CHAPTER 2 Preparation 2 In this chapter: Q Unpacking and Packing the Instrument Q Charging the Battery Pack Q Detaching and Re-Attaching the Battery Pack Q Setting Up the Tripod Q Centering Q Leveling Q Sighting Q Setting the Measurement Mode and Preparing the Target Q Measurement in Reflectorless mode Q Preparing the Reflector Sheet Q Setting Up the Prism Reflector Total Station Nivo Series Instruction Manual 7
Preparation Unpacking and Packing the Instrument Note – Handle the Nivo series instrument gently to protect it from shocks and excessive vibration. Unpacking To unpack the instrument, grip the carrying handle and gently remove the instrument from the carrying case. Packing To pack the instrument back into the carrying case, please refer to the figure on the right.
Preparation 2 C WARNING – Charge and use the rechargeable Lithium-ion battery only in strict accordance with the instructions. Charging or using the battery in unauthorized equipment can cause an explosion or fire, and can result in personal injury and/or equipment damage. To prevent injury or damage: – Do not charge or use the battery if it appears to be damaged or leaking. – Charge the Lithium-ion battery only in a product that is specified to charge it.
2 Preparation Power jack Charge indicator 0 This will read 5V, 4A Calibration indicator 0 Case “Top” edge Calibration button Calibration indicator 1 Charge indicator 1 Battery Slot 0 Battery Slot 1 Applying Power • Plug in the charger to the supplied AC adapter to turn the unit on. The power input must be 5 V with at least 4 A of current capability. Each battery may take up to 2 A while charging. Charging a battery 10 • Simply slide a battery into either battery slot to begin charging.
Preparation 2 Conditioning / calibrating a battery • Battery calibration is necessary about once every 6 months or more often if desired. Calibration insures the reported battery charge remaining is accurate. • Hold down the calibration button on the unit and then insert a battery while holding the calibration button to begin a battery calibration. Only the battery which was inserted while the button was depressed will begin calibration.
2 Preparation Detaching and Re-Attaching the Battery Pack Detaching the battery pack C CAUTION – Avoid touching the contacts on the battery pack. 1. If the instrument is turned on, press [PWR] to turn it off. 2. Turn the battery box release knob counterclockwise, open the battery box cover and remove the battery pack from the battery box. Inserting the battery pack Before inserting the battery pack, clear any dust or other foreign particles from the battery contacts. 2 3 1 C 12 1.
Preparation 2 Setting Up the Tripod C CAUTION – The tops of the tripod ferrules are very sharp. When handling or carrying the tripod, take care to avoid injuring yourself on the ferrules. 1. Open the tripod legs enough to for the instrument to be stable. 2. Locate the tripod directly over the station point. To check the tripod’s position, look through the center hole in the tripod head. 3. Firmly press the tripod ferrules into the ground. 4. Level the top surface of the tripod head. 5.
2 Preparation 5. Use the electronic level to level the instrument. For detailed instructions, see Leveling, page 15. 6. Look through the optical plummet to check that the image of the station point is still in the center of the reticle mark. 7. If the station point is off center, do one of the following: – If the station point is slightly off center, loosen the tripod mounting screw and then center the instrument on the tripod. Use only direct movement to center the instrument. Do not rotate it.
Preparation 2 Centering using a plumb bob 1. Set up the instrument on the tripod. For detailed instructions, see Setting Up the Tripod, page 13. 2. Hang the plumb line on the hook of the tripod mounting screw. 3. Adjust the length of the plumb line so that the tip of the plumb bob is at the height of the station point. 4. Loosen the tripod mounting screw slightly. 5.
2 Preparation Sighting When you sight the instrument, you aim the telescope at the target, bring the target image into focus, and align the image with the center crosshairs of the reticle. Center crosshairs To sight the instrument: 1. Adjust the diopter: a. C Aim the telescope at a blank area, such as the sky or a piece of paper. WARNING – Never look at the sun through the telescope. If you do, you may damage or lose your eyesight. b. 2.
Preparation 2 Setting the Measurement Mode and Preparing the Target The Nivo series has two measurement modes: Prism mode (Prism) and Reflectorless mode (N-Prism). These modes can be changed at any time by holding down the [MSR1] or [MSR2] key for one second. For more information, see Measurement settings, page 48. To set the measurement mode depending on the target you want to measure, see the following table.
2 Preparation • When measuring a short distance, incline the prism slightly so that the EDM can ignore unnecessary reflections on the prism surface, as shown below. Hold the prism securely in place and do not move while taking measurements. In Prism mode, in order to avoid false measurements on objects other than the prism or reflector-sheet, targets that are less reflective than the prism or reflector sheet are not measured. Even if you start a measurement, measured values are not displayed.
Preparation 2 Preparing the Reflector Sheet The reflector sheet can be used for measurements in Prism mode. Assemble the reflector sheet as shown below.
2 Preparation Setting Up the Prism Reflector 1. Assemble the prism reflector as shown below. Target plate for single prism Target pole Height adjustment adapter is not used. Tribrach adapter 15 Prism C Tribrach W30/W30b Tiltable single prism holder Triple prism holder Tripod 2. Adjust the height of the tribrach adaptor (see page 21). 3. If necessary, change the direction of the prism (see page 21). 4. Set the prism constant (see page 21). 5.
Preparation 2 Adjusting the height of the tribrach adapter The tribrach adapter has a height adjustment adapter. To use the prism reflector with a Nivo series instrument, remove the height adjustment adapter as shown in the Figure below. The height adjustment adapter will be used with other Nikon Total Stations. Height adjustment adapter Height adjustment adapter clamp screw Changing the direction of the prism The prism mounted on the tribrach adapter can be rotated to face in any direction.
2 Preparation Setting the position of the target plate If using a single prism, make sure that the target plate is aligned with the tribrach adapter and the prism. To set the position of the target plate: 22 1. Use the two set screws supplied to attach the target plate to the single prism holder. 2. Move the target plate within the screw holes until the apex of the wedge pattern is aligned with the vertical axis of the prism and the tribrach adapter.
CHAPTER 3 Getting Started 3 In this chapter: Q Turning the Instrument On and Off Q Changing Regional Configuration Pre-sets Q Display and Key Functions Q List Display Q Inputting Data Q Jobs Q Measuring Distances Total Station Nivo Series Instruction Manual 23
3 Getting Started Turning the Instrument On and Off Turning on the instrument 1. To turn on the instrument, press [PWR]. The startup screen appears. It shows the model name, current temperature, pressure, date, and time. The display automatically changes to the electronic level after 2 seconds. If you have entered your name or your company’s name in the Owner’s detail field, the text from this field appears on the start-up screen. To set the Owner’s detail field, go to MENU > Settings > Others.
Getting Started • You rotate the alidade • You tilt the telescope 3 Changing Regional Configuration Pre-sets To provide easier configuration for common regional settings, you can quickly configure the Nikon total station to a pre-set combination of default regional settings. The Regional Configuration screen appears only after the language configuration is complete, the instrument has rebooted. 1. Follow the steps in Language, page 113.
3 Getting Started Category Rec Others Setting Europe International United States Temp °C °C °F Press mm Hg mm Hg In Hg Store DB RAW&XYZ RAW&XYZ RAW&XYZ Data Rec Internal Internal Internal XYZ disp Fast Fast Fast 2nd Unit None None None Split ST No No No CD Input Owner’s Detail Blank Blank Blank The default regional configuration pre-set is “United States” settings. For more information, see Settings, page 108.
Getting Started Key Function 3 Details page 81 Records measured data, moves on to the next screen, or confirms and accepts the entered data in input mode. You have the option to record the measurement as a CP record instead of an SS record, if you hold this key down for one second in the Basic Measurement Screen (BMS). The instrument outputs the current measurement data (PT, HA, VA, and SD) on the COM port if you press this key in the BMS or in a Stakeout observation screen.
