Release Version 1.0.
© Copyright 2012, Rayence Co. Ltd. All pages of this document contain proprietary and confidential information of Rayence Corporation and are intended for exclusive use by Rayence Corporation personnel or customers. Copying, disclosure to others or other use is prohibited without the express written authorization from Rayence Corporation. Please report any violations of requirement to Rayence Corporation.
Attention For improvement of product performance, supplementation, or follow-up of information; the contents of this manual are subject to change without separate prior notice. Please note that our company has neither responsibility for any accidents nor obligation to do free repair service for any damage of the equipment due to user's mistake, which resulted from failure to follow the contents in this manual. Make sure to be familiar with the safety precautions and usage procedures.
Contents Introduction ........................................................................................ 1 Intended use ...................................................................................... 1 Product features ................................................................................ 2 Product components ......................................................................... 3 Components Description................................................................... 5 Warning ..
Introduction Overview The 1012WCA is a wireless digital X-ray flat panel detector that can generate images of any part of the body. The wireless LAN((IEEE 802.11a/g/n) communication feature improves the operability, and high-speed processing. This X-ray imaging system consists of a scintillator directly coupled to an a-Si TFT sensor. It makes high-resolution, high-sensitive digital images. Intended use For U.S.A.
Product features Wi-Fi (802.11a/g/n) Based on a-Si TFT active matrix Compact (18mm thickness) and light weight (Typ. 3.15kg) Limiting Resolution : 3.
Product components Photo Medical Image Processing Unit Item Part Name Quantity Detector SD1012WCA 1 Handle - 1 Battery pack RB37WH 2 Battery charger RC120W 1 Charger adapter PMP120-13-3 1 AGI RA001A 1 3 R-M-011-121030
Cables Item Part Name Length Quantity Link cable VRH076A 6m 1 PInterface cable VRH017A 8m 1 1.8m 1 1.
Components Description The detector is designed to capture radiographic images. Captured images are transmitted to PC using the wireless/wired connection. (1) Detector 1. LED indicator: The LED indicates the state of detector. i. Green on : Power on. ii. Blue blinking : Wireless Connection. iii. Orange on : Low battery. 2. Handle : Hold this handle when carrying the sensor unit. It is removable. (Horizon / Vertical) 3. Handle unlock-lever : This is an unlock-lever to remove handle. 4.
and power supplying. Connect the detector to PC and SMPS(not provided) using Link cable. 5. CFRP(Carbon Fiber Reinforced Plastic) : The part of the patient’s body to which an image is to be taken should be placed against this plate. 6. Power button : Power on / Power off switch. 7. Shock sensor : Detector has built-in 4 Shock sensors. It detects and records impact and mishandling of fragile 8. Battery unlock-lever : This is an unlock-lever to remove battery. (2) Battery & Charger 1.
(3) AGI 1. Trigger connector: This is a connector for synchronization between detector and generator. Connect the AGI to the generator using P-interface cable.. 2. USB connector: This is a connector for communication between AGI and PC. Connect the AGI to the PC using USB cable.
Warning Make sure to observe the following right Environment of Use and Storage Follow the specified process of operational instructions written in this manual for the safety of the users and patients. Does not use or store the instrument near any flammable chemicals such as thinner, benzene, etc. Also, this instrument is not a category AP or APG equipment. If chemicals are spilled or evaporate, it may result in fire or electric shock through contact with electric parts inside the instruments.
When Problem Occurs Should any of the following occur, immediately turn OFF the power of each instruments, unplug the power supply cord from the AC outlet, and contact Rayence representative or distributor. When there is smoke, odd smell or abnormal sound. When liquid has been spilled into the instrument or a metal object has entered through an opening. When the instrument has been dropped and it is damaged.
Caution Environment of Use and Storage Do not install the instrument in a location with the conditions listed below. Otherwise, it may result in failure or malfunction, cause fire or injury. Close to facilities where water is used. Where it will be exposed to direct sunlight. Close to air-conditioner or ventilation equipment. Close to heat source such as a heater. Prone to vibration. Insecure place. Dusty environment. Saline or sulfurous environment.
Handling Do not spill liquid or chemicals onto the instrument or, in cases where the patient is injured, allow it to become wet with blood or other body fluids, as doing so may result in fire or electric shock. In such situation, protect the instrument with disposable covering as necessary. Wipe the CFRP plate of the sensor unit with ethanol or glutaraldehyde solution to disinfect it each time a different patient uses the instrument, in order to prevent infection.
Marking and labeling symbols Symbols Meaning Caution : “Attention, see instructions for use” Manufacturer Date of manufacture Serial number WEEE : Waste Electrical and Electronic Equipment Authorized representative in the European Community CE symbol grants the product compliance to the European Directive for Medical Devices 93/42/EEC as a class Ⅱa device. Authorized by Notified Body SGS (code no.
Labels Detector label Battery Pack Battery Charger 13 R-M-011-121030
Notes for Using the Detector Preparing Fully charge the battery pack. Charge the battery on the day of examination or on the previous day. Battery slowly discharges even of not in use. The battery pack may have expired if it discharges immediately after being fully charged. You can purchase an optional battery pack to replace an exhausted one. Be sure to fully charge the battery before use. ※ The battery charger, RC120W is designed for the dedicated battery pack.
