Operating Instructions Safety Data Manager (SDM) Memo-Graph S System compatible data manager with a unique safety concept for critical applications. Compliant to the high FDA requirements laid down in the 21 CFR, Part 11. 8 BA 138R/09/en/08.04 No.
Memo-Graph S Overview Your new Safety Data Manager (SDM) has the Operating Instructions built-in. The device's simple control system enables you to commission for many applications, practically without any paper. Your SDM displays instructions at the push of a button directly on screen. This description is nevertheless delivered with the device - it is a supplement to the Operating Instructions built into the device.
Memo-Graph S CHANGE PROTOCOL CHANGE PROTOCOL Safety Data Manager Memo-Graph S Serial No.: see original name plate on unit Installed at (company / plant): DOCUMENT REVISION HISTORY Rev. # Comment Edited by Date Reviewed by Date 1.0 First release Kinzel 01/06/ 2002 Sabine Eisenmann 01/06/ 2002 1.1 Included description of Ethernet connection Kinzel 01/03/ 2003 Dieter Schmidt 01/03/ 2003 1.2 Corrected fault in connection diagram of RTD thermometers; changed Chapter 3.3.
Memo-Graph S Table of contents Table of contents Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 5.4 Brief overview . . . . . . . . . . . . . . . . . . . . . . . 2 5.5 5.6 5.7 1 Safety instructions . . . . . . . . . . . . . . . . 7 1.1 1.2 1.3 1.5 Notes on safety conventions and icons . . . . . . . . . . . Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . Installation, commissioning and operation . . . . . . . . Return . . . . . . . . . . . . . . . . . . . . .
Table of contents Memo-Graph S 8 Accessories . . . . . . . . . . . . . . . . . . . . 78 9 Trouble-shooting . . . . . . . . . . . . . . . . 79 9.1 9.2 9.3 9.4 Response of device to faults . . . . . . . . . . . . . . . . . 79 LED functions . . . . . . . . . . . . . . . . . . . . . . . . . . . 79 Searching for and eliminating faults . . . . . . . . . . . . 80 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81 9.4.1 Spare parts diagram . . . . . . . . . . . . . . . . . 81 9.4.
Memo-Graph S 1 Safety instructions 1 Safety instructions 1.1 Notes on safety conventions and icons The devices are safely built and tested according to state-of-the-art technology and have left the factory in perfect condition as regards technical safety. The devices comply with the relevant standards and guidelines as per EN 61010 “Safety requirements for electrical equipment for measurement, control and laboratory use”. They can, however, cause danger if used improperly or other than intended.
1 Safety instructions Memo-Graph S 1.4 " Operational safety Caution! Operation is only guaranteed as safe when the instructions and warnings in these Operating Instructions are observed: • Only operate the device when it has been installed. • Installation and connection require qualified experts. Please provide shock protection and connection in accordance with the valid safety regulations. • The protective earth connection must be made before all other connections.
Memo-Graph S 2 Identification 2 Identification 2.1 Device designation 2.1.1 Nameplate Compare the nameplate on the device with the delivery note and with the following diagram: 1 2 3 4 5 6 7 Figure 1: 1 2 3 4 5 6 7 Endress+Hauser Nameplate data for the “Safety Data Manager” Device designation Order code Device's serial number Order number Power supply data Power consumption data Patent no.
2 Identification Memo-Graph S 2.1.
Memo-Graph S 3 Installation 2.2 Scope of delivery • Device (with screw plug-in terminals for mains and signal connection, according to your order) • 2 jack screws • 1 screwdriver, blade width of 2.5 mm • PC operating and configuration software on CD-ROM • Delivery note • These Operating Instructions • Memory card / floppy disk, in case it has been ordered Anything missing? Then please inform your supplier. 2.
3 Installation Memo-Graph S 3.3 Installation instructions 3.3.1 Mounting kit To install the control panel, all you need is a screwdriver. 3.3.2 Figure 2: Panel mounting Panel mounting for version with die-cast or stainless steel front Make a panel cutout, 138+1 x 138+1 mm in size (acc. DIN 43700). Please observe the asymmetrical alignment of the front bezel to the casing. The installation depth of the device is approx. 211 mm without, or 232 mm with, rear panel or terminal cover. 1. 2. 3.
Memo-Graph S 4 Wiring " Endress+Hauser 4 Wiring 4.1 Quick wiring guide Caution! Before wiring, please compare the supply voltage with the information specified on the nameplate. If safe operation is no longer possible (e.g. visible damage) please place the device out of service and secure it against unintentional start-up.
