Operating Manual ALTAIR io360 Gas Detector Order No.: 10184747/00 Print Spec.: 10000005389 (E) CR 800000037948 MSAsafety.
WARNING! Read this manual carefully before using these devices. These devices will perform as intended only if used and maintained in accordance with the manufacturer’s instructions. Otherwise, the devices could fail to perform as designed and persons who rely on these devices for their safety could sustain serious personal injury or death. 1000 Cranberry Woods Drive Cranberry Township, PA 16066 USA Phone 1-800-MSA-2222 Fax 1-800-967-0398 For your local MSA contacts please go to our website www.MSAsafety.
Contents 1 Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 1.1 1.2 1.3 1.4 1.5 1.6 1.7 1.8 1.9 2 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 2.1 2.2 2.3 2.3.1 2.4 2.5 2.6 3 Safety io . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5 Deployment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33 5.1 5.1.1 5.1.2 5.2 5.2.1 5.2.2 5.2.3 5.3 5.3.1 6 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38 6.1 6.1.1 6.1.2 6.1.3 6.1.4 6.2 6.2.1 6.2.2 6.2.3 6.2.4 6.3 6.4 6.5 6.5.1 6.5.2 6.5.3 6.6 6.6.1 6.7 7 US Bump Test . . . . . . . . . . . . . . . . . . .
8.8 8.9 8.10 8.11 8.12 8.13 8.14 8.15 8.16 8.17 8.17.1 8.17.2 9 Decommissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72 9.1 9.1.1 9.1.2 9.1.3 9.1.4 9.2 10 Detector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72 Powering Down . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Safety Regulations 1 Safety Regulations 1.1 Correct Use The detector(s) and HUB(s) are for use by trained and qualified personnel. The detector is designed to detect exposure of a worksite to select combustible and select toxic gases and vapors as well as low level of oxygen - details provided in section 11 "Technical Data" section. It is imperative that this operating manual be read and observed when using the detector(s) and HUB(s).
Safety Regulations 1.3 FCC Regulations United States This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications.
Safety Regulations WARNING! TThis device complies with Part 15 of the FCC rules. Operation is subject to the following two conditions: 1) This detector may not cause harmful interference. 2) This detector must accept any interference received, including interference that may cause undesired operation. Failure to follow this warning can result in serious personal injury or death.
Safety Regulations A bump test must be performed more frequently than daily if any of the following occur: • The device is subjected to high levels of contaminants. • The device is subjected to rain or water spray.
Safety Regulations Observe Proper Time for Display Stabilizing Allow sufficient time for detector to display accurate reading. Response times vary based on the type of sensor being utilized. Observe Proper Battery Maintenance WARNING! Follow all instructions related to use and handling of batteries. Improper use and handling can result in battery leakage, battery venting, explosion or fire.
Safety Regulations Be Aware of Environmental Conditions A number of environmental factors may affect the sensor readings, including changes in pressure, humidity and temperature. Pressure and humidity changes also affect the amount of oxygen actually present in the atmosphere. Detector(s) exposed to environmental conditions must have a SensorShield installed.
Safety Regulations 1.5 Warranty Item Chassis and electronics All sensors unless otherwise specified Low energy combustible Sensor Oxygen Sensor HUB Warranty Four years Four years One year Two years Four years This warranty does not cover filters, fuses, batteries, etc. Certain other accessories not specifically listed here may have different warranty periods. This warranty is valid only if the product is maintained and used in accordance with Seller's instructions and/or recommendations.
Safety Regulations 1.9 Magnet disclaimers The detector contains a Neodymium magnet for mounting. MSA recommends the following:. Keep magnets away from • electronic devices • storage devices • identification cards • credit cards WARNING! Magnets can have an adverse effect on medical devices such as pacemakers, implantable cardio defribrilators (ICD) and other implantable medical devices. Persons with such devices should maintain sufficient distance between the medical device and the magnet.
