LCR20 HART Level Radar Technical Information ISO9001
With regard to the supply of products, the current issue of the following document is applicable: The General Terms of Delivery for Products and Services of the Electrical Industry, published by the Central Association of the Electrical Industry (Zentralverband Elektrotechnik und Elektroindustrie (ZVEI) e.V.) in its most recent version as well as the supplementary clause: "Expanded reservation of proprietorship" Worldwide Pepperl+Fuchs Group Lilienthalstr.
Level Radar LCR20 Contents 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 1.1 Content of this Document . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 1.3 Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 1.2 2 Terms and Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 2.1 2.
Level Radar LCR20 Contents 8 Electrical Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 8.1 Cable Assignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 8.3 Power Consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 8.2 8.4 8.5 8.6 8.7 8.8 9 Supply Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Level Radar LCR20 Contents 13 Mechanical Construction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 13.1 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 13.3 Materials . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 13.2 13.4 14 Connecting Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Level Radar LCR20 2021-06 Contents 6
Level Radar LCR20 Introduction 1 Introduction 1.1 Content of this Document This document contains information that you need in order to use your product throughout the applicable stages of the product life cycle. These can include the following: • • • • • • • • Product identification Delivery, transport, and storage Mounting and installation Commissioning and operation Maintenance and repair Troubleshooting Dismounting Disposal Note This document does not substitute the instruction manual.
Level Radar LCR20 Introduction 1.2 Safety Information Target Group, Personnel Responsibility for planning, assembly, commissioning, operation, maintenance, and dismounting lies with the plant operator. Only appropriately trained and qualified personnel may carry out mounting, installation, commissioning, operation, maintenance, and dismounting of the product. The personnel must have read and understood the instruction manual and the further documentation.
Level Radar LCR20 Introduction Symbols for Certain Types of Information and Graphics Permitted Procedures, processes or actions that are permitted. Forbidden Procedures, processes or actions that are forbidden. 1, 2, 3 ... Views 2021-06 A, B, C, ...
Level Radar LCR20 Terms and Abbreviations 2 Terms and Abbreviations 2.1 Document Types BA Document type Manual KA Document type Brief Instructions TI Document type Technical Information SI Document type Instruction manual 2.2 Abbreviations PN Nominal pressure MWP MWP (Maximum working pressure/max. process pressure) The MWP can also be found on the nameplate.
Level Radar LCR20 Product Life Cycle 3 Product Life Cycle 3.
Level Radar LCR20 Measuring principle 4 Measuring principle The device is a "downward-looking" measuring system, which functions according to the timeofflight (ToF) method. It measures the distance from the reference point R to the product surface. Radar pulses are emitted by an antenna, reflected off the product surface and received again by the radar system. 100% R D E D F L 0% Figure 4.
Level Radar LCR20 Input 5 Input 5.1 Measured Variable The measured variable is the distance between the reference point and the product surface. The level is calculated based on E, the empty distance entered. 5.2 Measuring Range Maximum measuring range • • Device with 40 mm (1.5 inch) antenna: 15 m (49 foot) Device with 80 mm (3 inch) antenna: 20 m (66 foot) Installation requirements • • • • • • Recommended tank height > 1.5 m (5 foot) for media with low εr value Open channel minimum width 0.
Level Radar LCR20 Output 6 Output 6.1 Output Signal Output signal 4 to 20 mA An 4 to 20 mA interface is used for measured value output and to power to the device. 6.2 Digital Output HART® • • Signal encoding; FSK ±0.5 mA over current signal Data transmission rate; 1200 Bit/s Bluetooth® Wireless Technology (available as an optional extra) The device has a Bluetooth® wireless technology interface and can be operated and configured via this interface using the P+F Level app. • • • 6.
Level Radar LCR20 Protocol-Specific Data, HART 7 Protocol-Specific Data, HART Manufacturer ID 0x005D Device type ID 0x5DD3 HART specification 7.0 Device description files (DTM) Information and files under: • • www.pepperl-fuchs.com www.hartcomm.org HART load Min. 250 HART device variables Assignment of HART device variables is fixed and cannot be changed. • • • • Measured values for PV (primary variable) Level linearized Advanced diag.
