Root™ Operator's Manual
Root operating instructions provide the necessary information for proper operation of Root. There may be information provided in this manual that is not relevant for your system. Do not operate Root without completely reading and understanding the instructions in this manual.
The use of the trademarks PATIENT SAFETYNET and PSN is under license from University HealthSystem Consortium. All other trademarks and registered trademarks are property of their respective owners. www.masimo.
Contents About this Manual---------------------------------------------------------------------------- 7 Product Description and Intended Use ------------------------------------------------- 9 Product Description ---------------------------------------------------------------------- 9 Intended Use ------------------------------------------------------------------------------ 9 Safety Information, Warnings, and Cautions---------------------------------------- 11 Safety Information, Warnings, and Cautions --
Root Contents Trend Settings --------------------------------------------------------------------------- 53 Profiles ------------------------------------------------------------------------------------ 54 Iris ------------------------------------------------------------------------------------------ 54 Layout ------------------------------------------------------------------------------------- 55 Sounds ------------------------------------------------------------------------------------ 57 Alarm Interface
Root Contents Replacing the Fuses ------------------------------------------------------------------- 88 Power-On Self Test -------------------------------------------------------------------- 88 Nurse Call Setting Connections----------------------------------------------------- 88 Battery Test ------------------------------------------------------------------------------ 90 Repair Policy----------------------------------------------------------------------------- 90 Return Procedure----------------------
About this Manual This manual explains how to set up and use Root. Important safety information relating to general use of the product appears in this manual. Read and follow any warnings, cautions, and notes presented throughout this manual. The following are explanations of warnings, cautions and notes. A warning is given when actions may result in a serious outcome (for example, injury, serious adverse effect and, death) to the patient or user.
Product Description and Intended Use Product Description Root is a patient monitoring and connectivity platform to transform care. It offers multiple high-impact innovations for broad applications across the continuum of care.
Safety Information, Warnings, and Cautions The following section lists warnings, cautions, notes, and safety information for Root. Root is to be operated by qualified personnel only. The manual, accessories, directions for use, all precautionary information, and specifications should be read before use. Safety Information, Warnings, and Cautions WARNING: Do not use Root if it appears to be damaged. Contact your biomedical engineering department for guidance before using Root.
Root Safety Information, Warnings, and Cautions General Information, Warnings and Cautions WARNING: When placed on a flat surface, the device should be secured with a mounting system recommended by Masimo. WARNING: Do not touch, press or rub the touchscreen with abrasive cleaning compounds, instruments, brushes or rough-surface materials, or bring them into contact with anything that could scratch the touchscreen.
Root Safety Information, Warnings, and Cautions Compliance Information, Warnings and Cautions WARNING: In accordance with international telecommunication requirements, the frequency band of 2.4 GHz and 5.15 to 5.25 GHz is only for indoor usage to reduce potential for harmful interference to co-channel mobile satellite systems. CAUTION: Disposal of product: Comply with local laws in the disposal of the instrument and/or its accessories.
Root Safety Information, Warnings, and Cautions Note: This Class A digital apparatus complies with Canadian ICES-003. Note: Use Root in accordance with Environmental Specifications section in this manual. www.masimo.
Root Safety Information, Warnings, and Cautions Electrical Information, Warnings and Cautions WARNING: Root may be used during electrocautery, but this may affect the accuracy or availability of the parameters and measurements. WARNING: Root may be used during defibrillation, but this may affect the accuracy or availability of the parameters and measurements. WARNING: Electrical shock hazard: Do not remove the back panel of the device. This could cause injury to personnel or device damage.
Root Safety Information, Warnings, and Cautions CAUTION: Do not under any circumstances remove the grounding conductor from the power plug. CAUTION: Do not use extension cords or adapters of any type. CAUTION: Only use the AC power cable provided by Masimo. Using a different AC power cable could cause damage to Root. Check the power cord and plug to ensure that it is intact and undamaged. CAUTION: Use a grounded outlet for proper grounding. A hospital-grade outlet is required.