Description 2 Description 2.1 Safety io Safety io, LLC (a subsidiary of MSA The Safety Company) developed the Grid industrial safety software platform so that, among other things, ALTAIR io360 users can proactively and easily manage and monitor their detectors as well as their associated networks and data logs. For more information, visit http://www.safetyio.com. 2.
Description 2.3 Detector The ALTAIR io360 Gas Detector will be referred to as the detector throughout the manual. The detector is an area monitor suitable for indoor or outdoor use, to monitor toxic and combustible gases, and oxygen levels of a work site. A variety of applications include fence lines and perimeters, confined spaces, and targeted areas where leaks could occur. Up to four gases, via three sensors can be monitored (one TwoTox Sensor can detect two toxic gases with a single sensor).
Description 2.3.1 Internal Features Fig. 3 1 Battery type warning 2 Batteries Fig. 4 1 US Fig.
Description Feature Sensors Horn LED Light Ring Display Button D-Ring Magnet Batteries RFID Reader Description Detect gas levels which trigger alarm if applicable. The horn provides an audible alarm of 90 dB at 30 cm. Communicate detector operations e.g. alarm, maintenance, connectivity status, calibration, etc. Displays gas reading, connectivity status, battery levels, operation instructions, sensor health. Various set up and operational functions. See section 6 "Operation".
Description 2.4 MSA HUB The MSA HUB will be referred to as the HUB throughout the manual. The HUB is an industrial gateway that enables mesh connectivity with ALTAIR io360 Gas Detectors. It acts as a gateway to Safety io Grid and/or mobile devices for monitoring and management. Fig.
Description 2.5 ALTAIR io360 Gas Detector Network The detector is wireless enabled and can form a network with other detectors when a HUB is deployed into the system. When a network is created, the user can enable shared alarms. Shared alarms are the communication of an alarm or evacuation throughout the network of detectors from the detector/device that initiated the alarm or evacuation.
Description 2.6 Deployment Options A key feature of the ALTAIR io360 gas detection solution is the easy and intuitive setup and deployment. Throughout this manual, the word "setup" will refer to workflows and tasks associated with preparing detectors and HUB(s) to function as necessary for individual needs. The word "deploy" or "deployment" will refer to the process of placing ALTAIR io360 Gas Detectors and MSA HUB(s) on the worksite, entering an operational state and to begin monitoring an area.
Hardware Setup 3 Hardware Setup 3.1 Getting Started Setup must be performed in a fresh air environment to ensure proper function. For ease and convenience, consider having the following: 3.
Hardware Setup 3.3 3.4 Powering the Detector (1) Hold detector button for approximately 3 seconds. (2) Release button. MSA HUB Setup The HUB must be disconnected from the power supply prior to assembly. The power supply may be activated only after the HUB is fully assembled and connected. Connection or disconnection of energized non-intrinsically-safe circuits is only permitted in the absence of a hazardous atmosphere.
Hardware Setup To ensure the ingress protection rating IP67: • all seals must be undamaged and correctly fitted • all screws of the housing / housing cover must be tightened with a torque of 13 in-lb +/- 1 in-lb (1.5 Nm +/- 0.1 Nm) • only use MSA supplied cables • all connectors must be hand tightened • If the ethernet connection is not is use, it must be sealed with MSA supplied blind plug. IP ratings do not imply that the equipment will detect gas during and after exposure to these conditions.
Hardware Setup WARNING! • Do not install in hazardous location. • Do not install with the HUB energized. • Do not touch the circuit board assembly. • Observe electrostatic procedures. Failure to follow these warnings can result in serious personal injury or death. (4) Optional: Use four screws to fasten the wall brackets to the housing. (5) Close housing lid by tightening the lid screws with a torque of 13 in-lb +/- 1 in-lb (1.5 Nm +/- 0.1 Nm). (6) Close joint. (7) Install grounding Clamp.
Hardware Setup 3.5.2 Ethernet Cable Ethernet cable is option and not included. WARNING! Ensure the HUB is not energized. Perform only in a non-hazardous environment. Failure to follow these warnings can result in serious personal injury or death. (1) Unscrew the Ethernet blind plug with a slotted screwdriver. (2) Open housing by opening the joints and removing the two lid screws as shown. (3) Remove the cover panel by removing the four screws as shown.
Hardware Setup (6) Tighten the nut to a torque of 3.5 in-lb +/- 1 in-lb (0.4 Nm +/- 0.1 Nm). (7) Connect the Ethernet connector to the main board. (8) Fasten the two screws to install the mounting flange to the housing. (9) Install the cover panel. Torque screws to 3.5 in-lbs +/- 1 in-lb (0.4 Nm +/- 0.1 Nm). (10) Torque the screws to 13 in-lb +/- 1 in-lb (1.5 Nm +/- 0.1 Nm). US (11) Close the housing lid. The HUB is assembled. 3.5.3 WiFi The MSA HUB supports only WPA2 personal networks.
Hardware Setup 3.5.4 HUB LED Description Fig. 7 Pattern MSA HUB LEDs Name Power Status On/Off Description On - HUB is on. Configuration On/Off On - Mobile device is connected to the HUB via Bluetooth. Network Connection On/Off On - Connection to Safety io Grid is established. MSA Radio Connection On/Off On - Mesh radio is on. Error Case Blinking simultaneously Any module is in error state.
Software Configuration Setup 4 Software Configuration Setup This section details the software configuration setup steps. If using pre-configured kits, skip the steps in this section. 4.1 Mobile Device Recommendations The mobile device must have Bluetooth capability and Android 5.0 or Apple iOS 8 or newer operating systems. The app can be downloaded and installed from the Google Play store or the App Store. 4.
Software Configuration Setup 4.2.1 Detector Profile If using the factory detector profile, skip this section. Editing the detector profile is optional. See section 11 "Technical Data" for Detector profile specifications. 4.3 Connecting the Detector to the App (1) Press the detector button to ensure that Bluetooth is communicating. (2) Within the app, search for the detector. (3) Within the app,select detector to be connected. (4) When prompted, acknowledge connection on detector.
Software Configuration Setup 4.5 Configuring the Detector (1) With the detector(s) connected to the app, follow the app prompts. (2) Through the app dashboard, search for nearby detectors to start the configuration process. The app starts a process to setup a detector profile, which can be used on any detector that is accessed by the app. The following parameters can be configured: 4.6 • Latching vs.
Software Configuration Setup 4.6.3 Internet Connectivity The HUB can be connected to the internet for Grid connectivity with Ethernet, Wifi, and/or cellular. The HUB will, in that order, preferrably use Ethernet, Wifi, or cellular for internet connectivity. 4.7 Modifying a Network Connect the app to the HUB to: • Add detector(s) to network: max. 32 for 915 Mhz and 16 for 868 Mhz • Remove detector(s) from network • Edit network name • Enable/disable shared alarms 4.7.
Software Configuration Setup 4.7.4 Detector Special States During detector configuration changes, the detector will temporarily enter a special state. If LEDs are enabled, the LEDs will indicate the special state. The special states are the following: • Fault • Warm up • Configuration/Calibration • Factory Reset WARNING! Detector is not monitoring gas during special states. Failure to follow this warning can result in serious personal injury or death. 4.
Deployment 5 Deployment Deployment is the process of placing detectors and HUB(s) throughout the work site and to be used in an operational state. With the exception of the AC connector for the HUB, all other components are able to operate in hazardous locations. The detector and HUB must be oriented upright and should avoid direct sunlight if possible. Detectors must have SensorShield and SunShield installed during deployment. Detectors must be oriented as shown below: 5.
