Polytel® Data Transmission System USER MANUAL Version: R Polymap Wireless 310 S. Williams Blvd. Ste.
Polytel® Data Transmission System Table of Contents 1 INTRODUCTION......................................................................................................................................... 4 1.1 SYSTEM OVERVIEW ................................................................................................................................. 4 1.2 INTENDED USE AND USER ........................................................................................................................
Polytel® Data Transmission System APPENDIX I: IEC 60601 SECTION 6 ........................................................................................................ 21 8.1 PWA-08-01 ACCESS POINT ................................................................................................................... 21 8.2 PWR-08-0X REMOTE UNITS ...................................................................................................................
Polytel® Data Transmission System 1 Introduction 1.1 System Overview The Polytel® Data Transmission system is designed to allow patients with chronic conditions to be monitored remotely. It consists of a series of wireless medical devices, an Access Point designed to forward data to the data center, and of a monitoring service. The PWA-08-01 is an Access Point, or “hub” that collects information from medical devices in the home and forwards them over a regular telephone line.
Polytel® Data Transmission System Polytel PWR-08-01-xxxxx Remote Board Polytel PWR-08-01-xxxxx Remote Board Modem Polytel PWR-08-02-xxxxx Remote Board Polytel PWA-08-01 Access Point Telephone Line Polytel PWR-08-01-xxxxx Remote Board AC Adapter PWA-08-01-1 or PWA-08-01-2 Polytel PWR-08-03-xxxxx Remote Board Polytel PWA-08-03-xxxxx Remote Board Figure 1: Polytel® Data Transmission System Block Diagram: The core of the System is the Access Point connected to the Mains through an AC Adapter and connec
Polytel® Data Transmission System 2.3 PWR-08-01 Remote This accessory has an onboard antenna (see Figure 2), and is incorporated into hosts by device manufacturers. 2.4 PWR-08-02 Remote This accessory is incorporated into hosts by device manufacturers; it incorporates an external patch antenna for use in hosts that have metal shielding. 2.5 PWR-08-03 This accessory consists of a circuit board in a plastic housing intended to hold two batteries.
Polytel® Data Transmission System 3 Identification of major parts PWA-08-01-1 AC Adapter Polytel® PWA-08-01 Access Point Polytel® PWR-08-01 module Polytel® PWR-08-02 module with antenna Figure 2: Polytel® System components. Polytel® PWR-08-03 accessory The Access Point is powered by an AC Adaptor connected to the Mains Outlet, and is also connected to the household telephone line.
Polytel® Data Transmission System Figure 3: Please note the small indicator light located near the bottom right corner of the back side of the device. This allows the user to confirm whether the device is operating correctly Figure 4: Indicator lights on PWA-08-01 Access Point. Leftmost indicates power, then Radio, Modem, and Error. Figure 5: back side of PWA-08-01. This shows the power connector on the left, and the two (interchangeable) telephone connectors to the right.
Polytel® Data Transmission System 4 Operating Instructions 4.1 Installation and setup The effective radio range of the Access Point and the matching medical devices is around 330 feet (100 meters). Please choose a location in your house that is centrally located, or close to where the measuring devices will be used. CAUTION – To reduce the risk of explosion or fire do not use this equipment in presence of flammable Anaesthetic mixture with Air or with Oxygen or Nitrous Oxide.
Polytel® Data Transmission System A good rule of thumb is that the access point should be located either in the same or an adjacent room to that where the measuring device is located. You can confirm that the devices are working together correctly by looking for the blinking light patterns on the access point to change within a minute after a measurement. 4.3 A note on telephone lines The PWA-08-01 Access Point is designed to operate over “regular” or traditional telephone lines.
Polytel® Data Transmission System correctly connected. In normal operation, the leftmost, or POWER light, will remain on, and the second light (RADIO), will blink roughly once per second. During the data transfer, the RADIO light will blink in other patterns. The PHONE light is used to indicate usage of the telephone, and the ERROR light is used to indicate a data transfer error.
Polytel® Data Transmission System 4.9 Troubleshooting 4.9.1 PWA-08-01 Access Point The PWA-08-01 Access Point incorporates a red “ERROR” LED. This light indicates that the device has received data from a measuring device but has not yet been able to transmit it. This failure may be a normal occurrence (the phone number the unit dialed was busy and it’s waiting to try again) or it may indicate a failure that requires user interaction.
Polytel® Data Transmission System LED after line measurement answer (this sounds like a fax machine), the problem may be due to your telephone line. If you have VoIP service, ensure that “fax and modem support” is selected. Solid red Server-side If none of the above LED after configuration helps resolve your issue, measurement issues please contact your service provider. A number of different service-side errors can lead to the data being retained on the Access Point. 4.9.
Polytel® Data Transmission System WARNING: “Modifications made to the product, unless expressly approved by Polymap Wireless, LLC could void the user’s license for and the warranty of the device.” CAUTION: Risk of explosion if battery is replaced by an incorrect type. Dispose of used batteries according to the instructions. 5.2 Electrical Safety Only authorized maintenance staff may remove the covers from the PWR-08-03 Remote Device. Always refer servicing to Polymap Wireless-authorized personnel.
Polytel® Data Transmission System • • • • IEC 60601-1-4: Medical Electrical Equipment – Collateral standard: Programmable medical electrical systems UL 60601: Medical Electrical Equipment, Part 1 – General Requirements for Safety CSA: Medical Electrical Equipment, Part 1: C22.2601 No. 601.
Polytel® Data Transmission System 5.4.2 Nameplate and Symbols Model: PWA-08-01-1 RoHS 2002/95/EC Model: PWA-08-01-2 RoHS 2002/95/EC Input: 120V~, 60Hz, 14VA Output: 6V ---, 1.0A Input: 230V~, 50Hz, 14VA Output: 6V ---, 1.
Polytel® Data Transmission System Model: PWR-08-03 IEC/EN/UL/CAN/CSA 60601-1 60601-1-2 FCC ID: QYPPWR0801 IC: 4552A-PWR0801 Use two CR2 Lithium batteries Assembled in USA Mfg Date: S/N: Label for PWR-08-03 accessory Model: PWR-08-01 IEC/EN/UL/CAN/CSA 60601-1 60601-1-2 FCC ID: QYPPWR0801 IC: 4552A-PWR0801 Assembled in USA Mfg Date: S/N: Label for PWR-08-01 wireless module Model: PWR-08-02 IEC/EN/UL/CAN/CSA 60601-1 60601-1-2 FCC ID: QYPPWR0801 IC: 4552A-PWR0801 Assembled in USA Mfg Date: S/N: Label f
Polytel® Data Transmission System undesired operation.
Polytel® Data Transmission System 7 Technical Specifications 7.1 Electrical Input Requirements 7.1.1 PWA-08-01 Access Point The PWA-08-01 is powered by an AC adapter. Two different adapters are available: • 120V~ 60Hz, PWA-08-01-1 • 230V~ 50Hz. PWA-08-01-2 Both configurations have the same output rating: 6VDC, 1000ma (Maximum power, 6VA) WARNING: There is an additional Power source inside the Access Point: 3V Lithium Battery. 7.1.
Polytel® Data Transmission System electrical medical devices. To ensure electromagnetic compatibility (EMC) the device must be installed and operated according to the information provided in this manual. NOTE: the entire Polytel® Data Transmission System wit all its components has been tested to comply with IEC 60601-1-2:2001 requirements for EMC with other devices. NOTE: This equipment has been tested and found to comply with the limits for a Class B digital device , pursuant to Part 15 of the FCC Rules.
Polytel® Data Transmission System Appendix I: IEC 60601 Section 6 8.
Polytel® Data Transmission System 8-1500-R - 22 -
Polytel® Data Transmission System 8-1500-R - 23 -
Polytel® Data Transmission System 8.
Polytel® Data Transmission System 8-1500-R - 25 -
Polytel® Data Transmission System 8-1500-R - 26 -
Polytel® Data Transmission System Index Location of devices, 9 Metal enclosure, 9 Metal surface, 9 Nameplate, 16 Operation in France, 11 Patient Safety, 13 PBX lines, 10 Power Brick, 19 Regulatory Compliance, 14 Retries, 11 Safety, Electrical, 14 Safety, Patient, 13 Setup, 9 Symbols, 16 Telephone Cord, 9 Telephone lines, 10 Temperature Operating, 19 Trademarks, 20 Troubleshooting, 12, 13 Voice over IP lines, 10 VoIP, 10 Wall Adapter, 19 Warnings, 13 Access Point, 4 Availability, 20 Batteries, 19 Inserting