SlowJec plus [DSD-PLA-0100] User Manual | OS-UM-C020 | 2021.03.12 / Ver. 1.0 IMPORTANT NOTES For correct use of the device, make sure to read the user manual thoroughly before use.
SlowJec plus [DSD-PLA-0100] User Manual | User Manual | 2021.03.12, Ver. 1.0 Dear SlowJec plus customers, Thank you very much for using our product, SlowJec plus. For correct use and maintenance of the product, make sure to read the user manual thoroughly before use and also store the Product Warranty document in a safe location. Table of Contents 01. Overview 02. Product Features 03. Part Names and Functions 04. How to Use 05. Finished Use and Product Separation 06.
Classification Level of risk, damage or severity DANGER Likely to cause fatal damages. WARNING Likely to cause physical injuries or device malfunctions. CAUTION Likely to cause slight physical injuries or device malfunctions. NOTICE Items that a user must comply with for safety reasons. MEDICAL - GENERAL MEDICAL EQUIPMENT AS TO ELECTRICAL SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH "ANSI/AAMI ES 60601-1 (2005) + AMD (2012)" "CAN/CSA-C22.2 No.
※DANGER Only use a dedicated Lithium polymer battery for the product. Other types of battery cannot be used. ※ DANGER Utilisez uniquement une batterie lithium-polymère dédiée au produit. Les autres types de batterie ne peuvent pas être utilisés. ※WARNING -When injecting anesthetic into a patient by using the product, make sure to properly adjust the amount of injection depending on whether a patient is a pregnant woman, breast feeding woman, woman at childbearing age, infant, kid or elder.
-On the charging terminal of the main body and the cradle, do not place any other objects (ex. Metal, coin, wire, plastic and so on). ※ ATTENTION - Lors de l'injection d'anesthésique à un patient en utilisant le produit, assurez-vous d'ajuster correctement la quantité d'injection selon que le patient est une femme enceinte, une femme qui allaite, une femme en âge de procréer, un nourrisson, un enfant ou une personne âgée. - Connectez la cartouche authentique incluse dans l'emballage uniquement.
when the power is on, it can cause malfunctions. -Charge the battery if the battery level is low before use. -If chemical, solvent, or detergent comes in contact with the main body, it can cause decolorization or deformation, so please wipe it off immediately. -The product can be used indoors only.
(8) Pay attention to the following requirements for the environment and conditions of use. A. Environment: Do not use it in a place with high humidity such as a bathroom. B. Environment of use C. Room temperature: 10℃ ~ 35℃ Relative humidity: 10% ~ 75% Air pressure: 700hPa ~ 1060hPa Transportation/Storage environment Recommended to transport/store the product within -10℃ ~ 50℃. Relative humidity: 10% ~ 80% Air pressure: 500hPa ~ 1060hPa 03. Part Names and Functions 1.
2. Descriptions of the exterior of the main body A. Cartridge Assembly -Area where the cartridge is assembled B. Inject button -Press the button to start injection or change the injection speed. -Long press (for more than 1 sec) the button in initial condition to remove bubbles. C. Stop button -Press the button to stop injection. -Long press the button in initial condition to switch between manual and auto modes. D. Decelerate button -Press the button to reduce the injection speed.
H 3. Descriptions of the exterior of the cradle A. Complete charging LED (Green lamp) -Lit in green if the charging is completed. B. Charging LED (Orange lamp) -Lit in orange if the charging is still in progress. C. AC adapter connector -5V/2A adapter connector D. Holder -Place the main body in here. D C A, B 4. LCD screen Display unit Mode Descriptions Power On Displayed when power is On. Power Off Displayed when power is Off. Manual mode selection Displayed when in manual mode.
Bubble removal Displayed when bubbles are being removed. Low speed (Level 1) Displays the speed for manual mode injection. Mid speed (Level 2) Displays the speed for manual mode injection. High speed (Level 3) Displays the speed for manual mode injection. Auto Displayed when for auto mode injection. Initialization Displayed when initialization is in progress. Complete injection Displayed when injection is complete. Operation error Displayed when an error occurs.
K. Since it can cause harms to the device or users, users or unqualified technicians are never allowed to disassemble or assemble the device. L. For electrical safety, do not place a heavy object on top of the charging power cable or place the device near the source of fire at high temperature. M. If the device is operating abnormally, then stop using it immediately and turn off the power. 3. Charging A. Insert the AC adapter cable into the back of cradle. B.
※NOTICE -If the LED is not lit while the battery is charging, then it means that there is a problem in the internal circuit, so ask the seller to repair it for you ※ AVIS - Si la LED ne s'allume pas pendant la charge de la batterie, cela signifie qu'il y a un problème dans le circuit interne, alors demandez au vendeur de le réparer pour vous 4. How to mount the cartridge A. Insert the cartridge correctly into the front groove of the main body. B.
-Refer to the user manual for more about how to use a single-use needle. ※ AVIS - Reportez-vous au manuel d'utilisation pour en savoir plus sur l'utilisation d'une aiguille à usage unique. 5. How to dismount the cartridge A. Press the button at the bottom of cartridge. B. When the ample pops up, separate it with hands. C. You can disassemble the cartridge in the reverse order of assembly. Button ※CAUTION -Make sure to turn off the power before dismounting the cartridge.
A. Press the “ “ part of the cover grip (Standard type/Ring type) and slide it to separate it. Push Sliding 8. Control (Operation) A. If you press the Power button, you can hear the voice guide and see “p” on the display. B. Short press (for less than 1 sec) the Power button once to display the battery level. C. Long press (for more than 1 sec) the Decelerate button once in Standby mode to hear the voice guide.
start at level 1 speed and the speed will go up gradually if you repeat the step. B. When injection is complete, initialization will take place automatically after 3 seconds. C. Short press the Decelerate button during injection for gradual deceleration. (Ex: Level 3 → Level 2 → Level 1) D. Short press the Stop button during injection to stop injection. When stopped, short press the Inject button to resume injection at level 1 speed. 10. Auto mode A.
06. How to Store and Manage the Product after Use 1. Storage A. Make sure to charge it in the cradle after use. B. When you don’t charge or use the cartridge, keep it in a separate location. C. Store it at the room temperature with no direct sunlight. D. If it is left dormant for a long time, then remove the cradle plug and store it in a cool and dry place with no direct sunlight. 2. Washing A.
the autoclave equipment. ※ ATTENTION - N'effectuez pas d'autoclavage sur le corps principal ou le berceau. - N'essuyez jamais le corps principal avec un solvant tel qu'un acide fort ou du benzène. - Si vous effectuez un autoclavage avec l'appareil contenant du liquide, cela peut décolorer la surface ou affecter les pièces internes. Assurez-vous que le liquide ne pénètre pas dans l'équipement de l'autoclave.
2. If the battery does not get charged A. If the battery temperature is high or low → If the battery temperature is lower than 0℃ or higher than 40℃, charging can fail. Right after the charging is complete, the temperature of battery might rise little bit, but if it is overheated then contact your seller. B. If the main body is not correctly mounted in the cradle → Insert the main body correctly into the cradle slot. C.
Guides and declaration by the manufacturer - Electromagnetic field immunity The product is guaranteed to operate in the following electromagnetic field environment. Immunity level IEC60601 test level Conformity level Electromagnetic field environment - guide ±(2,4,6) 8kV contact Electrostatic ±(2,4,6) 8kV contact The floor should be discharge(ESD) ±(2,4,8) 15kV air IEC61000-4-2 ±(2,4,8)15kV air made of concrete or tile.
electromagnetic field 150kHz to 80MHz IEC61000-4-6 3V/m Radiative RF 80MHz to 2.7GHz portable 3V/m RF wireless communication device, then make electromagnetic field maintain IEC61000-4-3 separation sure to the greater than the recommended separation calculated based on the transmission frequency applied at every part of the product including the cable. d = 1.2 √P d = 1.2 √P 80MHz to 800MHz d = 2.3 √P 800MHz to 2.
There can be some interference near a device that emits nonionizing electromagnetic radiation (IEC60417- 1:2002 5140). Remark 1 : Ut is the AC voltage before the test level is applied. Remark 2 : A higher frequency range is applied in the frequency bands of 80MHz and 800MHz. Remark 2.
10. Etc. 1. Manufacturer : OSSTEM IMPLANT CO,. LTD. Chair Business 2. Address : 192, Haebong-ro, Danwon-gu, Ansan-si, Gyeonggi-do, 15428, Korea 3. Applied part a. Needle b. Ample 4. Battery : 3.7V 500mAh, Li-polymer 5. Adapter a. Rated input voltage: 100 – 240V AC, 1A, 50/60Hz b. Rated output voltage: 5V DC 2A 6. Steam Sterilization a. Type : Pre-vacuum b. Temperature : 121℃ c. Time : 15 min d. Limit of number : 100 cycle 11.
FCC Compliance Statement This device complies with part 15 of the FCC rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. FCC Interference Statement This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules.