FCC submission Product Manual October 21, 2004 Medic4All Ltd.
Medic4All TELEMEDICINE SOLUTIONS Operation Manual Wireless Blood Pressure Monitor – BP 102 The Blood Pressure Monitor is a wireless monitoring device. It measures, stores, and transfers blood pressure and heart rate data to a medical monitoring center using the MedicGate™ or MiniGate™. Please read this booklet carefully in order to achieve optimal performance of the Wireless Blood Pressure Monitor. Ref: OM-109v1 Copyright© 2003-2004, Medic4All AG, all rights reserved.
Warnings & Precautions ♦ Consult your local provider concerning any operational difficulties. Do not attempt to repair the Wireless Blood Pressure Monitor on your own. Any modifications made to the equipment void the device warranty and service contract obligations. ♦ Keep the Wireless Blood Pressure Monitor away from electro-magnetic emitting devices. ♦ Do not store the Wireless Blood Pressure Monitor in direct sunlight, at a high temperature, or in a location with excessive humidity.
3. Wrap the cuff around the upper arm about 2-3 cm (1 inch) above the elbow, as shown below. Place the cuff directly against the skin, as clothing may cause the Wireless Blood Pressure Monitor to record a faint heartbeat and result in an error. 1. Verify that the index on the arm cuff points within the proper fit range. Note: Constriction of the upper arm caused by rolling up a shirt sleeve may prevent the Blood Pressure Monitor from recording accurate readings. 2. Sit comfortably around a table.
. The micro pump will pressurize the cuff. Note The cuff will become very tight. Do not be alarmed, as this is normal. When pressurization is complete, the automatic exhaust mechanism gradually reduces the pressure in the cuff. When the measurement is complete, the buzzer sounds and the air is automatically released from the cuff. The systolic pressure, diastolic pressure, and pulse rate are displayed. 6. Press the START 1 or START 2 button to turn off the device.
Replacing Batteries Replace all four batteries with new ones, when is displayed on the display screen. To replace the batteries in the Wireless Blood Pressure Monitor: 1.Remove the battery compartment cover. Push the cover up and lift at the arrow indication. 2. Insert the new batteries into the Battery Compartment, taking extreme care that you observe the correct polarity: (+) and (-). Use only four “AA” Alkaline batteries. 3. Slide the battery cover down into place to close it.
Use only batteries that match the requirements specified in the Specifications section. Please take the batteries out if the device is not used for a long period (app. 2 months or more) Do not mix old and new batteries, it may shorten the battery life or cause the device to malfunction. Disposal of this product and used batteries should be carried out in accordance with the relevant national regulations.
Specifications Medical measurement Measurement method Measurement range Accuracy Radio Frequency communication Frequency Max transmission power Bandwidth Transmission range Oscillometric measurement Pressure: 20 – 280 mmHg Pulse: 40 – 200 beats / minute Pressure: +/- 3 mmHg or 2%, whichever is greater Pulse: +/- 5% 433.92 MHz (EU) 866.3 MHz (EU) 915 MHz (USA) 10 dbm (EIRP) at 433.92 MHz and 866.3 MHz. -1 dbm (EIRP) at 915 MHz 60 kHz max. 200 m (free space) Power Supply Batteries 4 x 1.
Complies with the CE standard under the Euro Medical Device Directive. FCC Part 15 Information This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the Federal Communication Commission (FCC) Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
Contents 1. 4 x 1.5v AA batteries 2. Arm cuff 3. Operation Manual 4. Wireless Blood Pressure monitor Warranty This product is warranted against defects in material and workmanship beginning from the date of shipment for a period of one year (the “Term”). Under the Term, the obligation of Telcomed Advanced Industries Ltd (the “Company”) will be limited to repair or replacement of the product (not including batteries, cables and spendable parts), according to the Company’s discretion.