3 Getting Started Key Function Details Executes the function that is assigned to the [USR1] key. In numeric mode, enters 1. In alphanumeric mode, enters P, Q, R, S, or 1. Executes the function that is assigned to the [USR2] key. In numeric mode, enters 2. In alphanumeric mode, enters T, U, V, or 2. page 37 Opens a window where you can enter a code. The default code value is the last code entered. In numeric mode, enters 3. In alphanumeric mode, enters W, X, Y, Z, or 3.
Getting Started 3 EDM measurement status When you are taking measurements, the EDM measurement status shows the mode that is being used. When you display observation data, the EDM measurement status shows the mode that was used when the data was collected. Reflectorless mode Communication port status (This is only available when the optional Bluetooth is installed. See Optional Bluetooth function, page 129.
3 Getting Started LCD backlight, Laser pointer, Beep sound and Contrast adjustment [ENT] / [4] [ESC] To turn on/off the 1. LCD backlight, 2. laser pointer or 3. beep sound, and to do the 4. contrast adjustment, press the illumination key and open the switch set up window shown above. Holding down the illumination key for one second also turns on/off the LCD backlight.
Getting Started 3 [DSP] key Use the key to change the current display screen or to change display settings. Switching between display screens When several display screens are available, the DSP indicator appears at the top left of the screen, and the screen indicator (for example, 1/4) appears at the top right. To move to the next available screen, press [DSP]. For example, if the DSP2 screen is currently displayed, press [DSP] to move to the DSP3 screen. The screen indicator changes from 2/4 to 3/4.
3 Getting Started 4. To save your changes, press the Save softkey. Alternatively, highlight the last item for DSP3 and press [ENT]. The DSP screens show the items you have selected. Except for the (none) item, you cannot display the same item on more than one line of the same screen. The items displayed in the DSP1, DSP2, DSP3, and DSP4 screens are also used in the corresponding Stakeout screens (SO2, SO3, SO4, and SO5). You can also customize the displayed items in Stakeout.
Getting Started 3 Quick code measurement mode 1. To activate Quick code measurement mode, press [MODE] in the BMS. The PT field shows the default point name. 2. Press any numeric key ([0] through [9]) to start measuring and recording points. A list of the numeric keys and their assigned feature codes appears on the right side of the screen. For example, when you press [6], the code assigned to 6 is selected, and the instrument starts a measurement. 3.
3 Getting Started Qcode observations To enter the Quick code observation routine, press the Qcode softkey. In this function, you can use the ten numeric keys to both select a feature code and shoot a point. To change the measurement mode for the Quick code observation, press the Sett softkey. In Quick code measurement, the Rec mode can only be set to Confirm or ALL. [HOT] key The HOT key menu is available on any observation screen. To display the HOT key menu, press [HOT].
Getting Started 3 Selecting the target set A target set specifies settings for the target type, the prism constant, and height of target. When you change the selected target set, all three settings are changed. You can use this function to quickly switch between two types of target, such as a reflector sheet and a prism. You can prepare up to five target sets. Press [HOT] to display the HOT menu. Then either press [3], or select Target and press [ENT]. A list of the five target sets appears.
3 Getting Started Setting the default point name To change the default point name, press [HOT] to display the HOT menu. Then press [5], or select Default PT and press [ENT]. This function is available from any observation screen. Modify the default point name for the next record. Press [ENT] to confirm the new default point name. The new point name is appears as the default PT name on the input screen.
Getting Started 3 Laser luminance adjustment Laser luminance can be set to 4 levels. Press [Max.] to increase the luminance of the laser plummet. Press [Min.] to decrease the luminance of the laser plummet. Press[ESC] / [ENT] to return to the bubble indicator window. The current setting of leveling compensators is indicated by header characters (:, #, :, and #) after field labels (such as HA, VA, SD, and HD) in observation screens. For more information, see Header characters, page 32.
3 Getting Started If an item on the list has an arrow (->) beside it, this item is a menu. If you highlight a menu item and then press [ENT], a sub-menu appears. The first item on the sub-menu ends with the text [MENU]. If you select this item, the whole menu is assigned to the [USR] key. To assign a specific function from the sub-menu, press [^] or [v] to highlight the function. Then press [ENT].