Chapter 1: Description Root can be used in the following ways: As a docking station and charger for Radical-7 As a bedside monitoring display for parameters on Radical-7 and MOC-9 modules As a connectivity gateway for standalone devices. Features Front View Ref. Feature Description 1 Docking Station Provides a docking station for the Radical-7. While docked, the Radical-7 can communicate monitored parameters and measurements.
Root Chapter 1: Description Ref. Feature Description 5 AC Power Indicator Shows an indication of AC power connection Root. 6 Radical-7 Charging Indicator Shows an indication of battery charge for the Radical-7 in the Docking Station. *Only the touchscreen version of the Radical-7 is able to communicate monitored parameters and measurements. All other versions can only charge in the docking station but not communicate with Root.
Root Chapter 1: Description Feature Description 4 Ethernet Port Provides a network connection to Root using an RJ-45 cable. 5 USB Ports (2) Provide USB 2.0 connectivity. 6 Power Entry Module Contains the input connector for a hospital grade AC power cord and the fuse holder. 7 Equipotential Ground Connector Provides electrical grounding for Root. 8 Iris Connectivity Ports (4) Provide connection for standalone devices. Side Views Right side Feature www.masimo.
Root Feature 1 Description System Status Provides an indication of alarms and system messages. See System Status Lights on page 63. Lights 2 Power Button 3 Chapter 1: Description MOC-9 Ports (3) www.masimo.com Places Root in Power On, Standby and Power Off modes. Provide connectivity for MOC-9 modules.
Chapter 2: Setting Up Unpacking and Inspection To unpack and inspect Root 1. 2. Remove Root from the shipping carton and examine it for signs of shipping damage or exposed electronics. Check all materials against the packing list. Save all packing materials, invoice and bill of lading. These may be required to process a claim with the carrier. If anything is missing or damaged, contact Masimo's Technical Service Department. See Return Procedure on page 91.
Root Chapter 2: Setting Up Power On The Power Button can be used for Power On, Standby, and Power Off. To Power On, press the Power Button for two (2) seconds until a single audible tone sounds. Once Root turns on, if no Radical-7 or MOC-9 module is connected, the Root display shows the following message: Please Connect a Device. The user is now able to connect Radical-7 and MOC-9 modules. www.masimo.
Root Chapter 2: Setting Up For information about Standby Mode and Power Off, See Standby and Power Off on page 65. Initial Battery Charging To charge the battery for the first time 1. Securely plug the AC power cord into power entry module. 2. Plug the hospital grade AC power cord into an AC power source. 3. Verify that Root's battery is charging by ensuring that the AC Power Indicator (1) is green and the Battery icon on the Status Bar (2) is solid green or has the charging symbol.
Root Chapter 2: Setting Up Radical-7 Connection It is recommended that Root be powered on before performing the steps below. Snap the Radical-7 into the Docking Station. If the Radical-7 is not yet turned on, press the power button on the Radical-7 to power it on. When properly connected, the Radical-7 Charging Indicator light will illuminate. An illuminated Radical-7 Battery icon will also appear in the Status Bar. Root display will show active measurements and parameters.
Root Chapter 2: Setting Up To connect to a Nurse Call System 1. 2. 3. 4. Identify the Nurse Call connection end (1/4 inch round female connector) of the cable. Insert the Nurse Call connection cable securely into the compatible port on Root. Depending on the connection type of the Nurse Call System, it may be necessary to orient the other end of the Nurse Call connection cable to fit correctly into the system connection. For more information, see Device Output on page 51. www.masimo.
Chapter 3: Operation The information in this chapter assumes that Root is set up and ready for use. This chapter provides necessary information for proper operation of the device. Do not operate Root without completely reading and understanding these instructions. www.masimo.
Root Chapter 3: Operation About the Main Screen The Main Screen consists of several features. The following shows the Main Screen when three different technologies are connected, Radical-7 (top) showing rainbow® parameters and measurements, Phasein module (middle) showing capnography measurements and SedLine module (bottom) showing brain function measurements. Ref. Feature Description 1 Status Bar Displays system status as well as icons that provide shortcuts to menu items or actions.