Level Radar LCR20 Electrical Connection 8 Electrical Connection 8.1 Cable Assignment 1 Figure 8.1 1 2 8.2 + - 2 Cable assignment Plus, brown wire Minus, blue wire Supply Voltage 10.5 to 30 V DC An external power supply is necessary. R [Ω] 500 0 10 10.5 30 U0 [V] Maximum load R, depending on supply voltage Uo of power supply unit 2021-06 Figure 8.2 20 21.
Level Radar LCR20 Electrical Connection Battery operation The sensor's Bluetooth® wireless technology communication can be disabled to increase the operating life of the battery. Potential equalization No special measures for potential equalization are required. Note Various power supply units can be ordered as an accessory from Pepperl+Fuchs. 8.3 Power Consumption Maximum input power: 675 mW Current consumption • • Maximum input current: <25 mA Maximum start-up current: 3.6 mA 8.
Level Radar LCR20 Electrical Connection 8.6 Connecting the Device 4 to 20 mA HART block diagram Connection of the device with HART communication, power source and 4 to 20 mA display 1 2 3 Y + + - - I mA Figure 8.3 Block diagram of HART connection Device with HART communication HART resistor Power supply 1 2 3 Note The HART communication resistor of 250 in the signal line is always necessary in the case of a low-impedance power supply. The voltage drop to be taken into account is: Max.
Level Radar LCR20 Performance Characteristics 9 Performance Characteristics 9.1 Reference Operating Conditions 9.2 Maximum Measured Error • • • • • Temperature = +24 °C (+75 °F) ±5 °C (±9 °F) Pressure = 960 mbar abs. (14 psia) ±100 mbar (±1.45 psi) Humidity = 60 % ±15 % Reflector: metal plate with diameter ≥ 1 m (40 inch) No major interference reflections inside the signal beam Typical data under reference operating conditions: DIN EN 61298-2, percentage values in relation to the span.
Level Radar LCR20 Performance Characteristics 9.3 Measured Value Resolution Dead band as per EN61298-2: • Digital: 1 mm (0.04 inch) • Analog: 4 μA 9.4 Response Time The response time can be configured.
Level Radar LCR20 Installation 10 Installation 10.1 Installation Conditions Installation types A B E Figure 10.1 A B C D E F C D F Wall, ceiling or nozzle installation Wall or ceiling mount, adjustable Mounted at front thread Mounted at rear thread Ceiling installation with counter nut Horizontal installation with reflector plate Shaft wall mounting Caution! The sensor cables are not designed as supporting cables. Do not use them for suspension purposes.
Level Radar LCR20 Installation Position for installation on a vessel BD A D Figure 10.2 • • • • • • Installation position on a vessel If possible install the sensor so that its lower edge projects into the vessel. Recommended distance A wall – nozzle outer edge: ~ 1/6 of the vessel diameter D. Under no circumstances should the device be mounted closer than 15 cm (5.91 inch) to the vessel wall. Do not install the sensor in the middle of the vessel. Avoid measurements through the filling curtain.
Level Radar LCR20 Installation Nozzle installation The antenna should be located out of the nozzle for optimum measurement. The interior of the nozzle must be smooth and may not contain any edges or welded joints. The edge of the nozzle should be rounded if possible. A L B L D Figure 10.3 L D L D D Nozzle installation 80 mm (3 inch) antenna 40 mm (1.5 inch) antenna Nozzle diameter Nozzle length A B D L The maximum length of the nozzle L depends on the nozzle diameter D.
Level Radar LCR20 Installation Device alignment for installation on a vessel Align the antenna vertically to the product surface. Align the eyelet with lug towards the vessel wall as well as possible. 90 90 ° ° • • 90 90 ° 90 ° Device alignment for installation on a vessel 2021-06 Figure 10.
Level Radar LCR20 Installation Beam angle D _ W = 2 × D × tan α 2 α W Figure 10.5 Relationship between beam angle , distance D and beamwidth diameter W The beam angle is defined as the angle , at which the power density of the radar waves reaches half the value of the maximum power density (3 dB width). Microwaves are also emitted outside the signal beam and can be reflected off interfering installations. Beam diameter W as a function of beam angle and distance D. 40 mm (1.
Level Radar LCR20 Installation Measurement in plastic vessels 2 1 Figure 10.6 1 2 3 3 Measurement in a plastic vessel with a metallic, interfering installation outside of the vessel Pipe, tubing Ladder Grate, railing If the outer wall of the vessel is made of a non-conductive material (e. g. GFR), microwaves can also be reflected by interfering installations outside of the vessel.
Level Radar LCR20 Installation Free-field measurement with flooding protection tube The flooding protection tube ensures the sensor measures the maximum level even if it is completely flooded. In free-field installations and/or in applications where there is a risk of flooding, the flooding protection tube must be used. The flooding protection tube can be ordered as an accessory or together with the device via the ordering option Accessory enclosed. 2 2 3 1 4 Figure 10.
Level Radar LCR20 Installation Installation with mounting bracket, adjustable The mounting bracket is available as an accessory. Figure 10.9 • • Installation with mounting bracket, adjustable Wall or ceiling installation is possible. Using the mounting bracket, position the antenna so that it is perpendicular to the product surface. Caution! There is no conductive connection between the mounting bracket and transmitter housing. Electrostatic charging possible.
Level Radar LCR20 Installation Installation of horizontal mounting bracket for sewer shafts The horizontal mounting bracket is available as an accessory. Figure 10.11 Installation of horizontal mounting bracket for sewer shafts Mounting in a shaft The pivoted mounting bracket is available as an accessory. A Figure 10.12 Mounting in a shaft, pivotable and adjustable Arm with wall bracket Pivotable and adjustable arm (e. g.
Level Radar LCR20 Environment 11 Environment 11.1 Ambient Temperature Range Measuring device: -40 to +80 °C (-40 to +176 °F) Note It may not be possible to use the bluetooth connection at ambient temperatures > 60 °C (140 °F). Operating the Device Outdoors in Strong Sunlight 1. 2. 3. Mount the device in the shade. Avoid direct sunlight, particularly in warm climatic regions. Use a weather protection cover. 11.2 Storage Temperature 11.3 Climate Class 11.
Level Radar LCR20 Process 12 Process 12.1 Process Temperature, Process Pressure p [ bar (psi) ] 3 (43.5) 0 -1(-14.5) Tp -40 Figure 12.1 0 (+32) +80 (+176) [ °C (°F) ] Permitted range for process temperature and process pressure Process temperature range -40 to +80 °C (-40 to +176 °F) Process pressure range, threaded process connection • • prel = -1 to 3 bar (-14.5 to 43.5 psi) pabs < 4 bar (58 psi) Process pressure range, UNI flange process connection • • prel = -1 to 1 bar (-14.5 to 14.
Level Radar LCR20 Mechanical Construction 13 Mechanical Construction 13.1 Dimensions 40 mm (1.5 inch) antenna with G1-1/2 or MNPT1-1/2 thread ø75 (2.95) 26 (1.02) B 20 (0.79) 28 (1.1) 112 (4.4) 140 (5.5) A G 1" NPT 1" G 1 1/2" NPT 1 1/2" Figure 13.1 A B Dimensions of G1-1/2 or MNPT1-1/2 process connection thread, engineering unit: mm (inch) Cable gland FNPT1/2 conduit 40 mm (1.5 inch) antenna with G2 or MNPT2 thread ø75 (2.95) B 26 (1.02) 20 (0.79) 28 (1.1) 112 (4.4) 140 (5.
Level Radar LCR20 Mechanical Construction 40 mm (1.5 inch) antenna with flooding protection tube 40 mm (1.5 inch) 264 (10.4) ø75 (2.95) ø89 (3.5) Figure 13.3 Dimensions of 40 mm (1.5 inch) antenna mounted with flooding protection tube, engineering unit: mm (inch) The flooding protection tube can be ordered together with the device via the ordering option Accessory enclosed. 80 mm (3 inch) antenna ø75 (2.95) 20 (0.79) 26 (1.02) B G 1" NPT 1" 9 (0.35) 206 (8.11) 28 (1.1) A ø115 (4.
Level Radar LCR20 Mechanical Construction 80 mm (3 inch) antenna with flooding protection tube 80 mm (3 inch) ø75 (2.95) 20 (0.79) 26 (1.02) B G 1" NPT 1" 317 (12.5) 28 (1.1) A ø98.2 (3.87) ø138.1 (5.44) Figure 13.5 A B Dimensions of 80 mm (3 inch) antenna mounted with flooding protection tube, engineering unit: mm (inch) Cable gland FNPT1/2 conduit 2021-06 The flooding protection tube can be ordered together with the device via the ordering option Accessory enclosed.
