MP6025958-022 / A WARNING: If you receive a critical error on your pump, the following screen displays and the pump sirens. Immediately disconnect from your insulin pump and discontinue use. Contact your local representative for assistance. Remember, your body still needs insulin while your pump is removed. It is important that you consult your healthcare professional to determine an alternate method of receiving insulin while your pump is removed.
MP6025958-022 / A Audio: Depending on your Audio Options settings, the pump emits an alarm tone, a continuous three-pulse-and-pause vibration pattern, or both the alarm tone and vibration. Alerts An alert makes you aware of a situation that may require your attention. An alert is less serious than an alarm. D E S When an alert occurs: Display: The pump displays a notification with a yellow icon and instructions.
MP6025958-022 / A Messages A message informs you about the status of your pump or if you need to make a decision. When a message occurs: Display: The pump displays a notification with a blue icon and instructions. Notification light: Does not illuminate or blink. Audio: Depending on the message, the pump emits a message tone, an alert tone, or no tone. Depending on your Audio Options settings, you may hear a tone, feel a one-pulse-only vibration, or hear a tone and feel a vibration.
MP6025958-022 / A Title and text Explanation Next steps Auto Suspend You have not pressed any buttons during the time specified in the Auto Suspend settings. • To clear the alarm and resume basal insulin delivery, select Resume Basal. The pump battery does not have enough power. • Select OK to clear the alarm. Insulin delivery suspended. No buttons pressed within time set in Auto Suspend. • Remove the old battery and insert a new AA battery. For details, see About batteries, on page 21.
MP6025958-022 / A Title and text Explanation Next steps Cannot connect device • You may be trying to connect a device that is not compatible with your pump. • Select OK to clear the alert. This device is incompatible with your pump. See User Guide. • You are trying to connect a transmitter to your pump but another transmitter is already wirelessly connected to your pump. Check settings Startup Wizard settings complete. Check and set up your other settings.
MP6025958-022 / A Title and text Explanation Next steps Delivery limit exceeded You have attempted to deliver more insulin than expected based on your Max Bolus and Max Basal settings. • Check your BG. You had the Fill Cannula screen displayed for 15 minutes. • To proceed and fill the cannula, select Fill. The battery was removed from the pump. • Insert a new AA battery. Delivery stopped. Check BG. See User Guide for more information. Insert battery Delivery stopped. Insert a new battery now.
MP6025958-022 / A Title and text Explanation Next steps Insulin flow blocked Your pump has detected that the basal or bolus insulin flow was blocked. • Check your blood glucose. Consider checking ketones and take an injection if needed. Check BG. Consider injection and testing ketones. Change reservoir and infusion set. • Remove your infusion set and reservoir. • Select Rewind to start the new reservoir process using a new infusion set and reservoir.
MP6025958-022 / A Title and text Explanation Next steps Insulin flow blocked Your pump has detected that the insulin flow was blocked and there is no insulin in the reservoir. • Check your blood glucose. Consider checking ketones and take an injection if needed. • Remove your infusion set and reservoir. • Select Rewind to start the new reservoir process using a new infusion set and reservoir. ■ Check BG. Consider injection and testing ketones. Estimated 0 U insulin in reservoir.
MP6025958-022 / A Title and text Explanation Next steps Insulin flow blocked Your pump has detected the insulin flow was blocked while filling the cannula. • Check your blood glucose. Consider checking ketones and take an injection if needed. Fill Cannula stopped. Remove infusion set from body. Change reservoir and infusion set. • Remove your infusion set and reservoir. • Select Rewind to start the new reservoir process using a new infusion set and reservoir.
MP6025958-022 / A Title and text Explanation Next steps Insulin flow blocked Your pump has detected the insulin flow was blocked while filling the tubing. Possible connection issue between the tubing and reservoir. • Remove the reservoir and select Rewind to restart the fill tubing process. Fill Tubing stopped. Remove reservoir and select Rewind to restart. • Disconnect tubing from reservoir. • If this alarm occurs again, use a new infusion set.
MP6025958-022 / A Title and text Explanation Next steps Low reservoir Your reservoir is low on insulin, according to the number of hours or units set in the Low Reservoir Reminder. • Select OK to clear the alert. A pump error has occurred, and you need to restart your pump. Your backup settings have been lost, but your current settings are unchanged. • Select OK to restart your pump. Your current settings are unchanged. Only your backup settings are lost. XX hours remaining. Change reservoir.
