Contents 1 Safety ........................................................................................................................................................................... 3 Trademarks ...................................................................................................................................................................... 3 General Safety ............................................................................................................................
CHAPTER 1 Safety Read this information before you start using your device. Trademarks Brand or product names are trademarks of their respective owners. The following brand orproduct names are trademarks of Honeywell: ⚫ ⚫ ⚫ ⚫ Honeywell Vertex™ Compliance Chemcassette® Teflon® Sola/Hevi-Duty General Safety Follow all installation and operational instructions to ensure the safe and reliable operation of thisunit. If this monitor is used in a manner not specified by Honeywell Analytics Inc.
Symbol Description Caution: hot surface Protective conductor terminal (ground terminal) EMC Considerations Your Honeywell Analytics continuous gas monitor has been designed to comply with Electromagnetic Compatibility (EMC) standards applicable at the time of its manufacturing. Thedesign includes filtering, shielding and bypassing techniques. At the time of certification, simulated customer Input/ Output (I/O) schemes were tested.
devices must be susceptible with the interference from legal communications or ISM radio wave radiated devices. Cabling At the very minimum, all cables should include a braided shield. Ideal results have been obtainedwith twisted pair cabling which has a foil shield surrounding each pair plus foil and 90% braid shielding around the bundle. In addition, ensure local electrical code requirements are met.
Connectors All qualification and certification of Honeywell Analytics products were achieved with high quality connectors, providing 360° shield coverage. These connectors generally had metal shells. Failure to properly secure the connector to the equipment will result in high emission levels. Also, poorly constructed, or improperly assembled connectors can be a high source of radiatednoise and provide a path for external signals into the monitor.
CHAPTER 2 Introduction Analyzer side panel — Exterior Vertex Compliance System 7 User Manual
1. Needle Valve for flow adjustment 2. Optics Block 3. RFID reader 4. Take-up reel 5. Tape guide roller 1. Analyzer Number Tag 2. Analyzer Status LED. 3.
Analyzer side panel — Interior 1. Analyzer Main Board 2. Sample pressure transducers 3. Sample flow transducers Analyzer — Front view 1. Analyzer Status LED 2.
Vertex Compliance System 10 User Manual
Clean the Optics Clean Chemcassette optics annually or whenever optics verification error occurs. Compressed air is required 1. Open the Optics Block Gate. 2. Remove the Chemcassette. 3. Open the Vertex Cside panel. 4. Remove tubing (shown in photo) one at the time and blow out with compressed air. 5. Re-secure side panel and reload the Chemcassette. NOTE Remove and clean one port at a time to insure proper orientation of tubing. Do not remove the capillary tubing (microtubes).
3 Additional Information Specifications Size 24.4” x 13” x 6” (62x33x15cm) Weight ≤ 27lbs (12kg) Operating voltage 24VDC Max power consumption 72Watt Certificate UL/IEC 61010-1 FCC NCC Transport time Less than 50 seconds up to 325ft with thin wall tube (0.190”ID) Sample linetubing 1/4 in. (6.35 mm) O.D. x 0.190in. (4.83 mm) (Thin wall) Tubing length Up to 400ft (120m) with thin wall OPERATING CONDITIONS Temperature 59°F to 95°F (15°C to 35°C) Humidity 20-80% RH Altitude -1000 ft.