RecoveryAir JetBoots Integrated Pneumatic Compression System Fully wireless compression boots. Your RecoveryAir JetBoots come with two compression boots, each with integrated pumps. Plus a 36W charger, a splitter charging cord, and a carrying pouch. See Warnings on pg. 10 for important safety instructions.
Getting started 1. To get started, put your boots on and get into a comfortable position. 2. Once ready, turn ON the RecoveryAir JetBoots integrated pump, by pressing the power button. 3. After your boots are ON and have automatically paired, you can adjust the pressure and time settings can be adjusted directly from the integrated pumps before your treatment begins. More settings can be made from the Therabody app. (If your boots are not automatically paired, see section Troubleshooting on page 17.) 4.
Feature Callouts Buttons Power Button Start/Pause/Stop Button Time Button Pressure Button 3
Settings (Adjustable Range) Indicators Pressure Setting Warning Indicator Adjustable Pressure Range (25-100 mmHg) in increments of 25 mmHg.
Smart Features The RecoveryAir JetBoots also connect to the Therabody app via Bluetooth, giving you control of your session from the convenience of your smartphone. Not only can you easily access the Recover and Warm-up treatment options, but you can also adjust the pressure and time. Scan the QR code to get started.
RecoverAir JetBoots Programs About negative gradient for all programs: The RecoveryAir JetBoots’ unique true negative gradient of pressure sequentially travels up the limb from the foot toward the heart in four internal overlapping chambers. The spiraling overlap of chambers safely maximizes circulation. • • The pressure can be the same in two consecutive chambers, but the pump won’t allow a back chamber to have a higher pressure than a front chamber. ●The minimum pressure level of each chamber is 20 mmHg.
Charging the RecoveryAir JetBoots You can use your JetBoots while they are charging. Also, you can charge one of the boots at the time directly with the power adapter. 1. After turning off the integrated pump, locate the power charging port found at the front right side of the foot of the boot. 2. Connect one side of the splitter to the charging port. 3. Do the same for the second boot. 4.
Superior, hygienic design It’s important to take proper care of your RecoveryAir JetBoots compression boots because the longer they last, the longer you will feel the benefits. Each boot has non-porous medical grade material that helps fight bacteria and provide a resistant surface. Our internal overlapping chambers provide a smooth surface to clean, this means no wiping under chamber flaps where bacteria can build up. This also means, no smells.
Traveling with your RecoveryAir JetBoots The RecoveryAir JetBoots’ components are all packed into a sleek, compact design for on-the-go recovery, making traveling even more convenient. When traveling with your RecoveryAir JetBoots, we recommend: • P ● ack your boots, including the integrated pump, in the provided carrying pouch and simply fold to fit into your travel bag or carry on luggage. • T ● he integrated pump eliminates the need to pack the pump separately.
Important Safety Information Safety Warnings and Precautions Warning: • • • • • • • • • • • Explosion Hazard: Do not use the RecoveryAir JetBoots in the presence of flammable gases, including flammable anesthetics. Electric Shock Hazard: Do not allow liquid to enter any part of the RecoveryAir JetBoots pump. Do not immerse in water or liquid. To clean, follow instructions found on page 8. Electric Shock Hazard: To prevent electric shock, do not open the pump. Do not attempt to service the pump yourself.
Important Safety Information Caution • • • • • • • • • • • • • • • • • • 11 Do not operate the RecoveryAir JetBoots while operating a vehicle. Do not store or transport the RecoveryAir JetBoots beyond the specified temperature, humidity, and atmospheric pressure range. Do not use the RecoveryAir JetBoots beyond the specified temperature range: 10 to 30°C (50 to 86°F). Do not use the RecoveryAir JetBoots beyond the specified humidity range: 30%rH–75%rH, noncondensing.
Labels The following labels and symbols appear on the pump, garments and/or packaging. Label IP 22 Description Location Degree of protection against ingress of water On garment label Read instructions before use On garment label Read instructions before use (for China only) On garment label Class III equipment (only in the EU per IEC60335-2-32) Equipment relying on limitation of voltage to extra-low-voltage (ELV) values as provision for basic protection and with no provision for fault protection.
Label 13 Description Location Read instructions before use On garment label Level of protection type BF applied part On garment label Double insulation On DC power adapter Direct current On garment label Therabody, Inc. 6100 Wilshire Blvd.
Label UDI : (01) 00810036056908 (10) 2143 (21) 00001 Description Location Unique Device Identification (UDI) On garment label and package, and garment bag and package Separate collection for waste electrical and electronic equipment On garment label In accordance with Directive 2014/35/EU electrical equipment designed for use within certain voltage limits, and Directive 2014/30/ EU electromagnetic compatibility On garment label Serial number of the console On console base Fragile, handle with car
Label 15 Description Location This side Up On package Keep away from sunlight On package Transportation & storage humidity limitation On package Transportation & storage atmospheric pressure limitation On package Transportation & storage temperature limitation On package
Label Description Location Do not wash On garment tag Do not dry-clean On garment tag Do not tumble dry On garment tag Do not bleach On garment tag Do not iron On garment tag 16
Indications for Use RecoveryAir is indicated for the temporary relief of minor muscle aches and pains, and for the temporary increase in circulation to the treated areas in people who are in good health. RecoveryAir simulates kneading and stroking of tissues by using an inflatable garment. Note • • • • The pump is “multi-voltage” and can be used for travel. An appropriate socket adapter must be used in countries with incompatible wall outlets. Use only RecoveryAir garments with the RecoveryAir pump.
Symptom Possible Cause Corrective Action One garment inflates but the second one does not. The second garment does not receive air. Examine its hoses for kinks, twists and folds. The pump stops working, the Status Indicator turns On in yellow. Hose is not connected properly to garment or pump, or prong plug is not inserted into unused air outlet. Examine and fasten all air connections. When treating only one limb, always plug the unused air outlet with the prong plug provided with the pump.
Warranty For full warranty information, please visit www.therabody.com/warranty. To request a copy of the warranty by mail, you may send a request to the following address: Therabody - WarrantyAttn: Customer Service 6100 Wilshire Blvd., Los Angeles, CA 90048 Please note, this is not a return address or a retail location. No products or packages will be accepted at this location.
1. Report Online at: www.accessdata.fda.gov/scripts/medwatch/index.cfm?action=reporting.home 2. Consumer Reporting Form FDA 3500B. Follow the instructions on the form to either fax or mail it in for submission. For help filling out the form, see MedWatchLearn. The form is available at: www.fda.gov/ downloads/aboutFDA/reportsmanualsforms/forms/ucm349464.pdf 3. Call FDA at 1-800-FDA-1088 to report by telephone. 4. Reporting Form FDA 3500 commonly used by health professionals.
installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation.
22
VL Born in Los Angeles, CA. Designed for everybody.