USER & SERVICE MANUAL DIGITAL FLAT PANEL X-RAY DETECTOR ▐ Edge Air 1417 / 1417WCC
Preface Doc. No. : OK-RND-USM005-100 / OK-RND-SVM005-100 Rev. No.: draft Date : 04.13.
Preface Preface Please note that this information is for proper use and safety of the equipment. The following symbols may indicate a hazardous situation in which, if not heeded, may result in serious injury or even death to the user or others, or damage to the equipment. Used to emphasize essential information. Be sure to read this information to avoid incorrect operation. Indicates warning and safety instructions. If not adhered to, it could result in death or serious injury to the user or others.
Contents Contents Preface ...................................................................................................................... iii Contents .................................................................................................................... iv PART I. User & Installation Manual 1. 2. 3. 4. 5. 6. Safety Information ......................................................................................... 8 1.1 Safety Standard .........................................
Contents 1. Overview ....................................................................................................106 2. FPD Manager Instruction (IP, SSID Set Up / Firmware, FPGA Update) .............107 2.1 Detector IP Address Set Up .................................................................107 2.2 SSID, PSK (Pre-Shared Key) Set Up .....................................................109 2.3 Firmware, FPGA Update ......................................................................
Contents 6 / 126 Edge Air 1417 / 1417WCC
PART I. User & Installation Manual PART I. User & Installation Manual 1. Safety Information........................................... 8 2. Product Introduction and Specification ........ 23 3. Installation and Calibration ........................... 37 4. Usage ..............................................................83 5. Maintenance ................................................... 95 6. Warranty ......................................................
PART I.User & Installation Manual 1. Safety Information 1.1 Safety Standard 1.1.1 Medical Device Classification Item Description Classification by protection type against Electric Shock Class I or Internally Powered Equipment Classification according to the degree of protection against ingress of water IPX66 Mode of operation Continuous electromagnetic Environment of Use This equipment is not suitable for use in the presence of flammable anesthetic or oxygen.
PART I. User & Installation Manual 2. Radio Frequency compliance FCC (For USA) FCC ID: 2ALPQRYRWM001A • 5.15- 5.25 GHz band is restricted to indoor operations only. • Host device of the approved module shall be marked with the following item: • Compliance with FCC requirement 15.407(c) Data transmission is always initiated by software, which is the passed down through the MAC, through the digital and analog baseband, and finally to the RF chip. Several special packets are initiated byte MAC.
PART I.User & Installation Manual Le present appareilest conforme aux CNR d'Industrie Canada applicables aux appareilsradioexempts de licence.L'exploitationestautorisee aux deux conditions suivantes : (1) l'appareil ne doitpasproduire de brouillage, et (2) l'utilisateur de l'appareildoit accepter tout brouillageradioelectriquesubi, meme si le brouillageest susceptible d'encompromettre le fonctionnement. RED Notice (European Union) The product compliance to the Directive 2014/53/EU as Radio Equipment.
PART I. User & Installation Manual Immunity Test Basic EMC standard or test method Operating mode Test leve/requirement Harmonic Current Emission EN 61000-3-2 Wired mode Class A Voltage change, Voltage fluctuations and Flicker Emission EN 61000-3-3 IEC 61000-3-3 Wired mode Pst: 1 Plt: 0.65 Electrostatic Discharge Immunity EN 61000-4-2 IEC 61000-4-2 Wired & Wireless mode Radiated RF Electromagnetic Field Immunity EN 61000-4-3 IEC 61000-4-3 Wired & Wireless mode 3 V/m 80 MHz-2.
PART I.User & Installation Manual 1.
PART I. User & Installation Manual Symbols Descriptions Read carefully Manufacturer Date of manufacture Serial number WEEE : Waste Electrical and Electronic Equipment Authorized representative in the European community. Batteries Marking (Battery Directive 2006/66/EC) CE symbol grants the equipment compliance to the European Directive for Medical Devices 93/42/EEC as a classⅡa device and 1999/5/EC. Authorized by Notified Body SGS (code no.
PART I.User & Installation Manual 1.3 Warning Environment of Use and Storage Follow the specified process of operational instructions written in this manual for the safety of the users and patients. Do not use or store the detector near any flammable chemicals such as thinner, WARNING benzene, etc. Also, this detector is not a category AP or APG equipment. If chemicals are spilled or evaporate, it may result in fire or electric shock through contact with electric parts inside the detector.
PART I. User & Installation Manual Maintenance and Inspection For safety reasons, be sure to turn off the power of the detector when the following inspections are going to be performed. Otherwise, it may result in electric shock. When the detector is going to be cleaned, be sure to turn off the power of each detector, and unplug the power cable from the AC outlet. Do not use any type of solvent, such as benzene. Otherwise, fire or electric shock may WARNING result.
PART I.User & Installation Manual 1.4 Caution Environment of Use and Storage Do not install the detector in a location with the conditions listed below. Otherwise, it may result in failure or malfunction, cause fire or injury. CAUTION Do not position the equipment so that it is difficult to disconnect the power cord. Close to facilities where water is used. Where it will be exposed to direct sunlight. Close to air-conditioner or ventilation equipment.
