Fill the Pump with insulin
1. Remove another refill syringe (IN2) from a refill kit package.
2. Firmly attach the two-way stopcock to the refill syringe and attach an
18 gauge needle to the stopcock. Label the syringe IN2.
3. Draw 20 mL of INSULIN (two vials) into the IN2 syringe.
4. Remove the 18 gauge needle and expel all air bubbles in the syringe.
5. Close the stopcock.
6. Retract the plunger until it locks into place. This should be no further
than 55 mL. Do not go beyond the vent hole. Press the lock into the
plunger groove to be sure it is firmly secured.
7. Shake vigorously for a minimum of 30 seconds to degas the INSU-
LIN.
8. Point the syringe tip upward and slowly open the stopcock to vent the
syringe.
9. Release the locking ring on the refill syringe.
10. Expel the air in the syringe and carefully observe it to ensure no air
bubbles remain in the syringe. If air bubbles are noted, repeat steps
5 - 10.
11. Fill the hub of the refill needle with INSULIN and attach it to the
stopcock.
12. Prime the needle and close the stopcock.
13. Place a sterile beaker or sterile barrier towel on the scale and tare it
(position balance at 0.)
14. Weigh the refill syringe and record the weight on the Implant Work-
sheet.