User's Manual

DXS-F1
SUPERVISED
LONG RANGE
FALL DETECTION
PENDANT
TRANSMITTER
Operation
Instructions
PRODUCT DESCRIPTION
The Model DXS-F1 is a battery powered, miniature, water-resistant*, supervised
fall detection pendant transmitter designed for use with Linear PERS Consoles.
The transmitter is supplied with an adjustable lanyard with a breakaway release.
When a person’s fall is detected by the transmitter’s internal accelerometer, or
when the pendant’s button is pressed, a digitally coded wireless signal is sent to
the companion Console. The signal indicator on the transmitter lights green when
the unit is activated. When activated using the pendant’s button, the transmitter
can send a signal for up to 30 seconds per activation. To insure sending a
complete signal, press the pendant’s button for a minimum of two seconds.
The DXS-F1 transmitter is factory set to a unique code, so no fi eld coding
is required. Consoles must be programmed to the transmitter’s code before
system testing and operation. Refer to the Console instructions for details on
programming.
After the fi rst activation, the transmitter will send a status signal every two hours.
Every two hours the transmitter also tests for a low battery. If the battery tests low,
a low battery signal is sent. If the transmitter is activated with a low battery, the
signal indicator will light red.
The transmitter is sealed to provide water-resistance. The unit’s battery is not
replaceable. The transmitter should be disposed of properly at the end of its battery
life. Please contact your service provider for environmentally responsible options.
* NOTE: Water resistance not tested by UL
FEATURES
ENABLING STATUS & PROGRAMMING SYSTEM TESTING LIMITED WARRANTY
This Nortek Security & Control product is warranted against defects in material and workmanship for twenty
four (24) months. This warranty extends only to wholesale customers who buy direct from Nortek
Security & Control or through Nortek Security & Control’s normal distribution channels. Nortek Security & Control
does not warrant this product to consumers. Consumers should inquire from their selling dealer as to the nature
of the dealer’s warranty, if any. There are no obligations or liabilities on the part of Nortek
Security & Control LLC for consequential damages arising out of or in connection with
use or performance of this product or other indirect damages with respect to loss of
property, revenue, or pro t, or cost of removal, installation, or reinstallation. All implied
warranties, including implied warranties for merchantability and implied warranties for tness, are valid only
until the warranty expires. This Nortek Security & Control LLC Warranty is in lieu of all other
warranties express or implied.
All products returned for warranty service require a Return Product Authorization Number (RPA#). Contact
Returns at 1-855-546-3351 for an RPA# and other important details.
IMPORTANT !!! Radio controls provide a reliable communications link and ll an important need in portable
wireless signaling. However, there are some limitations which must be observed.
For U.S. installations only: The radios are required to comply with FCC Rules and Regulations as Part 15 devices. As
such, they have limited transmitter power and therefore limited range.
A receiver cannot respond to more than one transmitted signal at a time and may be blocked by radio signals that
occur on or near their operating frequencies, regardless of code settings.
Infrequently used radio links should be tested regularly to protect against undetected interference or fault.
A general knowledge of radio and its vagaries should be gained prior to acting as a wholesale distributor or dealer,
and these facts should be communicated to the ultimate users.
This device complies with part 15 of the FCC Rules and Industry Canada license exempt RSS standard(s).
Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2)
this device must accept any interference received, including interference that may cause undesired operation.
Le présent appareil est conforme aux CNR d’Industrie Canada applicables aux appareils radio exempts
de licence. L’exploitation est autorisée aux deux conditions suivantes : (1) l’appareil ne doit pas produire
de brouillage, et (2) l’utilisateur de l’appareil doit accepter tout brouillage radioélectrique subi, même si le
brouillage est susceptible d’en compromettre le fonctionnement.
Changes or modifications not expressly approved by the manufacturer could void the user’s authority to operate the equipment.
Les changements ou modi cations non approuvés expressément par la partie responsable de la conformité
pourrait annuler l’autorité de l’utilisateur à faire fonctionner l’équipement.
INSTRUCTIONS FOR
THE INSTALLER
ACTIVATE THE TRANSMITTER TO PERMANENTLY
ENABLE STATUS REPORTING
THE INDICATION THAT STATUS REPORTS ARE BEING
SENT IS THE SIGNAL INDICATOR BLINKING 3 TIMES
AFTER EACH TRANSMITTER ACTIVATION
PROGRAM THE TRANSMITTER INTO THE CONSOLE BY PLACING THE CONSOLE
INTO PROGRAM MODE, THEN SEND A SIGNAL TO THE CONSOLE FROM THE
TRANSMITTER. REFER TO THE UL LISTED PERS-2400B, PERS-3600B, O
R
PERS-4200 CONSOLE'S PROGRAMMING INSTRUCTIONS FOR DETAILS.
PLACE THE CONSOLE IN "TEST" MODE AND ACTIVATE THE
TRANSMITTER. VERIFY THAT THE SIGNAL WAS RECEIVED
AND THAT THE SYSTEM REPORTS AS REQUIRED.
IF THE SIGNAL INDICATOR
LIGHTS RED, THE BATTERY IS
LOW AND THE TRANSMITTER
NEEDS TO BE REPLACED
IMPORTANT ! TEST SYSTEM WEEKLY !
THE MANUFACTURER STRONGLY RECOMMENDS THAT YOU TEST THIS
EQUIPMENT FREQUENTLY. OVER A PERIOD OF TIME, NO ELECTRONIC
PRODUCT IS BETTER THAN THE INSPECTION AND MAINTENANCE IT
RECEIVES. THEREFORE, FROM THE TIME OF INSTALLATION, IT IS
ABSOLUTELY NECESSARY TO TEST THE SYSTEM AT LEAST ONCE A WEEK.
THE PENDANT IS DESIGNED
TO BE WORN WITH THE BUTTON
TOWARDS THE CHEST TO ALLOW
EASY THUMB PRESS ACTIVATION
AND PROVIDE THE WEARER WITH
A DISCRETELY STYLED DEVICE
ACTIVATION
BUTTON
2-COLOR
SIGNAL INDICATOR
LANYARD
BREAKAWAY
RELEASE
LENGTH
ADJUSTER
GREEN = BATTERY OK
RED = BATTERY LOW
INTERNAL
FALL DETECTION
ACCELEROMETER
PRINTER’S INSTRUCTIONS:
INSTR,INSTL,DXS-F1 - P/N: 10006532 X1 - INK: BLACK - MATERIAL: 20 LB. MEAD BOND - SIZE: 8.500” X 5.500” - TOLERANCE: ± .125” - FOLDING: 3 FOLD ALTERNATE, FINISH WITH LOGO SHOWING - SCALE: 1-1
USA & Canada (800) 421-1587 & (800) 392-0123
(760) 438-7000 - Toll Free FAX (800) 468-1340
www.nortekcontrol.com
BY
Copyright © 2015 Nortek Security & Control LLC 10006532 X1

Summary of content (1 pages)