Getting Started 3 When you select an option from this screen, the change is applied immediately, and the selected data type appears. Press [ESC] to return to the previous observation screen. To change the type of data that is assigned to [DAT], go to MENU > 1sec-Keys > [DAT]. For more information, see [DAT] key settings, page 133.
3 Getting Started If the last character of the previous point name is alphabetic, the default point name is the last point name. When the cursor is in a PT (point) field, there are several ways to specify a point, or input coordinates. Entering an existing point When you enter a known point name or number, the coordinates of that point are displayed briefly. A short beep sounds before the next screen appears or the next field is selected.
Getting Started 3 The input coordinates are used in the calculation. They are not saved in the database. Specifying a wildcard (*) If you include an asterisk (*) when you enter a point or code name, a list of points that match the entered text appears. Use [^] or [v] to move the cursor to the point that you want to use. Then press [ENT]. If the Page Up or Page Down icons are displayed, use [<] or [>] to page up or page down the list.
3 Getting Started When you move the cursor to a field, the current or default value appears in inverted text (this is the default “Replace All” input mode). Press [>] to change the input mode to Overwrite mode and highlight the first character. Press [<] to move the cursor the end of the string. Entering a point from the stack The point stack is a list of recently used points. To display the stack, press the Stack softkey when the cursor is in the PT field.
Getting Started 3 Entering a code The CD (Code) field always defaults to the last code used. You can change the selected code on the input point screen, or you can press [COD] in the BMS. For more information, see [COD] key, page 33. You can use numeric or alphanumeric names up to 16 characters long to identify codes. Entering a code directly To enter a code directly, press [MODE] to change the input mode to alphanumeric or numeric mode. Then use the keypad to enter the code.
3 Getting Started A layer has an arrow at the end of the code label. If you highlight a layer in the list and then press [ENT], the codes and layers in that layer are displayed. When you return to the input screen, the selected code is entered in the CD field. Items are shown in alphabetic order. You can also use the first character search to quickly find a code. For more information, see Advanced feature: Searching for a code by using the first character, page 44.
Getting Started 3 Entering values in feet and inches When either US Survey Feet (US-Ft) or International Feet (I-Ft) is selected as the distance unit, you can enter and display distances, HIs, HTs, and coordinate values either in decimal feet, or in feet and inches For more information, see Unit, page 111, and Others settings, page 112. To enter values in feet and inches in an input screen, enter the elements, separated by periods (.), in the following format: [.] [.] [.
3 Getting Started Creating a new job 1. Press [MENU] to open the MENU screen. 2. Press [1] to open the Job Manager. 3. Press the Creat softkey to open the Create Job screen. 4. Enter the job name. 5. Press the Sett softkey to check the job settings. You cannot change a job’s settings once you have created the job. 6. Press [ENT] in the last field of the Job Sett screen to create the new job.
Getting Started 3 Measuring Distances Sighting a prism reflector C WARNING – Never look at the sun through the telescope. If you do, you may damage or lose your eyesight. C WARNING – Precautions should be taken to ensure that persons do not look directly, with or without an optical instrument, into the beam. C WARNING – Laser beam path should be located well above or below eye level wherever practicable.
3 Getting Started If the average count is set to 0, measurements are taken continuously until you press [MSR1], [MSR2], or [ESC]. Each time a measurement is taken, the distance is updated. If the average count is set to a value from 1 to 99, the averaged distance appears after the last shot. The field name SD changes to SDx to indicate the averaged data. To change the height of target (HT), temperature, or pressure, press [HOT]. For more information, see [HOT] key, page 34.
Getting Started 3 The symbol then constantly runs from left to right over the prism constant in the display. When an N-prism measurement is taken, the icon appears in the status bar (above the battery icon). Incorrect Target settings may result in measurements outside the precision and intervals specified for the instrument. If a prism target is aimed in the N-prism mode, the warning message "Signal High! → Try Prism Mode" will be displayed because of the excessive light reflection.