Level Radar LCR20 Mechanical Construction 178 (7) 80 mm (3 inch) antenna with slip-on flange 3 inch/DN80 ø75 (2.95) Figure 13.6 ø115.6 (4.55) ø156.2 (6.15) ø200 (7.87) 9 (0.36) 18.5 (0.73) 8x ø21 (0.83) Dimensions of 80 mm (3 inch) antenna with slip-on flange 3 inch/DN80, engineering unit: mm (inch) The slip-on flange can be ordered via the ordering option Process connection front side. 80 mm (3 inch) antenna with slip-on flange 4 inch/DN100 30.5 (1.2) 20 (0.79) Figure 13.7 188 (7.
Level Radar LCR20 Mechanical Construction 80 mm (3 inch) antenna with slip-on flange 6 inch/DN150 30.5 (1.2) 20 (0.79) Figure 13.8 188 (7.4) 8x 23 (0.91) ø143 (5.63) ø98 (3.86) ø240 (9.45) ø241.3 (9.5) ø285 (11.2) Dimensions of 80 mm (3 inch) antenna with slip-on flange 6 inch/DN150, engineering unit: mm (inch) The slip-on flange can be ordered via the ordering option Process connection front side. G 1" 8 (0.31) 1 2 (0.08) 2 (0.
Level Radar LCR20 Mechanical Construction 13.2 Weight Weight (incl. 5 m (16.4 foot) cable) • • 13.3 Device with 40 mm (1.5 inch) antenna: approx. 2.5 kg (5.5 lb) Device with 80 mm (3 inch) antenna: approx. 2.8 kg (6.2 lb) Materials A 4 3 2 1 B 5 6 7 8 4 3 2 5 6 7 1 8 9 9 Figure 13.10 A B 1 2 3 4 5 6 7 8 9 13.4 Device with 40 mm (1.
Level Radar LCR20 Operability 14 Operability 14.1 Operating Concept • • • 14.2 4 to 20 mA, HART Menu guidance with brief explanations of the individual parameter functions in the operating tool Optional: P+F Level (app) via Bluetooth® wireless technology Operation via Bluetooth® Wireless Technology Figure 14.1 1 2 3 14.
Level Radar LCR20 Certificates and Approvals 15 Certificates and Approvals 15.1 CE Mark The measuring system meets the legal requirements of the applicable EU directives. These are listed in the corresponding EU Declaration of Conformity along with the standards applied. Pepperl+Fuchs confirms successful testing of the device by affixing to it the CE mark. 15.2 RoHS 15.
Level Radar LCR20 Certificates and Approvals 15.5 Explosion-Protected Smartphones and Tablets Only mobile end devices with Ex approval may be used in hazardous areas. Pepperl+Fuchs offers mobile end devices for use in hazardous areas. Note For further information, refer to the internet at www.pepperl-fuchs.com. 15.
Level Radar LCR20 Certificates and Approvals Astronomy Stations Country Name of the station Latitude Longitude Finland Metsähovi 60°13'04" North 24°23'37" East Germany France Great Britain Effelsberg Tuorla Plateau de Bure Floirac Cambridge Damhall Knockin Poland Russia Pickmere Sweden Switzerland Spain Krakow Fort Skala Dmitrov Kalyazin Penc Table 15.
Level Radar LCR20 Certificates and Approvals 15.8 FCC/Industry Canada This device complies with Part 15 of the FCC Rules [and with Industry Canada licence-exempt RSS standard(s)]. Operation is subject to the following two conditions: 1. This device may not cause harmful interference. 2. This device must accept any interference received, including interference that may cause undesired operation.
Level Radar LCR20 Certificates and Approvals FCC/Industry Canada IDs Tank level-probing radar • HVIN: LCR20X • FCC ID: IRELCR2XKT • Industry Canada ID: 7037A-2KT Level-probing radar: • 15.9 HVIN: LCR20+R7X, LCR20+R8X • FCC ID: IRELCR2XKL • Industry Canada ID: 7037A-2KL Japanese Radio Law Japanese Radio Law and Japanese Telecommunications Business Law Compliance This device is granted pursuant to the Japanese Radio Law (電波法) and the Japanese Telecommunications Business Law (電気通信事業法).
