Operating Manual GALAXY GX2 Automated Test System Order No.
WARNING Read this manual carefully before using the instrument. The instrument will perform as designed only if it is used and maintained in accordance with the manufacturer's instruction. Otherwise, it could fail to perform as designed and persons who rely on this instrument for their safety could sustain serious personal injury or death. © MINE SAFETY APPLIANCES COMPANY 2012 - All Rights Reserved This manual is available on the Internet at: www.msasafety.
MSA AUER Contents Contents 1 2 Safety Regulations ................................................................................................................... 5 1.1 Correct Use ....................................................................................................................... 5 1.2 Product Warranty ............................................................................................................... 6 Introduction ..................................................
Contents MSA 4.6 Instrument Configuration ................................................................................................. 39 Instrument Setup ............................................................................................................. 40 Sensor Setup ................................................................................................................... 42 Save Settings ..........................................................................................
MSA AUER Safety Regulations 1 Safety Regulations 1.1 Correct Use WARNING Read this manual carefully before using the instrument. The instrument will perform as designed only if it is used and maintained in accordance with the manufacturer's instruction. Otherwise, it could fail to perform as designed and persons who rely on this instrument for their safety could sustain serious personal injury or death.
MSA Safety Regulations FCC Warning Statements Changes or modifications not expressly approved by the manufacturer could void the user's authority to operate the equipment. Industry Canada (IC) Warning Statements The installer of this radio equipment must ensure that the antenna is located or pointed such that it does not emit RF field in excess of Health Canada limits for the general population; consult Safety Code 6, obtainable from Health Canada's website www.hc-sc.gc.ca/rpb.
MSA AUER Safety Regulations Exclusion of Consequential Damage Purchaser specifically understands and agrees that under no circumstances will seller be liable to purchaser for economic, special, incidental or consequential damages or losses of any kind whatsoever, including but not limited to, loss of anticipated profits and any other loss caused by reason of nonoperation of the goods.
MSA Introduction 2 Introduction Congratulations on purchasing the GALAXY GX2 Automated Test System, the next generation Test Stand and instrument management system from MSA. This system is used exclusively with the ALTAIR® family of gas detectors. This manual uses the terms instrument and gas detector to represent that entire line of ALTAIR gas detection instruments. In this manual the user will learn to install and configure the GALAXY GX2 Test Stand and optional attachments, and test gas detectors.
MSA AUER Introduction The Test Stand is a standalone unit that can accommodate one gas detector of the ALTAIR family. However, each Test Stand contains the plumbing and electronic ports to simultaneously connect a total of 10 Test Stands and four Cylinder Holders (electronic or non-electronic). For applications where multiple Cylinder Holders are connected, the Test Stand Expanded Solenoid option must be ordered.
MSA Introduction ware, and configuration options anticipate the user's needs and provide superb efficiency in this next generation Automated Test System. GALAXY GX2 System Features and Options The GALAXY GX2 automatically identifies the type of gas detector inserted into the Test Stand. Based on user-defined settings, the Test Stand then performs bump tests and/or calibrates the instrument. The data collected from each test event is stored to a memory card [ chapter 2.
MSA AUER Introduction 3 4 5 1 6 2 Fig. 3 Test Stand Physical Features 1 Touch Screen Display 4 SD Card Port 2 Status LED 5 USB Port 3 Test Stand to Test Stand connector 6 Gas plugs SD or SDHC Memory Card Slot A memory card port is provided on the Test Stand for calibration and bump record storage.
MSA Introduction Test Stand is Stand-Alone If the GALAXY GX2 is not networked to a PC interface the memory card will save each calibration and bump test record. Incomplete records will not be saved. If networked and e-mail alerts are configured, the Test Stand will generate an email alert once the memory card reaches 90% and again at 99% capacity. Once a memory card reaches full capacity, the Test Stand will be in fault and prohibit any unit from performing tests in that bank, until corrected.