MP6025958-022 / A Title and text Explanation Next steps Max Fill reached You have exceeded the number of units expected to fill the tubing. By now, insulin should be at the end of the tubing. • Remove the reservoir. There is no reservoir in the pump or the reservoir is not properly locked into place. • Select Rewind. The internal power source in your pump is unable to charge. Your pump is operating on the AA battery only. • Select OK to clear the alert. 4X.X U.
MP6025958-022 / A Title and text Explanation Next steps Pump error Your pump encountered an error and will restart. Your pump settings will return to factory default values. • When the pump restarts, follow instructions on the pump display. Delivery stopped. Current settings cleared. Pump restart needed. Select OK to restart and then reenter your settings. See User Guide. • After restart, check settings and re-enter values as needed.
MP6025958-022 / A Next steps Pump error Your pump encountered an error but a restart is not necessary. The issue is resolved. Your settings are not changed. • Select OK to resume basal delivery. Your pump has encountered a problem and has restarted. Your settings have not been changed. • Select OK to continue. Battery life is low and will be exhausted within 30 minutes. • Select OK to clear the alert. Delivery stopped. Settings unchanged. Select OK to continue. See User Guide.
MP6025958-022 / A Title and text Explanation Next steps Replace battery now Insulin delivery has stopped due to low power. Battery was not replaced after the Low battery Pump alert. Replace the battery immediately to resume insulin delivery. For details, see Removing the battery, on page 23. Your reservoir level is estimated at 0 units. • Select OK to clear the alert. A normal bolus delivery has been • Check the message to see how much of the bolus was actually Delivery stopped.
MP6025958-022 / A Title and text Explanation Next steps Resume Dual bolus? The Now portion of a Dual Wave bolus delivery has been interrupted because the pump battery was removed. If it is within 10 minutes since this interruption, you can resume this bolus. • Check the message to see how much of the Dual Wave bolus was actually delivered. The Square Wave bolus delivery was interrupted. If it is within 10 minutes since this interruption, you can resume this bolus.
MP6025958-022 / A Title and text Explanation Next steps Wrong key pressed Wrong button selected while unlocking the pump. • Select OK to clear the alert. Try again • Try to unlock the pump. Select the highlighted key when prompted. For more information, see Unlocking your pump, on page 26.
MP6025958-022 / A Title and text Explanation Next steps Alert on low XXX mg/dL Your SG value is at or below your specified low limit. • Select OK to clear the alert. Your SG value is at or below your specified low limit, and the pump has suspended insulin delivery due to a Suspend on low or Suspend before low event. • Select OK to clear the alert. Low sensor glucose. Check BG. Basal delivery resumed at XX:XX after suspend by sensor. Check BG.
MP6025958-022 / A Title and text Explanation Next steps Basal delivery resumed Your pump is resuming basal insulin delivery two hours after a Suspend before low or Suspend on low event occurred. • Select OK to clear the alert. Your pump is resuming basal insulin delivery two hours after a Suspend before low or Suspend on low event occurred. • Your pump has resumed basal insulin delivery; however, your SG value is still at or below your low limit. Maximum 2 hour suspend time reached. Check BG.
MP6025958-022 / A Next steps Calibrate now A BG meter reading is needed immediately to calibrate your sensor so that you can continue receiving sensor glucose readings. • Take a BG meter reading and enter for calibration. After you receive a Calibrate now alert, it takes up to 15 minutes after you calibrate for your pump to receive SG values. For details, see Calibrating your sensor, on page 183. Check BG and calibrate sensor. • If you are unable to calibrate now, you can use the Snooze feature.
MP6025958-022 / A Title and text Explanation Next steps Cannot find sensor signal The pump has not received a signal from the transmitter. • Disconnect and reconnect your transmitter and sensor. Disconnect and reconnect transmitter, then select OK. Notice if transmitter light blinks. • See if the light on your transmitter blinks when connected to the sensor. You may need this information for troubleshooting later. • Select OK. Your pump searches for your sensor.
MP6025958-022 / A Title and text Explanation Next steps Check connection The pump fails to detect the transmitter and is unable to receive sensor signal. • Select OK to clear the alert. Ensure transmitter and sensor connection is secure, then select OK. • If your sensor is fully inserted, select Yes. If your sensor is not fully inserted, select No. ■ • If your sensor was not fully inserted, insert a new sensor.
