Instruction Manual of WatchBP Pro © 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro
operational sequence Actions LCD display * Power –Up the WatchBP Pro *Power –Up the Bluetooth *First time switch on Description z Locate the black power switch on the back side. z Turn it on z The green indicator light on the top right front of the WatchBP Pro will illuminate and remain lit. z Locate the silver BT switch on the back side.
*Cuff connection *Press Start/Stop button *Auto zero reset * Fuzzy inflation z Binding left-arm cuff on your left arm and place chest-piece of stethoscope on your brachial artery correctly. z LCD full segment displays for 2 seconds with backlight on z Display shows “0” when device z in auto zero reset. z Pumping starts, and cuff pressure increases to working pressure (for fuzzy inflation).
*Catch the SYS pressure. *Catch the DIA pressure * Measurement finishes *Display measured values z Press “M” button when a faint, rhythmic tapping or thumping sound is heard, which stands for your SYS reading. z Press “M” button when your diastolic reading reaches and thumping sound stops, this is DIA reading. z When measuring finishes, cuff z A long beep deflates completely and a long beep sounds.
z *Memory reading Device auto memorizes one set z of measured values. Press M button, LCD shows AUE/AUS, then memory readings, with backlight on simultaneously *Withdraw from readings z Memory SP/DP and MAP/PP displays one after the other. z Backlight remains on 40 sec. z Press Start/Stop button or put z the device untouched for 5’ to terminate measurement display. First Visit mode (3 times measurement in succession, on both arms).
*Press Start/Stop button or remote control. *60 sec count down before measurement. * Auto zero reset * First turn fuzzy inflation. *First turn deflation. © 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro z LCD full segment displays for 1 second with backlight on z Beeper sounds 5 seconds before z the 1st measurement. z Press Start/Stop button or remote control to skip over count down and begin 1st turn measuring immediately.
*Pulse detected z Detect the pulse, “♥” icon start z to flash and a beep is heard for every heartbeat. Beeper each pulse *First turn measurement finishes z After first turn measuring, cuff z deflates completely, and a long beeper sounds A long “B”, and 2 short z After another two close beeps, device comes into 60 sec count down. z Beeper sounds 5 seconds before z the 2nd measurements. z Press Start/Stop button or remote control to skip over count down and begin 2nd turn measuring immediately.
*2nd turn deflation. z Pumping stops, pressure in z deflation by 4~6 mmHg/sec. *Pulse detected z Detect the pulse, “♥” icon start z to flash and a beep is heard for every heartbeat. A beep each pulse *2nd turn finishes z After 2nd turn measuring, cuff z deflates completely, and a long beeper sounds A long “B”, and 2 short z After another two close beeps, device comes into 60 sec count down. z Beeper sounds 5 seconds before z the 2nd measurements.
*3rd turn fuzzy inflation. z Pumping starts & cuff pressure increases to working pressure (for fuzzy inflation). *3rd turn deflation. z Pumping stops, pressure in deflation by 4~6 mmHg/sec. *Pulse detected z Detect the pulse, “♥” icon start z to flash and a beep is heard for every heartbeat. A beep each pulse *The 3rd turn measuring finishes z After 3rd turn measuring, cuff z deflates completely, and a long beeper sounds A long “B”.
z LCD displays values averaged from 3 times measurement with backlight on simultaneously. z Left arm BP values will flash 5 times by default if difference between L and R values are less than 10 mmHg, otherwise, the higher (L or R) flashes 5 times. z Display of SP/DP, following which, MAP & PP shows on. z SP/DP and MAP/PP values display one after the other till device auto off. z Backlight remains on 40 sec. *Data transmission.
* Memory reading--3rd turn measured values display. z Press M button again, LCD shows a “3” then displays of the 3rd turn measured values. z © 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro SP/DP and MAP/PP displays one after the other.
*Memory reading--2nd turn measured values display. z Press M button again, LCD shows a “2” then displays of the 2nd turn measured values. z © 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro SP/DP and MAP/PP displays one after the other.
*Memory reading--1st turn measured values display. *Withdraw from readings z Press M button again, LCD shows a “1” then display of the 1st turn measured values. z SP/DP and MAP/PP displays one after the other. z Press Start/Stop button or put the device untouched for 5’ to terminate measurement display. Follow up mode (3 times measurement in succession, optional on left or right arm).