3 50 Getting Started Total Station Nivo Series Instruction Manual
CHAPTER 4 Applications 4 In this chapter: Q HA Reset and Angle Operations Q Station Setup Q Stakeout Q Program Key Q Recording Measurement Data Q Measuring Offsets Total Station Nivo Series Instruction Manual 51
4 Applications HA Reset and Angle Operations To open the Angle menu, press [ANG] in the BMS. To select a command from this menu, either press the corresponding number key, or press [<] or [>] to highlight the command and then press [ENT]. Setting the horizontal angle to 0 To reset the horizontal angle to 0, press [1] or select 0-Set in the Angle menu. The display returns to the Basic Measurement Screen (BMS).
Applications 4 The averaged horizontal angle appears. This value is fixed until the process is finished or cancelled. HRx = HR¦ ÷ N HA = BSAz + HRx (normalized) HRx is not updated even if the instrument is moved. 6. Press [ENT] to store the foresight as a CP record. Check the PT, HT, and CD values. Then press [ENT] to record. In repeat angle measurement, the HA is replaced by HR¦. The number of repeat angles appears at the top of the screen (for example, N= 5).
4 Applications Setting up a station with known coordinates or azimuth 1. Press [1] or select Known in the Stn Setup menu. 2. Enter a point name or number in the ST field. 3. – If the input point number or name is an existing point, its coordinates are displayed and the cursor moves to the HI (Height of instrument) field. – If the point is new, a coordinate input screen appears. Enter the coordinates for the point. Press [ENT] after each field.
Applications 4 Sighting the backsight by entering coordinates Xb X0 Instrument height X Z Backsight point (Xb, Yb, Zb) Station point (Xi, Yi, Zi) Y0 Y Yb 1. To enter coordinates for the backsight point (BS), press [1] or select Coord in the Backsight screen. 2. Enter the point name. If the point exists in the job, its coordinates are shown. 3. If you intend to take a distance measurement to the BS, enter the height of target in the HT field. 4. Sight the BS.
4 Applications Sighting the backsight by entering the azimuth angle X X0 Instrument height Z Backsight point Azimuth Station point (Xi, Yi, Zi) Y0 1. To enter the azimuth angle to the backsight point, press [2] or select Angle in the Backsight screen. 2. If there is no point name for the BS, press [ENT] on the BS field. 3. In the AZ field, enter the azimuth angle to the BS point. If you press [ENT] without entering a value in the AZ field, the azimuth is automatically set to 0°00'00". 4.
Applications 4 Setting up a station using multiple point resection A resection sets up the station using angle/distance measurements to known points. X0 Y1 Height 2 Known point 1 (X1, Y1, Z1) Instrument height X1 X2 Height 1 X Z Y0 Known point 2 (X2, Y2, Z2) Station point (Xi, Yi, Zi) Y Y2 You can use a maximum of 10 points in a resection. Measurements can be distance and angle, or angle only. Calculation starts automatically when enough measurements are taken.
4 Applications 4. Sight PT1 and press [MSR1] or [MSR2]. 5. To proceed to the next point, press [ENT]. 6. Enter the second point (PT2) and its height of target. 7. Measure to PT2 and press [ENT]. When the instrument has enough data, it calculates the station (STN) coordinates. – If more than the minimum required data is available, a standard deviation screen appears. – To take measurements to strengthen geometry of the resection, press the Add softkey.
Applications 4 BS defaults to the first observed point. 11. To change the BS, press the Change softkey. 12. Select the BS point that you want to use and press [ENT]. 13. To finish the resection setup, move the cursor to the BS field and press [ENT]. The minimum data required for a resection is either three angle shots, or one angle shot and one distance shot. If you use a distance shot, the distance between the target points must be greater than the measured distance.
4 Applications Setting up the station quickly without coordinates The station point (ST) in this function defaults to a new point number. For the new point, MP (0, 0, 0) is stored as the coordinates. When the ST is manually changed to a known point name, the station is set up on the coordinates of the known point. 1. ST HI BS AZ To enter Quick Station setup, press [3] or select Quick in the Stn Setup menu.
Applications 4. 4 Sight the BM point and press [MSR1] or [MSR2]. The updated station coordinates are displayed. You can change the HI in this screen. 5. To record the updated STN, press [ENT]. When the HI is changed, the Z coordinate is updated before the station is recorded. You must complete a station setup before you use the Remote Benchmark function. Checking and resetting the backsight direction You must complete a station setup before you use the BS check function.
4 Applications Base XYZ function: Base XYZ does not store a ST record, so the BS Check cannot check the backsight when you enter a station using Base XYZ. To store raw data, use one of the other functions in the Stn Setup menu. This function does not store an ST record in the job. You can use this function without an open job. If there is an open job when you use this function, a CO record is stored to indicate that the instrument’s base coordinates have changed. 1.
Applications 3. 4. 4 Choose how you want to define a known line: – To define the line by entering P2 coordinates, press [1] or select By Coord. – To define the line by entering the azimuth, press [2] or select By Angle. If you select By Angle, the azimuth input screen appears. Enter the angle value and press [ENT]. A measurement screen appears. 5. Sight P2 and press [MSR1] or [MSR2] to take a measurement. Press [ENT].
4 Applications Sample records CO, Temperature:20C Pressure:1013hPa Prism:0 … ST,9005, ,265, ,1.2350,150.40300,150.40300 F1,265,1.6040,79.0010,90.30150,89.35260, F1,200,1.4590,50.2300,269.4035,93.50110, CO, P1-P2 HD=122.0350 VD=0.5600 Stakeout FiLL R OUT CUT IN L IN OUT Height of target R L FILL CUT To display the Stakeout menu, press [S-O].
Applications 4 Specifying the stakeout point by angle and distance 1. To display the input screen for the distance and angle to the target, press [1] or select HA-HD in the Stakeout menu. 2. Enter the values and press [ENT]. HD dVD HA Horizontal distance from station point to stakeout point Vertical distance from station point to stakeout point Horizontal angle to stakeout point If you press [ENT] without entering HA, the current HA is used. 3.
4 Applications Using [DSP] to switch between display screens Press [DSP] to switch between the Stakeout display screens. The following screens are available: S-O1 S-O2 S-O3 S-O4 dHA← R← OUT↑ CUT↑ HA VA SD HA VD HD HL V% HD S-O5 S-O6 S-O7 S-O8 X Y Z dX dY dZ rSD rVD rHD HD VD SD The S-O8 screen is only available if the secondary distance unit is set. For more information, see Others settings, page 112. Every time you press [DSP], the next screen appears.
Applications 4 If several points are found, they are displayed in a list. Use [^] or [v] to move up and down the list. Use [<] or [>] to move up or down one page. 3. Highlight a point in the list and press [ENT]. The delta angle and the distance to the target are shown. 4. Rotate the instrument until the dHA is close to 0°00'00". Press [MSR1] or [MSR2]. dHA HD 5. Difference in horizontal angle to the target point Distance to the target point Ask the rodman to adjust the target position.
4 Applications If you selected a point from the list, the display returns to the list, unless all points have been selected. Press [ESC] to return to the point input screen. Advanced feature: Specifying a stakeout list by range input 1. To input points by range, press the Fr/To softkey in the PT field. 2. Enter the start point (Fr) and the end point (To). The range between Fr and To must be less than 1001 points. If existing points are found between Fr and To, a point list appears.
Applications 3. 4 Enter the total stake number in the Span total field. The observation screen for the first stake (from the instrument) appears. 4. Sight the prism and press [MSR1] or [MSR2]. 5. Use [^] or [v] to change the guide point. You can calculate and guide up to double the number of the stakes. 6. To record the point as an SO record, press [ENT].
4 Applications 2. Enter the first point (P1) of the line. If you press [ENT] without entering a PT name, you can enter temporary coordinates which are not recorded in the job. Alternatively, press the MSR softkey to measure a point. 3. Enter the second point (P2) of the line. 4. Enter offsets to the line. Press [ENT] in a blank field to enter the value 0.0000.
Applications 4 Program Key To display the Programs menu, press [PRG]. Measuring distance and offset values along a specified line 1. Press [1] or select 2Pt RefLine in the Programs menu. 2. Enter the first point for the reference line. Alternatively (to enter the point by measuring), press the MsrPT softkey. Direct Measurement screen Press the MsrPT softkey to display a direct measurement screen. Sight the target and press [MSR1] or [MSR2]. The Record PT screen appears.
4 Applications 5. Sight the prism or reflective sheet and press [MSR1] or [MSR2]. Sta Horizontal distance from P1 to the measure point along the P1-P2 line Horizontal offset from the P1-P2 line to the measured point Vertical offset from the P1-P2 line to the measured poin O/S dZ Using [DSP] to switch between display screens Press [DSP] to switch between the Stakeout display screens.
Applications 4 Measuring distance and offset values on the arc-curve 1. Press [2] or select Arc RefLine in the Programs menu. 2. Enter the start of the curve point (P1) and the azimuth of its tangent line (AZ1). 3. To enter P1 by direct measurement, press the MSR softkey. 4. Choose a method to define the arc. P2 can be any point on the tangent line that is to exit the curve. 5. In the radius (Rad) field, enter a positive value for a clockwise curve.
4 Applications If the curve length (Len) is too large for a circle of the given radius, it is shortened. To ... Press ... switch between display screens [DSP] change HT [HOT] record points [ENT] Using [DSP] to switch between display screens Press [DSP] to switch between the Stakeout display screens. The following screens are available: ARC1 ARC2 Sta O/S dZ ARC3 X Y Z HA VA SD ARC4 HA VD HD ARC5 HD VD SD The ARC5 screen is only available if the secondary distance unit is set.
Applications 4 Remote distance measurement This function measures the horizontal distance, vertical distance, and slope distance between two points.
4 Applications 3. Sight the second point and press [MSR1] or [MSR2]. The distances between the first and second point are displayed. rSD rVD rHD Slope distance between two points Vertical distance between two points Horizontal distance between two points 4. To change display screens, press [DSP]. rAZ rV% rGD Azimuth from first point to second point Percentage of grade (rVD/rHD) × 100% Vertical grade (rHD/rVD) :1 5.
Applications 4 Difference between “1:Cont.” and “2:Radial” P3 P2 P3 P4 P2 P4 P1 P1 1:Cont. (The preceding two points are subject to calculation.) 2:Radial. (Calculations are made with reference to the first point.) Measuring remote elevation A prism is required only at the sighting (target) point Arbitrary point Sighting point Station point 1. To enter the Remote Elevation Measurement function, press [5] or select REM in the Programs menu. 2. Enter the height of target. 3.
4 Applications 4. Loosen the vertical clamp, and turn the telescope to aim at an arbitrary point. The difference in elevation (Vh) appears. You can use an REM measurement to update the height of target. Take a measurement to the prism, sight the bottom of the prism pole, and press [ENT]. Measuring distance and offset values on the vertical plane Baseline 1. To enter the 2-Pt Reference Plane function, press [6] or select V-Plane in the Programs menu. 2. Enter two points to define the plane.
Applications 5. 4 Enter the second point on the vertical plane. Press [ENT]. Once the plane is defined, the calculated Sta and dZ values are updated as you move the telescope. No distance measurement is required. Sta Horizontal distance from P1 to the target point along the baseline Vertical distance from P1 to the target point dZ Using [DSP] to switch between display screens Press [DSP] to switch between the plane display screens.
4 Applications Measuring distance and offset values on the slope Target P2 (X2,Y2,Z2) b a Stn (X,Y,Z) P3 (X3,Y3,Z3) P1 (X1,Y1,Z1) 1. To enter the 3-Pt Reference Plane function, press [7] or select S-Plane in the Programs menu. 2. Enter three points to define the slope plane. To enter the point by direct measurement, press the MSR softkey. If you press [ENT] in a blank field, an input screen for temporary coordinates appears. These coordinates will not be stored.
Applications 4 Once the plane is defined, the calculated a and b values are updated as you move the telescope. No distance measurement is required. a Distance between P1 and the point that is perpendicular to the target point along the P1P2 line Length of the perpendicular line from the target point to the P1-P2 line b Using [DSP] to switch between display screens Press [DSP] to switch between the plane display screens.