MP6025958-022 / A Title and text Explanation Next steps No calibration occurred The transmitter was unable to receive the calibration BG meter readings from the pump. • Select OK to clear the alert. Confirm sensor signal. Calibrate by XX:XX. • Check the status bar on your pump to ensure that your pump has a signal from your sensor. If there is no sensor signal, see My pump cannot find the sensor signal, on page 234.
MP6025958-022 / A Title and text Explanation Next steps Rise Alert Your SG value has been rising as fast or faster than your preset Rise Alert Limit. • Select OK to clear the alert. Sensor alert occurred when Alert Silence is on. • Select OK to clear the alert. Sensor glucose rising rapidly. • Check the Alarm History screen to see which alerts were silenced. For more information about accessing the Alarm History screen, see Alarm History, on page 122.
MP6025958-022 / A Title and text Explanation Next steps Sensor signal not found The pump has still not received a signal from the transmitter. When you reconnected the transmitter to the sensor, did you see a blinking green light on the transmitter? Did transmitter light blink when connected to sensor? • Select Yes or No and follow the instructions on the screen. Sensor signal not found See User Guide.
MP6025958-022 / A Title and text Explanation Next steps Suspend before low Your SG value is falling. Insulin delivery is suspended according to your Suspend before low setting and your SG is approaching your specified low limit. • Select OK to clear the alert. Your SG value is at or below the low limit you specified. • Select OK to clear the alert. Transmitter battery depleted The battery in the transmitter needs to • Select OK to clear the alert. Recharge transmitter now. be recharged.
MP6025958-022 / A Title and text Explanation Next steps Connect Device? The CareLink USB software is attempting to connect to your pump in preparation for data download. • Select Yes to allow connection, only if you are expecting or performing a data download. The download of pump data is taking • Select OK to clear the alert. Device with SN is trying to connect to your pump. Allow connection? Download slow Insulin delivery not affected. CareLink download may take longer than usual.
MP6025958-022 / A troubleshooting D E S E R A E L
MP6025958-022 / A D E S E R A E L
MP6025958-022 / A ■ This chapter contains procedures and information to help you understand and address conditions that might occur with your pump. For a list of alarms, alerts, and messages that may appear on your pump, see Pump alarms, alerts, and messages, on page 202. D E S Troubleshooting pump issues A E L WARNING: If you receive a critical error on your pump, the following screen displays and the pump sirens. E R Immediately disconnect from your insulin pump and discontinue use.
MP6025958-022 / A What is a Check Settings alarm? This alarm occurs when a condition causes your pump to reset to factory settings. You see this alarm after your pump guides you through re-entering the Startup Wizard settings. The Check Settings alarm is letting you know that other settings may have been cleared or reverted to factory default values. Review any settings that you have not already set in Startup Wizard and re-enter the values, if necessary.
MP6025958-022 / A Menu > Utilities > Self Test For details, see Self Test, on page 145. 6. If the Self Test does not complete successfully, or if you are concerned about your pump, call your local representative for assistance and check your BG. I cannot get to the Manage Settings screen ■ If you go to Menu > Utilities > Manage Settings, a message appears telling you that the feature is not normally accessible and to consult your user guide.
MP6025958-022 / A 17:00 BG 110 mg/dL Active Insulin Bolus 1.0 U Basal The Status screen appears. 2. D E S From the Status screen, you can select the type of status information you want to view. For example, to see a quick status of your pump and recent insulin deliveries, go to Quick Status. For details, see Status screens, on page 31. A E My pump is asking me to L enter my settings E R Certain pump errors can clear your settings and return them to their factory default values.
MP6025958-022 / A • Basal patterns After you enter your pump settings, you have the option of entering the following Bolus Wizard settings: • Carb ratio or Exchange ratio • Insulin sensitivity factor • BG target ■ To enter your pump settings: Begin entering your settings by selecting your language. Click Next to go to each new screen. 2. When the Select Time Format screen appears, select a 12-hour or a 24-hour time format. 3.
MP6025958-022 / A To enter your Bolus Wizard settings: 1. When your pump shows a list of settings for the Bolus Wizard, make sure you have the values you need before continuing. 2. Depending on the Carb Unit you set earlier, either the Carb Ratio or the Exch Ratio screen appears. Enter your carb ratio or exchange ratio by entering the End time and the Rate. You can adjust your carb or exchange ratio at any time. For details, see Changing your carb or exchange ratio, on page 69. 3.
MP6025958-022 / A Note: If you are using Alert Silence and currently have all sensor alerts silenced, your pump does not display the troubleshooting screens. Any glucose alerts you received are displayed in the Alarm History screen. 1. Make sure your setup meets the following requirements: • Your transmitter is fully charged. You have only one transmitter connected to your pump. Delete the current transmitter that is connected to your pump before continuing.
