Technical Information Proline Promag 50W, 53W Electromagnetic Flow Measuring System Flow measurement in water or wastewater applications Application Electromagnetic flowmeter for bidirectional measurement of liquids with a minimum conductivity of ≥5 µS/cm: • Drinking water • Wastewater • Wastewater sludge • • • • Flow measurement up to 110,000 m3/h Fluid temperature up to +80 °C Process pressures up to 40 bar Fitting lengths to DVGW/ISO Application-specific lining materials: • Polyurethane and hard rubb
Proline Promag 50W, 53W Function and system design Measuring principle Faraday’s law of induction states that a voltage is induced in a conductor moving in a magnetic field. In electromagnetic measuring, the flowing medium corresponds to the moving conductor. The induced voltage is proportional to the flow velocity and is detected by two measuring electrodes and transmitted to the amplifier. Flow volume is computed on the basis of the pipe's diameter.
Proline Promag 50W, 53W Input Measured variable Flow rate (proportional to induced voltage) Measuring range Typically v = 0.01...10 m/s with the specified measuring accuracy Operable flow range Over 1000 : 1 Input signal Status input (auxiliary input): U = 3…30 V DC, Ri = 5 kΩ, galvanically isolated. Configurable for: totalizer(s) reset, measured value suppression, error-message reset.
Proline Promag 50W, 53W PROFIBUS PA interface: • Transmission technology (Physical Layer): IEC 61158-2 (MBP), galvanically isolated • Profile version 3.0 • Current consumption: 11 mA • Permissible supply voltage: 9...
Proline Promag 50W, 53W MODBUS interface: • Transmission technology (Physical Layer): RS485 in accordance with ANSI/TIA/EIA-485-A: 1998, galvanically isolated • MODBUS device type: Slave • Adress range: 1...
Proline Promag 50W, 53W Power supply Electrical connection Measuring unit A C B d d g b g b a a a b d/(g) (d) PROFIBUS PA* FOUNDATION Fieldbus* HART* f – 27 + 26 – 25 + 24 – 23 + 22 – 21 + 20 N (L-) 2 L1 (L+)1 d e PA(–)/FF(–) PA(+)/FF(+) – + f – + – + c 27 26 25 24 23 22 21 20 N (L-) 2 L1 (L+) 1 b N (L-) 2 L1 (L+) 1 e c b PROFIBUS DP** MODBUS RS485** PROFIBUS DP* A (RxD/TxD-N) 27 B (RxD/TxD-P) 26 – 25 + 24 f – 23 + 22 – 21 + 20 d d e g A (RxD/TxD-N) B (RxD/TxD-P) – + f – + – + c
Proline Promag 50W, 53W Terminal assignment, Promag 50 Terminal No.
Proline Promag 50W, 53W Terminal No.
Proline Promag 50W, 53W Cable specifications remote version Coil cable: • 2 x 0.75 mm2 PVC cable with common, braided copper shield (Ø approx. 7 mm) • Conductor resistance: ≤ 37 Ω/km • Capacitance: core/core, shield grounded: ≤ 120 pF/m • Permanent operating temperature: –20…+80 °C • Cable cross-section: max. 2.5 mm2 Signal cable: • 3 x 0.38 mm2 PVC cable with common, braided copper shield (Ø approx. 7 mm) and individually shielded cores. • With Empty Pipe Detection (EPD): 4 x 0.
Proline Promag 50W, 53W Supply voltage 85…260 V AC, 45…65 Hz 20…55 V AC, 45…65 Hz 16…62 V DC PROFIBUS PA and FOUNDATION Fieldbus Non-Ex: 9...32 V DC Ex i: 9...24 V DC Ex d: 9...32 V DC Power consumption AC: <15 VA (including sensor) DC: <15 W (including sensor) Switch-on current: • max. 13.5 A (< 50 ms) at 24 V DC • max. 3 A (< 5 ms) at 260 V AC Power supply failure Lasting min.
Proline Promag 50W, 53W Metal, ungrounded piping In order to prevent outside influences on measurement, it is advisable to use ground cables to connect each sensor flange to its corresponding pipe flange and ground the flanges. Connect the transmitter or sensor connection housing, as applicable, to ground potential means of the ground terminal provided for the purpose. Caution! Also, observe company-internal grounding guidelines.