4 Applications To record the point, press [ENT] on the last field. When recording sideshots, stakeout records and control shots from the Rept function, you can choose to store raw data only, XYZ data only, or both. For more information, see Recording, page 112. If HA or VA is moved after you take a measurement but before you press [ENT], the angle recorded is the angle shown when [ENT] is pressed. In an angle-only record, SD is always recorded as 0.0000.
Applications 4 Sample output records through COM port When the Ext.Comm field is set to NIKON: TR PN: PT8 SD:000066626 HA:003856010 VA:008048500 HT:0000061757 (TR PN: point name SD HA VA HT; when ACK is returned, PN is incremented.) When the Ext.Comm field is set to SET: 0006662 0804806 0394324 97 (SD VA HA Chk-SUM) Measuring Offsets Measuring taped offsets 1. To enter the taped offset function, press [1] or select Tape in the Offset menu.
4 Applications You can enter any combination of taped offset distances to specify the point. 4. To go to the recording PT screen, press [ENT] in the last field. The calculated coordinates are shown. 5. Enter a PT (and CD) value. 6. Press [ENT] to record the point. Raw data is also recalculated, based on the taped offset value. Measuring angle offsets 1. To enter the angle offset function, press [2] or select Angle in the Offset menu.
Applications 4 Two-prism pole 1. To enter the two-prism pole function, press [3] or select 2Prism Pole in the Offset menu. 2. Sight the first prism and press [MSR1] or [MSR2]. 3. Sight the second prism and press [MSR1] or [MSR2]. 4. Enter the distance between the second prism and the target point. Alternatively, if you do not need QA information, you can leave the distance between the first and the second prism blank. 5. If you do enter a P1-P2 distance, the QA screen appears.
4 Applications Extending a line by horizontal angle offset 1. To enter the line extension (by HA) function, press [4] or select +Line by HA in the Offset menu. 2. Sight the first prism (or target) and press [MSR1] or [MSR2]. The display moves to the next screen. 3. Sight the second prism (or target) and press [MSR1] or [MSR2]. 4. Sight the alternative place on the same vertical line as the required target point. 5. To calculate the coordinates and the raw data of the target point, press [ENT].
Applications 4 The calculated point (TGT) is stored as a SS record. Measurements to the first and second target (P1 and P2) are stored as comment records (PT1 and PT2). The last record records the angle measurement to the ALT (vertically offset point from the actual target point). Entering a horizontal distance after an angle-only shot This function is useful when the instrument is very close to the point and it is difficult to take a measurement using the EDM. 1.
4 Applications Calculating a corner point 1. To enter the corner point function, press [6] or select Corner in the Offset menu. 2. Take a distance measurement to the first prism (or target) on the wall. Press [MSR1] or [MSR2]. 3. Sight a second point on the same wall with the first point measurement. Press [MSR1] or [MSR2]. 4. Sight the first point on the second wall. Press [MSR1] or [MSR2]. 5. If the two walls are at right angles, press the softkey to calculate the corner point by three points.
Applications 4 Sample records SS, 58, 0.0000, 48.3304, 169.19165, 82.02470, 10:52:37, FLOOR2 CO, PT1, 1.0080, 48.3020,169.19165, 83.58565 CO, PT2, 1.0080, 48.3155,128.54250, 85.42440 CO, O/S MSR:40 0.0000 0.0000 169.20370 87.02340 The calculated corner point is stored as an SS record. The next three or four comment records are measured points. For example: CO, Point name (fixed to PT1, PT2 etc.), HT, SD, HA, VA. Measuring circle offsets 1.
4 Applications The instrument calculates and records the center of the circle. It also calculates the coordinates of the center point and the radius of the circle. 5. To record the point, press [ENT] or the OK softkey. Sample records SS,71,1.5000,37.0518,32.08380,81.06510,11:51:48, CO, PT1, 0.0000, 0.0000,47.05350, 83.58560 CO, PT2, 0.0000, 0.0000, 29.53010, 83.58560 CO,O/S MSR:71 1.5555 36.5418 38.28360 81.06510 CO,Radius of circle 0.356 CO,Input +SD:0.