MP6025958-022 / A • If the transmitter light did not blink, you need to charge your transmitter. When your transmitter is charged, start your sensor. For details, see Starting the sensor, on page 182. • If your transmitter light blinked, but you still have no sensor signal, continue to the next step. 5. Move your pump closer to your transmitter and select OK. It can take up to 15 minutes for your pump to find the sensor signal. 6.
MP6025958-022 / A After a Suspend before low or Suspend on low event occurs, there is a period of time when the suspend functionality is unavailable. This time will vary depending on whether or not you respond to the suspend event. Typically, the suspend features will be unavailable for 30 minutes after your basal insulin delivery is resumed. For details, see When Suspend before low is unavailable, on page 159 or When Suspend on low is unavailable, on page 162. No SG values are available.
MP6025958-022 / A D E S A E L E R 238 Chapter 13
MP6025958-022 / A maintenance D E S E R A E L
MP6025958-022 / A D E S E R A E L
MP6025958-022 / A ■ maintenance Maintenance Read your warranty statement included with your pump for information about what is covered during your warranty period. Cleaning your pump Caution: Never use organic solvents, such as lighter fluid, nail polish remover, or paint thinner to clean your pump. Never use lubricants with your pump. When cleaning your pump, be sure to keep the reservoir compartment dry and away from moisture.
MP6025958-022 / A Cleaning your transmitter Always refer to your transmitter user guide for instructions on cleaning the transmitter. Storing your pump Storage mode allows you to safely place your pump in storage while not in use. Note: If you place your pump in storage mode, it is important to insert a new AA battery for 8 to 12 hours every six months to recharge the internal power source. WARNING: Active insulin is now cleared.
MP6025958-022 / A A Pump Error message appears. 2. Select OK. Your pump displays a Power Loss alarm. 3. Select OK. The Time & Date screen appears. Enter the current Time, Time Format and Date. 5. Select Save. ■ 4. 6. Select OK. Make sure that all of your settings, such as basal rate, are set as desired. If you need to, reapply your last saved settings by using the Restore Settings option as instructed in Restoring your settings, on page 143.
MP6025958-022 / A D E S A E L E R 244 Chapter 14
MP6025958-022 / A product specifications and safety information D E S E R A E L
MP6025958-022 / A D E S E R A E L
MP6025958-022 / A Product specifications This section provides detailed information on product specifications.
MP6025958-022 / A Minutes from Audio Audio and alarm or Vibration vibration alert 0 Audio Audio and vibrate Vibrate 1 Audio Audio and vibrate Vibrate 2 Audio Audio and vibrate Vibrate 3 Audio Audio and vibrate Vibrate 4 Audio Audio and vibrate Vibrate 5 Audio Audio and vibrate Vibrate 6 Audio and vibrate Audio and vibrate Audio and vibrate 7 Audio and vibrate Audio and vibrate Audio and vibrate 8 Audio and vibrate Audio and vibrate Audio and vibrate 9 Audio and vi
MP6025958-022 / A High Sensor Glucose 1312 Hz, followed by 1410 Hz, 1500 Hz, 1619 Hz, 1722 Hz Low SG 1722 Hz, 1619 Hz, 1500 Hz, 1410 Hz, 1312 Hz Lost SG 1485 Hz, followed by 1395 Hz, 1320 Hz, 1395 Hz Message tone 1655 Hz Reminder tone 934 Hz Fill tubing tone 1850 Hz Bolus delivery cancellation tone 1485 Hz, followed by 1655 Hz and 1485 Hz Loading complete tone 934 Hz Reservoir loading in progress tone 1850 Hz Easy Bolus activation 1045 Hz Easy Bolus step 1 increment 1175 Hz Easy Bolus
MP6025958-022 / A Basal patterns Maximum of 8 patterns. Each pattern covers a 24 hour period and can have up to 48 rates. Rates are set in 30 minute increments. Basal pattern names Fixed names: Basal 1, Basal 2, Basal 3, Basal 4, Basal 5, Workday, Day Off, Sick Day Increments • 0.025 units per hour for basal amounts in the range 0 to 0.975 units • 0.05 units per hour for basal amounts in the range 1 to 9.95 units • 0.