PART I. User & Installation Manual Maintenance and Inspection For safety reasons, be sure to inspect the detector before using it. In addition, carryout a regular inspection at least once a year. If the detector is defective, do not disassemble the detector randomly. Maintenance of CAUTION the detector should be done by an authorized service provider. Please contact OSKO Inc. Customer Service team or authorized agent. Be sure to check the user’s manual for replaceable components.
PART I.User & Installation Manual Do not subject cells or batteries to mechanical shock. In the event of a cell leaking, do not allow the liquid to come into contact with the skin or eyes. If contact has been made, wash the affected area with copious amounts of water and seek medical advice. Equipment should be designed to prohibit the incorrect insertion of cells or batteries and should have clear polarity marks.
PART I. User & Installation Manual Always purchase the correct cell or battery for the equipment. Keep cells and batteries clean and dry. Wipe the cell or battery terminals with a clean dry cloth if they become dirty. Secondary cells and batteries need to be charged before use. Always use the correct charger and refer to the manufacturer’s instructions or equipment manual for proper charging instructions. Do not leave a battery on prolonged charge when not in use.
PART I.User & Installation Manual 1.5 Safety Information Preparation Be sure to connect the cables to the proper connectors. Otherwise, the detector may malfunction or may be damaged. The power supply provided by OSKO Inc. is designed for the detector from OSKO Inc.. Please contact OSKO Inc., if any other type of power supply is needed to be used. Be sure to fully charge the battery before use. Charge the battery on the day of examination or on the previous day.
PART I. User & Installation Manual Be sure to check the detector daily and confirm that it works properly. Sudden heating of the room in cold areas will cause condensation to form on the detector. In this case, wait until condensation disappears before performing exposure. If the detector is used with condensation formed on it, problems may occur in the quality of the detector.
PART I.User & Installation Manual 1.6 Label and Location of Attachment Refer to the back of the device for details. 1.7 Summary of usability specifications Medical purposes Provision and reading of disease and injury diagnostic images Patient groups No patient population exists who uses or is in contact with the device. Patient population for the X-ray images read is not specified.
PART I. User & Installation Manual 2. Product Introduction and Specification 2.1 Product Features The Edge Air 1417 / 1417WCC are a wireless digital flat panel detector that has been designed for a faster, more streamlined approach to digital radiography systems. The Edge Air 1417 / 1417WCC detectors utilize a combination of propriety TFT glass and high quality scintillators, which along with a pixel pitch of 140 microns and assures delivery of sharp, high quality images.
PART I.User & Installation Manual 2.2 Product Components 2.2.1 Basic Components Detector 1EA AC power cord 1EA (110V or 220V) 2.2.2 Installation CD 1EA Manual File 1EA Manual 1EA Optional Components Power supply1EA AC power cord 2EA (110V or 220V 1.
PART I. User & Installation Manual 2.3 Part Names and Functions 2.3.1 Detector 1. Wireless Antenna 2. Battery Unlock button This is an unlock button to remove battery 3. OLED window 4. Mode select button & LED indicator Mode Change Press the 1 second Indication the status of detector Name of LED Status of LED Status of product Orange on Station mode(wireless) Green on AP mode(wireless) Off Wired mode Blinking Orange and Green alternatively Sleep mode Mode 5.
PART I.User & Installation Manual 2.3.2 Battery & Mobile charger 1. Battery unlock level This is an unlock-level to remove battery 2. Power connector Connects to the mobile charger adapter 3. LED Indicator Display battery charging status. 4. LED Color Battery Status Green Fully charged Orange Charging Red Error Battery : Rechargeable Lithium Ion battery(Charging Time-3 hrs) • In the diagram above, the box shows where the remaining battery percentage is displayed.
PART I. User & Installation Manual 2.3.3 • Battery warranty period: 6 months • If detector is not used for long period time (More than 15 days), battery should be removed from detector. Otherwise, it will be discharged. Stand Battery & Charger (Optional) 1. Battery : Rechargeable Lithium Ion battery(Charging Time-3 hrs) • In the diagram above, the box shows where the remaining battery percentage is displayed.
PART I.User & Installation Manual 2.3.4 Power Supply (Optional) 1. Power Plug Connector Connects to the AC power cord 2. LAN Connector Ethernet port for transmitting an image/command between the detector and PC 3. Link Connector Used for charging the battery while the detector is in use (Connect the detector and power supply) 4. LED Indicator Display status of the power supply. 5. Color Status Power Status Green On Power on Fuse T3.15 AL 250V 6. Switch Power On/Off switch 7. 2.3.
PART I. User & Installation Manual 2.4 Part Specifications 2.4.1 Detector Parameter Spec. Unit Sensor Type Amorphous Silicon with TFT (Single panel) - Scintillator Type CsI:TI - Total Pixel Matrix 2500 X 3052 Pixels Total Pixel Area 350.0 X 427.3 mm Pixel Pitch 140 μm Effective Pixel Matrix 2440 X 2992 Pixels A/D Conversion 14 / 16 bits Data Transfer 1Gbps Ethernet 802.
PART I.User & Installation Manual 2.4.2 • Wireless antennas: The module adopts the latest 802.11ac Dual-Band technology (2.4 GHz and 5 GHz). The transmitter of the module is powered by host equipment (Detector). The antennas are 3 printed-dipole antennas. • Wireless module: The Rayence RWM001A 802.11 a/b/g/n/ac half mini PCI-e module is implemented. It supports 3T3R (3 transmit 3 receive) MIMO technology, which delivers throughput up to 1300Mbps.
PART I. User & Installation Manual 2.4.5 Dual Battery Charger (Optional) [Model name: RC120WA] Parameter 2.4.6 Spec. Size 200 x 100 x 150 mm Weight 0.9 Kg Input 20 VDC Output 12.6 VDC Dual Battery Charger Adapter (Optional) [Model name: PMP120-13-3] Parameter 2.4.7 Spec. 160 x 76 x 43.7 mm Weight 0.8 Kg Input 100-240VAC, 47~63Hz, 1.4~0.6A - Output 20VDC, Max 6.0A - Power Supply (Optional) [Model name: RP003A] Spec. Unit Dimension 188 X 92 X 41.5 mm Weight 0.
PART I.User & Installation Manual 2.5 Environmental Requirements 2.5.1 PC Requirement Item Detail CPU At least Intel Pentium IV HT with 2.8GHz, Intel Core Duo / Core 2 or comparable AMD Dual Core processor RAM At least 4GB of RAM requirement (4GB for 32 BITS OS and 8GB for64 BITS OS recommended) Capacity of Disk Drive At least 500GB for application and archiving. Recommended 500GBfor applications and secondary drive of 1TB for image archiving. Dual 10/100/1000 network card system required.
PART I. User & Installation Manual 2.6 Dimensions (Unit: mm) 2.6.
PART I.User & Installation Manual 2.6.2 Battery 2.6.3 Mobile charger 2.6.
PART I. User & Installation Manual 2.6.5 Power Supply (Optional) 2.6.
PART I.User & Installation Manual 2.6.7 Dual Battery Charger Adapter (Optional) 2.6.
PART I. User & Installation Manual 3. Installation and Calibration 3.1 Installation 3.1.1 Software Installation 1. Insert the CD that comes with the Detector. 2. Install "setup.exe" from "\Release Davinci_version" and click "Next". 3. Choose the model from the Detector Type list and click "Next".
PART I.User & Installation Manual 4. Click “Next”. 5. Select “MultiByte-Character Set” and click “Next”. Choose UNICODE if console SW is supporting UNICODE. If Multibyte - Character set or UNICODE is not installed correctly, images will not be properly acquired.
PART I. User & Installation Manual 6. Click “Install”. 7. Click “Finish”.
PART I.User & Installation Manual 3.1.2 Install battery Attach the battery (RB37WHA) to the detector as below. • Battery removal Push the battery unlock button. 3.1.3 Mode Selection Mode Description Automatically detects X-ray radiation without integration of the generator and detector. Detects X-ray radiation by sending and receiving sync signals through the integration between generator and detector.
PART I. User & Installation Manual Auto Trigger & Station Mode Follow instructions from 3.1.4 Product Set Up 1. Auto Trigger & Station Mode. Auto Trigger & AP Mode Follow instructions from 3.1.4Product Set Up 2. Auto Trigger & AP Mode. Manual Trigger & Station mode Follow instructions from 3.1.4 Product Set Up 3. Manual Trigger & Station Mode.
PART I.User & Installation Manual Manual Trigger & AP mode Follow instructions from 3.1.4 Product Set Up 4. Manual Trigger & AP Mode. Auto Trigger & Wired Mode Follow instructions from 3.1.4 Product Set Up 5. Auto Trigger & Wired Mode.
PART I. User & Installation Manual Manual Trigger & Wired Mode Follow instructions from 3.1.4 Product Set Up 6. Manual Trigger & Wired Mode.
PART I.User & Installation Manual 3.1.4 Product Set Up 1. Auto Trigger & Station Mode Product Set up Connect the cable 1. Connect the wireless AP and PC with the LAN cable.
PART I. User & Installation Manual PC Set up 1. Set up the Network as below. • Desktop > Network Icon > Right click > Properties > Change Adaptor Settings • Control Panel > Network and Sharing Center > Change Adaptor Settings 2. To use station mode, right click "Local Area Connection" and click Properties. 3. Double click “Internet Protocol Version 4 (TCP/IPv4)”.
PART I.User & Installation Manual 4. Select “Obtain an IP address automatically” and click "OK". Wireless AP (Access Point) Set up Please use certified wireless AP and follow each manufacture’s setup manual. 1. Set up wireless AP as below. • SSID: Griffon • Internal network • • • IP address: 2.2.2.1 • Subnet mask: 255.255.255.0 • Dynamic IP allocation range: 2.2.2.2 ~ 2.2.2.
