X2 Synergy SERIES AS X2.PLUS Analytical Balances PS X2 Precision Balances WLC X2 Precision Balances BRIEF USER MANUAL IMMU-16-S-03-03-21-EN www.radwag.
CAUTION: This is only a brief version of the user manual, it does not provide all product-related information. Prior to moisture analyzer operation it is recommended to read the full user manual version that is to be found on a CD delivered along with the device. The full user manual version describes data crucial for settings and device operation. User manual no.: IMMU-16-14-02-16-ENG If you are reading this, it means that you are bound to achieve success.
CONTENTS 1. General Information .................................................................................... 5 1.1. Dimensions .......................................................................................... 5 1.2. Connectors........................................................................................... 6 1.3. Intended Use ........................................................................................ 7 1.4. Precautions ................................................
-4-
1. GENERAL INFORMATION 1.1.
WLC X2 series 1.2. CONNECTORS X2 balances 1. 2. 3. 4. 5. 6.
AS X2.PLUS balances 1. power supply seat 2. DB9/M connector (external buttons) 3. COM 1 connector (e.g. printer) 4. USB 2 type B connector (e.g. computer) 5. USB 1 type A connector (e.g. computer keyboard) 6. RJ45 Ethernet connector 1.3. INTENDED USE X2 series balances are designed to provide accurate measurement of weighed loads, performed under laboratory conditions. 1.4.
(1)this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. NOTE: The grantee is not responsible for any changes or modifications not expressly approved by the party responsible for compliance. Such modifications could void the user’s authority to operate the equipment.
1.5. WARRANTY CONDITIONS A. RADWAG feels obliged to repair or exchange all elements that appear to be faulty by production or by construction, B. Defining defects of unclear origin defects and means of their elimination can only be realized with assistance of manufacturer and user representatives, C. RADWAG does not bear any responsibility for defects or losses resulting from unauthorized or inadequate performing of production or service processes, D.
2. TRANSPORT AND STORAGE 2.1. DELIVERY CHECKLIST Upon delivery it is necessary to check the package, make sure that your package bears no signs of damage. 2.2. PACKAGE Keep all package elements should your device be transported in the future. Remember that only original packaging can be used for shipping purposes. Prior packing uncouple any cables, remove any separable components (weighing pan, shields, inserts). Pack the device components into an original packaging.
AS X2.PLUS • • remove a transport lock (1) – gently press the transport lock and turn it accordingly to instruction, keep the transport lock should your balance be transported in the future. Install components following the above diagram: • bottom insert (2), • centring ring [embossment side up] (3), • weighing pan (4), • draft shield (5). CAUTION! AS balance with d=0,01/01mg is equipped with 2 weighing pans (standard and open-work). Remember to calibrate the balance upon changing the pan.
WLC X2, 128x128 mm • • WLC X2, 195x195 mm remove tape protecting the grounding spring, located on one of the rubber mandrels (1) install components following the above diagram: • weighing pan (2), • glass draft shield (3). 3.3. SETTINGS It is necessary to level the balance prior connecting it to the mains. To level the balance turn its feet until an air bubble takes central position. The balance shall firmly rest on a surface, each of the feet must be supported. X2.
• • Level the balance by turning its feet left/right, follow the pictograms, level marker moves towards the centre of the circle. When the marker takes central position, its colour turns from red to green – the balance is correctly levelled. Go to the home screen. 3.4. MAINTENANCE ACTIVITIES 1. Disassembly a weighing pan and other detachable components (the components differ depending on a balance type – see Unpacking and Installation section).
Cleaning draft shield panes: Select dissolvent depending on a dirt. Never soak the glass panes in alkaline solutions since they interact with glass and may cause damage. Do not use abrasive substances. For organic dirt use acetone first, next use water or detergent. For other than organic dirt use diluted acid solutions (soluble salts of hydrochloric or nitric acid) or base solutions (ammonium or sodium base).
AS X2.PLUS series balances; disassembly steps: Detach the front part of the frame and take out the front pane. Detach the back part of the frame and take out the top pane. Remove the back pane of the chamber. Remove the left and the right side pane. Carefully disassemble the weighing pan, the draft shield and the bottom insert. Clean the weighing chamber and the panes. All the operations have to be done carefully.
3.5. POWERING THE DEVICE Balance can be connected to the mains only with a power adapter that comes standard with the particular model. Nominal power supply of the power adapter (specified on the power adapter data plate) should be compatible to the power from the mains. Plug the balance to the mains – connect the power adapter to the socket, next connect its connector to port located at the back of the balance housing.
5.
6. WEIGHING MODE HOME SCREEN The home screen of balance software can be divided into 4 sections: • Top section displaying data on active working mode (pictogram and name), metrologically important data and button enabling selection of functions available for a particular working mode. The top bar displays the following information: Working mode name and symbol. Symbol informing that wireless communication is on. Symbol informing that communication with a USB flash drive is on.
7. OPERATING BALANCE MENU Operation of balance software menu is intuitive and uncomplicated. The touch panel makes the software operation easy. Pressing a function key, a soft key or an area on the display initiates an assigned function or process Touch screen operation is also possible while wearing gloves. 7.1. ENTERING BALANCE MENU In order to enter balance menu, press button.
8. WEIGHING OPERATION Load a weighed object on a balance weighing pan. On stabilization of weighing result, indicated by stability marker visible on the left side of balance display, read the measurement result.
8.2. LOGGING Full access to user parameters and to editing databases requires logging as an operator with permissions level. The logging procedure should be carried out on each switching on of the balance.
8.4. WEIGHING UNIT SELECTION Change of weighing unit is carried out by pressing the weighing unit icon visible next to the value of measurement result, or by clicking key (if displayed in an information section). Clicking the unit triggers its replacement, the clicked unit is replaced with the unit that is next on the list of available units. Another option for unit replacement is selecting a particular unit out of the units list, to view the list click key (if displayed in an information section).
8. ADJUSTMENT In order to ensure the highest weighing accuracy, it is recommended to periodically introduce a corrective factor of indications to balance memory, the said factor must be referred to a mass standard. In other words, balance adjustment shall be performed from time to time.
8.9. USER ADJUSTMENT User adjustment is carried out with an optional standard of mass ranging between 0,3 Max and Max. User adjustment and external adjustment procedures are likewise with one exception, before user adjustment start, a message box for entering mass of a standard used for user adjustment is opened. CAUTION! User adjustment is possible for balances that are not a subject to conformity assessment (verification).
Peak Hold Means of operation: max temporary indication occurring in course of the weighing process is hold on a display. Formulations Means of operation: by mixing specified ingredients you can obtain particular mixture, in order to program given formulation you have to specify weight of particular ingredients. Particular working modes settings feature specific functions. The functions enable adapting mode operation to your individual needs.
10. DATABASES Balance software features the following databases < >: Products (5 000 products) Users (100 users) Packaging (100 packaging types) Customers (1 000 customers) Formulations (100 formulations consisting of 25 ingredients maximally) Formulations reports (500 reports) Ambient Conditions (10 000 records) Weighings (50 000 records) Alibi (512 000 records) CAUTION! It is not possible to edit some of the databases for a balance cooperating with E2R software.
13. COMMUNICATION PROTOCOL General information A. A character based communication protocol balance-terminal is designed for establishing communication between a RADWAG balance and a peripheral device via RS-232C serial interface. B. It consists of commands sent from a peripheral device to the balance and responses from the balance. C. Responses are sent from the balance on each receipt of a command as a reaction to a specific command. D.