MSA AUER 2.3 Introduction Ethernet Interface Two Ethernet Interfaces are provided on the rear of the GALAXY GX2 [ Fig. 6]. The ports allow for the connection and communication distribution between multiple Test Stands. If networking to a computer, one USB port on the Master Test Stand is used to communicate with the MSA Link Pro software application. Fig. 6 2.
MSA Introduction 2.5 Cylinder Holder (electronic) The electronic Cylinder Holder can accommodate one test gas cylinder and includes: 1 2 3 Fig. 7 1 Cylinder Holder Physical Features Light Band 2 Multi-Color Light Band Indicates gas bottle functionality. - - - - 14 Green indicates a bottle is completely functional, and gas parameters are within pressure and expiration date limits. Yellow indicates low calibration gas, or gas is nearing its expiration date.
MSA AUER 2.6 Introduction Optional Printer The Printer can print calibration and bump result receipts and calibrations stickers for gas detectors [ Fig. 7]. The Receipt/ Sticker Printer uses a USB cable to connect to the port of the farthest right Test Stand [ Fig. 3]. Labels are available in two formats [ chapter 7.2]: - Format #1: Calibration Sticker only (2 cm x 2 cm square label). - Format #2; Receipt and Sticker combined. DU MM Fig. 8 2.
MSA Introduction Fig. 9 2.8 Multi-Unit Charger (Altair 4/4X and 5/5X only) Special Conditions for Use with Reactive Gases If using a Test Stand and Cylinder Holder for calibrating or bump testing a gas Detector configured with a chlorine or ammonia sensor: - Place the chlorine or ammonia gas cylinder in the Cylinder Holder closest to the Test Stand. - Only one reactive gas Test Stand is allowed in a bank.
MSA AUER 3 Installation Installation The GALAXY GX2 is a simple to install system that can be desktop or wall mounted. Setup requires simple tools and just a few minutes of time.
MSA Installation This manual provides steps for desktop and wall mounting. The following sections describe the proper installation for various GALAXY GX2 configurations: 3.1 18 - Connect Units in a Bank (Test Stands and Cylinder Holders) [ chapter 3.1] - Connect a Test Gas Source Without a Cylinder Holder (optional) [ chapter 3.2] - Network Test Stands (optional) [ chapter 3.3] - Removing Gas Seal for Certain ALTAIR and ALTAIR Pro Instruments [ chapter 3.4] - Desktop Mounting [ chapter 3.
MSA AUER Installation (4) If connecting multiple Test Stands, remove the white gas plugs [ Fig. 3] from all units except the farthest right Test Stand. If using ammonia or chlorine test gas, read the restriction found under chapter 2.8 “Special Conditions for Use with Reactive Gases” regarding the white plugs. (5) Continue adding Test Stands to the right and Cylinder Holders to the left [ Fig. 10].
MSA Installation 2 1 DUMMY! Will be replacing this photo - mid March Fig. 11 3.4 Test Stand Ethernet connections 1 MSA Link Pro port 2 Test Stand to Test Stand port (2) If connecting the bank to a computer with the MSA Link Pro software, use a customersupplied ethernet cable and connect via the Master Test Stand ethernet port #1 shown above.
MSA AUER 3.5 Installation Desktop Mounting (1) Place the GALAXY GX2 on a flat, stable surface. (2) Insert the power supply into the GX2 power jack [see chapter 4.1 for first-time power up initialization]. The Test Stand display pivots for ease of viewing. (3) US If using MSA Cylinder Holders, the left most Cylinder Holder should have a fresh air filter attached to the top port.
MSA Installation 22 (4) If not using MSA Cylinder Holders, the left most Test Stand should have a fresh air filter attached to the top port. (5) Ensure the white plugs are inserted into the five gas connection fittings on the right side of the farthest right test stand.
MSA AUER 3.6 Installation Wall Mounting When wall mounting the Test Stand (and applicable Cylinder Holder), MSA recommends steps 1-5 from the Desktop Mounting section [ chapter 3.5] above be completed prior to installing on the Din Rail. For a large configuration of Test Stands and Cylinder Holders, MSA recommends several people be used for installing or uninstalling from the DIN rail. US (1) Complete steps 1-5.
