User MANUAL OAKWORKS DTPM 300 Imaging Table ® made in the USA with US & imported parts www.oakworksmed.com · 717.235.
© Copyright 2010 OAKWORKS® Medical Equipment, a division of OAKWORKS®, Inc. Notice The information contained in this document is subject to change without notice and should not be construed as a commitment by Oakworks®, Inc. This manual was originally drafted and approved in the English language. Printed in U.S.A. All rights are reserved.
Table of Contents Product Use Description ........................................................................................................... 1 Contradictions........................................................................................................................... 1 Important Safety Instructions Symbol Identification.............................................................................................................. 1 Safety Instructions ...................................
Product use description / important safety instructions product use description The Oakworks® Fluoroscopy table with Trendelenburg feature is a radiographic table intended for use with mobile or compact stationary C-arm Fluoroscopy Systems where the x-ray generator is located below the table top. It is ideally suited for pain management imaging and therapeutic procedures.
Important Safety instructions important safety instructions caution Read and Save These Instructions Improper use of this device can cause injury. Be sure to read all operating instructions prior to use. Duty Cycle: Max. 10% or 2 min. continuous use at full load then 18 min. without use. Use caution when changing the table position to avoid collision with other moving or stationary items. The Foot control must remain on the floor.
Important Safety instructions WARNING To reduce the risk of burns, fire, electric shock or injury to persons: 1. Unplug this furnishing from the electrical outlet before cleaning. 2. Unplug from outlet before adding or removing parts. 3. Never operate this furnishing if it has a damaged cord or plug, if it is not working properly, if it has been dropped or damaged, or dropped into water. Contact Oakworks Customer Service before use. 4. Keep the cord away from heated surfaces. 5.
Product Description DTPM300 Polycarbonate Top Polycarbonate Top Electronic Lift Towers Locking Casters Optional T-Rails Standard Specifications options Motion 1 Powered 24”-40” (61-101 cm.) Height Range 5” (12.5 cm.) Locking Casters Motion 2 Powered ±14° Trendelenburg/ Reverse Trendelenburg 6”, 20”, 78” (15.3 cm, 50.8 cm, 198.1 cm) or custom T-Rails Lengths Motion 3 Manual 23” (58 cm.
installation The Imaging tables come completely assembled and ready to use. Plug the cord into a functioning outlet that is rated for the table. (see Grounding below) Arrange the power cord and control cords so that they will not create a tripping hazard and where the controls are located to your liking and are conveniently accessible. Be sure access to plug is not blocked for disconnecting the table from power.
directions for use The Oakworks® Fluoroscopy Table with Trendelenburg positioning offers a variety of positioning capabilities for diagnostic x-ray imaging and imaging during therapeutic procedures. DANGER Should the use of this device create a circumstance under which the patient could be overexposed to ionizing radiation, discontinue use immediately and determine an alternative radiology table or alternative x-ray generating source to use.
directions for use Hand Control Operations caution When lowering the chair or using the Trendelenburg functions, make sure there is nothing that can impede motion (like stools, cabinets, accessory parts, cleaners, etc.) When operating the table’s controls, be sure to observe all cautions and warnings. Reverse Trendelenburg: Trendelenburg: - activates tilt up to 14º - positioning stops when table is level; press again for complete reverse.
directions for use manual longitudinal travel operation caution When operating the table’s locking knob be sure to avoid pinching. Keep your fingers and other materials clear from the pinch point. Both locking knobs must be securely tightened at all times - Travel can occur if the locking knobs are not secured by inadvertently pushing on the end of the table or by tilting the table. The table is designed to allow travel in either the left or the right direction as shown.
directions for use Top Pad Removal Remove the Top Pad by pulling the flaps at the underside of the table on either end until the Velcro® detaches. Replace the pad by first centering on the table and then pressing the flaps in place, anchoring the flaps to the Fastener Velcro® as shown. (applies to both the 1” (2.5 cm.) & 2” (5 cm.) pads) Underside of Tabletop Fastener Velcro® Top pad flap emergency stop All electric tables are equipped with an emergency stop switch.
directions for use carbon fiber arm board (accessory) caution Do not place undue weight or downward pressure on the Carbon Fiber Arm Board. It is a positioning device for the arms and should not be used as leverage to get on or off the table. Injury can occur. DO NOT exceed 30 lbs. (13.6 kg.) of load on the Support Section. Injury can occur. Use This radiolucent device can be used with any flat support surface where ionizing radiation (x-ray) is utilized for imaging.
directions for use spine positioning system ii (accessory) For directions on the use of the Spine Positioning System, see publication “Spine Positioning System” Part number MMMMNUP0003 Contoured Torso Support Pad Crescent Face Pad Radiolucent frame with adjustable face rest Contoured Torso Wedge 8” x 22” x 2” (20cm x 56cm x 5cm) Large rectangular adjuster pad 7” x 12” x 1½” (18cm x 31cm x 4cm) small rectangular adjuster pad 8” (20 cm.
reference usable imaging area Centered for a-p view: [16.57 CM] [16.57 CM] 12.01" [30.52 CM] 37.93" [96.33 CM] NO IMAGE No NO image IMAGE No image 19.88" [50.48 CM] 12.01" [30.51 CM] Centered Oblique 30º view from vertical 6.52" [16.57 CM] 6.52" [16.57 CM] 12.01" [30.52 CM] 37.93" [96.33 CM] NO IMAGE No NO image IMAGE No image 16.34" [41.50 CM] 12.01" [30.51 CM] Centered caudal 30º view from vertical 7.02" [17.84 CM] 7.85" [19.93 CM] 34.49" [87.61 CM] NO IMAGE 7.56" [19.
reference offset for a-p view: No image No image offset caudal 30º view from vertical No image No image 13
cleaning & disinfection / inspections / warranty cleaning & disinfection warning Before cleaning with any liquid cleaner be sure to unplug the power cord from the outlet. Use a 10% sodium hypochlorite (bleach) solution or Recommended Disinfectants on all surfaces. Clean all sides of each upholstered section. Follow the directions on the disinfectant and wipe off excess.
model number & serial number model & serial numbers Model Number Serial Number Power Rating US 120 V 60 Hz 4 A Class 1 Equipment Duty Cycle: MAX 2 minutes ON / 18 minutes OFF www.oakworks.com 923 East Wellspring Rd. New Freedom, PA 17349 pn 70537-3 (Dual Tower Fluoro Table - US) rev. A The model number and serial number are located on the frame of the top.
specifications product specifications Materials of Construction This product contains no latex. Top Frame Extruded aluminum with powdercoated or anodized surfaces. Table Top .5” (1.3 cm.) Polycarbonate Patient Comfort 1” (2.5 cm.) upholstered foam pad Table Length 78” (198 cm.) Table Width 24” (61 cm.) Height Range 24”-40” (61-102 cm.) standard Table Top Travel Range ±11.5” (29 cm) Trendelenburg Tilt ±14° Weight 390 lbs. (177 kg.) Shipping Weight 526 lbs. (239 kg.
specifications Guidance and Manufacturer’s Declaration - Electromagnetic Emissions The table is intended for use in the electromagnetic environment specified below. The customer or the user of the table should assure that it is used in such an environment.
specifications Guidance and manufacturer’s declaration - electromagnetic immunity The table is intended for use in the electromagnetic environment specified below. The customer or the user of the table should assure that it is used in such an environment.
specifications Guidance and manufacturer’s declaration - electromagnetic immunity The table is intended for use in the electromagnetic environment specified below. The customer or the user of the table should assure that it is used in such an environment.
this page is intentionally left blank 20
User MANUAL OAKWORKS DTPM300 Imaging Table ® Contact Information: Oakworks® Inc. 923 East Wellspring Road New Freedom, PA 17349 Phone: 717-235-6807 FAX: 717-235-6798 www.oakworksmed.com 3034177 CONFORMS TO: ANSI/AAMI STD ES60601-1 IEC STD 60601-1 3rd Edition IEC STD 60601-1-2 3rd Edition IEC STD 60601-1-6 3rd Edition iec std 60601-2-32 Certified to CAN/CSA STD C22.2 No.