User Manual v.0 Shoulder Pacemaker Control 1.1.2.
KIT SPM01K2 The product configuration SPM01K2 kit contains: • 1 electrostimulator (REF: SPM02); • 1 receiver (REF: GTWY02); • Software Shoulder Pacemaker control 1.1.2.0; • accessories and consumable material as in the SPM01K1 kit with stand-alone operation.
SHOULDER PACEMAKER CONTROL SOFTWARE INSTALLATION No procedure is required for installing the Shoulder Pacemaker Control software on the PC. Just copy the program folder from the installation key to the target PC in a location of your preference and launch the start link called "SPC_v1.1.2.
PATIENT PREPARATION Refer to the instructions for use provided in the KIT regarding the correct and safe positioning of the electrodes on the patient.
ACCESS PROCEDURE When starting the software, the user will be asked to enter USERNAME and PASSWORD Once you have filled in the fields, click on LOGIN.. In case the compiled values are incorrect, an error message will be shown.
LAUNCH THE SHOULDER PACEMAKER CONTROL SOFTWARE 6
LAUNCH THE SHOULDER PACEMAKER CONTROL SOFTWARE Choose the desired language (Italian / English) by pressing the corresponding flag 7
START RECEIVER CONFIGURATION • Select the COM port if not recognized automatically • Press CONNECT 8
START RECEIVER CONFIGURATION To confirm the receiver configuration press OK 9
START DEVICE CONFIGURATION Enter the file path where you want to save the file automatically generated during the communication of the device with the receiver (it may contain error messages) If this value is not filled in, an error message appears that prevents you from continuing with the procedure. The path to the file is stored and proposed again at the next program start.
START DEVICE CONFIGURATION • Enter the ID number of the device provided in the kit • Press CONFIGURE 11
START DEVICE CONNECTION The successful configuration of the device is evidenced by the INACTIVE / IDLE message in the window at the bottom of the screen TURN ON THE DEVICE MANUALLY AND ENTER THE WIRELESS STIMULATION PROGRAM (W) Refer to the instructions for use provided in the KIT regarding the selection procedure of the W program 12
START DEVICE CONNECTION The successful connection of the device is shown by the CONFIGURED message in the window at the bottom of the screen 13
START DEVICE COMMUNICATION To start communication with the device, press START 14
START DEVICE COMMUNICATION The successful communication is demonstrated by the indication of the battery charge percentage and the appearance of the ACQUISITION IN PROGRESS message 15
SET OF STIMULATION PARAMETERS Enter the stimulation parameters: • Stimulation amplitude (V) • Pacing Pulse Duration (μs) • Pacing pulse frequency (Hz) 16
SET OF STIMULATION PARAMETERS Settable values for the parameters: • Stimulation amplitude: 0-150V • Stimulation pulse duration: 0-200μs • Stimulation pulse frequency: 1-100Hz 17
SET OF STIMULATION PARAMETERS Confirm the set of parameters by clicking on UPDATE PARAMETERS If the entered values are out of range, an error message will be shown that will prevent 18 you from continuing with the procedure
OTHER SETTINGS Press CLEAR if you want to clear the screen from previous messages 19
OTHER SETTINGS Press PAUSE to block communication between the device and the PC CAUTION: Pressing PAUSE DOES NOT stop stimulation, which continues with the last set of parameters set 20
STOP AND END OF THE SESSION To end the session, you must manually act on the stimulation device following the normal operating flow described in the user manual provided in the kit. It is necessary for the patient to manually operate the device button with a press to return to the menu selection mode and with a long press to turn off the device.
MANUFACTURER AND CONTACTS For more information, contact the NCS Lab support team: NCS Lab Srl, Via Pola Esterna 4/12, 41012 Carpi (MO)-Italy, +39 059 669813, info@ncs-company.