HCLH/SOP-YF-072-108 MT300MIX Gas Monitoring Series Operation and Installation Manual
B1-MT300MIX [Manufacturer]: Wuhan Huchuang Union Technology Co., Ltd. [Production Address]: No.1 Workshop, 1F, Building B10, Wuhan Hi-Tech Medical Device Park, No. 818 Gaoxin Avenue, East Lake Hi-Tech Development Zone, Wuhan, Hubei, China [After-Sales Service Unit]: Wuhan Huchuang Union Technology Co., Ltd.
B1-MT300MIX Contents 1 Overview .................................................................................................................................................................3 1.1 Operation Instructions of Manual .................................................................................................................3 1.2 Overview ......................................................................................................................................................
B1-MT300MIX 1 Overview 1.1 Operation Instructions of Manual 1.1.1 It is not allowed to print or disclose any content of this Manual, including pictures and audio products, under any name without the consent of Huchuang Union; 1.1.2 The equipment operator may copy some sections of this Operation Manual for internal use only, such as for instructing the user how to deal with emergencies. These sections are clearly listed in the catalogue of this manual; 1.1.3 Wuhan Huchuang Union Technology Co., Ltd.
B1-MT300MIX 1.4.1 MT300MIX device contains reusable materials, and its components can be recycled after being cleaned and sterilized. 1.4.2 During recycling and handling MT300MIX, it is recommended that the company's technical personnel dismantle it and recycle it according to different waste groups. 1.4.3 According to national regulations, the compositions of the main raw materials of MT300MIX shall be are shown in (Table 1).
B1-MT300MIX e c d c) Power Interface Equipped with a 5V/2A power adapter. d) Antenna e) Gas Outlet Refers to the gas outlet of the device. After the solenoid valve of the device is opened for the intake detection, excess gas will be discharged through the outlet. Do not block the gas outlet.
B1-MT300MIX i h h) i) Gas Inlet Power Switch j j) 2.2 Manual Ball Valve Refers to the manual valve of the gas inlet pipe; in case of equipment leakage alarm, the user can manually close this valve to ensure that the equipment will not leak.
B1-MT300MIX Sensor Detection resolution of O2 concentration 0.01% Detection accuracy of O2 concentration ±0.2% Detection range of O2 concentration 0%~30% Detection resolution of CO2 concentration 0.01% Detection accuracy of CO2 concentration ±0.
B1-MT300MIX 3.2 Operation Instructions of Buttons 3.2.1 Button Description a) b) c) d) a) UP key b) DOWN key c) OK key d) Menu key 3.2.
B1-MT300MIX Main Interface and Operation Instructions During Intake UP key: No function (disabled) DOWN key: No function (disabled) OK key: No function (disabled) Menu key: No function (disabled) Main Interface and Operation Instructions During Sampling after Intake UP key: No function (disabled) DOWN key: No function (disabled) OK key: No function (disabled) Menu key: No function (disabled) Note: 1. When the device is in the state of intake sampling, the control operation will be shielded; 3.2.
B1-MT300MIX Menu Function Description: "1. SET O2": O2 concentration calibration function; "2. SET CO2": CO2 concentration calibration function; "3. SET DATA FREQ": Setting function of data upload frequency; "4. SET SAMPLE FREQ": Setting function of sampling frequency; "5. SET VALVER TIME": Setting function of intake time of single sampling; "6.SET FLOW": Flow sensor calibration function; "7.SET PRESS": Calibration function of pressure sensor in rear of internal pressure reducing valve; "8.
B1-MT300MIX UP key: No function DOWN key: No function OK key: No function Menu key: The system returns to the setting interface when this key is pressed Interface and Operation Instructions Upon O2 Concentration Calibration Failure UP key: Press this button to add the value by 0.1% and press and hold this button to add the value by 0.1% continuously DOWN key: Press this button to reduce the value by 0.1% and press and hold this button to reduce the value by 0.
B1-MT300MIX CO2 Concentration Calibration Interface: Operation Instructions of Buttons: "+" key: Press this button to add the value by 0.1% and press and hold this button to add the value by 0.1% continuously; "-" key: Press this button to reduce the value by 0.1% and press and hold this button to reduce the value by 0.
B1-MT300MIX Menu key: The system does not save the calibration value and directly returns to the setting interface when this key is pressed Note: CO2 concentration calibration setting range: From 0% to 10%. 3.2.7 "3. SET DATA FREQ" Select "3.
B1-MT300MIX 3.2.9 "5. SET VALVE TIME" Select "5.
B1-MT300MIX Select "7.
B1-MT300MIX Select "9.
B1-MT300MIX HEPA gas filter Mini M-type hose clamp Note: 1) In this installation mode, a disc filter should be added at the front end of the gas intake of MT300MIX to prevent impurities from entering the equipment; 2) Hoops shall be installed at hose joints and tightened; 3) Pay attention to the pressure range of MT300MIX: 0.08MPA-0.2MPa; 4) Note that the total length of the pipe from tee to MT300MIX should not exceed 2 meters. 4.
B1-MT300MIX 3) Note that the total length of the pipe from the last incubator to MT300MIX should not exceed 2 meters. 5 Precautions 6 When MT300MIX is installed and used for the first time, because there is air inside the device, the air cannot be guaranteed to be emptied at one time, therefore, the concentrations tested first few times after the installation of the device will be inaccurate; The wireless transmission distance of MT300MIX is limited and should generally not exceed 3 rooms.
B1-MT300MIX which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: -Reorient or relocate the receiving antenna. -Increase the separation between the equipment and receiver. -Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. -Consult the dealer or an experienced radio/TV technician for help.
B1-MT300MIX Wuhan Huchuang United Technology Co., Ltd. Address: No.1 Workshop, 1F, Building B10, Wuhan Hi-Tech Medical Device Park, No.