Root Chapter 3: Operation About the Status Bar At the top of the Main Screen is the Status Bar with interactive icons. Each icon provides a shortcut to a menu item or an action on Root. An example is shown below. Ref. Feature 1 2 3 Alarm Silence Audio Pause Profiles Description Displays alarm status and mutes all audible alarms for Root, Radical-7, and MOC-9 modules. See Alarm Silence on page 60. Displays Audio Pause status and temporarily silences an alarm event. See Audio Pause on page 62.
Root Chapter 3: Operation Ref. Feature Description 7 Provides access to the Iris screen. The example shown above indicates that standalone devices are connected to Ports 1, 2 and 3 and the information is being sent to Patient SafetyNet or Connectivity Gateway. Iris The color of the icon matches the status colors of connected standalone devices on the Iris screen. See Iris on page 54. 8 Radical-7 Battery Displays charging status for Radical-7 and provides access to the Battery Radical screen.
Root Chapter 3: Operation Using the Touchscreen Interface Using the gestures described below, the user is able to customize the viewing experience, including displaying the highest priority parameters and measurements. The availability of navigation features is dependent on the connected Radical-7 and MOC-9 module(s). Action Illustration Example Description Press Touch and release. Action performed once finger is released. Slide Touch, move (left, right, up or down), and release.
Root Control Chapter 3: Operation Applicable Actions Description toggle Labeled Toggle Slide knob Switches between toggle states Press left or right of Quickly moves knob left or right toggle Spinner Slider Press label Quickly moves knob left or right Press center (focused) tile When closed, expands the spinner When open, collapses the spinner Swipe up or down When open, scrolls through spinner tiles Press unfocused tile When open, scrolls tile into center (focused) position Press anywher
Root Control Chapter 3: Operation Applicable Actions Description Press anywhere outside spinner When open, collapses spinner Button Press Performs action (as defined by the button description) Icon Menu Press tile Opens menu specified by tile Swipe left or right (anywhere) Scrolls icons left or right Press bottom indicator icon Quickly centers tile corresponding to indicator icon Press parameter or measurement When no parameter or measurement alarm is present, opens parameter or measuremen
Root Chapter 3: Operation Control Applicable Actions Description Trend Line Pinch in Zooms in Pinch out Zooms out Pan Changes time range Press y-axis Opens parameter or measurement trend menu Press '+' Increases time range Press '-' Decreases time range Press time label Resets time range to default Trend Zoom Alarm Silence Press icon Silences all alarms Audio Pause icon Press Enables Audio Pause Other Status Bar icons Press Opens relevant menu Back Arrow Press Exits menu, aban
Root Chapter 3: Operation To navigate to the previous screen, press the arrow at the top left corner of the touchscreen. To return to the Main Screen, at any time, press the Home Button at any time. The Home Button is always illuminated when Root is powered on. Understanding Windows Root creates a Window for Radical-7 and every MOC-9 module connected. Parameters or measurements can be expanded or minimized within a Window to customize view. Radical-7 Windows are shown in the examples below.
Root Chapter 3: Operation Feature Description 1 Window The area where all data from a docked Radical-7 or connected MOC-9 module are displayed. 2a Trend View Select to see Trend View of the parameters and icon measurements, if supported. 2b Analog View icon Select to see Analog View of the parameters and measurements, if supported. 3 Waveform Shows a parameter or measurement over time (only for Radical-7 and MOC-9 modules).
Root Chapter 3: Operation Using Trend View In Trend View, a parameter or measurement is displayed as a graph of its values over time. If a Window supports Trend View, then (Trend View icon) displays in the upper right corner. Press the icon to view parameter or measurement data in Trend View. The following diagram and table describe key features of a parameter's Trend Display in Trend View. Feature Description 1 Value Range Indicates current viewing of the parameter or measurement.
Root Chapter 3: Operation Using Analog View The Analog View shows parameter and measurement data as a needle pointing to graduations in a circular array around a dial. This view provides indications of change that can be interpreted at a quick glance. Analog View displays alarming and normal ranges of a parameter or measurement. These indicators can be used to alert clinicians to a patient's condition.
