Cleaning the WAM Transmitter and ECG Cable
1. Remove cables and remove the power source from device before cleaning.
2. For general cleaning, use a soft, lint-free cloth lightly moistened with a mild soap and water solution. Wipe
and air dry.
3. For disinfecting, wipe exterior with a soft, lint-free cloth using a solution of Sodium Hypochlorite (10%
household bleach and water solution) minimum 1:500 dilution (minimum 100 ppm free chlorine) and
maximum 1:10 dilution as recommended by the APIC Guidelines for Selection and Use of Disinfectants.
Alternatively, the following disinfectant wipes may be used:
Manufacturer: PDI
Part: SANI-CLOTH® HB
Order No.: Q85484
4. Use caution with excess liquid as contact with metal parts may cause corrosion.
5. Do not immerse cable ends or lead wires; immersion can cause metal corrosion.
6. Do not use excessive drying techniques such as forced heat.
WARNING: Prevent liquid from penetrating the device and do not attempt to clean/disinfect the device or
patient cables by submerging into a liquid, autoclaving, or steam cleaning. Never expose cables to strong
ultra-violet radiation. Do not sterilize the device or ECG cable with Ethylene Oxide (EtO) gas.
Periodic Maintenance
Check the WAM, and the lead wires before each use to ensure they are not damaged or broken.
Disposal of Waste Materials
The WAM uses one AA battery and disposable monitoring electrodes. Disposal must be in accordance with the
following procedures:
Battery: applicable disposal or recycling standards
Electrodes: normal waste