MP6025958-022 / A Fluid delivered/stroke • 0.25 μL (microliter) for 0.025 unit pump stroke • 0.5 μL for 0.05 unit pump stroke • 2.0 μL for 0.2 unit pump stroke Bolus Wizard feature default settings Default Limits Increments Carb units grams - - Insulin to carb (or exchange) ratio None 1–200 g/u (0.075–15.0 u/exch) 0.1 g/u for 1–9.9 g/u; 1 g/u for ratios of 10 g/u to 200 g/u (0.001 u/exch for 0.075–0.099 u/exch 0.01 u/exch for 0.10–9.99 u/exch; 0.
MP6025958-022 / A (food estimate) total bolus estimate = A B (correction estimate) C-D E + - active insulin where: A = food (grams) B = carb ratio C = current BG D = High BG Target E = insulin sensitivity Food estimate: Carb grams ÷ Carb ratio = Units of insulin Correction estimate: (Current BG - High BG Target) ÷ Insulin sensitivity - Active insulin = Units of insulin Total bolus estimate: D E S Food estimate + Correction estimate = Units of insulin 2.
MP6025958-022 / A Food estimate + Correction estimate = Units of insulin 3. If your current BG is between or equal to your High or Low BG Target, the total bolus estimate is based only on the food estimate. (food estimate) food (grams) ■ total bolus estimate = carb ratio Carb grams ÷ Carb ratio = Units of insulin Note: When the current BG is below the Low BG Target, an active insulin amount is not considered in the Bolus Wizard feature calculations. D E S Total bolus estimate = Food estimate 4.
MP6025958-022 / A Active Insulin Curve 100% 90% 8 Hour 80% 7 Hour 6 Hour 5 Hour % Insulin Remaining 70% 4 Hour 3 Hour 2 Hour 60% 50% 40% 30% 20% 10% 0% 0 1 2 3 4 5 6 7 8 Time (hours) D E S Graph adapted from Mudaliar and colleagues, Diabetes Care, Volume 22, Number 9, Sept. 1999, page 1501. A E L Carb ratios E R Maximum ratio settings 8 Range 1 to 200 grams/unit 0.075 to 15 units/exch Delivery accuracy • Delivery accuracy: ±5%.
MP6025958-022 / A Trumpet Curve for MMT−1511/1512 25 20 10 ■ product specifications and safety information Percent Error [%] 15 5 0 −5 −10 −15 −20 0 10 30 20 D E S 40 50 60 Overall Error Maximum Error Minimum Error 70 80 90 100 Observation Window [Minutes] A E L Easy Bolus E R Allows user to set up and deliver a Normal Bolus when the pump is in Sleep Mode. This is done using the and with the help of audio and vibration cues.
MP6025958-022 / A • Operating humidity range of the case: 20% to 90%. This requirement exceeds IEC 60601-1, subclause 7.9.3.1 (30% to 75%). Filling the infusion set and cannula • The cannula can be filled from 0.025 units to 5.1 units, in increments of 0.025 units. • The standard fill rate is 1.5 units per minute. The quick fill rate is 15 units per minute. • When filling the tubing, a warning occurs at 30 units, and thereafter at every 10 units.
MP6025958-022 / A Basal settings Limits Increments Max Basal Rate 2 U/h 0–35 U/h 0.025 U for 0.025–0.975 U/h 0.05 U for 1.00–9.95 U/h 0.1 U for rates of 10.0 U/h or more Basal Rate 0.000 U/h 0.000 U/h to Max Basal Rate setting 0.025 U for 0.025–0.975 U/h 0.05 U for 1.00–9.95 U/h 0.1 U for rates of 10.0 U/h or more Temp Basal Type Percent Percent, Rate N/A Temp Basal Percent 100% 0-200% 5% Temp Basal Rate Current basal rate 0.0 U/hr to Max Basal Rate ED S A E L E 0.025 U for 0.025–0.
MP6025958-022 / A Alert Alert range Increment type Default value Time First reminder occurs at 2 to 24 hours. Second reminder occurs one hour before empty. The 30 min 8 hours 1 unit 20 units second reminder is automatic and cannot be changed by the user. Units First reminder occurs at 5 to 50 units. Second reminder occurs at 50 percent of the remaining specified amount. The second reminder is automatic and cannot be changed by the user.
MP6025958-022 / A Percent temp basal The default value is 100 percent of basal programming. Range 0 to 200% Default 100% of basal programming Increment 5% ■ A single fault condition will cause the pump to suspend insulin delivery. Maximum infusion with a single fault condition is 0.2 units. Pump dimensions The MMT-1512 pump dimensions in inches are approximately 2.1 width x 3.34 length x 0.96 depth. D E S The MMT-1512 pump dimensions in centimeters are approximately 5.3 width x 8.5 length x 2.
MP6025958-022 / A Sensor default settings High sensor settings Item Default Limits Increments setting High SG alert limit 250 mg/dL 100 to 400 mg/dL 5 mg/dL Alert before high Off - - Alert on high Off - - Time before high 15 minutes 5 to 30 minutes 5 minutes Rise Alert Off - - Rise Limit Two up arrows • 1up arrow (1 mg/dL/min) • 2 up arrows (2 mg/dL/min) D E S • 3 up arrows (3 mg/dL/min) A E L • Custom limit (1.0 to 5.
MP6025958-022 / A Low sensor settings Item Default Limits Increments - - setting Resume basal alert Off ■ product specifications and safety information D E S A E L E R Product specifications and safety information 261
MP6025958-022 / A Guidance and manufacturer's declaration Guidance and Manufacturer's Declaration - Electromagnetic Emissions The MiniMed 640G insulin pump is intended for use in the electromagnetic environment specified below. The customer or the user of the MiniMed insulin pump should make sure that it is used in such an environment. Emissions Test Compliance RF emissions • 6 dB and 99% Bandwidths: Pass Test: 47 CFR Part 15, Subpart C Section 15.
MP6025958-022 / A Guidance and Manufacturer's Declaration - Electromagnetic Immunity The MiniMed 640G insulin pump is intended for use in the electromagnetic environment specified below. The customer or the user of the MiniMed insulin pump should assure that it is used in such an environment.
MP6025958-022 / A Guidance and Manufacturer's Declaration - Electromagnetic Immunity The MiniMed 640G insulin pump is intended for use in the electromagnetic environment specified below. The customer or user of the MiniMed insulin pump should assure that it is used in such an electromagnetic environment.
MP6025958-022 / A Guidance and Manufacturer's Declaration - Electromagnetic Immunity The MiniMed 640G insulin pump is intended for use in the electromagnetic environment specified below. The customer or user of the MiniMed insulin pump should assure that it is used in such an electromagnetic environment. IEC 61000-4-3 80 MHz to 800 MHz 10 V/m 80 MHz to 800 MHz 10 V/m 800 MHz to 2.
MP6025958-022 / A Guidance and Manufacturer's Declaration - Electromagnetic Immunity The MiniMed 640G insulin pump is intended for use in the electromagnetic environment specified below. The customer or user of the MiniMed insulin pump should assure that it is used in such an electromagnetic environment.
MP6025958-022 / A Recommended separation distances between the MiniMed 640G insulin pump and common household radio transmitters Household RF Transmitter Frequency Recommended Separation Distance (meter) Recommended Separation Distance (inch) Cordless Household 2.4 GHz 0.3 12 Cordless Household 5.8 GHz 0.3 12 TDMA-50 Hz (cell phone) 1.9 GHz 0.3 12 TDMA-50 Hz (cell phone) 800 MHz 0.3 12 PCS (cell phone) 1.9 MHz 0.3 12 DCS (cell phone) 1.8 MHz 0.3 12 GSM (cell phone) 900 MHz 0.
MP6025958-022 / A Recommended separation distances between portable and mobile RF communications equipment and the MiniMed 640G insulin pump The insulin pump is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
MP6025958-022 / A an energy scan. The pump will perform the energy scan if after 10 minutes no CGM transmitter signal has been received. If the channel change occurs the pump will send beacons on the new channel. product specifications and safety information In normal operation the CGM transmitter will transmit a packet every 5 minutes and retransmit the packet if the data is corrupted or missed. Radio frequency (RF) communications specifications Utilizes the IEEE 802.15.
MP6025958-022 / A Icon table Follow instructions for use Manufacturer Date of manufacture (year - month) Use by (year - month) Catalogue number Device serial number Storage temperature range Storage humidity range Magnetic Resonance (MR) unsafe D E S A E L Type BF equipment (protection from electrical shock) Radio communication E R Pump: Protected against the effects of continuous immersion in water (12 feet or 3.6 meters for 24 hours).
MP6025958-022 / A appendix a: end user software license agreement D E S E R A E L
MP6025958-022 / A D E S E R A E L
MP6025958-022 / A ■ End user software license agreement NOTICE TO USER: Certain portions of software contained in this product may be covered by the GNU General Public License, Version 2 or Version 3 (“Open Source”), which can be obtained through the GNU web site at www.gnu.org/copyleft/ gpl.html. The source code for any Open Source can be obtained, for a nominal fee to cover the cost of shipping and media, by contacting Medtronic MiniMed, Inc.