Level Radar LCR20 Certificates and Approvals 15.10 Other Standards and Guidelines • • • • • • • • • • • • • IEC/EN 55011 "EMC Emission, RF Emission for Class B". Industrial, scientific and medical equipment – Electromagnetic disturbance characteristics – Limits and methods of measurement IEC/EN 61000-4-2 EMC Immunity, ESD (Performance Criteria A).
Level Radar LCR20 Ordering Information 16 Device LCR20 Ordering Information This overview does not mark options which are mutually exclusive. Option with ** = multiple options can be selected Level radar Antenna, maximum measuring range 1 40 mm/1-1/2 inch, -40 ... 80 °C (-40 ... 176 °F), measuring range up to 15 m in liquids X Special version 2 80 mm/3 inch, -40 ... 80 °C (-40 ...
Level Radar LCR20 Ordering Information Approval NA Version for non-hazardous area EX ATEX II 1/2G Ex ia IIC T4...T1 Ga/Gb E1 CG C1 IA IB XX ATEX II 1G Ex ia IIC T4...T1 Ga CSA C/US General Purpose CSA C/US IS Cl. I Div. 1 Gr. A-D, AEx/Ex ia IIC T4 IEC Ex ia IIC T4...T1 Ga IEC Ex ia IIC T4...T1 Ga/Gb Special version Additional Options Service ** YY Special version Accessory enclosed ** RA UNI flange 2 inch/DN50/50, PP, front side, max.
Level Radar LCR20 Accessories 17 Accessories 17.1 Enclosed Accessories 17.1.1 Device-Specific Accessories These accessories can be ordered together with the device via the ordering option Accessory enclosed. Weather protection cover 90 (3.54) 90 (3.54) The weather protection cover can be ordered together with the device via the ordering option Accessory enclosed. Ø 98 (3.86) Figure 17.1 Material: PVDF Ø 98 (3.
Level Radar LCR20 Accessories Flooding protection tube 40 mm (1.5 inch) Suitable for use with devices with a 40 mm (1.5 inch) antenna and G1-1/2 process connection on front. The flooding protection tube can be ordered together with the device via the ordering option Accessory enclosed. 152 (5.98) G 1-1/2" ø 89 (3.5) Figure 17.2 Dimensions of 40 mm (1.
Level Radar LCR20 Accessories Flooding protection tube 80 mm (3 inch) Suitable for use with devices with a 80 mm (3 inch) antenna and Mounting customer side w/o flange process connection. The flooding protection tube can be ordered together with the device via the ordering option Accessory enclosed. 136.5 (5.37) ø 138 (5.43) ø 97.3 (3.83) Figure 17.
Level Radar LCR20 Accessories UNI flange 2 inch/DN50/50, PP The UNI flange can be ordered together with the device via the ordering option Accessory enclosed. 4× ° 90 °( =3 60 °) 20 (0.79) 19 (0.75) 90 A 120 (4.72) 125 (4.92) 165 (6.5) Figure 17.
Level Radar LCR20 Accessories UNI flange 3 inch/DN80/80, PP The UNI flange can be ordered together with the device via the ordering option Accessory enclosed. 8×4 5° ( =36 0°) 20 (0.79) 19 (0.75) 45° A 150 (5.91) 160 (6.3) 200 (7.87) Figure 17.
Level Radar LCR20 Accessories UNI flange 4 inch/DN100/100, PP The UNI flange can be ordered together with the device via the ordering option Accessory enclosed. 8×4 5° ( =36 0°) 20 (0.79) 19 (0.75) 45° A 175 (6.89) 190.5 (7.5) 228.6 (9.0) Figure 17.
Level Radar LCR20 Accessories 17.2 Further Accessories 17.2.1 Device-Specific Accessories These accessories can be ordered separately. Securing nut G1-1/2 10 (0.39) Suitable for devices with G1-1/2 and MNPT1-1/2 process connection. G1 1/2" DIN ISO 228 60 mm Figure 17.7 Material: PC Dimensions of securing nut, engineering unit: mm (inch) Order number: 52014146 Securing nut G 2 G2" 10 (0.39) Suitable for devices with G2 and MNPT2 process connection on front. 70 mm Figure 17.
Level Radar LCR20 Accessories Mounting bracket, adjustable 92.5 (3.64) 12 (0.47) ø 9 (0.35) 9 (0.35) 35 (1.38) 117.5 (4.63) 70 (2.76) 12 (0.47) Figure 17.9 Consists of: • • • • 9 (0.35) Dimensions of mounting bracket, engineering unit: mm (inch) 1 × mounting bracket, 316L (1.4404) 1 × mounting bracket, 316L (1.4404) 3 × screws, A4 3 × securing disks, A4 Order number: 71325079 Angle bracket for wall mounting 400 (15.7) 3 (0.12) Figure 17.10 120 (4.72) ø16 (0.6) 30 (1.2) 250 (9.
Level Radar LCR20 Accessories Cantilever with pivot Installation type sensor process connection rear side A B 1 3 2 Figure 17.11 1 Installation type sensor process connection rear side Installation with cantilever and wall bracket Installation with cantilever and mounting frame Cantilever Wall bracket A B 1 2 Mounting frame 3 Cantilever (short) with pivot, sensor process connection on rear 200 (7.87) M8 250 (9.84) 35 (1.4) Figure 17.12 15 (0.6) 6.5 (0.3) 2 (0.08) 22 (0.9) 35 (1.
Level Radar LCR20 Accessories Cantilever (long) with pivot, sensor process connection on rear 300 (11.8) M8 750 (29.5) 35 (1.4) Figure 17.13 15 (0.6) 6.5 (0.3) 3 (0.12) 22 (0.9) 35 (1.4) 105 (4.1) 75 (2.9) 75 (2.9) 100 (3.9) 50 100 (3.9) (2.0) 35 (1.4) 25 (1.0) 20 (0.8) 3 (0.12) 20 (0.8) 1085 (42.7) Dimensions of cantilever (long) with pivot for sensor process connection on rear, engineering unit: mm (inch) Weight: 4.5 kg (9.92 lb) Material: 316L (1.
Level Radar LCR20 Accessories Cantilever (short) with pivot, G1-1/2 sensor process connection on front 200 (7.87) M8 250 (9.84) 35 (1.4) Figure 17.15 15 (0.6) 6.5 (0.3) 2 (0.08) 22 (0.9) 50 (2.0) 105 (4.1) 75 (2.9) 75 (2.9) 100 (3.9) 50 (2.0) 100 (3.9) 35 (1.4) 25 (1.0) 20 (0.8) 2 (0.08) 20 (0.8) 585 (23) Dimensions of cantilever (short) with pivot for G1-1/2 sensor process connection on front, engineering unit: mm (inch) Weight: 1.9 kg (4.19 lb) Material: 316L (1.
Level Radar LCR20 Accessories Cantilever (long) with pivot, G1-1/2 sensor process connection on front 300 (11.8) M8 750 (29.5) 35 (1.4) Figure 17.16 15 (0.6) 6.5 (0.3) 3 (0.12) 22 (0.9) 50 (2.0) 105 (4.1) 75 (2.9) 75 (2.9) 100 (3.9) 50 100 (3.9) (2.0) 35 (1.4) 25 (1.0) 20 (0.8) 3 (0.12) 20 (0.8) 1085 (42.7) Dimensions of cantilever (long) with pivot for G1-1/2 sensor process connection on front, engineering unit: mm (inch) Weight: 4.4 kg (9.7 lb) Material: 316L (1.
Level Radar LCR20 Accessories Cantilever (short) with pivot, G2 sensor process connection on front 200 (7.87) M8 250 (9.84) 35 (1.4) Figure 17.17 15 (0.6) 6.5 (0.3) 2 (0.08) 22 (0.9) 62 (2.4) 105 (4.1) 75 (2.9) 75 (2.9) 100 (3.9) 50 (2.0) 100 (3.9) 35 (1.4) 25 (1.0) 20 (0.8) 2 (0.08) 20 (0.8) 585 (23) Dimensions of cantilever (short) with pivot for G2 sensor process connection on front, engineering unit: mm (inch) Weight: 1.9 kg (4.19 lb) Material: 316L (1.
Level Radar LCR20 Accessories Cantilever (long) with pivot, G2 sensor process connection on front 300 (11.8) M8 750 (29.5) 35 (1.4) Figure 17.18 15 (0.6) 6.5 (0.3) 3 (0.12) 22 (0.9) 62 (2.4) 105 (4.1) 75 (2.9) 75 (2.9) 100 (3.9) 50 100 (3.9) (2.0) 35 (1.4) 25 (1.0) 20 (0.8) 3 (0.12) 20 (0.8) 1085 (42.7) Dimensions of cantilever (long) with pivot for G1-1/2 sensor process connection on front, engineering unit: mm (inch) Weight: 4.4 kg (9.7 lb) Material: 316L (1.