Handling Handle the instrument carefully, as it may be damaged if something is hit against it, dropped, or receives a strong jolt. Handle Assembly A B ※ Insert the handle always in the same direction.
Handle Assembly Type 1 A Please confirm lever position. When the handle is locked, the lever should be same position with the right picture. If the handle is not locked, the detector can be dropped. Insert the handle in the same direction Handle Assembly Type 2 B Insert the handle in the same direction Please confirm lever position. When the handle is locked, the lever should be same position with the right picture.
Before Exposure Be sure to check the equipment daily and confirm that it works properly. Sudden heating of the room in cold areas will cause condensation to form on the instrument. In this case, wait until condensation disappears before performing exposure. If the instrument is used with condensation formed on it, problems may occur in the quality of the instrument.
Sleep Mode/Wake up Do not change parameters except “Control e” Control “e” parameter : sleep time - 0 : disabled (default) - 1 : 1 sec - 2 : 2 sec … 600 : 600 sec If you set sleep time, the detector goes to the sleep mode after “N” sec after acquiring image If you want to wake up the detector, Press the power button(less than 1 second) 18 R-M-011-121030
Limit of Load Uniform load: 150 kg over the whole area of sensor window. Local load: 100 kg on an area 40 mm in diameter.
Be sure to use the sensor unit on a flat place so it will not bend. Otherwise, the sensor may be damaged. Disinfection and Cleaning Do not spray the detector directly with disinfectants or detergents. Do not use anything other than neutral detergent for cleaning the cover of the instrument. Otherwise, the coating will be corroded. Others Be sure to reconnect the cables to the proper connectors. Otherwise, the instrument may malfunction or may be damaged.
Technical Features Mechanical Features Size 395 x 337 x 18 mm Weight 2.85 kg (incl. Battery pack) Encapsulation Material Mg Window Material Carbon fiber plate with 1.4 mm thickness Electrical Features Detector Sensor Type Amorphous Silicon with TFT (Single Panel) X-ray Converter CsI:Tl Total Pixel Number 2080 × 2560 pixels Active Pixel Area 25.9 cm × 32.0 cm Active Pixel Number 2040 × 2520 pixels Pixel Pitch 127 μm Limiting Resolution Max. 3.
Battery Size 232.5 x 132.8 x 7 mm Weight Typ. 0.3 kg Input 12.5 VDC Output 11.1 VDC Charging time Typ. 2.5 hours Capacity Typ. 3400 mAh The number of times being acquired image 600 images The Battery level can be displayed on the LED status of battery. If the battery level goes down under 25%, please charge the battery Battery level Display 75~100 % 50~75 % 25~50 % 0~25 % Charger Size 267.5 x 100 x 77 mm Weight Typ. 0.86 kg Input 20 VDC Output 11.1 VDC Adapter Size 160 x 76 x 43.
Wireless Spec Standard 802.11a/g/n compliance Peak Rate 300Mbps Frequency 2.4 GHz / 5 GHz Bandwidth 20MHz/40MHz MIMO 2x2 5.15~5.25 GHz band is restricted to indoor operations only. (for FCC) 5.15~5.350 GHz band is restricted to indoor operations only. (for CE) 5.15~5.350 GHz is indoor use only.
Environmental requirement Item Min. Temperature Typ. Max. Unit -10 50 ℃ Humidity 10 80 Pressure 70 106 Note % H.R. kPa 16ms, Shocks (Wrapping condition) 10G 1000times, in (25G) 6directions, non-Driving 10-150Hz, Vibrations (Wrapping condition) 2G 10Sweeps, (10G) 1min/Octave, XYZ axis ※ Regularly changed parts : Battery (warranty 6 months) PC Requirements Processor : At least Intel Pentium IV HT with 2.
Installation The Detector is composed of sensitive electronic parts and Portable Imaging processing unit must be installed in a way that enables the user to achieve optimal use components. It is recommended to use the product in a clean place and to exercise caution to ensure that it is not affected by dust or liquids. It is recommended to Use a dry and soft cloth to clean the detector housing. General interface Connection Power Connection A. Connect the battery pack or power cable to the equipment. B.
Wireless Communication A. AP Router(Line sharer) setting - SSID : Griffon - Internal network - IP address : 2.2.2.1 - Subnet mask : 255.255.255.0 - Dynamic IP allocation range : 2.2.2.2~2.2.2.254 - Pre-Shared Key(Password) : project302 - Authentication methods : WPAPSK or WPA2PSK - Password methods : TKIP/AES - AP IP : 2.2.2.
C. Checking Link Quality - After wireless connection is established, perform ‘Get Bright’ in ‘Calibration’ tap. - Check the value named ‘Wireless Signal’ in black log screen. Wireless Signal = Link Quality (Max.
Trigger Connection A. Connect the P-interface cable to the generator X-ray Generator Connection Make assurance doubly sure SIGNAL RATING before connection. Connect the P-interface cable between the AGI box and X-ray generator. A.
Operating description The window time can be changed.
Exposure Time(*clk) 1333333 is designate to 0.
IP set up [My Network Places] → [Properties] → [Local Area Connection] → [Properties] → [Internet Protocol (TCP/IP)] → [Use the following IP address] IP address : Obtain an IP address automatically IP address : Obtain an IP address automatically 31 R-M-011-121030