Operator's Manual

Alarms Resetting Alarms
TM80/TM70 Telemetry Monitor Operator’s Manual 6 - 9
The alarm pause time is two minutes. When the pause time expires, the alarm
paused status is automatically deactivated. You can also cancel the alarm paused
status by tapping [
Alarm Pause] in the quick keys area.
6.9 Resetting Alarms
You can acknowledge the on-going alarms by resetting the alarms. After being reset, the
alarm reset symbol is displayed in the message area.
When an alarm occurs, follow this procedure to reset the TM80/TM70’s alarm system.
Press the key to enter the main menu, and then tap [Alarm Reset] from the
[Commands] section.
OR
1. Flick your finger up at the bottom of the main screen to display the quick keys
area.
2. Tap the [Alarm Reset] quick key to reset the alarm system.
6.9.1 Resetting Physiological Alarms
After the alarm system is reset, physiological alarms give the following alarm indicators:
The alarm sound is silenced.
The symbol appears in the message area.
NOTE
When the TM80/TM70 is connected to the CMS and the function of
remotely pausing alarms is enabled at the CMS, alarms can be paused
either at the TM80/TM70 or at the CMS. For information on pausing
alarms at the CMS, refer to BeneVision Central Monitoring System Opera-
tor’s Manual.
NOTE
When the TM80/TM70 is connected to the CMS and the function of
remotely resetting alarms is enabled at the CMS, alarms can be reset
either at the TM80/TM70 or at the CMS. For information on resetting
alarms at the CMS, refer to BeneVision Central Monitoring System Opera-
tor’s Manual.