Level Radar LCR20 Accessories Mounting frame (short) for cantilever with pivot 3.2 (0.13) 60 (2.36) ø33.7 (1.3) 4 (0.16) Figure 17.19 130 (5.12) 100 76 (3.94) (2.99) 200 (7.87) 700 (27.6) 45 (1.77) 25 (0.98) 75 (2.95) 100 (3.94) 55 (2.17) 109 (4.29) 6.5 (0.3) 13 (0.5) 20 (0.8) 100 (3.94) 150 (5.91) Dimensions of mounting frame (short), engineering unit: mm (inch) Weight: 3.2 kg (7.06 lb) Material: 316L (1.
Level Radar LCR20 Accessories Mounting frame (long) for cantilever with pivot 3.2 (0.13) 60 (2.36) ø33.7 (1.3) Figure 17.20 25 (0.98) 75 (2.95) 100 (3.94) 55 (2.17) 109 (4.29) 4 (0.16) 130 (5.12) 100 76 (3.94) (2.99) 200 (7.87) 1400 (55.1) 6.5 (0.3) 45 (1.77) 13 (0.5) 20 (0.8) 100 (3.94) 150 (5.91) Dimensions of mounting frame (long), engineering unit: mm (inch) Weight: 4.9 kg (10.08 lb) Material: 316L (1.4404) Order number: 71452326 Wall bracket for cantilever with pivot 3.2 (0.1) ø33.
Level Radar LCR20 Accessories Ceiling mounting bracket 75 1.5 185 105 ø5.6 10 100 95 15 50 110± 2 12 10 10 5 ø35 15 155 Figure 17.22 Dimensions of ceiling mounting bracket, engineering unit: mm (inch) Material: 316L (1.
Level Radar LCR20 Accessories Pivoted mounting bracket for sewer channel The pivotable mounting bracket is used to install the device in a manhole over a sewer channel. 100 (3.94) 104 (4.09) 6.8 (0.27) 6.8 (0.27) 34 (1.34) 177 (6.97) 358 (14.1) Figure 17.23 Dimensions of the pivotable mounting bracket, engineering unit: mm (inch) Material: 316L (1.
Level Radar LCR20 Accessories 17.2.2 Communication-Specific Accessories HART Modem For intrinsically safe HART communication with PACTware via the USB interface. HART Loop Converter KFD2-HLC-Ex1.D.** This isolated barrier is used for intrinsic safety applications. The device is a HART loop converter that provides power to transmitters or can be connected to existing HART loops in parallel. The devcie used to evaluate and convert dynamic HART process variables to analog current signals or limit values.
Level Radar LCR20 Accessories 17.2.4 System Components KFD2-STC5-Ex1 This isolated barrier is used for intrinsic safety applications. The device supplies 2-wire and 3-wire SMART transmitters, and can also be used with 2-wire SMART current sources. The device transfers the analog input signal to the safe area as an isolated current value. Digital signals may be superimposed on the input signal in the hazardous or non-hazardous area and are transferred bidirectionally.
Level Radar LCR20 Documentation 18 18.1 Documentation The following document types are available on the Pepperl+Fuchs website: www.pepperl-fuchs.com → search → model number → product detail page → documents. Technical Information (TI) Planning aid The document contains all the technical data on the device and provides an overview of the accessories and other products that can be ordered for the device. 18.
Level Radar LCR20 Registered Trademarks 19 Registered Trademarks HART® Registered trademark of the FieldComm Group, Austin, Texas, USA Apple® Apple, the Apple logo, iPhone, and iPod touch are trademarks of Apple Inc., registered in the U.S. and other countries. App Store is a service mark of Apple Inc. Android® Android, Google Play and the Google Play logo are trademarks of Google Inc. Bluetooth® 2021-06 The Bluetooth® word mark and logos are registered trademarks owned by the Bluetooth SIG, Inc.
Level Radar LCR20 2021-06 Notes 69
Pepperl+Fuchs Quality Download our latest policy here: www.pepperl-fuchs.com/quality TI01267O/98/EN/07.20 www.pepperl-fuchs.