MSA Installation 3.7 SD or SDHC Memory Card Option The Test Stand can accommodate an SD or SDHC memory card, inserted in the port on the right hand side of the farthest right Test Stand. Only one memory card can be used in a bank. To install the memory card: (1) Ensure that memory card is not write-protected or locked. (2) Insert the memory card into the port, located on the right side of the Test Stand [ Fig. 13]. (3) Place the optional end cap over the port to protect the memory card.
MSA AUER 4 Setting Up the GALAXY GX2 Setting Up the GALAXY GX2 The GALAXY GX2 Automated Test System is shipped with the most common default options and therefore requires minimal set up. A table of all default settings is provided in the Galaxy GX2 Default Parameters section [ chapter 9] at the end of this manual. Configuration settings are described in this chapter and can be changed to meet your individual needs via the touch screen display.
MSA Setting Up the GALAXY GX2 Set Time and Date There are 3 tabs at the bottom of the screen - Date, Time Zone and Time. The user will need to configure each tab. The Date Setup screen automatically displays. (1) Enter the current month, day and year. The time setting is used on all calibration and bump test records. Test Stand time and date must be maintained for accurate record keeping.
MSA AUER 4.2 Setting Up the GALAXY GX2 Security Setup The Security Setup is the last screen to automatically display. Setting a numeric password prohibits unauthorized changes to the GALAXY GX2. If no password is desired, enter "0000" to disable the password feature. The Test Stand is shipped with the password disabled. (1) Enter a four-digit password for the GALAXY GX2. (2) Select Save. Retain a copy of the password for your records. (3) Select Home.
MSA Setting Up the GALAXY GX2 Test Mode Bump Only, Calibration Only, or Bump/Cal on Fail. The mode is selected in the Test Setup section [ chapter 4.5] of this manual. Gas Detector Datalog Download On or Off. This setting is described in the GALAXY GX2 Setup, Datalog section [ chapter 4.5] of this manual. Pressure Gauges Display test gas cylinder pressure from the electronic Cylinder Holder only. Selecting a pressure gauge displays gas cylinder details.
MSA AUER Setting Up the GALAXY GX2 Instrument Records Provides the most recent calibration and/or bump record for each instrument in the Test Stand bank. This button is only active on the Master Test Stand. All other Test Stands will report their data to the Master Test Stand for display. 4.4 General Setup The following settings can be changed from their default values after entering the password. Select GX2 Configuration from the Home screen. The Administrator screen displays. Fig.
MSA Setting Up the GALAXY GX2 Fig. 16 GX2 Configuration 1 Network Setup [ chapter 4.5] 5 GX2 Setup [ chapter 4.5] 2 Time/Date Setup [ chapter 4.1] 6 Backlight/Volume Setup [ chapter 4.5] 3 Test Setup [ chapter 4.5] 7 Languages [ chapter 4.1] 4 Security Setup [ chapter 4.2] The Time/Date Setup, Security Setup, and Languages Setup have been described in the Initial Setup section [ chapters 4.1 and 4.2].
MSA AUER Setting Up the GALAXY GX2 To return to the GX2 Configuration screen, select the back arrow W on the top left corner of the screen. The GALAXY GX2 will reduce the backlight intensity automatically after a period of inactivity. Either select a button or insert an instrument to return the intensity to the user selected level. Test Setup Select Test Setup on the GALAXY GX2 Configuration screen [ Fig. 16] to access the Test Setup screens.
MSA Setting Up the GALAXY GX2 Calibration Tab The user can set the calibration interval that is used to determine if a calibration is due. (1) Set the interval days by selecting either the up or down arrows on the calibration screen. Z (2) The default calibration interval is set at 30-days. Select Save. Bump Tab The user can set up the bump interval that is used to determine if a bump is due. (1) Set the interval days by selecting either the up or down arrows on the bump screen.
MSA AUER Setting Up the GALAXY GX2 GX2 Setup To access GX2 Setup, select GX2 Setup on the GALAXY GX2 Configuration screen [ Fig. 16]. GALAXY GX2 Tab The following options are available on this screen: - USB Drive Enable setting allows the user to gather data from the Memory card with a USB drive that can be inserted in the port on the right side of the unit. The default setting is Off (2). - Display psi or bar options display the units of pressure on the gauges on the home screen and pressure screen.
MSA Setting Up the GALAXY GX2 Datalogs are not written to the memory card because of potentially large file sizes. If the unit is not networked to the MSA Link Pro software application, instrument datalogs cannot be downloaded with the GALAXY GX2. In this situation the user can use the MSA IR dongle and the free MSA Link application to download datalogs from the instrument.
MSA AUER Setting Up the GALAXY GX2 The GALAXY GX2 processes gas cylinder information and instrument readings in ppm, %(volume) or %LEL. If using certified gas cylinders with the content specified in ppm, the most accurate gas readings from the instrument will be obtained with the instrument set to ppm. Conversions between ppm and mg/m3 are completed at standard pressure and temperature values.
MSA Setting Up the GALAXY GX2 If the Test Stand is NOT using Cylinder Holders, the gas connections on the left-most GALAXY GX2 Test Stand [ Fig. 17] are as follows from top: - Fresh Air - Cylinder 1 - Cylinder 2 - Cylinder 3 - Cylinder 4 Fig. 17 Gas inlets on the Test Stand If the Test Stand is using Cylinder Holders [ Fig. 10], they are set up as follows from right to left: - Cylinder 1 is the closest to the Test Stand. - Cylinder 2 is to the left of Cylinder 1.
MSA AUER Setting Up the GALAXY GX2 Changing Cylinders Changing test cylinders in the Cylinder Holders is a simple procedure. To change an RFID tagged test gas cylinder, using an electronic Cylinder Holder: (1) Unscrew and remove a cylinder with its attached RFID tag from the Cylinder Holder. (2) Screw a new test gas cylinder into the Cylinder Holder. The new cylinder's RFID tag will automatically populate the cylinder data fields on the Cylinder Setup screens.
MSA Setting Up the GALAXY GX2 If background gases could be present, connect a Zero Air Cylinder to any port on the Cylinder Holder. MSA provides an RFID-tagged Air cylinder for this purpose. This requires a calibration station with expanded solenoid capacity. Refer to the Introduction section [ chapter 2] for more information about the expanded solenoid option. Network Setup (Optional) The GALAXY GX2 can be networked to a PC with the MSA Link Pro software application.
MSA AUER (2) Enter the Static IP Address on the number keypad. Z (3) 4.6 The yellow highlighted cursor then moves to the Gateway field. Enter the Gateway on the number keypad. Z (5) The yellow highlighted cursor then moves to the Subnet field. Enter the Subnet on the number keypad. Z (4) Setting Up the GALAXY GX2 The yellow highlighted cursor then moves to the Subnet field. Select Save.
MSA Setting Up the GALAXY GX2 Fig. 18 Instrument Configuration 1 Instrument Setup 4 Load Settings 2 Sensor Setup 5 Update Settings 3 Save Settings Instrument Setup The user can set the instrument's datalog gathering functionality, alarms, calibration intervals, and identification through the 4 tabs on the Instrument Setup page. Select Instrument Setup [ Fig. 18] from the Instrument Configuration screen.
MSA AUER Setting Up the GALAXY GX2 Alarms Tab Select the Alarm tab to set the instrument alarm functionality. Fig. 20 Instrument Setup - Alarm - Audio Alarm Enable (3) sets the instrument to emit a beeping sound in an alarm condition. - LED Alarm Enable (3) sets the instrument to flash its LEDs in an alarm condition. - Vibrating Alarm Enable (3) sets the instrument to a vibrating alarm in an appropriate condition.
MSA Setting Up the GALAXY GX2 For proper operation, ensure the GALAXY GX2 and instrument Calibration Due intervals are the same value. If not selected, calibration occurs based on the gas detector Cal Interval. This can lead to more (or less, depending on setting) frequent instrument calibration and unwanted overdue indications. ID Tab Select the ID tab to set the instrument name, company, and department. Fig. 22 Instrument Setup - ID (1) Select User Name to assign a designation.
MSA AUER Setting Up the GALAXY GX2 (1) Select Sensor Setup from the Instrument Configuration screen [ Fig. 18]. Z (2) This screen displays the sensors currently installed in the instrument. Select the sensor that you wish to configure on the Sensor Setup screen. Options Tab Select the Options tab to set Span Value and enable or disable the sensor for that gas type. (1) Select Span Value to enter the calibration and bump span value.
MSA Setting Up the GALAXY GX2 Alarm Setpoint Tabs Select the LO, HI, STEL, and TWA tabs to enable each alarm, determine setpoints, or enable latching for each sensor. WARNING Using an incorrect alarm setting can seriously impair the gas detector's ability to warn the user of dangerous conditions. Extreme caution must be used when changing alarm levels. Ensure alarm changes are properly set prior to instrument use.
MSA AUER Setting Up the GALAXY GX2 Load/Delete Settings Configured instrument settings can be loaded into the GALAXY GX2 and applied to an instrument, provided the black USB key is inserted into the Master Test Stand as described in the USB Drives section [ chapter 2.4] of the System Features section. This feature is particularly useful when configuring a large number of instruments. Unused settings may be permanently deleted from the Test Stand.
MSA Setting Up the GALAXY GX2 4.7 Cylinder Configuration The user can configure the GALAXY GX2 to change assigned gases in each cylinder holder attached to the Test Stand(s). (1) To access Cylinder Configuration, navigate to the Administrator screen: 2 1 Fig. 23 1 Icon - Cylinder Configuration (2) Select Cylinder Configuration. Z 46 Administrator screen 2 Icon - Home Screen The Cylinder screen displays.
MSA AUER Setting Up the GALAXY GX2 Cylinder Setup Select Cylinder Setup. Z The Cylinder Setup screen displays for Cylinder 1, which is the cylinder closest to the Test Stand. Cylinder Tabs Cylinder data will populate automatically when using the electronic cylinder holder and a test gas cylinder with RFID tag. This feature is enabled when this button appears as shown: The Lot # and Expires fields cannot be manually entered. They are populated only when using an electronic cylinder holder.
MSA Setting Up the GALAXY GX2 (5) Set the Unit by selecting either the left or right arrows. (6) Set the Value by entering the correct value on the number keypad. (7) Set the Type by selecting either the left or right arrows. (8) Select Save. (9) Repeat steps 1 through 6 for each cylinder (maximum of 4 cylinders). WARNING Proper setup of non-RFID tagged gas cylinders is critical. Accurate calibration may not occur if improper cylinder values are entered.
MSA AUER 5 Using the GALAXY GX2 Using the GALAXY GX2 Once the GALAXY GX2 is set up and configured to accept a gas detector, the user must complete the following steps to perform instrument tests. 5.1 Turning on the System The system is turned on when the AC Power Module or Vehicle Power Adapter is plugged into the unit and a reliable power supply is provided. The GALAXY GX2 does not have an on/off button and is designed to remain powered on. 5.
MSA Using the GALAXY GX2 To insert the ALTAIR 5/5X: (1) Insert the instrument into the cradle, sliding the bottom over the charging connector. If your Test Stand has the optional charging feature, ensure the charging connector at the bottom of the cradle makes contact with the instrument contact. (2) Pivot the instrument in the cradle upwards, into the gas inlet sleeve (see arrow). The instrument will lock into place, once properly positioned.
MSA AUER Using the GALAXY GX2 Before inserting an ALTAIR or ALTAIR Pro into the Test Stand, press and hold the instrument Test button for one second. The screen on the gas detector should display Gas? prior to inserting it into the Test Stand (see photo below). (4) Insert the instrument into the Test Stand cradle. (5) The GALAXY GX2 will read the instrument configuration via the IR link up to 10 seconds. If the LED lights on the instrument flash red do not remove the instrument.
MSA Using the GALAXY GX2 If the instrument is removed while the Test Stand is performing a calibration or bump the instrument will default to the last valid and retained calibration parameters. For gas detectors with multiple sensors, any green colored calibration progress values are considered a valid calibration. Reminder: Instrument datalogs can only be downloaded when an Ethernet connection to MSA Link Pro is active or via an IR Dongle to MSA Link.
MSA AUER Using the GALAXY GX2 If the instrument fails the bump test while the mode is set to Bump/Cal on Fail the Test Stand will automatically calibrate the instrument. The Calibration Information screen displays two columns for the sensors: As Found and Target. As Found is what the Test Stand detected prior to calibration. The Target column specifies the span gas value that is being applied. If the sensor passes calibration, the Target value becomes the new calibration point.
MSA Using the GALAXY GX2 If performing a manual bump or calibration on an individual sensor in a multi-gas detector, always perform the manual test after the GALAXY GX2 test is complete. If using MSA Link Pro software, partial bump tests will result in the instrument being listed as "Overdue" for bump testing since not all sensors were tested. After Setup is complete: After initial installation and setup are complete, confirm the GALAXY GX2 is properly configured.
MSA AUER 5.5 Using the GALAXY GX2 GALAXY GX2 Always Ready Feature The Always Ready feature is a capability of the ALTAIR 4X and ALTAIR 5X with the following software versions: - ALTAIR 4X: v2.03 and higher - ALTAIR 5X: v1.11 and higher The feature permits the gas detector to be automatically calibrated on a user-defined interval. The most common use of this feature allows the user to configure the GALAXY GX2 to automatically calibrate an instrument prior to the start of work-shift.
MSA Using the GALAXY GX2 5.6 E-mail Notification If a network connection is active and e-mail notification is configured via the MSA Link Pro software, the Master Test Stand will send e-mails based on the events shown in the table below. The MSA Link Pro user can configure which day(s) of the week that e-mails are sent and for which conditions. Note that "EO" = Each Occurrence and "FO" = First Occurrence. Frequency 5.
MSA AUER 6 Troubleshooting Troubleshooting Use the following information to diagnose abnormal conditions. 6.1 US Instrument Will Not Initiate a Bump or Calibration (1) Ensure the instrument is turned on and errors are not displayed. (2) If using an ALTAIR or ALTAIR Pro, press and hold the button for one second. The screen on the gas detector must display Gas? prior to insertion into the Test Stand. (3) Ensure the instrument is seated securely in the Test Stand cradle.
MSA Troubleshooting 6.2 6.3 Instrument Fails Zero Calibration or Bump (1) Check the Fresh Air filter and replace if contaminated. (2) If not using zero gas, ensure that the atmosphere is free of any interfering gas. (3) If connecting test gas without a Cylinder Holder, ensure the gas tubing line is not connected to the top fresh air port on the left side of the farthest left unit. Instrument Fails Span Calibration or Bump (1) Verify the instrument span setting is the same value as the test gas.
MSA AUER 6.5 Troubleshooting Cylinder Data Does Not Display If using an electronic Cylinder Holder with RFID-tagged MSA gas cylinders, the general cylinder parameters automatically populate. If using a non-electronic Cylinder Holder or test gas from an independent source the user must manually enter cylinder data. Gas cylinder pressure will ONLY display with an electronic Cylinder Holder.
MSA Troubleshooting Cylinder Holder (number) If Status is Fail, then: -CRC Corrupt RFID tag. Enter cylinder information via manual setup screens. -Identifier 60 Test Stand (number) If Status is Fail, then: -IR Cycle power to the Test Stand. Ensure there is no dirt on the gas detector blocking the IR interface. If the problem persists, return to an Authorized MSA Service Center for repair.
MSA AUER 6.7 Troubleshooting System Error Messages The GALAXY GX2 can display a number of system errors that display on the Test Stand screen with a red error box. The error message must be acknowledged for Test Stand operations to resume. Memory Card Write Protect Occurs if the memory card cannot accept data. Ensure the switch on the side of the memory card is unlocked so it can accept data. Serial Number Required Occurs when the instrument is missing its serial number.