Root Chapter 3: Operation Ref. Feature 4 Parameter or Measurement Label www.masimo.com Description Indicates the name of the parameter or measurement.
Root Chapter 3: Operation Specific ranges of the Analog View are: Ref. Feature Color Description 1 Normal Range White Area of the display range where an alarm will not be triggered. 2 Caution Range Yellow Area of the display range that provides a caution indicator. 3 Alarming Range Red Area of the display range where an alarm will be triggered. Some ranges display as quarter circles, others display as half circles.
Root Chapter 3: Operation Customizing Windows Windows can be customized by expanding and minimizing parameters and measurements in both Trend View and Analog View. When a parameter is minimized, it is only displayed the Well with its Numeric Value and Parameter Label. When a parameter is expanded, it will be shown as either a Trend Display or Gauge. To expand a parameter or measurement Order Instruction Step 1 Press and hold the Numeric Value until it dims.
Root Chapter 3: Operation Minimizing a parameter or measurement Order Instruction Step 1 Press and hold the Numeric Value until it shrinks. Step 2 Drag the Numeric Value to the Well. Step 3 Release the Numeric Value. www.masimo.
Root Chapter 3: Operation Accessing Main Menu Options To access the Main Menu options At the bottom right corner of the touchscreen, press the Main Menu icon. The Main Menu options are: Device Settings See Device Settings on page 44. About See About on page 53. Trend Settings See Trend Settings on page 53. Profiles See Profiles on page 54. Iris See Iris on page 54. Layout See Layout on page 55. Sounds See Sounds on page 57. www.masimo.
Root Chapter 3: Operation Device Settings The Device Settings menu allows the user to view and customize settings for Root. The Device Settings options are: Localization See Localization on page 45. Wi-Fi See Wi-Fi on page 46. Ethernet See Ethernet on page 46. Bluetooth See Bluetooth on page 47. Battery Root See Battery Root on page 47. Battery Radical See Battery Radical on page 48 Brightness See Brightness on page 49. Access Control See Access Control on page 50.
Root Chapter 3: Operation Localization Use the Localization screen to view the current date and time and configure settings related to local time, language and geography. The user can also access the Localization screen by pressing the current time on the Status Bar. See About the Status Bar on page 29. Option Description Language Selects the language display for Root. English Date Format Sets the display format for current date. mm/dd/yy Time Sets the display format for current time.
Root Chapter 3: Operation Wi-Fi Use the Wi-Fi screen to enable or disable Wi-Fi connectivity. When Root is connected to a Wi-Fi network, the Wi-Fi icon on the Status Bar conveys the strength of the connection. See About the Status Bar on page 29. Option Description Wi-Fi Enables or disables Wi-Fi connectivity.
Root Chapter 3: Operation Bluetooth Use the Bluetooth screen to enable or disable Bluetooth connectivity. When Bluetooth connectivity is enabled, the Bluetooth icon will appear in the Status Bar. See About the Status Bar on page 29. Option Description Bluetooth Enables or disables Bluetooth connectivity. Factory Configurable Default Setting Settings Off On or Off Battery Root Use the Battery Root screen to view the specific percentage of charge on the battery.
Root Chapter 3: Operation Battery Radical Use the Battery screen to view the specific percentage of charge on the Radical-7's battery. The user can also access the Battery screen by pressing the Radical-7's Battery icon on the Status Bar. See About the Status Bar on page 29. Option Description State of Charge Provides a read-only display of battery level remaining. Battery Diagnostics Allows trained personnel to access battery diagnostic information www.masimo.
Root Chapter 3: Operation Brightness Use the Brightness screen to adjust the brightness of the Root display. Option Description Auto Brightness Allows automatic adjustment of Root's display brightness Off based on ambient light. On or Off Brightness Adjust the brightness level of the Root display by sliding the button (4 is brightest). 1, 2, 3, 4 www.masimo.
Root Chapter 3: Operation Access Control Access Control contains configurable options and settings that require a password. To enter Access Control 1. 2. 3. Press the key. When the numeric screen displays, enter the following numbers: 6 2 74 Asterisks (****) will be displayed. To undo an entry, press Backspace. Press Enter to access the password protected screen. Note: The password will have to be entered every time this screen is accessed.
Root Chapter 3: Operation Option Description Factory Default Setting Configurable Settings Save as Pediatric Saves current profile parameter as the Pediatric Profile. N/A Press Save to load. Save as Neo Saves current profile parameter as the Neonatal Profile. N/A Press Save to load. Factory Defaults Options are restored to factory N/A values. Press Restore. Device Output Device Output configures the Nurse Call feature on Root.
Root Chapter 3: Operation Note: The Nurse Call feature is disabled when Audio Pause is enabled and Nurse Call Trigger is set to Alarms. For more information about Audio Pause, see Audio Pause on page 62. www.masimo.
Root Chapter 3: Operation About Use the About screen to view the serial number as well as software and hardware version information about Root. These details may be helpful during troubleshooting. Option Description Serial Number Displays the serial number for the device. MCU 1 Displays software version number. Processor Displays processor version number. MCU 2 Displays software version number. Information about Radical-7 and MOC-9 modules will display in a separate list.
Root Chapter 3: Operation To configure trend settings of specific parameters and measurements, refer to the instructions for use or operator's manuals for Radical-7 and appropriate MOC-9 module(s). Profiles Use the Profiles screen to select patient type. Option Description Patient Type Identifies the Profiles setting in the device. Factory Configurable Default Setting Settings Adult Adult, Pediatric, Neonatal Root can be configured for various patient types by using the Profiles feature.
Root Chapter 3: Operation Layout Use the Layout screen to select sizing options for Windows and Trend Displays. Additional Settings There are different ways to display the parameters and measurements by changing the Layout Style. Note: This feature only applies to Trend View. See Using Trend View on page 37. Option Description Layout Style Controls the sizing of Trend Displays.
Root Chapter 3: Operation Fixed Ref. Description 1 A set number of Trend Displays can be shown at the same time and all Trend Displays are fixed in size. Every additional parameter or measurement expanded will replace an existing Trend Display. For more information about expanding parameters, see Customizing Windows on page 41. 2 Size of each Trend Display is fixed. Dynamic Ref. Description 1 Size of all Trend Displays decreases or increases to accommodate parameter(s) expanded or minimized.
Root Chapter 3: Operation Available Layout When only Radical-7 or a single MOC-9 module is connected to Root, the Available Layout will be 100%. When Radical-7 and/or multiple MOC-9 modules are connected, the user will have the option to select from several pre-configured layouts. Sounds www.masimo.
Root Chapter 3: Operation Use the Sounds screen to control the volume level of sounds and duration of audio pause for Root. Option Description Factory Default Setting Configurable Settings Alarm Volume Sets the alarm volume level. Highest volume Slide towards the left to decrease volume to silence. Pulse Tone Volume Sets the pulse tone volume level. Highest volume Slide towards the left to decrease volume to silence. 1, 2, 3 minutes, Permanent*, Permanent with Reminder*.
Root Alarm Source Chapter 3: Operation Example Explanation Parameter Level The example shown here is a PR alarm as the reading (142) exceeds the upper alarm limit (140). Note that the borders of both the PR Trend Display as well as the Window are illuminated red, and the explanation of the alarm is shown at the top of the Window (High PR > 140). Window Level The example shown here is a "High Impedance" alarm in the lower Window.
Root Alarm Source Chapter 3: Operation Example Explanation System Level The example shown here is a "Low Battery" alarm. Note that the border of the entire Root display is illuminated red, and the explanation of the alarm is shown in the Status Bar (Low Battery). For more details about specific alarms on Radical-7 and MOC-9 modules, refer to the instructions for use or operator’s manuals for Radical-7 and MOC-9 modules.
Root Icon Appearance Chapter 3: Operation Description Visual Alarms There are currently no active alarms, but at least one alarm has been and is still silenced. No There is currently at least one active alarm that Yes has not been silenced. There is currently at least one active alarm, but Yes all active alarms are silenced. www.masimo.
Root Chapter 3: Operation Audio Pause Audio Pause temporarily suspends all audible alarms on Root. When it is active, visual alarms are not impacted and will still display. The Audio Pause icon is located on the left side of the Status Bar – do not confuse with the Sounds icon on the right side of the Status Bar. By default, Audio Pause is inactive, and the icon appears in the following way: Audio Pause inactive To activate Audio Pause, press the icon.
Root Chapter 3: Operation Lights System Status Lights The System Status Lights provide visual indications of alarms and system messages. The lights will illuminate in different colors depending on the state of the device. To locate the System Status Lights, see Side Views on page 19. System Status Light Light Status Alarm Priority Indication None None Monitoring has not begun. Green None There is currently no active alarms. Yellow Low There is an active alarm of low priority.
Root Chapter 3: Operation Whenever Root is connected to an AC power source, this light will illuminate: Light Status Indication Green Root is connected to an AC power source. Off Root is not connected to an AC power source. Root Charging Indicator Whenever Root is connected to an AC power source, if not fully charged, its battery will charge. Light Status Battery Indication Green Battery is fully charged. Orange Battery is charging. Red Battery charging error.
Root Chapter 3: Operation When Root is connected to an AC power source, it is able to charge a correctly docked Radical-7. This is true, whether Radical-7 is powered on, in Standby Mode, or powered off. Conversely, when Root is not connected to AC power, it will not charge the Radical-7. The light status provides a visual indication of the battery condition: Light Status Battery Indication Green Battery is fully charged. Orange Battery is charging. Red Battery is unable to charge.
Chapter 4: MOC-9 Flexible measurement expansion is enabled through MOC-9. It can display parameters and measurements captured by third-party technologies in an all-in-one view on Root. When any MOC-9 module is connected, Root automatically creates a Window that displays all the data from that module. The example below shows the “SedLine” and “Capnography” Windows which display data from the SedLine brain function monitoring and Phasein capnography MOC-9 modules that are connected to Root.
Chapter 5: Iris Iris allows a variety of standalone devices to connect to Root. Patient data can be passed through Root to Patient SafetyNet or Connectivity Gateway, which can send the data to electronic health records. Below is an example of one way Root can be used in a network setting using Patient SafetyNet. Root receives and may display information from Radical-7, MOC-9 modules, as well as standalone devices. Ref. Description 1 - 4 Standalone devices connected via Iris (e.g.
Root Chapter 5: Iris Iris Connection Ports Use Iris Adapters and RJ-45 cables to connect standalone devices to Root. Port 1 Port 2 Port 3 Port 4 To connect a standalone device via an Iris Connectivity Port: 1. 2. Connect the RS-232 end of the Iris Adapter to the standalone device. Connect the RJ-45 end of the Iris Adapter to any of the four (4) compatible Iris Connectivity Ports on Root using a RJ-45 cable. www.masimo.
Root Chapter 5: Iris Iris Status Screen Information about Iris Connectivity Ports are displayed on this screen, which is accessible by selecting the Iris option on the Main Menu. Connection Status Color Description of Connection Green Standalone device is successfully connected to Root, and Root is successfully connected to a Patient SafetyNet or Connectivity Gateway.
Root Port 3 Chapter 5: Iris Port 4 Status and connection type are read-only and not configurable by the user. For more information on Iris connectivity, refer to the instructions for use or operator’s manual for the appropriate version of Masimo Patient SafetyNet. www.masimo.
Chapter 6: Messages The following messages are specific to Root: Message Explanation Next Step The internal battery needs Battery Charge is to be charged. System Low. Status Lights flash yellow. MOC-9 module disconnected (e.g. SedLine Disconnected) Radical-7 Disconnected Charge Root's battery using AC power. A MOC-9 module is disconnected from Root. Reconnect module or acknowledge message by pressing the Alarm Silence icon. Radical-7 is disconnected from Root.
Chapter 7: Troubleshooting Troubleshooting Radical-7 and MOC-9 Modules For information on troubleshooting values that are provided from Radical-7 and MOC-9 modules refer to their respective instructions for use or operator's manuals. Troubleshooting Root Symptom Root does not turn on. Possible Cause Correction Power Button not pressed long enough. Press Power Button for two (2) seconds. The battery may be depleted. Connect Root to AC power to charge battery.
Root Symptom Chapter 7: Troubleshooting Possible Cause Correction Iris screen does not display connection Connection error status for standalone devices. Unplug and replug the Iris Adapter. Iris screen does not display connection Connection error. status for standalone devices.
Root Symptom Chapter 7: Troubleshooting Possible Cause Root Charging Indicator illuminates Internal failure. red. Root requires service. See Return Procedure on page 91. Nurse Call does not Connection error communicate. www.masimo.com Correction 77 Unplug and replug Nurse Call connector. See Nurse Call Setting Connections on page 88.
Chapter 8: Specifications This chapter contains specifications of Root. For information on the specifications of Radical-7, MOC-9 modules, and standalone devices, refer to their respective instructions for use or operator's manuals. Electrical Root AC Power requirements 100-240 VAC~, 47-63 Hz, 65 W (Max) Fuses (2) 2 Amp, Fast Acting, Metric, (5x20mm), 250V Battery Type 10.
Root Chapter 8: Specifications 500 mbar to 1060 mbar Operating Altitude -1000 ft to 18,000 ft (-304 m to 5,486 m) Display Characteristic Description Type Backlit Active Matrix TFT LCD Resolution 1280 x 800 pixels Size 10.1 in (25.65 cm) Diagonal Color 24 bit RGB Touchscreen Characteristic Description Type Multi-Touch P-Cap Alarms Visual Alarm Priority System Status Light Color High Red Low Yellow www.masimo.
Root Chapter 8: Specifications Nurse Call Specifications The Nurse Call relays have the following electrical specifications per switch: Parameter Specification Max Voltage 36 VDC or 24 VAC peak Connectors Connector Type Number of Ports Ethernet 10/100 MBps 1 Nurse Call 1/4 inch round female 1 MOC-9 Masimo Connector 3 USB USB 2.0 2 Iris RS-232/RJ-45 4 Communication Specifications Communication Type Description WLAN Radio Tri Mode IEEE 802.
Root Chapter 8: Specifications Open System, Shared Key, Pre-Shared Key (PSK), Authentication 802.1X: LEAP, PEAP< TTLS, TLS, EAP-FAST Applicable Standards and Regulations Software Compliance EN 60601-1-4:1996 + A1: 1999 EMC Compliance EN 60601-1-2, Class A device ISO 9919, Clause 36 20 V/M radiated immunity Protection from Ingress of Liquids IPX1 per IEC 60529 Electrical and Mechanical Safety UL 60601-1: 2003 CAN/CSA 22.2 No. 601.
Root Chapter 8: Specifications Compatibility Standards FCC ID: VKF-RDS7 USA Model - RDS-7: FCC parts 15.207, 15.209, 15.247, and 15.407 IC:7362A-RDS7 Canada IC Model: RDS-7 RSS-210 EN 3000 328, EN 301 893, R & TTE Directive EN 301 489-17 Europe www.masimo.
Root Chapter 8: Specifications Symbols The following symbols are found on Root or its packaging and are defined below. Some of the interfaces and symbols are not available on all versions.
Root Symbols FCC ID, IC, IC Model Chapter 8: Specifications Definition Symbols Definition Federal Communications Commission (FCC) licensing Iris Connection Identifies unit has been registered as a radio device USB port Mark of conformity to European Medical Instrument Directive 93/42/EEC AC Current Authorized representative in the European community Nurse Call interface Wireless features can be used in member states with the restriction of indoor use in France www.masimo.
Chapter 9: Service and Maintenance This chapter contains information about cleaning, battery operation, performance verification, service, repair and warranty. Cleaning Root is a non-sterile and reusable device. To surface clean, wipe down the outer surface of Root using any of the following solvents: Cidex Plus (3.4% glutaraldehyde) 10% bleach solution 70% isopropyl alcohol solution Do not allow liquids to enter the interior of Root.
Root Chapter 9: Service and Maintenance Replacing the Fuses Should a power-related problem cause one or both of the fuses to fail, the fuse(s) will need to be replaced. The fuses are 1 amp, metric, fast-acting, 5x20mm, 250V. The fuses can be removed by hand or with a 5-millimeter or 3/16-inch screwdriver. To replace the fuse(s) 1. 2. 3. 4. 5. 6. 7. Power Off Root completely. Do not put in Standby Mode. See Standby and Power Off on page 65.
Root Chapter 9: Service and Maintenance disconnected. In addition, the Nurse Call Polarity can be inverted to accommodate various nurse call station requirements. Only qualified personnel should connect one of these two signals to a hospital’s Nurse Call system.
Root Chapter 9: Service and Maintenance Battery Test To conduct a Battery Test 1. 2. 3. 4. Fully charge Root by connecting it to AC power. Verify that the Root Charging Indicator is illuminated. When Root is fully charged, the Root Charging Indicator turns off. Power On Root and verify that the Root Battery Indicator icon on the Status Bar shows a full charge. Repair Policy Masimo or an Authorized Service Provider must perform warranty repair and service. Have the malfunctioning device repaired.
Root Chapter 9: Service and Maintenance Return Procedure Clean a contaminated or dirty device before returning, following instructions in Cleaning on page 87. Make sure the device is fully dry before packing. For customers in the United States, contact Masimo Technical Services at 800-326-4890 (select option 2) to obtain support and a RMA number (if required). All customers can also visit http://service.masimo.com for Technical Services contact information for their region.
Root Chapter 9: Service and Maintenance Sales & End-User License Agreement This document is a legal agreement between you (“purchaser”) and Masimo Corporation (“Masimo”) for the purchase of this Product (“Product”) and a license in the included or embedded Software (“Software”) except as otherwise expressly agreed in a separate contract for the acquisition of this Product, the following terms are the entire agreement between the parties regarding your purchase of this Product.
Root Chapter 9: Service and Maintenance Exclusions The warranty does not extend to, and Masimo is not responsible for, repair, replacement, or maintenance needed because of: a) modification of the Product or Software without Masimo’s written authorization; b) supplies, instruments or electrical work external to the Product or not manufactured by Masimo; c) disassembly or reassembly of the Product by anyone other than an authorized Masimo agent; d) use of the Product with Sensors or other accessories other
Root 3. 4. Chapter 9: Service and Maintenance publish, translate, release, distribute copies of, modify, adapt, translate, reverse engineer, decompile, disassemble, or create derivative works based on the Software or the written materials. Transfer Restrictions: In no event may Purchaser transfer, assign, rent, lease, sell, or otherwise dispose of the Product or the Software on a temporary basis.
Chapter 4 Index MOC-9 • 61 Chapter 5 A Iris • 63 Chapter 6 About • 40, 49 Messages • 67 About the Main Screen • 26 Chapter 7 About the Status Bar • 23, 26, 27, 42, 43, 44, 45 Troubleshooting • 69 Chapter 8 About this Manual • 7 Specifications • 71 AC Power Indicator • 23, 58 Chapter 9 Access Control • 41, 47, 53 Service and Maintenance • 77 Accessing Main Menu Options • 26, 40 Cleaning • 14, 77, 79 Communication Specifications • 73 Alarm Interface • 54 Alarm Silence • 27, 56 Alarms • 7
Root Index Power-On Self Test • 77 F Product Description • 9 Features • 17 Product Description and Intended Use • 9 Front View • 17 Profiles • 27, 40, 50 G R General Information, Warnings and Cautions • 12 Radical-7 Charging Indicator • 24, 28, 59 Guidelines for Setting Up • 21 Radical-7 Connection • 24 I Repair Policy • 77, 79 Initial Battery Charging • 21, 23 Replacing the Fuses • 69, 77 Intended Use • 9 Restrictions • 81 Iris • 28, 40, 50 Return Procedure • 21, 69, 70, 79 Iris Conn
Root Index U Understanding Windows • 33 Unpacking and Inspection • 21 Using Analog View • 34, 36 Using the Touchscreen Interface • 29 Using Trend View • 34, 51 W Warranty • 80 Wi-Fi • 27, 41, 43 www.masimo.