MP6025958-022 / A D E S A E L E R 274 Appendix A
MP6025958-022 / A glossary D E S E R A E L
MP6025958-022 / A D E S E R A E L
MP6025958-022 / A ■ glossary Glossary active insulin Bolus insulin that has been delivered by the pump and is still working to lower your blood glucose levels. active insulin adjustment The amount of insulin that is subtracted from your BG correction bolus to account for the active insulin that is D E S tracked by the Bolus Wizard. Active Insulin Time A Bolus Wizard setting that lets you set the length of time that bolus insulin is tracked as active insulin.
MP6025958-022 / A Alert on low An alert that occurs when your sensor glucose value reaches or falls below your low limit. Auto Suspend An alarm that you set to suspend insulin delivery and trigger an alarm if no buttons are pressed for a specified period of time. Clearing the alarm resumes insulin delivery. Awake mode A state in which the pump screen is on. Unless you are actively using another screen, your Home screen appears.
MP6025958-022 / A A feature that uses your individual Bolus Wizard settings to calculate an estimated bolus amount based on the BG values and carbs that you enter. These settings include Carb Ratio, Insulin Sensitivity Factor, BG Target Range, and Active Insulin Time. calibrate The process of using a meter blood glucose reading to calculate sensor glucose values. Calibration reminder A reminder you can set to let you know when your next calibration is due.
MP6025958-022 / A exchange ratio The number of insulin units that are needed to cover 1 carbohydrate exchange. The exchange ratio is based on your individual needs and is used to calculate bolus amounts. food bolus A dose of insulin you give to cover an expected rise in glucose levels from carbohydrates. High limit The value you set to determine when the pump will alert you of a high sensor glucose condition.
MP6025958-022 / A A reminder that a bolus was not delivered during time periods that you specify, often set around your meal times. Normal Bolus A type of bolus that provides an entire dose of insulin immediately. notifications All notifications are designed to get your attention and convey different types of information. They include alarms, alerts, reminders, and messages. occlusion A blockage or crimp of the cannula or tubing that prevents proper insulin flow.
MP6025958-022 / A Rewind A feature used when changing a reservoir. It returns the piston to its starting position and allows a new reservoir to be placed into the pump. sensitivity See insulin sensitivity factor. sensor (glucose sensor) The small part of the continuous glucose monitoring system that you insert just below your skin to measure glucose levels in your interstitial fluid.
MP6025958-022 / A transmitter A device that connects to a glucose sensor. The transmitter collects data measured by the sensor and wirelessly sends this data to monitoring devices.
MP6025958-022 / A D E S A E L E R 284 Glossary
MP6025958-022 / A index D E S E R A E L
MP6025958-022 / A D E S E R A E L
MP6025958-022 / A ■ Index accessories 15 active insulin about 71 adjustment 71 clearing settings 144 Home screen 26 injection warning 71 viewing amount 27 Active Insulin cleared alert 202 Active Insulin Time about 66 changing 71 Airplane Mode setting up 138 status bar 29 using 137 Alarm History 122 alarms about 199 audio options 201 Auto Suspend 203 Battery failed 203 Battery not compatible 203 Bolus not delivered 203 Bolus stopped 203 CGM 216 Critical pump error 204 Delivery limit exceeded 205 emergency
MP6025958-022 / A Cannot find sensor signal 220 CGM 216 Change sensor 220 Check connection 221 Check settings 204 Download slow 226 Lost sensor signal 221 Low battery Pump 209 Low battery transmitter 221 Low reservoir 210 No calibration occurred 222 notification light 201 overview 201 Possible signal interference 222 pump 202, 225 Reservoir estimate 214 Rise alert 223 sensor 216 Sensor alert occurred 223 Sensor expired 223 Sensor signal not found 224 SG value not available 224 silencing 193 Transmitter bat
MP6025958-022 / A bolus types DualWave 60 examples 60 Normal 59 Square Wave 60 Bolus Wizard Active Insulin Time 66 BG target 66 Carb ratio 65 Dual Wave 79 Exchange ratio 65 High BG warning 72 injection warning 71 Insulin Sensitivity Factor 65 Low BG warning 72 Max Bolus Exceeded warning 73 Normal bolus 73 Square Wave 77 warnings 72 CD E S A E L E R ■ index BG not received alert 218 BG target Bolus Wizard 66 setting up 70 Block Mode status bar 31 using 139 blood glucose Home screen 26 sensor calibratio
MP6025958-022 / A clearing active insulin 144 pump settings 143 Connect Device message 226 connecting pump, transmitter automatically 175 manually 178 connection, failure 178 Consumable infusion set 13 reservoir 13 continuous glucose monitoring about 151 airplane mode 137 alarms 216 alerts 216 calibrating sensor 183 connecting pump, transmitter 175 examples 164 High Settings 156 home screen 152 Low Settings 157 messages 216 sensor feature, turning on 167 sensor glucose reading 191 sensor graph 191 settings
MP6025958-022 / A D E S I icons Home Screen 28 sensor status 152 SmartGuard 155, 159, 162 suspend by sensor 155 infusion set about 93 best insertion sites 103 disconnect Quick-set 107 fill cannula alarm 205 fill tubing 101 inserting 103 reconnect Quick-set 107 reservoir alarm 209 reservoir and tubing 33 reservoir removal 93 rotating sites 104 setting up 93 type 13 Insert battery alarm 205 A E L E R inserting infusion set 103 reservoir 99 sensor 181 insulin basal 39 basal patterns 41 bolus 59 flow block
MP6025958-022 / A M maintenance 241 Manage Settings error alarm 210 managing, pump settings 142 Manual bolus Dual Wave 81 Easy bolus 84 Normal bolus 75 Square Wave 78 Max Basal alarm 205 rate 40 Max Basal Rate setting 41 Max Bolus alarm 205 exceeded 73 setting 62 warning 63 Max Fill reached alarm 210, 211 Menu screen about 32 accessing 32 audio options 33 display 21 event markers 33 history 33 insulin settings 33 reminders 33 reservoir and tubing 33 sensor settings 33 suspend delivery 33 utilities 33 messa
MP6025958-022 / A Quick-set infusion set disconnecting 107 reconnecting 108 R D E S A E L E R Q quick status 31 rate, basal about 39 Preset Temp 40 Temp 40, 47 rate of change arrows 152 red light 200, 201 reminders about 129 Bolus BG Check 130 Calibration 133 Low Reservoir 132 Menu screen 33 Missed Meal Bolus 131 Personal 129 Set Change 133 remote bolus 62, 112 Replace battery alarm 213, 214 reservoir about 93 compartment image 19 inserting 99 Loading incomplete alarm 209 Low Reservoir alert 210 Max
MP6025958-022 / A Resume Square bolus message 215 resuming suspended delivery 54, 163, 174 return to previous screen 21 RF communications 267, 268 Rise Alert responding to 223 SG setting 157 summary screen 120 Rise Limit SG setting 157 S saving power 34 pump settings 142 screen Active Insulin Time 71 Airplane Mode 138 Alarm History 122 Audio Options 139 Auto Suspend 139 Basal Pattern Setup 46 BG Check 130 BG Target 70 Block Mode 140 Bolus Increment 64 Bolus Speed 64 Bolus Wizard Setup 66, 71 Carb Ratio 69
MP6025958-022 / A D E S A E L E R Square Wave 60 about 76 Bolus Wizard 77 example 60 Manual bolus 78 resume message 214, 215 setting up 76 starting sensor 182 Startup Wizard about 24 re-entering settings 232 using 24 status bar about 28 Airplane Mode 29 audio mode 29 battery 28 Block Mode 31 CGM, with 152 CGM, without 26 Connection icon 29 icons 28 insulin amount 29 sensor calibration 30 sensor life 30 time 26 status screens about 31 notifications 31 pump 31 quick status 31 sensor 31 settings review 32
MP6025958-022 / A U overview 118 total carbs 118 total daily dose 118 viewing 117 supplies, ordering 15 Suspend before low about 158 examples 164, 165, 166 unavailable 159 suspend by sensor icon 155 Suspend on low about 161 examples 166 unavailable 162 suspended delivery, resuming automatically 163 manually 54, 174 suspending insulin delivery 54, 139, 158, 161 unavailable Suspend before low 159 Suspend on low 162 unlocking 26 utilities 33 V viewing history, pump settings 145 W wireless connection autom
MP6025958-022 / A ■ index D E S A E L E R Index 297
MP6025958-022 / A D E S A E L E R 298 Index
MP6025958-022 / A D E S E R A E L
MP6025958-022 / A D E S A E L E R 6025958-022_a
MP6025958 E L E S A