PART I. User & Installation Manual Set up SW 1. Turn on the detector 2. Choose Station Mode by pressing and holding the mode select button. LED LED Color Orange Station Mode(Wireless) Green AP Mode(Wireless) None Wired Mode Blinking Orange and Green alternatively Sleep Mode MODE select button 3. Mode Open “_vadav.lnk” from “C:\davinci”.
PART I.User & Installation Manual 4. After checking connectivity, click the “Calibration & Acquisition” tab and click "Edit". 5. Another window will now be opened as shown below. Select "Auto Trigger" from "Trigger Mode". If the "Window time" needs to be changed, type the value at "Window Time" from "Setting". (For change parameter settings, must click “Get Settings” button) In Auto trigger mode, be sure to set the "Window time" longer than an exposure time.
PART I. User & Installation Manual 2. Auto Trigger & AP Mode Product Set Up PC Set up 1. Set up the Network as below.
PART I.User & Installation Manual 2. To use AP mode, right click "Wireless Network Connection" and click Properties. 3. Double click “Internet Protocol Version 4 (TCP/IPv4)”. 4. Select “Obtain an IP address automatically” and click "OK". In automatically IP address, If Detector cannot connets to PC, select “Use the following IP address” and type "2.2.2.50" at the "IP address" and type "255.255.255.0" at the "Subnet mask".
PART I. User & Installation Manual Set up SW 1. Connect the Detector and turn on the power. 2. Choose AP Mode by pressing and holding the mode select button. LED MODE 3. LED Color Mode Orange Station Mode(Wireless) Green AP Mode(Wireless) None Wired Mode Blinking Orange and Green alternatively Sleep Mode Choose the SSID (detector's SN) from Wireless Network Connection list.
PART I.User & Installation Manual 4. Open “_vadav.lnk” from “C:\davinci”. Once the program is opened and the detector is connected, the LINK LED light will blink and the Detector Status will display detector information as below. Once the correct Sensor IP is put into the Davinci, it will automatically pull the parameter of the connected detector. Model type Edge Air 1417_140A Parameter Selected [75] Edge Air 1417_140A.par Default IP address for wireless connection is 2.2.2.
PART I. User & Installation Manual 5. After checking connection, click the “Calibration & Acquisition” tab and click "Edit". 6. Another window will open as below once the “Edit” button is pressed. Select "Auto Trigger" from "Trigger Mode". If "Window time" needs to be changed, type the value at "Window Time “from "Setting". (For change parameter settings, must click “Get Settings” button) In Auto trigger mode, be sure to set the "Window time" longer than an exposure time.
PART I.User & Installation Manual 3. Manual Trigger & Station Mode Product Set up (with AGI) Connect the cable (with AGI box and AP) 1. Connect the wireless AP and PC with the LAN cable. 2. Connect the AGI box and PC with the USB cable.
PART I. User & Installation Manual 3. Connect the AGI box and generator with the P-interface or trigger cable.
PART I.User & Installation Manual PC Set up 1. 2. 56 / 126 Set up the Network as below. • Desktop > Network Icon > Right click > Properties > Change Adaptor Settings • Control Panel > Network and Sharing Center > Change Adaptor Settings To use station mode, right click "Local Area Connection" and click Properties.
PART I. User & Installation Manual 3. Double click “Internet Protocol Version 4 (TCP/IPv4)”. 4. Select “Obtain an IP address automatically” and click "OK". Wireless AP (Access Point) Set up Please use certified wireless AP and follow each manufacture's setup manual. 1. Set up wireless AP as below. • SSID: Griffon • Internal network • • • IP address: 2.2.2.1 • Subnet mask: 255.255.255.0 • Dynamic IP allocation range: 2.2.2.2 ~ 2.2.2.
PART I.User & Installation Manual Set up SW 1. Connect the Detector and turn on the power. 2. Choose Station Mode by pressing and holding the mode select button. LED LED Color Orange Station Mode(Wireless) Green AP Mode(Wireless) None Wired Mode Blinking Orange and Green alternatively Sleep Mode MODE select button 3. Mode Open “_vadav.lnk” from “C:\davinci”.
PART I. User & Installation Manual 4. After checking connectivity, click the “Calibration & Acquisition” tab and click "Edit". 5. Another window will be opened as below once the “Edit” button is pressed. Select "Manual" from "Trigger Mode". If "Window time" needs to be changed, type the value at "Window Time" from "Setting".
PART I.User & Installation Manual 4. Manual Trigger & AP Mode Product Set Up (with AGI) Connect the cable (with AGI) 1. Connect the AGI box and PC with the USB cable.
PART I. User & Installation Manual 2. Connect the AGI box and generator with the P-interface or trigger cable.
PART I.User & Installation Manual PC Set up 1. 2. 62 / 126 Set up the Network as below. • Desktop > Network Icon > Right click > Properties > Change Adaptor Settings • Control Panel > Network and Sharing Center > Change Adaptor Settings To use AP mode, right click "Wireless Network Connection" and click Properties.
PART I. User & Installation Manual 3. Double click “Internet Protocol Version 4 (TCP/IPv4)”. 4. Select “Obtain an IP address automatically” and click "OK". In automatically IP address, If Detector cannot connets to PC, select “Use the following IP address” and type "2.2.2.50" at the "IP address" and type "255.255.255.0" at the "Subnet mask". It is recommended to use the IP of the product independently of each other as it can cause PC disconnect when two or more products use the same IP.
PART I.User & Installation Manual 4. Open “_vadav.lnk” from “C:\davinci”. Once the program is opened and the detector is connected, the LINK LED light from the detector will blink and the Detector Status will display panel information as below. Once the correct Sensor IP is put into the Davinci, it will automatically pull the parameter of the connected detector. Model type Edge Air 1417_140A Parameter Selected [75] Edge Air 1417_140A.par Default IP address for wireless connection is 2.2.2.
PART I. User & Installation Manual 5. After checking connectivity, click the “Calibration & Acquisition” tab and click "Edit". 6. Another window will be opened as below once the “Edit” button is pressed. Select "Manual" from "Trigger Mode". If the "Window time" needs to be changed, type the value at "Window Time" from "Setting".
PART I.User & Installation Manual 5. Auto Trigger & Wired Mode Product Set Up Connect the cable 1. Connect the power supply and PC with the LAN cable. 2. Connect the power supply and Detector with the Link cable.
PART I. User & Installation Manual PC Set up 1. Set up the Network as below. • Desktop > Network Icon > Right click > Properties > Change Adaptor Settings • Control Panel > Network and Sharing Center > Change Adaptor Settings 2. To use Wired mode, right click "Local Area Connection" and click Properties. 3. Double click “Internet Protocol Version 4 (TCP/IPv4)”.
PART I.User & Installation Manual 4. Select “Use the following IP address” and type "2.2.2.20" at the "IP address" and type "255.255.255.0" at the "Subnet mask". Set up SW 1. Turn on the power supply 2.
PART I. User & Installation Manual 3. Open “_vadav.lnk” from “C:\davinci”. Once the program is opened and the detector is connected, the LINK LED light will blink and the Detector Status will display detector information as below. Once the correct Sensor IP is put into the Davinci, it will automatically pull the parameter of the connected detector. Model type Edge Air 1417_140A Parameter Selected [75] Edge Air 1417_140A.par Default IP address for wireless connection is 2.2.2.
PART I.User & Installation Manual 4. After checking connectivity, click the “Calibration & Acquisition” tab and click "Edit". 5. Another window will be opened as below once the “Edit” button is pressed. Select "Auto Trigger" from "Trigger Mode". If "Window time" needs to be changed, type the value at "Window Time" from "Setting". (For change parameter settings, must click “Get Settings” button) In Auto trigger mode, be sure to set the "Window time" longer than an exposure time.
PART I. User & Installation Manual 6. Manual Trigger & Wired Mode Product Set up (with Power supply and AGI) Connect the cable (with Power supply and AGI) 1. Connect the power supply and PC with the LAN cable. 2. Connect the power supply and Detector with the Link cable.
PART I.User & Installation Manual 3. Connect the AGI box and PC with the USB cable. 4. Connect the AGI box and generator with the P-interface or trigger cable.
PART I. User & Installation Manual PC Set up 1. Set up the Network as below. • Desktop > Network Icon > Right click > Properties > Change Adaptor Settings • Control Panel > Network and Sharing Center > Change Adaptor Settings Set up SW ` 2. To use Wired mode, right click "Local Area Connection" and click Properties.
PART I.User & Installation Manual 3. Double click “Internet Protocol Version 4 (TCP/IPv4)”. 4. Select “Use the following IP address” and type "2.2.2.20" at the "IP address" and type "255.255.255.0" at the "Subnet mask". Set up SW 1. Turn on the power supply 2.
PART I. User & Installation Manual 3. Open “_vadav.lnk” from “C:\davinci”. Once the program is opened and the detector is connected, the LINK LED light will blink and the Detector Status will display detector information as below. Once the correct Sensor IP is put into the Davinci, it will automatically pull the parameter of the connected detector. Model type Edge Air 1417_140A Parameter Selected [75] Edge Air 1417_140A.par Default IP address for wireless connection is 2.2.2.
PART I.User & Installation Manual 76 / 126 4. After checking connectivity, click the “Calibration & Acquisition” tab and click "Edit". 5. Another window will be opened as below once the “Edit” button is pressed. Select "Manual" from "Trigger Mode". If "Window time" needs to be changed, type the value at "Window Time" from "Setting".
PART I. User & Installation Manual 3.2 Calibration In order to properly acquire images, calibration must be performed. Without calibration, optimum images cannot be acquired. OSKO Inc. recommends to warm up the detector for 5 minutes after turning on the power. Calibration should be performed at regular intervals, typically once every six (6) months, or whenever the central beam of the X-ray source has been moved relative to the Detector. 3.2.1 Auto Calibration Mode 1.
PART I.User & Installation Manual 78 / 126 4. After checking connection, click “Calibration & Acquisition” tab. 5. Click “Calibration”.