Proline Promag 50W, 53W Plastic pipes and isolating lined pipes Normally, potential is matched using the reference electrodes in the measuring tube. However, in exceptional cases it is possible that, due to the grounding plan of a system, large matching currents flow over the reference electrodes. This can lead to destruction of the sensor, e.g. through electrochemical decomposition of the electrodes. In such cases, e.g.
Proline Promag 50W, 53W Performance characteristics Reference operating conditions To DIN EN 29104 and VDI/VDE 2641: • Medium temperature: +28 °C ± 2 K • Ambient temperature: +22 °C ± 2 K • Warm-up period: 30 minutes Installation: • Inlet run >10 x DN • Outlet run > 5 x DN • Sensor and transmitter grounded. • Sensor centred relative to the pipe. Maximum measured error Promag 50: Pulse output: ± 0.5% o.r. ± 1 mm/s (o.r. = of reading) Current output: plus typically ± 5 µA Promag 53: Pulse output: ± 0.
Proline Promag 50W, 53W Operating conditions Installation conditions Installation instructions Mounting location Correct measuring is possible only if the pipe is full. Avoid the following locations: • Highest point of a pipeline. Risk of air accumulating • Directly upstream of a free pipe outlet in a vertical pipe. F06-5xxxxxxx-11-00-00-xx-000 Installation of pumps Do not install the sensor on the intake side of a pump.
Proline Promag 50W, 53W Partially filled pipes Partially filled pipes with gradients necessitate a drain-type configuration. The Empty Pipe Detection (EPD) function offers additional protection by detecting empty or partially filled pipes. Caution! Risk of solids accumulating. Do not install the sensor at the lowest point in the drain. It is advisable to install a cleaning valve.
Proline Promag 50W, 53W Orientation An optimum orientation helps avoid gas and air accumulations and deposits in the measuring tube. Promag, nevertheless, supplies a range of options and accessories for correct measuring of problematic mediums: • Electrode Cleaning Circuitry (ECC) to remove electrically conductive deposits in the measuring tube, e.g. in accretive mediums.
Proline Promag 50W, 53W Vibrations Secure the piping and the sensor if vibration is severe. Caution! It is advisable to install sensor and transmitter separately if vibration is excessively severe. Information on resistance to vibration and shock can be found on page 20. > 10 m F06-5xxxxxxx-11-00-00-xx-006 Foundations, supports If the nominal diameter is DN ≥ 350, mount the transmitter on a foundation of adequate load-bearing strength. Caution! Do not allow the casing to take the weight of the sensor.
Proline Promag 50W, 53W Inlet and outlet runs If possible, install the sensor well clear of fittings such as valves, T-pieces, elbows, etc. Compliance with the following requirements for the inlet and outlet runs is necessary in order to ensure measuring accuracy: • Inlet run ≥ 5 x DN • Outlet run ≥ 2 x DN F06-5xxxxxxx-11-00-00-xx-005 Adapters Suitable adapters to DIN EN 545 (double flange junction sections) can be used to install the sensor in largerdiameter pipes.
Proline Promag 50W, 53W Length of connecting cable Permissible cable length Lmax depends on the conductivity of the medium. A minimum conductivity of 20 µS/cm is required for measuring demineralized water.
Proline Promag 50W, 53W Degree of protection • Standard: IP 67 (NEMA 4X) for transmitter and sensor • Optional: IP 68 (NEMA 6P) for sensor in remote version Shock and vibration resistance Acceleration up to 2 g by analogy with IEC 60068-2-6. Electromagnetic compatibility (EMC) To EN 61326/A1 and NAMUR recommendation NE 21. Process conditions Medium temperature range The permissible medium temperature depends on the measuring-tube lining: • 0...+80 °C for hard rubber (DN 65...2000) • −20...
Proline Promag 50W, 53W Limiting flow The diameter of the pipe and the flow rate determine the nominal diameter of the sensor. The optimum velocity of flow is 2...3 m/s. The velocity of flow (v), moreover, has to be matched to the physical properties of the medium: • v < 2 m/s: for abrasive mediums such as potter’s clay, lime milk, ore slurry, etc. • v > 2 m/s: for accretive mediums such as wastewater sludge, etc.
Proline Promag 50W, 53W Flow characteristics (US units) Nominal diameter Pressure loss 22 Recommended flow rate Min./max. full scale value (v ~ 0.3 or 10 m/s) Factory settings Full scale value (v ~ 2.5 m/s) Pulse weighting (~ 2 pulse/s) Creepage (v ~ 0.04 m/s) [inch] [mm] 1" 25 2.5…80 gal/min 18 gal/min 0.20 gal 0.25 gal/min 1 1/4" 32 4…130 gal/min 30 gal/min 0.20 gal 0.50 gal/min 1 1/2" 40 7…190 gal/min 50 gal/min 0.50 gal 0.
Proline Promag 50W, 53W Measuring-tube specifications Endress+Hauser Nominal Diameter Pressure rating Inside diameter of measuring tube AS 4087 ANSI [lbs] AWWA JIS [inch] EN (DIN) [bar] AS 2129 [mm] Hard rubber Polyurethane 25 1" PN 40 − − Cl 150 − 20K − 24 32 − PN 40 − − − − 20K − 32 40 1 1/2" PN 40 − − Cl 150 − 20K − 38 50 2" PN 40 − − Cl 150 − 10K − 50 65 − PN 16 − − − − 10K 66 66 80 3" PN 16 Table E Cl.
Proline Promag 50W, 53W Mechanical construction Design / dimensions Dimensions: Wall-mount housing (non hazardous area and II3G / zone 2) E 90.5 (3.56) + 250 (9.84) Esc - 159.5 (6.28) mm (inch) 90 (3.54) 45 (1.77) 135 (5.32) 215 (8.46) 102 (4.02) 192 (7.56) 95 (3.74) 81.5 (3.21) 8 x M5 11.5 (0.45) 53 (2.09) 53 (2.09) 81 (3.19) 81 (3.19) > 50 (1.97) 11.5 (0.
Proline Promag 50W, 53W There is a separate mounting kit for the wall-mounted housing. It can be ordered from Endress+Hauser as an accessory. The following installation variants are possible: • Panel-mounted installation • Pipe mounting Panel-mounted installation 0 + – 00.5 .5 245 +0.5 – 0.5 21 10 ~1 A0001131 Ø 20...
Proline Promag 50W, 53W Compact version DN ≤ 300 F06-53Fxxxxx-06-00-xx-xx-000 DN L A B C K E EN (DIN) / JIS / AS* [mm] ANSI [inch] [mm] [mm] [mm] [mm] [mm] [mm] 25 1" 200 341 257 84 120 94 32 − 200 341 257 84 120 94 40 1 1/2" 200 341 257 84 120 94 50 2" 200 341 257 84 120 94 65 − 200 391 282 109 180 94 80 3" 200 391 282 109 180 94 100 4" 250 391 282 109 180 94 125 − 250 472 322 150 260 140 150 6" 300 472 322 150 260 140
Proline Promag 50W, 53W Remote version DN ≤ 300 129 163 C A B 102 143 E K L (DVGW) A0003219 DN L A B C K E EN (DIN) / JIS / AS* [mm] ANSI [inch] [mm] [mm] [mm] [mm] [mm] [mm] 25 1" 200 286 202 84 120 94 32 − 200 286 202 84 120 94 40 1 1/2" 200 286 202 84 120 94 50 2" 200 286 202 84 120 94 65 − 200 336 227 109 180 94 80 3" 200 336 227 109 180 94 100 4" 250 336 227 109 180 94 125 − 250 417 267 150 260 140 150 6" 300
Proline Promag 50W, 53W 227 187 207 168 Esc + E C A B - 160 Compact version DN ≥ 350 E K L F06-xxFxxxxx-06-05-xx-xx-000 DN L A B C K E EN (DIN) / AS 1) [mm] ANSI / AWWA 2) [inch] [mm] [mm] [mm] [mm] [mm] [mm] 350 14" 550 738.5 456.5 282.0 564 276 400 16" 600 790.5 482.5 308.0 616 276 450 18" 650 840.5 507.5 333.0 666 292 500 20" 650 891.5 533.0 358.5 717 292 600 24" 780 995.5 585.0 410.5 821 402 700 28" 910 1198.5 686.5 512.
Proline Promag 50W, 53W Remote version DN ≥ 350 163 129 C A B 102 143 E K L A0003220 DN L A B C K E EN (DIN) / AS 1) [mm] ANSI / AWWA 2) [inch] [mm] [mm] [mm] [mm] [mm] [mm] 350 14" 550 683.5 401.5 282.0 564 276 400 16" 600 735.5 427.5 308.0 616 276 450 18" 650 785.5 452.5 333.0 666 292 500 20" 650 836.5 478.0 358.5 717 292 600 24" 780 940.5 530.0 410.5 821 402 700 28" 910 1143.5 631.5 512.0 1024 589 750 30" 975 1143.5 631.