4 Wiring Memo-Graph S Analogue outputs (can be preset in set-up) RL= Cable resistance RL + x15 8 3 x16 RxD/TxD(+) RxD/TxD(_) Further units x16 + I out - 0...20 mA, 4...20 mA - x15 RS 485 RL= Cable resistance RL + x15 8 3 x16 RxD/TxD(+) RxD/TxD(_) Further units x16 x16 5 1 6 To PC: Cable with 9 pol. Sub-D connector 9 7 14 13 25 2 3 5 To PC: Cable with 25 pol. Sub-D connector 5 6 9 Rel.1 + + + + - N L 44 42 41 Rel.5 Rel.4 Rel.3 Rel.2 Ch.2 Ch.
Memo-Graph S 4 Wiring " Caution! Depending on the order, slots 1 or 2 can be fitted with different boards (with analogue and/or digital inputs/outputs). Please carry out connection in accordance with terminal diagram. 4.2 " Terminal diagram Caution! If high-energy transients occur when using long signal cables, we recommend connecting a suitable overvoltage protection (e.g. E+H HAW 561 (51003570) and HAW 560 (51003571)).
4 Wiring Memo-Graph S Optional open collector output* (NPN Transistor): 98 Collector 99 Emitter ! Note! The opening and closing function (= activation and deactivation of the relay coil or the open collector output) in case of a limit value can be defined in the set-up “Set-up - Signal settings - Digital outputs (Relay / OC)” Optional digital inputs on power supply board: 91 Digital input 1 92 Digital input 2 93 Digital input 3 94 Digital input 4 95 Digital input 5 96 Digital input 6 97 Di
Memo-Graph S 4 Wiring 4.2.3 Multifunction output board with relay (slot 2) Analogue outputs: The first digit (x) of the three-digit terminal number corresponds to the associated channel (1.. to 8..: channels 1 to 8, or A.. to H..: channels 9 to 16).
4 Wiring Memo-Graph S ! B52 Relay 16, contact 2 C51 Relay 17, contact 1 C52 Relay 17, contact 2 Note! The opening and closing function (= activation and deactivation of the relay coil) in case of a limit value can be defined in the set-up “Set-up - Signal settings - Digital outputs (Relay / OC)” 4.2.
Memo-Graph S 4 Wiring Relay outputs* on digital board(s): ! Digital board 1 Digital board 2 151 Relay 6, contact 1 751 Relay 12, contact 1 152 Relay 6, contact 2 752 Relay 12, contact 2 251 Relay 7, contact 1 851 Relay 13, contact 1 252 Relay 7, contact 2 852 Relay 13, contact 2 351 Relay 8, contact 1 951 Relay 14, contact 1 352 Relay 8, contact 2 952 Relay 14, contact 2 451 Relay 9, contact 1 A51 Relay 15, contact 1 452 Relay 9, contact 2 A52 Relay 15, contact 2 551 R
4 Wiring Memo-Graph S Signal 9-pin RS 232 to device 9-pin RS 232 to PC 25-pin RS 232 to PC 9-pin RS 232 to modem 25-pin RS 232 to modem TxD 2 2 3 3 2 RxD 3 3 2 2 3 GND 5 5 7 5 7 Screen Housing (Housing) (Housing) (Housing) (Housing) Optional interface: alternatively RS 485 or Profibus DP (bus monitor) Pin RS 485 PROFIBUS-DP 3 RxD/TxD (+) RxD/TxD (+) RxD/TxD (-) RxD/TxD (-) Screen Screen 4 5 8 9 Housing 4.2.
Memo-Graph S 4 Wiring 4.3 Connecting a device to the Ethernet (TCP/IP) In principle, all devices equipped with an internal Ethernet interface can be integrated into a PC network (TCP/IP Ethernet). The device(s) can be accessed by any of the PCs in the network using PC software. It is not necessary to install driver software ("COM redirection") on the PC because the PC software has direct access to the Ethernet.
4 Wiring Memo-Graph S 4.3.2 Operating location The operating location of the device should be selected such that the network-specific maximum permitted cable length of 100 metres is not exceeded. Please ensure that all connections are inserted only when the end devices are switched off. 4.4 Ingress protection class The device with stainless steel front meets all the requirements of ingress protection IP65 (NEMA 2x).
Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning 5 Operation / adjusting device settings Set-up to commissioning 5.1 Basics on set-up directly at the device • The function of the operating keys is described in the fields directly above the corresponding keys on the screen. Empty fields mean that the corresponding keys currently have no function. • Access to the set-up is released ex works and can be locked in various ways: – Entering a 4-digit release code (see Chapter 5.
5 Operation / adjusting device settings - Set-up to commissioning Checking the rear panel cover OFF Rear panel cover mounted Independent of whether mounted or not mounted Memo-Graph S ID and password system activated Logged in No Function not available Yes - No Administrator Yes User No Function not available No - No Administrator No; Administration possible User No Function not available Yes - No Administrator Yes User No Yes No Yes Yes ON No Set-up changes permitted N
Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning The main menu: ! Note! If individual chapters are missing, your device is not equipped with this option. The individual parameters are summarised in the main menu in chapter: Basic settings see Chapter 5.7.1 – General parameters (date, time, release code ...) Signal settings see Chapter 5.7.
5 Operation / adjusting device settings - Set-up to commissioning Memo-Graph S Input principle: 1. Begin changing each parameter with “↵”. 2. Use “↓” or “↑” to leaf through values, symbols, selection lists. 3. If the parameter is set correctly, use “↵” to confirm. ! Note! – Any settings that are displayed in grey can not be selected / can not be changed (only notes, or option not available/not activated). – With the factory setting “0000” (delivery status), configuration is possible at any time.
Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning 5.6 Set-up using ATA flash card Save the device settings on the PC onto the ATA flash card using the operating and readout software. This set-up file can be adopted into the device, if released, under “Extras - ATA flash - Load ATA flash set-up”. ! " ! Note! In order to be able to use this function, the device must already be registered in the PC database.
5 Operation / adjusting device settings - Set-up to commissioning Memo-Graph S Set-up -> Basic settings Time NT -> ST Time at which the time is moved forward by 1 hour on the day of the changeover from normal time to summertime. Format: hh:mm Date ST -> NT Day in autumn on which the changeover from summertime to normal time takes place. Format: DD.MM.YY Time ST -> NT Time at which the time is moved back by 1 hour on the day of the changeover from summertime to normal time.
Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning Set-up -> Basic settings -> Operating modes LED operating mode Namur “NE44”: the signalling through the front-mounted LEDs complies with NAMUR guideline NE 44. Green LED lights up: power supply OK Red LED lights up: measurement signal failure. Red LED flashes: maintenance required, e.g. messages to be acknowledged, calibration .... Namur “NE44+”: same as Namur “NE44” + limit values.
5 Operation / adjusting device settings - Set-up to commissioning Memo-Graph S Basic settings: ATA flash change sub-menu Settings that specify when/how the filling level is reported to an ATA flash card used as stack memory. Set-up -> Basic settings -> ATA flash change Warning at xx % Warns before ATA flash card is 100 % full. The internal (ring) memory is still written on during the change or when the ATA flash card is full.
Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning 5.7.2 Signal settings Settings, groupings and combinations for analogue and digital inputs and outputs. ! Note! Values can only be changed if the authorised person is logged in, otherwise the values can only be looked at. Signal settings: Analogue inputs sub-menu: View or change settings/limit values of connected measuring points for analogue measured variables.
5 Operation / adjusting device settings - Set-up to commissioning Memo-Graph S Set-up -> Signal settings -> Analogue inputs -> Analogue input x Zoom end Same as “Zoom start”. However, enter the upper value of the required section here. Example: Transmitter 5-9 pH. Enter “9” here. Offset Factory setting “0”. Set value is added for further use (display, storage, limit value monitoring) for real measured input signal.
Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning Signal settings -> Analogue inputs -> Analogue input x: Profibus DP sub-menu (Option Listener / Monitor): Settings only required if the measured values for this channel are to be acquired using the data interface (and not by an analogue input). ! Note! This option can only be seen if the device has been ordered with the “Profibus DP Monitor” option and has been activated under “Interfaces - Data interface - DP bus monitor”.
5 Operation / adjusting device settings - Set-up to commissioning Memo-Graph S In the analogue inputs / signal menu the “from data interface” option appears, i.e. the input signal is sent via Profibus to the SDM. Signal settings -> Analogue inputs -> Analogue input x: Limit value x sub-menu: Settings only required if limit values are to be monitored for this channel. If a limit value type has been selected, it is marked as activated by “Limit value created”.
Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning Set-up -> Signal settings -> Analogue inputs -> Analogue input x -> Limit value x Hysteresis type “percentual %”: set hysteresis in %. “absolute”: specify hysteresis in the set process unit (e.g. in °C, bar, ...). Hysteresis value The limit value status is not deactivated again until the signal is back to the normal range again by the value set here.
5 Operation / adjusting device settings - Set-up to commissioning Memo-Graph S Set-up -> Signal settings -> Digital inputs -> Digital input x Function Activation of the input triggers the set function. The digital inputs are high-active, i.e. the selected effect is achieved by activation with +12 to +30 VDC. See also Chapter 4.2 “Connections/terminal diagram”.
Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning Set-up -> Signal settings -> Digital inputs -> Digital input x Effect of the control input Only when used as control input: triggers the allocated control function on the device when activated. “only record”: the control input has no function, it is just recorded. “External intermediate analysis”: as long as the control input is activated all active channels are analysed, i.e. the analysis of min./max.
5 Operation / adjusting device settings - Set-up to commissioning Memo-Graph S Set-up -> Signal settings -> Digital inputs -> Digital input x Continued: effect of the control input “Reset batch number”: if “Autom. incrementation (6 characters)” has been selected under “Basic settings - Operating modes - Batch operating mode”, a new batch number is generated with each new batch start, in which this number is incremented by 1.
Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning Set-up -> Signal settings -> Digital inputs -> Digital input x Display Counter readings are saved at determined intervals (e.g. intermediate counter, daily counter, monthly counter, total / year counter, product counter... - see setting in Chapter 5.7.3 “Signal analysis”). Here, the counter type is selected that is to be displayed constantly.
5 Operation / adjusting device settings - Set-up to commissioning Memo-Graph S Only when using the digital input as on/off events function: Set-up -> Signal settings -> Digital inputs -> Digital input x Function On/off events Designation Description of the function or measuring point name carried out with this input. Designation ‚H' Description of the status when the control input is activated. 5 characters. Logically high = +12 to +30 V.
Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning Detailed settings in the set-up under Signal settings - Analogue combinations: Set-up -> Signal settings -> Analogue combinations -> Mathematics channel x Formula Individual channels can be mathematically combined with each other and calculated with functions (g) / factors (a or b) / constants (c).
5 Operation / adjusting device settings - Set-up to commissioning Memo-Graph S Set-up -> Signal settings -> Analogue combinations -> Mathematics channel x Zoom end Here, enter the highest value that the mathematical combination can produce. Example: y1 measuring range = 0...10, a = 5, y2 measuring range = 0...20, b=3, k=4, Addition of the channels according to the formula: y = (10*5) + (20*3) + 4. Result: enter “114”. Data interface Not used: the data interface is not used. Ext.
Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning Set-up -> Signal settings -> Digital combinations -> Combination x Function “Addition/Subtraction”: for impulse counter. “Sum/average”: via impulse or operating time counter. “Logical combination”: for control input, logging in/out, operating time and events + operating time ! Note! Only those channels that it makes sense to combine can be selected. Combinations are processed like real signals, i.e.
5 Operation / adjusting device settings - Set-up to commissioning Memo-Graph S Set-up -> Signal settings -> Digital combinations -> Combination x Description “H” Description of the status when the combination is logically true (“H”). Only for logical combination. Description “L” Description of the status when the combination is logically incorrect (“L”). Only for logical combination. Events text L->H Description of the status change from logically incorrect (“L”) to logically true (“H”).
Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning Signal settings: Analogue outputs (option) sub-menu Settings only required if analogue outputs are to be used. Set-up -> Signal settings -> Analogue outputs -> Analogue output x Output signal Can be selected per channel. See also “Connections / terminal diagram” Reference channel Here, the input channel that is to be changed is selected.
5 Operation / adjusting device settings - Set-up to commissioning Memo-Graph S Set-up -> Signal settings -> Digital outputs (Relays/OC) Relay 2 (3, 4, 5) in LV Signal settings -> Digital outputs, Position 2-5: Function of the relay in case of a limit value, i.e. when it is activated. “Closing”: in case of a LV the corresponding connections (shown in brackets) are shorted, opened in idle status.
Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning Set-up -> Signal settings -> Signal groups -> Group x Memory cycle Time for measured value memory cycle = cycle for updating display in normal operation (no limit value is exceeded or limit values are not monitored). This provides the visual range for the graphic display (see value in brackets).
5 Operation / adjusting device settings - Set-up to commissioning Memo-Graph S Set-up -> Signal analysis Synchronous time Time at which signal analyses are completed. Example of daily analysis: the daily analysis is completed at the set time, i.e. it contains the analysis of the measured values over the last 24 hours. Zero reset The analyses are reset with this function (e.g. reset after completing the commissioning of a system) ! Note! • All previous (commissioning) signals are discarded.
Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning Products: Product x sub-menu Viewing or changing settings for the selected product. Set-up -> Products -> Product x Product active A product's parameters, such as the limit values, are only monitored when the product is activated. The product name and product code are also saved in the events list. ! Note! Only activated products can be selected during normal operation.
5 Operation / adjusting device settings - Set-up to commissioning 5.7.5 Memo-Graph S Texts 40 different texts (22 characters) can be entered. Settings only required if you are using the text version (“Texts” softkey). Here, enter the texts that are to be output during operation. 5.7.6 Administration The system administration is carried out here. The administrator has to be defined first. Users (19) can then be registered or deleted. The user's rights can be allocated in the “Create ID” menu.
Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning Administration: Create ID sub-menu Set-up -> Administration -> Create ID New ID Enter a unique user identification. This identification must only exist once in the system. The first ID entered automatically receives administration rights. Name Enter the name of the new user. Access rights Here, enter the new user's access rights. ! Note! The user can operate the device.
5 Operation / adjusting device settings - Set-up to commissioning Memo-Graph S Interfaces: RS232 / RS422 / RS485 sub-menu Settings only required if you use the device's interfaces (operation using PC, serial data readout, modem operation, etc.). Set-up -> Interfaces -> RS232 / RS422 / RS485 Unit address Every device used in series must have its own address (00...99). This is required for identifying the PC software.
Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning Interfaces: Data interface (option) sub-menu Selecting the data interface via which communication is made. Settings only required if you are communicating via an external Profibus module. Set-up -> Interfaces -> Data interface Communication with Not used: the data interface is not used. Ext. DP slave module: communication via the externally-connected DP slave module.
5 Operation / adjusting device settings - Set-up to commissioning Memo-Graph S Service: General sub-menu Information for service technicians, e.g. for questions about the device/device fault Set-up -> Service -> General SW version Device software version. Please state this code if you have questions about the device. Last power on Please state this code if you have questions about the device. Last C-assertion Please state this code if you have questions about the device.
Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning Service: Calibration sub-menu Factory calibration for clock module and rear wall temperature. " Caution! Do not change this, otherwise the device will malfunction. Calibration only possible using a PC calibration routine. Must only be carried out by qualified skilled personnel.
5 Operation / adjusting device settings - Set-up to commissioning Memo-Graph S 5.9 Communication via serial interfaces / modem 5.9.1 RS 232 The RS 232 serial interface can be accessed from the front (3.5 mm stereo jack plug under the ATA flash drive) (not available for stainless steel front) and from the rear (9 pin Sub-D socket). " Caution! • Simultaneously using the front and the rear-mounted RS 232 is not permitted. Device malfunction.
Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning 5.9.4 Commissioning a modem link In principle, any conventional modem with AT command set can be used for data transfer between your device with RS 232 interface and the supplied PC software. ! Note! Data transmission between an analogue modem and an ISDN terminal is not possible.
5 Operation / adjusting device settings - Set-up to commissioning ! Memo-Graph S Note! You can obtain the system parameters from your relevant network administrator. The following system parameters must be set: 1. 2. 3. IP address Subnet mask Gateway 5.10.1 Menu: SETUP – Miscellaneous All parameters that affect the device's operating system are configured in the setup - miscellaneous menu. The system parameters are input in the „RS232 / Ethernet“ menu.
Memo-Graph S 5 Operation / adjusting device settings - Set-up to commissioning Please be aware that this number is not arbitrary, rather it must be defined dependent on the network address of the TCP/IP network. The input form corresponds to syntax (e.g. 172.016.231.005). Complete the input with "↵ accept". 5.10.4 Subnet mask allocation The subnet mask must be input if the device is to establish connections in another sub-network.
5 Operation / adjusting device settings - Set-up to commissioning 3. Memo-Graph S Now enter the IP address. The port address is 8000. Entering the device address and the release code is optional. Confirm the input with "Continue" and start the transfer with OK. The connection is now established and the device is saved in the device database.
Memo-Graph S 6 Guaranteeing the requirements of 21 CFR 11 6 Guaranteeing the requirements of 21 CFR 11 6.1 General Prior to introducing electronic signatures, a letter, signed by hand, must be sent to the Office of Regional Operations (HFC-100) 5600 Fishers Lane Rockville, MD 20857 USA informing it of the company's intention to use electronic documents/signatures in the future. Administrators and users must be trained to 21 CFR 11 or must already have appropriate experience in this area.
6 Guaranteeing the requirements of 21 CFR 11 6.2 Important settings at the device 6.2.1 Basic settings/operating modes Memo-Graph S • ATA operating mode: Recommendation for stand-alone applications: stack memory Recommendation for cyclic serial PC read-out: ring memory • Encryption: On • Cover: Checking Note: After commissioning and qualification of the expected device functions/system validation, cover the rear of the device with a terminal cover/rearside cover.
Memo-Graph S 6 Guaranteeing the requirements of 21 CFR 11 6.2.3 Administration Set the password length and validity period etc. in accordance with your application-related risk potential. • Administrator password: Recommended: min. 7 characters • User password: Recommended: min. 5 characters • Password valid for: Recommended: 000 (depends on application) 6.2.4 Administration/create ID • New ID: Unique ID (may only occur once in the system, e.g.
6 Guaranteeing the requirements of 21 CFR 11 6.2.5 Memo-Graph S Analogue inputs • Select the number of decimal places to suit the measuring range of the sensor/transmitter used. Note: The accuracy of the device is not increased by selecting more decimal places. 6.3 Important PC software settings Certain functions are required in the associated PC software in order to meet the requirements of 21 CFR 11.
Memo-Graph S 6 Guaranteeing the requirements of 21 CFR 11 6.3.2 Activating the automatic read-out function and autom. data storage A reliable, fully functional interface connection between the device and the PC software is required. This should be verified before activating automatic batch print-out.
6 Guaranteeing the requirements of 21 CFR 11 6.3.4 Memo-Graph S Activating automatic batch print-out A reliable, fully functional interface connection between the device and the PC software and a fully functional, trouble-free printer driver are required. This should be verified before activating automatic batch print-out.
Memo-Graph S Endress+Hauser 6 Guaranteeing the requirements of 21 CFR 11 67
6 Guaranteeing the requirements of 21 CFR 11 68 Memo-Graph S Endress+Hauser
Memo-Graph S 7 Procedure in operation 7 Procedure in operation Version with die-cast front Figure 7: 1 2 3 4 5 6 7 8 ! : Version with stainless steel front: Version with die-cast front and version with stainless steel front Colour screen Function LEDs Drive doors Socket for front-mounted RS 232 operating interface (only for IP54 die-cast front) ATA flash memory card Eject button for ATA flash memory card Write LED for ATA flash Operating keys / Softkeys Note! On the version with “IP 65 stainless
7 Procedure in operation Memo-Graph S Functions of the LEDs “NamurNE44” setting (see “Set-up”): - Green LED lights up: power supply OK - Red LED lights up: measurement signal failure - Red LED flashes: maintenance required, e.g.
Memo-Graph S 7 Procedure in operation Checking product limit values: Product - Information Product - Carrying out recalibration: Product - Recal.
7 Procedure in operation Memo-Graph S Saving device settings / saving onto ATA flash card: Extras - ATAFlash - Save set-up to ATAFlash Reading in device settings / loading ATA flash card: Extras - ATAFlash - Load set-up from ATAFlash Saving IDs / saving to ATA flash card: Extras - ATAFlash - Save IDs to ATAFlash Reading in IDs / loading from ATA flash card: Extras - ATAFlash - Load IDs from ATAFlash Checking memory capacity: Extras - Hardware Checking ATA flash card technical data: Extras - Hardware - ATA
Memo-Graph S 7 Procedure in operation • The user who presently has responsibility is displayed in normal operation in the screen's header. If nobody is logged in and has responsibility, “Logged out” appears. Changing password • Select Function - Change password. • Then select your ID and enter you current password (or the initialisation password obtained from the administrator). • Now enter your new password and confirm it by re-entering it.
7 Procedure in operation Memo-Graph S 7.2.3 Enter comment / text The device can save pre-definable or free texts together with the automatically-produced messages. A quality controller can, for example, use this to check and sign for the production. ! Note! • The text entry can be carried out at any time, even when a production procedure is not running. • The text is not saved until a unique ID/password combination (“Electronic signature”) has been entered.
Memo-Graph S 7 Procedure in operation Group display Select how the currently-selected group is to be displayed: • Plot - time axis is plotted horizontally, signals use the entire display range. • Plot in zones - time axis is plotted horizontally, every signal is displayed in its own display zone, signals do not overlap. • Waterfall - time axis is plotted vertically, signals use the entire display range.
7 Procedure in operation Memo-Graph S Search The ring memory can be searched through according to different search criteria. The result is displayed in a list. You can search for events, a time or a certain measured value. Overview Here, there is an overview of the current device status. Contrast Adjusts the angle of view (up / down), in order to achieve the optimum contrast - depending on the mounting location. ! Note! This setting has no effect on the brightness/service life of the back lighting.
Memo-Graph S 7 Procedure in operation The following functions are available: • Update ATA flash: closes the current block in the memory and saves it onto the card including the most recently saved values independent of the data quantity. • Save set-up to ATA flash or load set-up to ATA flash: this copies all device settings (except IDs of the users / administrators approved for the device) onto the card or reads in the device's set-up data into a new device (important, e.g. in case of replacement).
8 Accessories Memo-Graph S 8 Accessories Accessories included in delivery see Chapter 2.2 Designation Order code Ethernet module, RS 232, 230 VAC for DIN top hat rail mounting incl. interface cable RSG12A-E2 Ethernet module, RS 232, 115 VAC for DIN top hat rail mounting incl. interface cable RSG12A-E3 Ethernet module, RS 485, 230 VAC for DIN top hat rail mounting incl. interface cable RSG12A-E4 Ethernet module, RS 485,115 VAC for DIN top hat rail mounting incl.
Memo-Graph S 9 Trouble-shooting Spare parts structure for the Safety Data Manager's device software Software A Standard software Operating language A B C D E F G H I J L RSG12A1- A German English French Italian Spanish Dutch Danish American English Polish Russian Swedish ⇐ Order code 9 Trouble-shooting 9.1 Response of device to faults Your Safety Data Manager informs you of faults or incorrect entries using plain text on the screen. 9.
9 Trouble-shooting Memo-Graph S 9.3 Searching for and eliminating faults Problem Analogue input shows “---” Display does not function No LED lights up LED lights up ATA flash drive does not function Digital input does not function Relays do not function Analogue outputs do not function Configuration is locked 80 Cause Remedy The signal lines are incorrectly connected or not connected. Please check the connections. The input signal does not correspond to the configured signal.
Memo-Graph S 9 Trouble-shooting Problem Cause Modem transmission does not function Remedy Modem at Memo-Graph S has not been initialised Initialise modem with PC software Wrong cable between modem and Memo-Graph S Use original cable RSG12A-S2 Device address or release code of Memo-Graph S and PC software do not match Set identical device address and release code in Memo-Graph S and PC software Etrhernet connection does not function IP address, subnet mask or gateway are not set correctly Check
9 Trouble-shooting Memo-Graph S 9.4.2 82 Spare parts list Pos. Order No. HOUSING AND HOUSING PARTS 1 50084747 Casing without front bezel 2 RSG12X-HA Front bezel (die-cast) IP54 2 RSG12X-HB Front bezel (die-cast) IP54, neutral 2 RSG12X-HC Front bezel (stainless steel) without door IP65 2 51003843 Front bezel (die-cast) varnis.
Memo-Graph S 9 Trouble-shooting Pos. Order No. ELECTRONICS AND CONDUCTOR BOARDS 14 RSG12X-EE PROFIBUS DP interface module 93.75kBit/s, until device no. 44542041 14 RSG12X-EF PROFIBUS DP interface module 45.45kBit/s, until device no. 44542041 RSG12X-LA Retrofit-kit for drive ATA-Flash on front (without card, from unit no. 44542041) RSG12X-LB ATA-Flash card backside (from unit no.
9 Trouble-shooting Memo-Graph S 9.5 Repairs/return The device must be packed in protective packaging for later reuse or in case of repair. The original packaging offers the best protection. Repairs must only by carried out by your supplier's service organisation or by skilled personnel. ! Note! When sending for repair, please enclose a note with a description of the error and the application. 9.6 Disposal Please observe local regulations. 9.7 1. 2. 3. 4. 5. 6. 7.
Memo-Graph S 10 Technical data 10 Technical data 10.1 Function and system design Measuring principle Electronic acquisition, recording and archiving of analogue and digital input signals. Measuring system The connected analogue measuring points are measured at a scan cycle of 125 ms. Galvanic isolation channel-channel: 60 VP. Damping adjustable at 0...999.9 seconds per analogue input, system base damping negligible.
10 Technical data Memo-Graph S Description Mesurement range Signal resolution/accuracy Thermocouples Type T (Cu-CuNi): -270 to +400 °C / -454 to 752 °F 0.05 K / 0.25 % of measurement range from -200 °C / -328 °F Type U (Cu-CuNi): -200 to +600°C / -328 to 1112 °F 0.1 K / 0.25 % of measurement range from 0 °C / 32 °F Type W3 (W3Re/W25Re): 0 to +2315 °C / 32 to 4199 °F 0.2 K / 0.25 % of measurement range Type W5 (W5Re/W26Re): 0 to +2315 °C / 32 to 4199 °F 0.2 K / 0.
Memo-Graph S 10 Technical data 10.3 Output values 10.3.1 Multifunction output board (socket 2) Analogue outputs 4 or 8 analogue outputs, each galvanically isolated from all electric circuits (test voltage 500 VAC) Output range: 0 - 10 V, 1 - 5 V, 0 - 20 mA, 4 - 20 mA Accuracy: 0.25 % of range (at 1-5 V 0.5 %) Resolution: 0.025 % (at 1-5 V 0.06 %) Temperature drift: < 0.05 % /K of range Output ripple: < 10 mV effective Response time: max.
10 Technical data Memo-Graph S Given that this is an interface for industrial use, a screened cable (CAT5) must be used. If a PC is to be connected directly (without Hub or Switch) then a so called cross-over cable must be used. PROFIBUS DP connection (option) “Bus monitor” function - Multi-channel PROFIBUS display and registration (Without influence on PROFIBUS system) as with conventionally-connected components.
Memo-Graph S 10 Technical data Ingress protection class Front ingress protection: for die-cast front with door: IP 54 (EN 60529, Cat. 2) for stainless steel front without door: IP 65 Rear ingress protection: IP 20 (EN 60529, Cat. 2) Electromagnetic compatibility (EMC) EN 61326, NAMUR recommendation NE21: - ESD (Electrostatic discharge): EN 61000-4-2, level 3 (6/8 kV) - Electromagnetic interference fields: EN 61000-4-3: Level 3 (10 V/m); additional deviation at 180 MHz: 0.7 °C (1.
10 Technical data Memo-Graph S Panel mounting dimensions for version with IP 65 stainless steel front and rear panel or terminal cover, rear-mounted ATA flash: Figure 10: Panel mounting dimensions with stainless steel front View of terminal (1) or rear panel cover (2): Note: the ATA flash drive is only mounted on the rear side for version with IP65 stainless steel front. Figure 11: Terminal or rear panel cover Installation depth approx. 211 mm / 8.31“ incl.
Memo-Graph S Field housing 10 Technical data IP65 field housing dimensions: 254 mm (10 “) 320 mm (12.6“) 320 mm (12.6 “) Figure 13: Field housing dimensions Weight VDM with front in stainless steel or metal die-cast: approx. 3.5 kg / 7.72 lb VDM built-in in desk top housing: approx. 6.4 kg / 14.
10.9 Data storage Selectable memory cycle per group (standard or event storage) 1s/2s/3s/5s/10s/15s/30s/1min/2min/3min/6min >= 4 year buffering for program/measured value memory (internal memory component: 2048 k SRAM) using integrated lithium battery (ambient temperature 15 to 25°C / 59 to 77°F); Cyclic copy of measured data for archiving to ATA flash memory card (max. 128 MB), selectable as stack or ring memory; resolution according to the selected memory cycle.
Memo-Graph S 10 Technical data 10.10 Certificates and approvals CE mark The measuring system meets the legal requirements of the EC directives. The manufacturer confirms that the device has been successfully tested by applying the CE mark. Electronic recording / electronic signature FDA 21 CFR 11: Meets the requirements of the “Food and Drug Administration” for electronic recording / electronic signature 10.11 Accessories For Accessories, see Chapter 8 10.
Index Memo-Graph S Index Symbols "Extras" key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . "Group" key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . "Login" key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . "Product" key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . "Set-up" key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . "Texts" key . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Memo-Graph S Last power on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . LCD running time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . LED operating mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . Lighting out after ... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Limit value x . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Limit value, analogue . . . . . . . . . . . . . . . . . . . . . . . . . . . . Limit values . . . . . .
Index Memo-Graph S Working temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . 88 Z Zero reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Zoom display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Zoom end . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Zoom start . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
www.endress.com/worldwide BA 138R/09/en/08.04 51006049 FM+SGML 6.