*Cuff and connection blue tooth z Binding left & right arm cuffs on your both arms correctly. z Push blue-tooth switch on the back side of device (ON/OFF sculpted) to “ON” for computer connection if remote control. z Hold down M button for 3 sec, memory shows with backlight on, till L or R flashes. z Press M button to choose that preferred arm for measurement, and press Start/Stop button for confirmation. *Press Start/Stop button or remote control.
z Pumping starts & cuff pressure increases to working pressure (for fuzzy inflation). z Pumping stops, pressure in deflation by 4~6 mmHg/sec. *Pulse detected z Detect the pulse, “♥” icon start z to flash and a beep is heard for every heartbeat. *First turn measurement finishes z After first turn measuring, cuff deflates completely and device comes into 60 sec count down. *60 sec count down z Beeper sounds 5 seconds z before the 2nd measurements.
* Auto zero reset z Display shows “0” for 2 sec z when device in auto zero reset. * 2nd turn fuzzy inflation. z Pumping starts & cuff pressure increases to working pressure (for fuzzy inflation). *2nd turn deflation. z Pumping stops, pressure in deflation by 4~6 mmHg/sec. *Pulse detected z Detect the pulse, “♥” icon start z to flash and a beep is heard for every heartbeat.
*2nd turn measurement finishes z After 2nd turn measuring, cuff deflates completely. z If a difference of 5mmHg exists between 1st&2nd time measured sys or dia pressure, device auto comes into 60 sec count down and takes the 3rd measurement. *Measuring finishes with a long beep and average values display if difference between 1st & 2nd turn measured sys or dia pressure is less than 5 mmHg. *60 sec count down z Beeper sounds 5 sec before the z 3rd turn measurement.
*3rd turn deflation. z Pumping stops, pressure in deflation by 2~4 mmHg/sec. *Pulse detected z Detect the pulse, “♥” icon start z to flash and a beep is heard for every heartbeat. A beep each pulse *3rd turn finishes measurement z After 3rd turn measuring, cuff z deflates completely, and a long beeper sounds A long “B”. * Display of average values z LCD displays values averaged from 3 times measurement with backlight on simultaneously. z Display of SP/DP, following which, MAP & PP shows on.
*Data transmission. z Through blue tooth, measured values are submitted to computer for further analysis. *Memory reading--average values display. z Device auto memorizes 3 times measured values and their average readings. z Press M button, LCD shows an “A” then average reading, with backlight on simultaneously. z Averaged SP/DP and MAP/PP displays one after the other. z Backlight remains on 40 sec.
* Memory reading--3rd turn measured values display. z Press M button again, LCD shows a “3” then displays of the 3rd turn measured values. z © 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro SP/DP and MAP/PP displays one after the other.
*Memory reading--2nd turn measured values display. z Press M button again, LCD shows a “2” then displays of the 2nd turn measured values. z © 2007 Microlife Corporation Document: Instruction Manual of WatchBP Pro SP/DP and MAP/PP displays one after the other.
z Press M button again, LCD shows a “1” then display of the 1st turn measured values. z SP/DP and MAP/PP displays one after the other. *Withdraw from readings z Press Start/Stop button or put the device untouched for 5’ to terminate measurement display. *Memories deletion. z Hold down M button for 7 sec z till 3 short beepers are heard and bars appear on LCD with *Memory reading--1st turn measured values display. backlight on simultaneously, which means memory readings are deleted successfully.
* Battery low indication z When voltage inside batteries is lower to 4.8V, LCD shows low battery icon and batteries begin to be charged if adaptor is connected. *Turn off the device. z Once voltage inside batteries is lower to 4.6V, LCD shows low battery icon, then batteries must be charged again. z Switch off O/I button at the back side of the unit z Memories won’t be stored once the device is turned off. Federal Communications Commission (FCC) Statement 15.
connected. -Consult the dealer or an experienced radio/TV technician for help. Operation is subject to the following two conditions: 1) this device may not cause interference and 2) this device must accept any interference, including interference that may cause undesired operation of the device. FCC RF Radiation Exposure Statement: This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment.