator position The device is designed to be operated within arm's length. An operator should position their line of sight within an angle of 30 degrees from a plane perpendicular to the screen.
Displayed values Value Pressure sensor at air outlet: Mask pressure Flow derived values: Leak Tidal volume Respiratory rate Minute ventilation Ti I:E ratio Value Pressure measurement1: Mask pressure2 Flow and flow derived values1: Flow Leak2 Tidal volume2,3 Respiratory rate2,3 Minute ventilation2,3 Range Display resolution 3–30 cm H2O 0.1 cm H2O 0–120 L/min 0–4000 mL 0–50 BPM 0–30 L/min 0.1–4.0 sec 1:100–2:1 Accuracy1 1 L/min 1 mL 1 BPM 0.1 L/min 0.1 sec 0.1 ±0.
Device without humidification and SlimLine air tubing / Device with humidification and SlimLine air tubing Pressure (cm H2O] 10 BPM 15 BPM 20 BPM 4 0.5 / 0.5 0.5 / 0.5 0.8 / 0.8 8 0.5 / 0.5 0.5 / 0.5 0.8 / 0.8 12 0.5 / 0.5 0.5 / 0.5 0.8 / 0.8 16 0.5 / 0.5 0.5 / 0.5 0.8 / 0.8 20 0.5 / 0.5 0.5 / 0.5 0.8 / 0.8 Pressure accuracy - bilevel Maximum dynamic pressure variation according to ISO 80601-2-70:2015.
Symbols The following symbols may appear on the product or packaging. Read instructions before use. use. Manufacturer. Catalogue number. weight. Indicates a warning or caution. European Authorised Representative. Serial number. Device number. Batch code. On / Off. Device Protected against finger sized objects and against dripping water when tilted up to 15 degrees from specified orientation. equipment. Follow instructions before Humidity limitation. China pollution control logo 1.
Limited warranty ResMed Ltd (hereafter 'ResMed') warrants that your ResMed product shall be free from defects in material and workmanship from the date of purchase for the period specified below.
8010 cover.
ResMed Ltd Distributed by 1 Elizabeth Macarthur Drive ResMed Corp 9001 Spectrum Center Boulevard San Diego CA 92123 USA Bella Vista NSW 2153 Australia ResMed (UK) Ltd 96 Jubilee Ave Milton Park Abingdon Oxfordshire OX14 4RW UK See ResMed.com for other ResMed locations worldwide. Air10, AirCurve, ClimateLine, SlimLine, HumidAir, SmartStart and VPAP are trademarks and/or registered trademarks of the ResMed family of companies. For patent and other intellectual property information, see ResMed.com/ip.