PART I. User & Installation Manual 6. Once "Start” is pressed, the program automatically gets a Dark frame and the acquired Dark frame is stored in “C:\davinci\cal”. A Calibration Point file will be created automatically. 7. After acquiring the Dark frame, shoot an X-ray when the “Acquiring bright frame” window popup. Depending on the dark level of the product, 1~2 additional calibration points may be created.
PART I.User & Installation Manual 8. “Acquiring bright frame” is closed after radiation is detected, and the program will show if the detected radiation is within acceptable range. When PASS is displayed Shoot X-ray with same technique when “Acquiring bright frame” is popped up. 9. When FAIL is displayed Adjust technique to get acceptable value and shoot again. For each Calibration point, four images must successfully be acquired.
PART I. User & Installation Manual 3.2.2 Manual Bad Pixel Map Set Up 1. Click “Calibration”. 2. Click "View Frames".
PART I.User & Installation Manual 3. Set Manual bad pixel map (BPMM) as below. No. 1 2 3 4 Overview At Figure 1, choose S from the list of M (❶) and click "BPM". Check if BPMM window is popped up as Figure 3. Choose "Bad" from Pixel list (Figure 3 - ❷). Choose either “row” or “col” from Figure 3 - ❸. Put the coordinate of pixel to set bad pixel at Figure 3 - ❹. If bad pixel is a line, put the range as below at Figure 3 – ❺. From 5 6 7 8 4.
PART I. User & Installation Manual 4. Usage 4.1 Set Up 4.1.1 Product Connectivity 1. Connect the detector and turn on the power. 2. Open “_vadav.lnk” from “C:\davinci”. 3. Once the detector is connected, detector information is displayed in Detector Status and Link &Ready are checked as below. If "Detector Status" does not show anything, please refer to 3.1 Installation in Part.1 User & Installation Manual to connect the detector properly.
PART I.User & Installation Manual 4.1.2 4.1.3 Image Set Up 1. In order to rotate or flip an image, use the option of "Flip and Rotate" as shown below. 2. In order to change the size of an image, use "Crop Rows and Columns" as below. 3. Click "Apply" to save. Multi Detector Set Up Refer to 3 Multi Detector Set Up in Part.2 Service Manual for Multi-Detector Setting.
PART I. User & Installation Manual 4.2 Image Acquisition 4.2.1 Product Connection 1. Connect the detector and turn on the power. 2. Open “_vadav.lnk” from “C:\davinci”. 3. Once the detector is connected, information of the detector is displayed in Detector Status and Link & Ready & Calibration is checked as below. If "Detector Status" does not show anything, please refer to 3.1Installation in Part.1 User & Installation Manual to connect the detector properly.
PART I.User & Installation Manual 4.2.2 86 / 126 Image Acquisition 1. Click the "Calibration & Acquisition” tab and type the name of the image inside the box below. After naming the image, click “Get Image”. 2. Shoot an X-ray once the "Acquiring bright frame" window pops up. 3. An acquired image will be stored in “C:\davinci\I.2440x2992” and the name of the file will be "(typed name from Step 1).raw”. 4. The format of the stored file is 16 bit little-endian order.
PART I. User & Installation Manual 4.3 View Images 1. Click “View Images” 2. Another window will be popped up as below.
PART I.User & Installation Manual Pixel value at certain level Choose "S" from marked box. Profile for horizontal line Choose "R" from marked box.
PART I. User & Installation Manual Profile for vertical line Choose "C" from the marked box. 4.4 Additional Function 4.4.1 Battery Remain Once you click "Get Image" under the "Calibration & Acquisition” tab, the Status window will show how much battery remains.
PART I.User & Installation Manual 4.4.2 Wireless signal Strength Once you click "Get Image" under the "Calibration & Acquisition” tab, the Status window will show the strength of the wireless signal. Before you check the wireless signal in the AP mode, the detector's serial number should be entered at the "SSID".
PART I. User & Installation Manual 4.4.3 Sleep Mode 1. Click “Edit” under the Calibration & Acquisition tab. 2. Under the “Sleep Timeout” setting, enter a designated time for the detector to go into Sleep Mode. (For change parameter settings, must click “Get Settings” button) Sleep Mode does not apply when set to zero. Power consumption is reduced by 40% when Sleep Mode is used. 3. To turn off Sleep Mode, attempt to acquire an image or press the power button on the detector just once.
PART I.User & Installation Manual 4.4.4 Preview 1. Click “Edit” under the Calibration & Acquisition tab. 2. After checking the Preview and 4x4 Binning, a 4x4 binned image appears which allows for a quicker image preview. (For change parameter settings, must click “Get Settings” button) By unchecking 4x4 Binning, a normal image preview appears. By unchecking Preview, a full frame image appears.
PART I. User & Installation Manual 4.4.5 Recent Frame The last acquired image can be opened by clicking "Recent Frame" under the "Calibration & Acquisition” tab. 4.4.6 Restore Connection When the connection between the detector and PC is lost, the connection can be made again by clicking "Restore Connection" under the "Calibration & Acquisition” tab.