Deployment A detector must be within the range of the HUB in order to be a part of the network. Under ideal line of sight conditions, the range is up to: • 915 MHz radio: • • HUB to detector: X meters (X feet) • Detector to detector: X meters (X feet) 868 MHz radio: • HUB to detector X meters (X feet) Detectors that are out of range of the HUB are able to make up to two hops through neighboring detectors to extend the range.
Deployment 5.2 Environmental Considerations 5.2.1 SunShield The SunShield is required on detectors to avoid exceeding maximum operating temperature and/or false alarms. The SunShield is securely held in place by the magnet on the back of the detector as shown below. With the SunShield installed, the D-ring and magnet can still be used for installation. Fig.
Deployment 5.2.2 HUB The HUB placement: • Ambient temperature range: -20°C to +60°C • Humidity Range: 10 to 95 % r.H. (non-condensing) Avoid exposure to weather to reduce excessive wear. Powering the HUB Option 1: (1) Connect Ethernet (optional). (2) Connect Power to HUB assembly. (3) Connect power to mains supply in non-hazardous location. Option 2: (1) Power the HUB in a non-hazardous location. (2) Move assembly into hazardous location.
Deployment 5.2.3 Detector Mounting WARNING! Do not use attachment methods containing spark producing materials including but not limited to magnesium. Failure to follow this warning can result in serious personal injury or death. Place the detector using the D-ring or the magnet. Attachment methods include but limited to carabiners, S-hooks, and/or cable ties. It is the user's responsibility to choose an attachment method suitable in a potentially explosive atmosphere. Fig.
Operation 6 Operation Observe proper time for gas reading to stabilize. Response times vary based on the type of sensor being utilized. 6.1 Detector 6.1.1 Interfaces The detector is operated by a combination of the button, the Calibration Cap, the app, and the Safety io Grid. 6.1.2 Button The single button on the face of the detector controls the following: Result Powers on Start State Off Action (1) Hold button approximately 3 seconds. Powers off On (2) (1) Release.
Operation 6.1.
Operation 6.1.4 Detector LED Definition Use the combination of the LEDs and detector display to evaluate the state of the detector.
Operation 6.2 Sensor Alarms If the gas concentration reaches or exceeds the alarm set point, the detector will alarm. There are two alarm levels: low alarm and high alarm. 6.2.1 Sensor Missing Alarm The detector alerts the Sensor Missing alarm if the detector detects that the sensor is not properly installed in the detector or is not functional. During power on, if a sensor is detected as missing, the detector display prompts the user of "Sensor Change?".
Operation 6.2.4 Exceed High and Low of Sensor When a gas reading exceeds the full scale range of the sensor, the detector displays can "XXX" or "+++" in place of the actual reading. When a gas reading falls below a defined under-range threshold, the detector displays " ---" in place of the actual reading. For combustible sensors, the sensor reading is in in locked alarm. The sensor reading is in over-range. The sensor reading is in under-range.
Operation 6.4 Monitoring Combustible Gases The detector has been optimized for a response to Methane. The detector can monitor these concentrations in ambient air: Gas Type Methane Propane Hydrogen Acetylene Response Factor 1 1.4 1.1 1.2 The detector displays the combustible gas concentration in % LEL or % Vol CH4 on the Measuring screen until another screen is selected or the detector is turned off. WARNING! A combustible gas reading of "100" or "5" indicates the atmosphere is above 100 % LEL or 5.
Operation 6.
Calibration and Bump 7 Calibration and Bump WARNING! Detectors will not be performing hazardous gas sensing operations while calibrating. Failure to follow this warning can result in serious personal injury or death. The detector must equilibrate to room temperature before bump or calibration. Calibration Certification All applicable inspections, testing, and calibrations were performed using NIST traceable equipment, where available, in accordance with MSA's ISO 9001 Certified Quality System. 7.
Calibration and Bump CSA requires (per 22.2 NO. 152) that combustible sensor sensitivity be tested before each day's use on a known concentration of methane equivalent to 25 to 50 % of full scale concentration. ACCURACY MUST BE WITHIN 0 to +20 % OF ACTUAL. Correct accuracy by performing the calibration procedure described in section 7.2.1 "Calibration". 7.1.4 Equipment See section 13.1 "Detector Spare Parts and Accessories" for ordering information of these components: 7.
Calibration and Bump (4) Tap the Bump RFID tag in Calibration Cap to the button. Once prompted, a detector will display 'Install Cap', at which time Install the calibration cap. (5) Snap the Calibration Cap onto the detector. (6) Insert tab on Calibration Cap into slots on detector (3 places). (7) Press Calibration Cap until it seats onto detector. a) If needed, connect one end of the tubing to the Calibration Cap. b) If needed, connect other end of tubing to the cylinder regulator.
Calibration and Bump After bump completes, the detector will: Pass: • Beep • Display green LEDs • Display check mark and “Bump Pass“ text Fail: • Beep • Display red LEDs • Display “X“ and “Bump Fail“ text If the detector fails, the detector will automatically attempt a second bump test. (10) Turn off gas. WARNING! Do not continue operation of a detector that is not able to pass a bump test. Failure to follow this warning can result in serious personal injury or death. (11) Remove Calibration Cap.
Calibration and Bump (13) Install SunShield on detector. 7.2.1 Calibration Calibration must be performed using a flow regulator with a flow rate set to 0.25 liters per minute. Under normal circumstances MSA recommends calibration before each deployment. Please check your national legislation. Calibrate at room temperature. Allow detector(s) to stabilize for 30 minutes before calibration. Stated calibration performance is for room temperature. Equipment For Part Numbers, see section 13.
Calibration and Bump Calibration Steps (1) Remove SunShield. (2) Remove SensorShield with the detector exposed to fresh air. NOTE: If uncertain, use a bottle that contains fresh air. (3) Press button on the detector to enter interactive mode. (4) Tap the Cal RFID tag in Calibration Cap to the button.
Calibration and Bump Once the Cal RFID tag is read, a Zero calibration will initiate Screen will display 'Zero Test' and LEDs will flash blue. NOTE: Do not install the Calibration Cap during Zero Calibration! After Zero calibration completes, the detector will: Pass: Check mark and 'Install Cap' text are displayed. Fail: • Beep • Display red LEDs • Display 'X' and 'Zero Fail' text During zero calibration the O2 sensor is also span calibrated to 20.
Calibration and Bump (4) Open valve on regulator to allow gas to flow to the detector. The duration of the application of the calibration gas is controlled by the detector. The detector will indicate that calibration span has successfully passed and that the gas can be removed.
Calibration and Bump (7) Install SensorShield on the detector. (8) Install SensorShield. WARNING! A SensorShield must be installed on detector(s) in all situations where the detector(s) may be subject to conditions that could lead to blockage of the sensor path (including, but not limited to, rain, water spray, dusty or dirty environments, snow and ice storms or icing conditions). Failure to do so may result in erroneous gas readings or inability to detect a gas event.
Maintenance and Service 8 Maintenance and Service WARNING! Do not use silicone-type lubricants in assembling the detector and do not allow silicone vapors to be drawn into the flow system while in operation. Silicone can desensitize the combustible gas sensor, thereby giving erroneously low readings. Use only genuine MSA replacement parts when performing any maintenance procedures on the detector. Substitution of components can seriously impair performance.
Maintenance and Service 8.3 Detector If an error occurs during operation of the detector, use the displayed error codes to determine appropriate next steps. While detector case is open, do not touch any internal components with metallic /conductive objects or tools. Damage to the detector can occur 8.3.1 Routine Cleaning for Detector Clean the exterior of the detector regularly using only a damp cloth. Do not use cleaning agents, as many contain silicone, which will damage the combustible sensor.
Maintenance and Service 8.8 Batteries Observe Proper Battery Maintenance WARNING! Follow all instructions related to use and handling of batteries. Improper use and handling can result in battery leakage, battery venting, explosion or fire. Use only Tadiran model number TL-5920 and Tadiran model SL-2770. MSA Part number 10201430. Use of another battery may present a risk of fire or explosion. When replacing batteries of a detector, do not mix old and new batteries.
Maintenance and Service (1) Power off the detector. (2) Press the button for approximately 3 seconds and release. (3) Remove the housing screws (6 places) with a 5 mm hex key. (4) Separate the front and back housing. a) If needed, gently insert the slotted screwdriver into the slots shown. Do not exceed more than ¼" of penetration. b) Do not pry where the shield attaches. c) Rotate screwdriver apart slightly at each appropriate pry location. US (5) Gently pull batteries out of the cavities (3x).
Maintenance and Service (6) Install new batteries into the cavity oriented such that the battery matches the orientation marked in the cavity. (7) Assemble front and back housing. Install screws until resistance starts. (8) Finish tightening screws in the pattern as shown below. Tighten screws to a torque of 10 in-lb +/- 1 in-lb (1.1 Nm +/- 0.1 Nm). (9) Dispose of used batteries promptly. Observe local disposal regulations. US Do not disassemble and do not dispose of in fire.
Maintenance and Service 8.9 Gasket Replacement Observe all maintenance precautions. The gasket shall not be replaced in a hazardous area. (1) Power off the detector. Press the button for approximately 3 seconds and release. (2) Remove the housing screws (6 places) with a 5mm hex key. (3) Separate the front and back housing. a) If needed, gently insert the slotted screwdriver into the slots shown. Do not exceed more than ¼" of penetration. b) Do not pry where the shield attaches.
Maintenance and Service (5) Remove battery holder retaining screws. (6) Separate the front housing from the battery holder. (7) Remove old gasket. (8) Install new gasket by orienting the open ring is positioned around the opening for the horn. Always replace the gasket when servicing the horn or sensors. (9) Assembly front housing to battery holder. (10) Install screws to torque of 3 in-lb +/- 0.5 in-lb (0.3 Nm +/- 0.1 Nm).
Maintenance and Service (12) Assembly front and back housing. Install screws until resistance starts. (13) Finish tightening screws in the pattern as shown below. Tighten screws to a torque of 10 in-lb +/- 1 in-lb (1.1 Nm +/- 0.1 Nm). (14) Dispose of gasket following local regulation. Do not reuse the gasket. WARNING! The detector must pass calibration before continuing use. Failure to follow this warning can result in serious personal injury or death.
Maintenance and Service 8.10 Sensors The only routine maintenance item is the sensing element itself, which has a limited lifetime. The detector will indicate when the sensor is near end of life through the detector display When the Sensor Life & Health status is "Warning", there is a limited amount of time to replace the sensor before it will no longer function. When a sensor is no longer capable of sensing it will go into maintenance and LEDs will flash yellow.
Maintenance and Service Physical Sensor Configurations WARNING! Place Sensor plug into sensor slot if removing a sensor and not re-installing a sensor. Failure to follow this warning can result in serious personal injury or death. In order to detect combustible gas, both the low power combustible sensor and XCell Combustible sensor must be installed and enabled. 8.
Maintenance and Service (2) Remove the housing screws (6 places) with a 5mm hex key. (3) Separate the front and back housing. a) If needed, gently insert the slotted screwdriver into the slots shown. Do not exceed more than ¼" of penetration. b) Do not pry where the shield attaches. c) Rotate screwdriver apart slightly at each appropriate pry location. (4) Gently remove batteries. (5) Remove battery holder retaining screws. (6) Separate front housing from battery holder.
Maintenance and Service (9) Carefully align the new sensor contact pins with the sockets on the printed circuit board and pressing it firmly in place. a) Ensure tab on sensor aligns with groove at top of holder - applies to XCell sensors only. b) Insert the Xcell combustible sensor as shown. c) Insert the O2 sensor as shown. d) Insert the toxic sensor as shown. e) Insert the low power combustible sensor as shown. WARNING! Place Sensor plug into sensor slot if removing and not re-installing a sensor.
Maintenance and Service (14) Finish tightening screws in the pattern as shown. Tighten screws to a torque of 10 in-lb +/- 1 in-lb (1.1 Nm +/- 0.1 Nm). (15) Install SensorShield. (16) Install SunShield. When replacing O2 sensor allow sufficient time for warm up before use. 8.14 Horn Replacement Observe all maintenance precautions. The horn shall not be replaced in a hazardous area. (1) Power off the detector, Press the button for approximately 3 seconds and release.
Maintenance and Service (3) Separate the front and back housing. a) If needed, gently insert the slottted screwdriver into the slots shown. Do not exceed more than ¼" of penetration. b) Do not pry where the shield attaches. c) Rotate screwdriver apart slightly at each appropriate pry location. (4) Remove battery holder retaining screws. (5) Separate front housing from battery holder. (6) Remove old sensor gasket. Dispose of sensor gasket following local regulations. (7) Gently remove the horn.
Maintenance and Service (10) Install screws with a torque of 3 in-lb +/- 0.5 in-lb (0.3Nm +/- 0.1Nm). (11) Install new batteries into the cavity oriented such that the battery matches the orientation marked in the cavity. See reference and warnings in section 8.8 "Batteries". (12) Assembly front and back housing. Install screws until resistance starts. US (13) Finish tightening screws in the pattern as shown. Tighten screws to a torque of 10 in-lb +/- 1 in-lb (1.1 Nm +/- 0.
Maintenance and Service 8.15 Scope of Delivery Pack the detector in its original shipping container with suitable padding. If the original container is unavailable, an equivalent container may be substituted. WARNING! Do not use silicone-type lubricants in assembling the detector and do not allow silicone vapors to be drawn into the flow system while in operation. Silicone can desensitize the combustible gas sensor, thereby giving erroneously low readings.
Maintenance and Service 8.17.2 Factory Reset If the device access password has been lost and no further access is possible, a factory reset can be performed to reset all values to their defaults. (1) Power off the HUB. (2) Unmount the HUB. a) If necessary, move to a non hazardous location. (3) Open housing by opening the right joints and unscrew the lid screws using a Torx T20. (4) Power the device. (5) Use a thin object non-conductive (e.g. a pencil) to push the reset button for 10 seconds.
Maintenance and Service (6) Power off the HUB. (7) Close housing lid by tightening the lid screws with a torque of 13 in-lb +/- 1 in-lb (1.5 Nm. +/- 0.1 Nm). (8) Close joint. (9) Install grounding Clamp. a) Remove grounding clamp. b) Insert grounding cable and re-tighten grounding clamp with a torque of 17 in-lb +/- 1 in-lb (2 Nm +/- 0.1 Nm). (10) Mount HUB to its final location. Use four screws to fasten the wall brackets to the surface and apply a torque of 45 in-lb +/- 5 in-lb (5 Nm +/- 1 Nm).
Decommissioning 9 Decommissioning 9.1 Detector 9.1.1 Powering Down (1) Hold the button for approximately 3 seconds. The detector turns off. (2) Follow the steps in section 8.8 "Batteries" to remove and store batteries. (3) Wrap battery terminals with electrical tape for storage. (4) Assembly front and back housings together. 9.1.
Decommissioning 9.2 HUB WARNING! • Do not connect or disconnect power and/or Ethernet connections to or from the HUB in a hazardous location. Connect or disconnect only in a non-hazardous area. • Do not remove, replace or disconnect the HUB while the circuit is live. • Do not open the HUB when the power supply is energized. • Do not substitute components. Substitution of components may impair intrinsic safety. • Do not open the HUB in a hazardous location.
Decommissioning (8) Tighten screws and close hinge. (9) Remove antenna. The HUB is decommissioned. 9.2.1 Storage Conditions When not in use, store the HUB in a safe, dry place between 18 °C (64 °F) and 30 °C (86 °F). The relative humidity range of 15 % - 90 % relative humidity, non-condensing environment. US For extended storage, remove batteries and store separately in a safe, dry place.
Troubleshooting 10 Troubleshooting 10.1 Detector The health indicator on the detector display of the combustible and low power combustible sensor are combined on the combustible screen. Use the app to determine which sensor has the health warning. Record the fault number and all other information on the detector display during fault to provide guidance to MSA customer support if needed.
Technical Data 11 Technical Data 11.1 Detector Weight Dimensions (L x W x H) Alarms Display Battery type 27.2 oz / 771 g (detector with battery and clip) 5.2 x 5.6 x 2.7 inches/132 x 142 x 69 mm Visible LEDs and 90 dB at 30 cm horn LCD display Primary Non-rechargeable Lithium battery. Only replace with Tadiran model number TL-5920 and Tadiran model SL-2770 Detector run time 50-60 days at 25 °C (77 °F) combustible sensor installed 80-85 days at 25 °C (77 °F) toxic/O2 only.
Technical Data 11.2 MSA HUB Dimensions (L x W x H) Temperature range Humidity Ingress protection Power dissipation Power Supply Ethernet Bluetooth Wifi Cellular Radio SIM Card Slot 5.2 x 5.6 x 2.7 inches/132 x 142 x 69 mm -4°F -to 140°F (-20°C to +60°C) for intrinsic safety Storage: 64 °F to 86 °F (18°C to 30°C) 10 to 90 % r.H.
Technical Data 11.3 Factory Set Alarm Thresholds and Set Points Check the monitor or calibration certificate for exact alarm levels as they vary depending on national or corporate regulations. Sensor Combustible (%LEL) O2 (% vol) H2S (ppm) CO (ppm) Low Alarm 10 19.5 5 25 Sensor Combustible (% LEL) O2 (% vol) H2S (ppm) CO (ppm) High Alarm 20 23.0 10 100 Default Calibration Set point 50 15.
Technical Data 11.4 Gas Performance Specification 11.4.1 Combustible Gas Range Resolution Response Time Reproducibility 0 to 100 % LEL or 0 to 5 % vol 1 % LEL or 0.05 % vol 90 % of final reading in less than or equal to 15 sec (methane) (normal temperature range) 3 % LEL, 0 % to 50 % LEL reading or 0.15% vol, 0.00% to 2.50% vol (normal temperature range) 5 % LEL, 50 % to 100 % LEL reading or 0.25% vol, 2.50% to 5.00% vol (normal temperature range) 5 % LEL, 0 % to 50 % LEL reading or 0.25% vol, 0.
Technical Data 11.4.4 Carbon Monoxide Cross Sensitivity Test Gas Applied Concentration (PPM) Applied Hydrogen Sulfide (H2S) Carbon Monoxide (CO) Nitric Oxide (NO) Nitrogen Dioxide (NO2) Chlorine (Cl2) Hydrogen Cyanide (HCN) Ammonia (NH3) Toluene Isopropanol Hydrogen (H2) 40 100 50 11 10 30 25 53 100 100 CO Channel % Cross-sensitivity 0 100 84 0 0 -5 0 0 -8 48 Data is presented as the indicated output in PPM which will result from the application of the test gas. 11.4.
Patents 12 Patents • US 8826721 • US 7959777 • US 8702935 • US 8790501 • Additional patents pending US ALTAIR io360 81
Ordering Information 13 Ordering Information To order parts, go to msasafety.com 13.
Ordering Information WARNING! Repair or alteration of the detector or HUB beyond the procedures described in this manual or by anyone other than a person authorized by MSA, could cause the unit to fail to perform properly. Use only genuine MSA replacement parts when performing any maintenance procedures described in this manual. Substitution of components can seriously impair performance of the unit, alter intrinsic safety characteristics or void agency approvals.
For local MSA contacts, please visit us at MSAsafety.