USER MANUAL The WRM-Installation manual RDS-32™ SURVEY METER
NOTE This document and the information herewith are copyrighted material as mentioned in the Copyrights Act. No part of this document may be copied without a written authorization from the manufacturer. The manufacturer withholds the right to make changes in the contents of this document without prior notice. If you have any comments or suggestions for additions concerning either the products or this document, please submit your feedback as instructed at the end of the manual.
General FCC Compliance Statement This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: 1. This device may not cause harmful interference, and 2. This device must accept any interference received, including interference that may cause undesired operation. Modifications: Any modifications made to this device that are not approved by Mirion Technologies Oy may void the authority granted to the user by the FCC to operate this equipment.
General 1 CONTENTS 2. General ........................................................................................... 3 3. RDS-32 meter preparation ............................................................. 3 RF-module preparation ...................................................................................... 3 4. Final assembly and the functionality testing ............................... 6 The RF-module installation ..........................................................................
General 2. GENERAL This document describes the installation of the 915 MHz WRM-radio (XB Pro 900 HP) in the RDS-32 -meter. 3. RDS-32 METER PREPARATION RF-MODULE PREPARATION The preparation is as follows: NOTE! The antenna is fragile and must not be bent. The RF-module contacts (1-10) are cut to a length of 2,5 mm as shown in Figure 1. Figure 1. The cutting and trimming of the contacts of the RF-module contacts. 1.
RDS-32 me Figure 2. The positioning of the dual-sided tape. 2. Attach the super-conductor 21638023 with the paper side upwards to the module as shown in Figure 3.The soldering can be trimmed. Figure 3. The attaching of the super-cpnductor to the RF-module. 3. The Flex print JRG 20566322 (version B) is soldered askew to the cut connectors of the RF-module as shown in Figure 4.
RDS-32 me Figure 4. The soldering of the flex print to the RF-module. 4. The flex print is soldered to the super-conductor as shown in Figure 4. 5. The protectice paper is removed from the super-conductor. The Mirion FCC-label is attached as shown in Figure 5.
Final ass 4. FINAL ASSEMBLY AND THE FUNCTIONALITY TESTING THE RF-MODULE INSTALLATION The RF-module is installed on top of the KRO-board on the HV-convertor’s interference shield as follows: 1. The interference shield’s surfaces are cleaned with rubbing alcohol to remove any grease residues. 2. The askew edge of the interference shield is covered with two layers of polyethene adhesive tape 21220033 as shown in Figure 6. Figure 6.
Final ass 4. The RF-module is attached on top of the interference shield as shown in Figure 8. When viewed from the side, the contacts of the RF-module must not come in contact with the interference shield’s slanted sides (Figure 9.). The RF-module must be pressed for a while to ensure that the adhesive tape has attached itself properly. The antenna is positioned as shown in the pictures below. Antenna type is: 24AWG Full length: 75mm, Gain: 1.
Final ass THE INSTALLATION OF THE KRO-BOARD IN THE RDS-32 CASING The installation of the component board in the casing is performed as in the assembly instruction 2096 7154. The proper positioning of the RF-module antenna must be ensured as in Figure 10. Figure 10. The positioning of the RF-module antenna within the RDS-32 casing.
Calibrati 5. CALIBRATION AND TESTING CALIBRATION AND TESTING The calibration and testing are carried out as in the instruction 2096 7154. THE PRODUCTION CLIMATE TESTS The climate tests are performed according to instruction 2096 7177. THE TESTING OF THE RF-MODULE 5.3.1 THE RF-MODULE TESTING The testing of the RF-module is performed after the production climate tests.
Finalizat 15. Measure the currency from the external output. The currency must be at a level of 1 ... 2mA with spikes of 10mA in between. 16. Remove the external power input from the Binder connector. 17. Set the transmission interval at ’30s’ from the meter’s menu and switch the meter off thereafter. 18. Once the radio receiver has been connected to the computer and the settings in the TERMINAL-program are correct, the phases 1-6 and 13—17 can be repeated. 6.
REFERENCE REFERENCES The Digi-Internation manual: Digi International XB Pro 900 HP Manual 1792196 The RDS-32 manual: User Manual RDS-32 Survey Meter The Maximum Permissible Exposure report 11
APPENDIX APPENDIX A: AGENCY CERTIFICATIONS FOR S3B HARDWARE FCC (United States) Certification / ISED Certification The XBee-PRO® 900HP/XBee-PRO® XSC RF Module complies with Part 15 of the FCC rules and regulations. Compliance with the labeling requirements, FCC notices and antenna usage guidelines is required. In order to operate under 's FCC Certification, RF Modules/integrators must comply with the following regulations: 1.
APPENDIX The Omni-directional antenna below has been approved for use with this module when installed into the host device. The antenna Gain with Cable loss is less than 1.9 dBi. FCC-approved Antennas WARNING: The host device can only use the approved monopole antenna. The specifications and installation instructions of the antenna are described in Chapter 4. Portable Applications and SAR Testing The module can be used at distances closer than 20cm to all persons.
FEEDBACK FEEDBACK We are continuously working hard at producing correct and easy-to-read technical documents. However, complex systems are often difficult to explain or understand and therefore mistakes or inadequacies may occur occasionally in the documentation process. To correct these errors, we would like to hear your opinion on this document. You can submit your feedback on our website www.mirion.com filling out the contact form.