Honeywell BW™ Icon Portable Multi-Gas Detector
TOC Introduction 5 Product Description 5 Safety 5 Standards and Certifications 5 What's in the Box 7 Overview 7 Operations 9 Activate the Detector 9 Self Test 9 Deactivate the Detector 9 Common Button Operations 10 Bluetooth Pairing 10 Calibration 11 Bump Test 19 Zero Calibration 21 Capture Real Time Reading 22 Set the Detector via Device Configurator 22 Maintenance 23 Charge the Battery 23 Firmware Update 26 Update the Firmware from Local Drive 26 Additional Inform
DataLogs and Event Logs 37 Alarms 38 Replacement Parts 40 Security Information 41 Contact Us 43 Glossary BW Icon 46 3 User Manual
CHAPTER 1 Introduction Learn what you need to know about the Honeywell BW™ Icon Gas Detector before operating. Product Description The Honeywell BW™ Icon & Honeywell BW™ Icon+ gas detectors warns of hazardous gas at levels above user-defined alarm setpoints. The detector can monitor up to four gases at a time. Safety CAUTION • The detector is a personal safety device. It is your responsibility to respond properly to the alarm.
ATEX: Sira 20ATEX2012X I M1 Ex ia op is I Ma, -40°C ≤ Tamb ≤ 60°C (with IR sensor installed) II 1G Ex ia op is IIC T4 Ga, -40°C ≤ Tamb ≤ 60°C (with IR sensor installed) I M1 Ex ia I Ma, -40°C ≤ Tamb ≤ 60°C II 1G Ex ia IIC T4 Ga, -40°C ≤ Tamb ≤ 60°C RE-D Directive 2014/53/EU EMC Directive 2014/30/EU ROHS Directive (EU) 2015/863 amending 2011/65/EU IP: IP66, IP68 (1.2 meters for 45 minutes) Contains FCC ID: SU3RMBLED Contains IC: 20969-RMBLED CSA C22.2 No.
What's in the Box 1 Honeywell BW™ Icon gas detector 1 Battery (factory-installed) 1 USB charger 1 Calibration cap 1 Klick Fast stud 1 Quick Reference Guide 1 Tubing Overview 1 LED Alarm 6 Sensor 2 IntelliFlash 7 Clip 3 Display 8 Battery 4 Button 9 Charging Port 5 Beeper BW Icon 7 User Manual
User Interface Alarm One — Shows when alarm one is breached and gas highlighted next to sensor Alarm Two — Shows when alarm two is breached. Alarm two will over write any alarm one status. Time Weighted Average — Settable in Safety Suite DC for each toxic sensor. Short Term Exposure Limit — Settable in Safety Suite DC for each toxic sensor. Bump — Shows when bump is due and you can configure to have a count down. Calibration — Shows when bump is due and you can configure to have a count down.
CHAPTER 2 Operations Learn what you can do with your Honeywell BW™ Icon Detector, from commissioning to Calibration. Activate the Detector To turn the detector on, press, and hold the button for four seconds. LEDs light and the instrument vibrates and beeps. The detector performs a Self Test and Sensors warm-up. During the Self Test, the icons light and the IntelliFlash flashes in amber. During the Sensor warm-up, the sensors LEDs flash clockwise.
Common Button Operations Feature Operation Power On 4-second hold Power Off 4-second hold Enter or Exit menu Double-Press Switch Menu (bump, cal, & BLE) Single-Press Initiate selected 3-second hold Acknowledge latched alarm 1-second hold Bluetooth Pairing The user can pair the Honeywell BW™ Icon to a mobile device via built-in Bluetooth Low Energy (BLE).
BLE Status Description Pairing The BLE icon flashes every second, Tee IntelliFlash flashes every five seconds, and two shot beeps. Succeed The BLE flashes in blue every 15 seconds. The IntelliFlash flashes every five seconds in green. Fail The BLE icon is solid red, and two short beeps. Calibration Perform a calibration to adjust the sensitivity levels of sensors and ensure accurate responses to gases.
CAUTION Move to a normal atmosphere (20.9% v/v O2) that is free of hazardous gas. Use 50% LEL for test gas. Details for calibration and maintenance: l l l Recommendations for calibration of the equipment on a routine basis including the maximum time interval between calibrations. Calibrate the apparatus before first-time use and then on a regular schedule, depending on use and sensor exposure to poisons and contaminants.
2. Attach the hose. 3. In your mobile device, open the Device Configurator app 4. In your BW Icon: l Double press to enter the menu. l Single press until the BLE icon is displayed. l Hold press 3 secs to initiate the pairing mode. 5. In the Device list screen from the Device Configurator app, select the BW Icon Serial Number to start pairing. 6. In your Mobile device, tap on the Menu button and then select Remote Calibration 7.
8. Enter the Operator Name, and then Tap START. The instrument will start Zero calibration; the IntelliFlash flashes amber, and the Calibration icon lights in blue. 9. After the Zero calibration is completed, use any of the following three ways to select the gas cylinder, and then click START. 10. Select the sensor that you want to calibrate and then tap START.
11. Open the cylinder valve by turning the pressure regulator knob counterclockwise. Follow onscreen instructions to know when to apply gas. The IntelliFlash lights amber.
12. After calibration is complete, a report is displayed. Tap the Return Arrow button to exit the report and go back to the remote calibration main screen. The detector will start purge, and the slots LEDs will flash in yellow clockwise. After purge is completed, the instrument will back to the regular mode.
Procedure to calibrate the detector via the SSDC Calibrate the BW Icon via the Safety Suite Device Configurator (SSDC). 1. Turn On the BW Icon. Place the cap over the detector, and then press down on both tabs to snap it into place. Wait a few minutes to sensors warm up. 2. Attach the hose.
3. Connect the instrument to the PC via the IR link. 4. Log in to SSDC with an authorized user account. For further information, refer to the SSDC User Manual. 5. Click on the Device List View tab, the SSDC scans for connected devices or you can click Refresh to browse manually. 6. Select the connected detector and then click Start Bump/Cal. 7. In the Start Bump/Calibration Test window, do the following: l Select Calibration; l Select the calibration sensor.
Bump Test The detector can be tested in two ways: • Apply gas from a cylinder to the sensors manually through the calibration cap, and using the Safety Suite Device Configurator (SSDC) software or the Device Configurator (DC) app. • Use an IntelliDoX module. For further reference see the IntelliDoX User Manual. CAUTION Move to a normal atmosphere (20.9% v/v O2) that is free of hazardous gas. Use 50% LEL for test gas.
2. Attach the hose. 3. In your mobile device, open the Device Configurator app 4. In your BW Icon: 5. l Double press to enter the menu l Single press until the BLE icon is displayed l Hold press 3 secs to initiate the pairing mode. In the Device list screen from the Device Configurator app, select the BW Icon Serial Number to start pairing. 6. In your Mobile device, tap on the Menu button and then select Bump Test. 7. Enter the Operator Name, and then Tap Save.
Capture Real Time Reading 1. Pair your BW Icon with a mobile device. 2. In your mobile device, open the Device Configurator app. 3. Tap Menu 4. Tap Measurements 5. Tap Start Recording. . Set the Detector via Device Configurator 1. Pair the BW Icon with the Device Configurator App on your mobile device. 2. Tap the menu button 3. Tap Device Setup 4. Tap Download, to get the configuration table. 5. BW Icon Confirm the value of the setting: 1 to turn on, 0 to turn off.
CHAPTER 3 Maintenance Charge the Battery You can charge the battery via an IntelliDox module, the charger adaptor & USB Charger, and the Cradle Charger. Note: The lithium battery may require 5 hours to full capacity. While charging, the battery icon will flash amber once per second. The time needed to charge will increase if the apparatus is activated. The detector may be warm during charging; this is normal. To preserve the life of the battery, deactivate the device when not in use.
l l Dispose of used lithium cells immediately. Do not disassemble and do not dispose of in fire. Do not mix with the solid waste stream. Spent batteries must be disposed of by a qualified recycler or hazardous materials handler. Keep lithium cells away from children. Battery Capacity Indicator Status Duration Indication or Alarm Battery low Less than 7 days IntelliFlash, the battery icon, and the sensor LEDs flash in amber every 5 seconds.
Status Percentage Indication or Alarm Notes when charging with an IntelliDoX: l l l l The auto-power is on if the device is inserted into an IntelliDox module. If the battery is low, the device will auto-power on after the battery capacity exceeds. The auto-power is off if there is no communication with the IntelliDox for five minutes. For further information, refer to the IntelliDoX user Manual. Charge the battery via the USB Charger 1. Press and hold the button to deactivate the detector. 2.
Firmware Update 1. Open the Device Configurator app on your mobile phone. 2. Tap Menu 3. Tap Firmware 4. Tap Update 5. Tap YES to implement the Firmware update, and wait until Update Successfully is displayed. 6. Tap OK.
CHAPTER 4 Additional Information Learn from about strategic information related to the Honeywell BW™ Icon Detector. Sensor Poisons and Contaminants Several cleaners, solvents, and lubricants can contaminate and cause permanent damage to sensors.
Sensor Specifications Gas Type Measuring Range Resolution Measuring Unit New Sensor Warm Up Time Working Temperature CO 0~2000 ppm 1 ppm ppm, mg/m3, umol/mol 0.5h -40°C to +60°C H2S 0~200 ppm 0.1 ppm ppm, mg/m3, umol/mol 0.5h -40°C to +60°C O2 0~30% VOL 0.1% VOL %VOL 12h -40°C to +60°C NDIR-CH4 0~100% LEL 1% LEL %LEL/%VOL SO2 0~150 ppm 0.
General Specifications BW Icon BW Icon + Size 108.2mm x 61.5mm x 43.2mm(4.29" x 2.44" x1.7 ") with Alligator Clip 108.2mm x 61.5mm x 37.8 mm(4.29" x 2.44" x 1.49") with Klick Fast Stud Weight 185g with Alligator Clip,169g with Klick Fast Stud Appearance Colour Yellow, Dark Gray Working Temperature -40°C to +60°C Working Humidity 0%~95% IP Rating IP 66 IP 68, 45min@underwater 1.
Action Time Out Auto exit error screen and power off 5 seconds Auto skip error message screen and enter warmup 5 seconds Auto exit menu and turn off Icon LED 6 seconds Auto exit Force bump and Calibration 30 seconds Exit auto detected span gas 60 seconds Pairing, bump, and calibration result display 5 seconds BLE pairing timeout 60 seconds Troubleshooting Problem Battery icon blinks for 5 seconds when press button to power on.
Problem Cause Solution light for 5 seconds Sensor bay and IntelliFlash Sensors fail light for 5 seconds Bump icon lights for 30 seconds. Detector alarms after startup sequence Replace the sensors Bump overdue and must carry out bump testing before use. Hold the button for 3 seconds to start the bump testing or detector will auto power off after 30 secs. Sensor not stabilized SPE O2 sensor: Wait for at least 10 min before power on.
Problem Cause Solution peak gas exposure reading. Alarm setpoints are set incorrectly. Define the alarm setpoint in Device Configurator. Sensors require calibration. Carry out calibration. Missing or faulty sensors. Replace the sensors. Battery temperature is out of acceptable range. Move to lower temperature ambient to charge the battery. Battery indicator doesn’t display when charging. Battery is depleted. Charge the battery for 8 hours.
DataLogs and Event Logs DataLogs The detector records various information to create a report. The detector is capable of storing 45 days of data. When the memory is full, the detector replaces the oldest datalogs with the most recent datalogs. Event Logs The detector records a maximum of 50 gas alarm, maintenance events, and error conditions.
Alarms A gas detected event supersedes anything. When a gas alarm occurs, even when displaying other behavior, the device goes back to the home screen and displays appropriate gas behavior. When more than one alarm occurs on one sensor, the highest priority is displayed: Over Range > High > STEL, TWA, Low, Negative. When more than one sensor alarms, the alarm status is displayed as multi-alarm no matter what kind of gas alarms they are.
Alarm type from high priority to low Description STEL STEL icon lights in red. Left and Right alarm LEDs flash every second. Alarmed sensor LED flashes twice every second. Beeps and vibrates. TWA TWA icon lights in red. Left and Right alarm LEDs flash every second. Alarmed sensor LED flashes twice every second. Beeps and vibrates. Low Alarm 1 icon lights in red. Left, Right, and sensor alarm LEDs flash every second. Beeps and vibrates. Negative Calibration icon flashes in red every second.
Alarm Setpoint Item Resseting Silencing Low Alarm decease to below 10%LEL to Device Configurator or IR Link Default High Alarm 20%LEL Keep alarm until reading decease to below 20%LEL Silence when transmitting data to Device Configurator or IR Link Default TWA N/A Default STEL N/A Over Alarm >100%LEL Keep alarm until reading increase over 100%LEL Silence when transmitting data to Device Configurator or IR Link Gas Type NDIRCH4 Setting Resolution 1% LEL Low Alarm Setting Range 5~60% High
Security Information This manual provides additional information for the customer and organization related to identification and risk management associated with the use of the system in connected infrastructure.
Honeywell, its affiliates, and licensors liable for any loss of data, loss of profit, or any loss or damage, whether direct, indirect, incidental, special, or consequential, however arising, as a result of accessing or using the software. So long as this provision is enforceable in Customer’s jurisdiction, the foregoing limitations, exclusions, and disclaimers apply to the fullest extent permitted by law even if any remedy fails of its essential purpose.
Contact Us Europe, Middle East, Africa Life Safety Distribution GmbH Toll-Free 00800 333 222 44 Middle East +971 4 450 5800 Middle East +971 4 450 5852 (Portable Gas Detection) gasdetection@honeywell.com Americas Honeywell Analytics Distribution Inc. Tel: +1 847 955 8200 Toll free: +1 800 538 0363 detectgas@honeywell.com Asia Pacific Honeywell Analytics Asia Pacific Tel: +82 (0) 2 6909 0300 India Tel: +91 124 4752700 analytics.ap@honeywell.com Technical Services EMEA: HAexpert@honeywell.com US: ha.us.
www.honeywellanalytics.
© 2020, August 17
Glossary M My Term My definition BW Icon 46 User Manual