MSA Maintenance (Cleaning and Part Replacement) 7 Maintenance (Cleaning and Part Replacement) 7.1 Corrective Maintenance See the listing below for common replacement parts associated with the GALAXY GX2. If you have any questions regarding the equipment, any procedure in this manual, or the location of the nearest MSA Repair Center and source of replacement parts, please call MSA Customer Service.
MSA AUER US Maintenance (Cleaning and Part Replacement) Accessories Part No.
MSA Maintenance (Cleaning and Part Replacement) GALAXY CALIBRATION CYLINDERS 64 Type Gas Content Size PN Quad 1.45% Methane, 15.0% O2, 60ppm CO, 20ppm H2S ECONO-CAL 10048280 Quad 1.45% Methane, 15.0% O2, 60ppm CO, 20ppm H2S RP CYLINDER 10045035 Quad 2.5% Methane, 15.0% O2, 60ppm CO, 20ppm H2S ECONO-CAL Quad 2.5% Methane, 15.0% O2, 60ppm CO, 20ppm H2S RP CYLINDER 10048890 3-GAS 2.5% Methane, 15.0% O2, 20ppm H2S ECONO-CAL 3-GAS 2.5% Methane, 15.
MSA AUER US Maintenance (Cleaning and Part Replacement) Type Gas Content Size C5H12 0.375% Methane, 15.
Maintenance (Cleaning and Part Replacement) 7.3 MSA Fresh Air Filter Replacement The disposable fresh air filter is located on the tubing attached to the upper port on the left side of the farthest left unit. The filter requires periodic replacement, based on the unit's frequency of use and environmental cleanliness. Visual inspection of the fresh air filter will indicate contamination. Replace as needed. Fig. 25 Fresh air filter Pump Alarms may indicate the need to replace the fresh air filter.
MSA AUER 7.4 Maintenance (Cleaning and Part Replacement) O-Ring Seal Replacement (Cylinder Holder) Each month, visually inspect the O-ring on the fitting and clean if necessary. Replace if any damage is found. (1) Remove the damaged O-ring. (2) Clean the fitting with a water-dampened cloth and apply a light film of mineral oil to the new O-ring. NOTICE Do not use a silicone-based lubricant on this O-ring. (3) 7.5 Place the new O-ring onto the fitting.
Maintenance (Cleaning and Part Replacement) MSA Fig. 26 Fig. 27 Fig.
MSA AUER 7.6 Maintenance (Cleaning and Part Replacement) Rubber Insert Replacement (ALTAIR Gas Detector) The rubber inserts may degrade in the GALAXY GX2 over time. Pull the degraded insert from the Test Stand. Replace the insert and ensure that it is firmly seated. Fig. 29 Fig.
MSA Technical Specifications 8 Technical Specifications Specification Part Range Operating Temperature GALAXY GX2 0-40º C Power Input Power Module 100 - 240 VAC, 47 - 63 Hz Optional Vehicle Module 9-32 VDC Test Stand only < 6.0 W Power Consumption With Cylinder Holder Physical Characteristics Test Stand Height: 11.80" (299.72 mm) Width: 6.50" (165.10 mm) Depth: 7.90" (200.66 mm) Material: Acrylonitrile Butadiene ABS Cylinder Holder Height: 11.80" (299.72 mm) Width: 6.50" (165.
MSA AUER 9 Galaxy GX2 Default Parameters Galaxy GX2 Default Parameters Refer to the listing below for the default settings for each display screen for configuring the GALAXY GX2. The settings can be changed for optimal performance in the user's facility.
MSA in Europe [ www.msa-europe.com & www.msa-gasdetection.com ] Northern Europe Southern Europe Eastern Europe Central Europe Netherlands MSA Nederland Kernweg 20 1627 LH Hoorn Phone +31 [229] 25 03 03 Fax +31 [229] 21 13 40 info@msaned.nl France MSA GALLET Zone Industrielle Sud 01400 Châtillon sur Chalaronne Phone +33 [474] 55 01 55 Fax +33 [474] 55 47 99 message@msa-gallet.fr Poland MSA Safety Poland ul.