TABLE OF CONTENT I. Terminology................................................................................................................................................................................ 3 II. Used Icons Significance ......................................................................................................................................................... 3 III. Product diagram .............................................................................................
XIV. Regulatory information ..................................................................................................................................................... 18 3. Notice to Users: ................................................................................................................................................................. 18 4. FCC Statement: .............................................................................................................................
III.
IV. SAFETY RULES Warning Please read this user’s guide very carefully before use. Advise all recommendations to end-user and keep informed of all risks endured using the system. CTRLER/CHARGER must be plugged using equipment in compliance with INTO standard 60950-1 via USB2 port with the provided cable. In order to fulfill safety requirements for electro-medical systems and equipments, power must be supplied by a transformer in compliance with the EN standard 60950.
It is highly recommended not to connect any peripherals (modem, printers…) other than the MOTIONCONTROLLER on the computer during the use of the system. The MPOD has a Li Ion battery and therefore must be recycled according to current standards. It must be returned to the manufacturer. V. INTRODUCTION Thank you for purchasing our motion sensing product MPOD. This user manual was written to familiarize you with the use of this system. Read it carefully in order to rightfully use the device.
Photo of MotionPod™ Side view of MPOD scheme Mechanical fixing base FIGURE 3 : MOTIONPOD™ On the top of MPOD you can see the electronic components with an imprinted name and logo of the manufacturer (Movea,), the brand name of the product (MotionPod). The sensor button indicates the exact orientation of the MPOD when positioning it properly on the body. Please pay attention to these elements as they are useful to establish the MPOD correctly in the application that is provided with your system.
Connected on the PC via USB port cable (provided), it allows: • Data reception transmitted from the MPOD by radio frequency (2.45 GHz Bluetooth Band) • Charging of the MPOD FIGURE 5 : MOTIONCONTROLLER™ The CTRLER features a wireless link between the MPOD and the computer where you run your applications. The data goes to the computer via USB cable. 3. MPOD CHARGER FIGURE 6: MPOD CHARGER The CHARGER and the CTRLER have the same casing.
VII. USAGE 1. RECHARGING YOUR MOTIONPOD™ The CTRLER must be connected to a computer in compliance with IEC/EN 60950-1 standard. In order to fulfill safety, power must be supplied by a transformer in compliance with EN standard 60950. Only the MPOD is in contact with the patient environment. First connect the CTRLER to the USB port of your computer using the provided cable. Once the CTRLER is connected, the light indicator “pwr” goes on and gives a green continuous light (see Figure 2).
FIGURE 7 : THE MPODFIXED ON THE CTRLER, READY FOR RECHARGE While charging, an orange light indicator located at the back of the MPOD switches on. It switches off as soon as the recharging process is complete. Theoretically, charging duration is 3 hours. When the MPOD is recharged after a very long period of no use, it is possible that the orange light indicator takes a few minutes before switching on.
To fix MPOD on the body patient, use solutions provided by the product supplier. Be sure not to tighten up excessively. The MPOD is provided with a mechanical interface to attach bracelets, belts or straps. This is the only interface that the user must use to adapt non-magnetic bracelets provided by the manufacturer. The MPOD should not be in contact with any other material that can cause measurement distortion. MPOD must however remain attached on the body.
o o o o o o Keys : 15 cm While using the MPOD for measurement or calibration (see further below), see to it that there isn’t any object likely to disturb measurements (ferrous objects, watches, mobile phone…) MPOD should be used far from motorized chairs or beds. Only small magnets are accepted in the same room (i.e. whiteboard magnet), but must be kept far from the MPOD (>50 cm), as they might cause a serious decrease in the systems accuracy.
Temperature Shock resistance Magnetic field survivability Maximum rate of magnetic field tolerated before misscalibration Between -20°C à 35°C (For 6months maximum) 45°C (For 1 month maximum) 4600 g for a time of 0,5 ms 10 000 Gauss 20 Gauss MPOD transmits data to the CTRLER via radio signal. The transmitted signal is in the Bluetooth band. MPOD is tested in CEM. However, the presence of other radio transmitting devices can lead to interferences with MPOD™.
. Be sure to maintain the MPOD in the same position, in a sphere of 5cm of diameter. Carry out slow movements without acceleration, nor shock X. CLEANING 1. CLEANING THE MOTIONCONTROLLER™ The CTRLER/CHARGER are made of ABS plastic. Frequent disinfection or cleaning is not needed because they’re not in contact with the body. Disconnect the CTRLER/CHARGER before cleaning. Use a soft and dry cloth. 2.
Liquid container Wet cleaning napkins FIGURE 8 : WELL ADAPTED DISINFECTION PRODUCTS HEXANIOS G+R Before using the MPOD, it is recommended to avoid skin contact by applying biocompatible CE certified bandage (example: Urgoderm band, see figure 7) FIGURE 9 BIOCOMPATIBLE BANDAGES TO PUT BETWEEN THE SKIN AND MPOD XI. SIGNS AND LABELS 1. MOTIONPOD™ The MPOD is the part of the system in contact with the subject.
2. MOTIONCONTROLLER™ The label of the CTRLER figures at the bottom of the case Dimension~ 30 x 33 mm 3.
XII. MAINTENANCE The MPOD system does not require any particular maintenance other than cleaning after each use. Manufacturer calibration is carried out at the time of manufacture of the MPOD™. Follow instructions on section VIII for re-calibration. The MPOD should not be opened. Battery replacement operation should only be carried out by supplier’s qualified personnel. XIII.
XIV. REGULATORY INFORMATION 3. NOTICE TO USERS: This equipment has been tested and found to comply with the limits for Class B digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications.