Table of Contents 1. 2. 3. 3.1 3.2 3.3 3.4 3.5 4. 4.1 5. 5.1 5.2 5.3 5.4 5.5 5.6 6. 7. 8. 9. General notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Classification and intended use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Installation and commissioning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1. General notes This manual provides you with an overview of the necessary steps for commissioning and operating the computer system. Please read them even if you are familiar with the operation of computer systems. The term "device" is used below to refer to the OR-PC® 21 or OR-PC® 24 manufactured by ACL GmbH for Dräger. 2.
3. Installation and commissioning The device was thoroughly cleaned before leaving the factory. However, due to the transport conditions and the transport packaging, neither a sterile nor a disinfected product can be assumed upon delivery. For the initial cleaning, please observe the instructions provided in Section 5.1. EN First check the device and accessories for completeness by comparing with the enclosed delivery bill.
When establishing an electrical connection with other devices and when operating the device on a portable multiple socket outlet, ensure compliance with IEC 60601-1 in the currently valid version. To maintain conformity with the standard, so-called "decoupled interfaces" are provided for this case for the RJ-45 port (LAN1) and, depending on the configuration, one or both RS232 ports. Data can be transmitted via these galvanic isolators without a direct electrical connection.
3.2 Device environment, safety and handling instructions Please observe the safety instructions compiled below to avoid fire, electric shocks, injuries and damage to the device.
The device is intended for operation in an electromagnetic environment with controlled RF disturbances. Electromagnetic interference can be avoided by observing the distances to mobile HF telecommunication devices (transmitters) specified below to prevent performance degradations. These minimum distances depend on the output power of the communication device but must not be closer than 30cm between portable HF telecommunication devices and any point of the unit or its cables.
Electromagnetic immunity test Test level High frequency electromagnetic fields acc. 61000-4-3 IEC 60601-1-2:2014 table 4, home care; IEC 60601-1-2:2014 table 9 ESD (electrostatic discharge) acc. IEC 61000-4-2 IEC 60601-1-2:2014 table 4; IEC 60601-1-2:2014 table 8 Electrical fast transients / bursts acc. IEC 61000-4-4 IEC 60601-1-2:2014 Table 5 Surge acc. IEC 61000-4-5 IEC 60601-1-2:2014 Table 5 Conducted disturbances induced by RF fields acc.
3.4 BIOS The BIOS can be called up with the "Del" key when starting the device. The BIOS settings of the device are password-protected by default to prevent malfunctions caused by improper settings. BIOS password ex works: asdf9 3.
4. Operation The standby button can be used to switch the device on and off. This function corresponds to the soft start switch of conventional PCs. This does not disconnect the power supply from the mains. EN Your device has a capacitive keypad. It is enough to touch the glass pane on the key icon to trigger the function of the key. To avoid unintentional triggering of the keys during wipe disinfection, the keys must be pressed for at least 0.5 sec.
5. Maintenance 5.1 Care and hygiene instructions The device cannot be sterilized. Before starting disinfection on the running device, use the touch shutdown key to avoid unintentional incorrect entries. All surface disinfectants on the RKI list are approved for cleaning (see: "List of disinfectants and disinfection methods tested and approved by the Robert Koch Institute," Federal Health Gazette 12/2013.) The complete list can also be viewed on the RKI website or requested from ACL.
5.2 Maintenance intervals and wearing parts All devices are designed for 24/7 use. The resulting load places the highest demands on the components and materials used.
5.4 In-house repairs The manufacturer guarantees that the device and accessories have been carefully checked before leaving the factory. The manufacturer is only responsible for safety-related properties within the scope of statutory regulations if all work on the product is carried out by service personnel authorized by the manufacturer and the device and accessories are used exclusively in accordance with their intended use. For this reason, you will find warranty seals on or inside your device.
5.5 Transport notes for ACL devices To avoid transport damage, please use only the original ACL transport packaging. If you need suitable transport packaging, we will be happy to send you suitable materials for your device at short notice at cost price. Improper packaging may result in damage to individual components. Therefore, ensure careful handling during transport and loading operations. Ensure compliance with the transport conditions as described in the device manual and on the package label.
6. Manufacturer contact We are at your disposal at the following address: ACL GmbH Apelsteinallee 5 04416 Markkleeberg Germany Phone: Fax: +49 (341) 23078 - 60 Monday to Friday 8:00 am to 5:00 pm CET +49 (341) 23078 - 99 Email: service@acl.de Internet: www.acl.de/en 7. Warranty and guarantee We will be happy to provide you with all information on warranty conditions, types of warranty, repair procedures and exclusions of liability upon request. Our contact details can be found in Section 6. 8.
EN ALWAYS A STEP AHEAD 17