PART I.User & Installation Manual 4.4.7 Image Storage function 1. Press the "Mode select button" twice (within 2 seconds) and click one more time to go to "storage mode". Mode status can be checked by the OLED window of Detector 2. Press the mode select button for 3 seconds to check the status of storaged images.
PART I. User & Installation Manual A. Mobile Device can connect to Detector’s AP Mode B. Connect to "http://192.168.1.80:3000". Sign in with password ‘1234’ C. Ater enter to worklist menus, click ‘+’ icon and create worklist info. D. Can see all worklist for will acquire image. Click ‘Capture’ button. E. In detector ready status, can take X-ray image.
PART I.User & Installation Manual F. Shot X-ray Show image and can check correct Select Accept/Suspend button G.
PART I. User & Installation Manual 4.4.8 OLED function 1. OLED is operating when detector is turned on. 2. Use the Mode select button and Power button can be used OLED function as follows.
PART I.User & Installation Manual 4.4.9 GPR function 1. GPR(Grid Pattern Removal) is an option to remove grid line artifacts when using an analog grid 2. Applicable grid : 103/line/inch, 85Line/inch Setting GPR Set vadav as below No.
PART I. User & Installation Manual GPR Parameter Select a parameter in consideration of the used grid specification & detector model. 1417 model(Edge Air 1417 / 1417WCC) use to Automatic parameter recommended 1717 model(Edge Air) use to Vertical Parameter recommended.
PART I.User & Installation Manual 5. Maintenance 5.1 Cleaning 5.2 1. Clean the detector with IPA (Isopropyl-alcohol) when it is contaminated. 2. Before cleaning the detector, turn off the power and separate the battery. 3. Wear waterproof gloves to protect your hands from direct contact with IPA or any other liquid. 4. Do not pour or spray IPA directly on the detector. Use fabric or soft cloth moistened with IPA to clean. 5. Avoid getting IPA or any other liquid into the detector. 6.
PART I. User & Installation Manual 5.3 Replaceable Parts and Instruction of Replacement 5.3.1 Fuse: T3.15 AL 250V Replacing the Fuse 1. Press the fuse as below and pull the fuse box. 2. Pull the fuse and replace with another fuse. 5.3.2 Power cord: H05VV-F 0.75SQ * 3C 5.3.3 Ethernet Cable: UTP 4PR 24AWG (CAT.
PART I.User & Installation Manual 5.4 Disposal or Recycling Follow local governing ordinances and recycling plans regarding the disposal or recycling of device components. Disposal of old Electrical & Electronic Equipment (Application in the European Union and other European countries with separate collection system.) This symbol indicates that this product shall not be treated as household waste.
PART I. User & Installation Manual 6. Warranty 6.1 Warranty If Buyer promptly notifies OSKO Inc. or Seller regarding any parts that fail to perform as specified under normal usage during the Warranty Period and OSKO Inc. determines that such failure resulted from a defect in materials or workmanship during the Warranty Period, then OSKO Inc., at its option, shall repair, rebuild or adjust the affected parts. OSKO Inc.
PART II. Service Manual PART II. Service Manual 1. Overview ......................................................106 2. FPD Manager Instruction (IP, SSID Set Up / Firmware, FPGA Update) .............................107 3. Multi Detector Set Up...................................115 4. Troubleshooting ...........................................
PART II.Service Manual 1. Overview This service manual gives additional instructions for setting up the detector.
PART II. Service Manual 2. FPD Manager Instruction (IP, SSID Set Up / Firmware, FPGA Update) 2.1 Detector IP Address Set Up 1. Turn on the power of detector and connect with PC. (IP address: 2.2.2.100) [Connect as wired mode (IP address: 2.2.2.101) with Link cable or wireless mode (IP address: 2.2.2.100).] 2. After the power of detector is on, open "FPD_Manager.exe". 3. Type detector's current IP address at "Detector IP Address" from "Current Setting" as below.
PART II.Service Manual 108 / 126 4. Select "IP change" and type the IP address. Click "Start Setting". 5. Click "OK" once the message below pops up. 6. Click "OK" again once the message below pops up. 7. Turn off the power of the detector and after 5 seconds, turn the power back on.
PART II. Service Manual 2.2 SSID, PSK (Pre-Shared Key) Set Up 1. Turn on the power of the detector and connect it to the PC. (IP address: 2.2.2.100) [Connect as wired mode (IP address: 2.2.2.101) with Link cable or wireless mode (IP address: 2.2.2.100).] 2. After the power of detector is on, open "FPD_Manager.exe". 3. Type the detector's current IP address in "Detector IP Address" from "Current Setting" as below. Detector's Ethernet Controller is operated with Second IP address, 192.168.124.80.
PART II.Service Manual 5. Select "Station Mode Setup" and "AP Mode Setup". Type the SSID and PSK, and click "Start Setting". Station Mode Setup: When external AP is being used, use the same SSID and PSK as applied. AP Mode Setup: When built-in AP is being used, use the same SSID and PSK as applied. Frequency (Ghz): Default Channel is set to 11n/2.4G/Auto. (802.11n/2.4Ghz/ACS). The ACS (Automatic Channel Selection) feature is setting automatically channel for good AP Mode connection quality.
PART II. Service Manual 2.3 Firmware, FPGA Update 1. Turn on the power of the detector and connect it to the PC. [Connect as wired mode (IP address: 2.2.2.101) with Link cable or wireless mode (IP address: 2.2.2.100).] 2. After the power of detector is on, open "FPD_Manager.exe". 3. Type the detector's current IP address in "Detector IP Address" from "Current Setting" as below. Detector's Ethernet Controller is operated with Second IP address, 192.168.124.80.
PART II.Service Manual 4. Select "Firmware/FPGA Update" and click "Select" to browse the Firmware and FPGA. Once the files are selected, click "Start Setting". Firmware File: File extension is either Davinci or tar. FPGA file: File extension is bin. 5. Click "OK" once the message below pops up. 6. Turn off the power of detector and after 5 seconds, turn the power back on. Detector IP address might be changed to 192.168.1.80 after updating depends on Firmware.
PART II. Service Manual 2.4 Set Windows Firewall to use FPD_Manager (For Win 7) FPD_Manager would not be performed properly if Windows Firewall blocks FPD_Manager. Please follow the steps below in this case. 1. ‘Control Panel’ -> ‘Windows Firewall’ -> ‘Allow an app or feature through Windows Firewall’ 2. Check "Name", "Private" and "Public" if FPD_Manager program is already on the list. Click "Allow another app..." when FPD_Manager program is not on the list.
PART II.Service Manual 114 / 126 3. Select the program and add if it is already on the list. Click "Browse" when the program is noon the list. 4. Browse and open FPD_Manager program and repeat step ②.
PART II. Service Manual 3. Multi Detector Set Up 1. Open “_vadav.lnk” from "C:\davinci”. 2. Click "Edit" under the "Settings" tab. 3. Select "Detector #1". Choose the product model and type the IP address.
PART II.Service Manual 116 / 126 4. Click "Edit" from "Detector #1". Set up the "Trigger Mode" and "Setting". (For change parameter settings, must click “Get Settings” button) 5. Select "Detector #2". Choose the product model and type the IP address.
PART II. Service Manual 6. Click "Edit" from "Detector #2". Set up the "Trigger Mode" and "Setting". (For change parameter settings, must click “Get Settings” button) 7. If a third detector is being used, select "Detector #3". Choose the product model and type the IP address.
PART II.Service Manual 8. Click "Edit" from "Detector #3". Set up the "Trigger Mode" and "Setting". (For change parameter settings, must click “Get Settings” button) 9. Choose the detector from the "Number" option and perform the calibration. The calibration folder is named according to the third and fourth numbers of the IP address. (e.g. C:\Davinci\CAL_ 01_80) For further instructions on calibration, please refer to 3.2Calibration in Part.1 User & Installation Manual.
PART II. Service Manual 4. Troubleshooting If any problem occurs during the usage of the product, please use this chapter as a troubleshooting guide. Follow the instructions to resolve the problem. If the problem is not resolved, please contact our OSKO Inc. Customer Service team (E-mail: service@oskomedical.com). 4.1 LAN Connection Issue 1. Wireless Mode 1. Check the power Make sure the remaining battery percentage is above 25%. Check that the power of the detector is on. 2.
PART II.Service Manual 3. Check PC Set up Make sure that the "Obtain an IP address automatically" is selected from "Internet Protocol Version 4 (TCP/IPv4)". 2. Wired Mode 1. Check the power Check the link cable and the power cord are connected properly Check that the power of the detector is on. 2. Check PC Set up Make sure that the IP address is set to “2.2.2.
PART II. Service Manual 4.2 4.3 Lost IP Address (Use one of the methods below) 1. Use a second IP address (192.168.124.80) and change the IP address 2. Press the “Reset” button to reset the IP address. (Default IP : 2.2.2.100) When using a reciprocating bucky If there is image noise casued by the electromagnetic, follow the instructions below. 1. Open “_vadav.lnk” from "C:\davinci”. 2. Click the "Calibration & Acquisition" tab. 3. Click "Edit".
PART II.Service Manual 4. Change "window time" from 2 to 4 sec and click “OK” (For change parameter settings, must click “Get Settings” button) The Window time default value is 0.
PART II. Service Manual 4.4 Auto Trigger Mode Follow these instructions when the panel auto triggers on its own and/or unintentionally acquires blank images. 1. Open “_vadav.lnk” from "C:\davinci”. 2. Click the "Calibration & Acquisition" tab. 3. Click "Edit". 4. Change "Auto Trigger Threshold" from 5 to 100 and click "OK".
PART II.Service Manual Supplement.1 Generator Specification All the medical generators are avabile to use. The specifications of the generators avalible are below. kV Range mA Range mAs Range 40 ~ 150 kV 1.0 ~ 1000 mA 0.1 ~ 1000 mAs If you need further information for generators, please contact your OSKO Inc. representative.
OSKO Inc. 8085 NW 90th Street, Miami, Florida, 33166 U.S.A. Tel: +1-305-599-7161 E-Mail: service@oskomedical.com www.OSKOMEDICAL.