Proline Promag 50W, 53W Ground disk (DN 25...300) H ØB 6.5 15 Ø Ø di 10 2 ØD A0003221 DN 1) di B D H EN (DIN) / JIS / AS 4) [mm] ANSI [inch] [mm] [mm] [mm] [mm] 25 1" 26 62 77.5 87.5 32 − 35 80 87.5 94.5 40 1 1/2" 41 82 101 103 50 2" 52 101 115.5 108 65 − 68 121 131.5 118 80 3" 80 131 154.5 135 100 4" 104 156 186.5 153 125 − 130 187 206.
Proline Promag 50W, 53W Weight Weight data in kg Nominal diameter Compact version Remote version (without cable) Sensor 7.3 7.3 5.3 5.3 5.3 6.0 32 1 1/4" 8.0 7.3 – 6.0 5.3 – 6.0 40 1 1/2" 9.4 8.3 9.4 7.4 6.3 7.4 6.0 50 2" 10.6 9.3 10.6 8.6 7.3 8.6 6.0 65 2 1/2" 12.0 11.1 – 10.0 9.1 – 6.0 80 3" 14.0 12.5 14.0 12.0 10.5 12.0 6.0 100 4" 14.7 16.0 12.7 14.0 6.0 125 5" 21.5 21.0 – 19.5 19.0 – 6.0 150 6" 25.5 24.5 25.5 23.5 22.5 23.
Proline Promag 50W, 53W Materials Transmitter housing: • Compact housing: powder coated die-cast aluminium • Wall-mounted housing: powder coated die-cast aluminium Sensor housing: • DN 25…300: powder-coated die-cast aluminium • DN 350…2000: painted steel (Amerlock 400) Measuring tube: • DN < 350: stainless steel 1.4301 or 1.4306/304L; non-stainless flange material with Al/Zn protective coating • DN > 300: stainless steel 1.
Proline Promag 50W, 53W Material load diagrams Caution! The following diagrams contain material load curves (reference curves) for various process connections relating to the fluid temperature. But the maximal permissible fluid temperature always depends on the lining material of the sensor and/or of the sealing material (Page 20).
Proline Promag 50W, 53W Flange connection to ANSI B16.5 Material: A105 [bar] [psi] 60 870 50 725 Class 300 40 580 30 435 20 290 Class 150 145 10 0 0 -60 -40 -20 0 40 20 60 80 100 120 140 160 180 [°C] A0003226 Flange connection to ANSI B16.
Proline Promag 50W, 53W Flange connection to JIS B2238 Material: RSt37-2 (S235JRG2) / H II / 1.0425 [bar] 30 20K 20 10K 10 0 -60 -40 -20 0 20 40 60 80 100 120 140 160 180 [°C] A0003228 Flange connection to AS2129 Table E or AS4087 Cl. 14 Material: A105 / RSt37-2 (S235JRG2) / St44-2 (S275JR) [bar] 25 20 15 10 5 0 -60 -40 -20 0 20 40 60 80 100 120 140 160 [°C] F06-xxFxxxxx-05-xx-xx-xx-010 Fitted electrodes Measuring, reference and EPD electrodes: • Standard with: 1.
Proline Promag 50W, 53W Human interface Display elements • Liquid-crystal display: backlit, two lines (Promag 50) or four lines (Promag 53) with 16 characters per line • Custom configurations for presenting different measured-value and status variables • Totalizer: Promag 50: 2 totalizers Promag 53: 3 totalizers Operating elements Unified operation concept for both types of transmitter: Promag 50: • Local operation with three push buttons (–, +, E) • Quick Setup menus for straightforward commissioning P
Proline Promag 50W, 53W Certificates and approvals Ex approvals Information about currently available Ex versions (ATEX, FM, CSA) can be supplied by your Endress+Hauser Sales Centre on request. All explosion protection data are given in a separate documentation which is available upon request. CE mark The measuring system is in conformity with the statutory requirements of the EC Directives. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.
Proline Promag 50W, 53W Ordering information The Endress+Hauser service organisation can provide detailed ordering information and information on the order codes on request. Accessories Various accessories, which can be ordered separately from Endress+Hauser, are available for the transmitter and the sensor. The Endress+Hauser service organisation can provide detailed information on request.