alarms and messages 5 Alarm identification The Rad-87 visually and audibly indicates alarm conditions that the system detects. Audible alarms may be silenced, without affecting the operation of visual alarms. Three levels of alarm priority are implemented: high, medium and low priority. The following table outlines the alarm priority specifications.
5 alarms and messages Alarm indication An alarm condition caused by an out-of-limit parameter/measurement is indicated by: ■ Audible alarm tone ■ Flashing Out-of-limit parameter/measurement label and value ■ System Status Light flashes red for high priority alarms, flashes yellow for medium priority alarms or shows solid yellow for low priority alarms ■ Red flashing Alarm Bell for high priority alarms Alarm limits CAUTION: TO ENSURE THAT ALARM LIMITS ARE APPROPRIATE FOR THE PATIENT BEING MONITORED, CHECK
alarms and messages 5 ALARM LIMIT: USER CONFIGURABLE SETTINGS (CONTINUED) SETTING SpHb High Limit SpHb Low Limit* PI High Limit PI Low Limit* RANGE The SpHb high alarm limit can be set anywhere between 2 g/dl and 24.5 g/dl, then “----” (Off). Adjust in increments of 0.1 g/dl step size between 2 g/dl and 20 g/dl and 0.5 g/dl step size between 20 g/dl and 24.5 g/dl. When SpHb is placed in the “----” (Off) setting, the SpHb High Limit alarm is disabled.
5 alarms and messages SINGLE ALARM When an alarm is activated, the display shows the screen containing the parameter/measurement in alarm status. The number value and the label (name) for the parameter/measurement in alarm status flash, an audible alarm activates, the Alarm Bell flashes red for high priority alarms and the System Status Light is solid yellow for low priority alarms, flashes yellow for medium priority alarms or flashes red for high priority alarms.
alarms and messages 5 ALARM SILENCE Audible alarms may be silenced, while visual alarms remain active. The alarm silence function is controlled by pressing the Alarm Silence Button. The Alarm Bell and the System Status Light provide visual feedback when the Rad-87 audible alarms are silenced.
5 alarms and messages SYSTEM STATUS LIGHT While monitoring a patient and an alarm condition occurs, an audible alarm activates and the System Status Light shows solid yellow for low priority alarms, flashes yellow for medium priority alarms or flashes red for high priority alarms. Pressing the Alarm Silence Button (one time) silences the alarm tone for 120 seconds (default).
5 alarms and messages MESSAGES The Rad-87 Pulse CO-Oximeter will indicate other data or system errors. Message conditions for the Rad-87 follow: DISPLAY TYPE SCROLLING ZEROS Pulse Search PULSE BAR (SIQ) TURNS RED Low Signal IQ SOLUTION Wait for found pulse. (This Search should occur whenever a sensor is first applied to a patient). 1. Rule out occlusion of blood flow. 2. Verify placement of sensor. 1. Rule out occlusion of blood flow. 2. Attempt to warm patient.
5 alarms and messages Messages continued DISPLAY TYPE No Sensor Connected Sensor off patient SINGLE BATTERY LEVEL INDICATOR FLASHES (WITH AUDIBLE ALARM) 5-8 SOLUTION Connect sensor to cable. 1. Reattach sensor to patient. 2. Verify proper sensor placement. Battery level too low Connect device to AC Power to charge the battery. No Cable Connected Connect appropriate cable to unit.
troubleshooting 6 Troubleshooting The following chart describes what to do if the Rad-87 system does not operate properly or fails. PROBLEM TYPE DEVICE DOES NOT POWER ON Low battery/ not plugged into AC power supply Connect the AC Power Cord to the Rad-87 and to an AC outlet. Make sure that the AC Power Indicator light is on. BATTERY RUN-TIME IS SIGNIFICANTLY REDUCED Low battery Contact Technical Services or your local Masimo representative.
troubleshooting 6 Troubleshooting continued PROBLEM POSSIBLE CAUSE(S) RECOMMENDATION SpO2 VALUES DO NOT CORRELATE WITH CLINICAL ASSESSMENT OR ABGs. Low perfusion or sensor displacement. Check for error messages. See section 5 Messages for recommended corrections. Check placement of sensor or if it is too tight. Reapply sensor or select a new site. Set to MAX sensitivity and confirm that the sensor is securely on the patient. Refer to sensor Directions For Use.
specifications 7 Rad-87 specifications PERFORMANCE Measurement Range SpO2: SpMet: SpCO: SpHb SpOC Pulse Rate: Perfusion Index: Pleth Variability Index: Accuracy: Arterial Oxygen Saturation Accuracy1 Saturation No Motion Adults, Infants, Pediatrics Saturation No Motion2 Adults, Infants, Pediatrics Neonates* Motion3 Adults, Infants, Pediatrics, Neonates Low Perfusion4 Adults, Infants, Pediatrics, Neonates Pulse Rate Accuracy5 Pulse Rate: No Motion Adults, Infants, Pediatrics, Neonates Motion4 Adults, Infant
7 specifications ELECTRICAL AC Power requirements: Power consumption: Batteries Type: Capacity (battery life): Charging time: 100 - 240 VAC, 47-83 Hz 15 VA max. Sealed lead acid up to 4 hours6 8 hours ENVIRONMENTAL Operating Temperature: Transport/Storage Temperature: 41°F to 104°F (5°C to 40°C) -40°F to 158°F (-40°C to +70°C)7 Operating Humidity: 5% to 95%, non-condensing Operating Altitude: 500 mbar to 1060 mbar pressure -1000 ft to 18,000 ft (-304 m to 5,486 m) PHYSICAL CHARACTERISTICS 8.2” x 6.
specifications 7 PHYSICAL CHARACTERISTICS (CONTINUED) Output Interface Serial RS-232 Wireless Radio (if installed) Nurse Call Philips VueLink, RadNet, Patient SafetyNet Compliance Safety Standard for Medical Equipment 802.11 a/b/g EC 60601-1-2nd Edition UL 60601-1 CAN/CSA C22.2 No.
7 specifications Serial interface specifications The digital interface for serial communication is based on the standard RS-232 protocol. The Rad-87 Pulse CO-Oximeter by default always outputs ASCII 1 text data through the serial port, unless the user selects a different output mode in the Output menu. To interface with the Rad-87 and receive serial text data, simply connect a serial interface cable with a serial output connector located on the back of the Rad-87.
specifications 7 SERIAL PRINTER SETUP To print the SpO2 and pulse rate data in ASCII 2 format on a serial printer, simply connect the serial printer to the serial port and set output mode to ASCII 2. Once serial communication is established, the Rad-87 will automatically start printing the ASCII 2 text data. WARNING: ALL EXTERNAL DEVICE CONNECTIONS TO THE RS-232 SERIAL PORT MUST BE IEC-60950 COMPLIANT.
sensors & patient cables 8 Introduction This section covers the use and cleaning of Masimo sensors and patient cables. Before use of any sensor, carefully read the sensor’s Directions for Use. Use only Masimo sensors and cables with the Rad-87 Pulse CO-Oximeter. Other transducers, sensors and cables may affect Rad-87's performance. Tissue damage can be caused by incorrect application or use of a sensor, for example by wrapping the sensor too tightly.
8 sensors & patient cables Masimo Rainbow Sensors Masimo Rainbow sensors must be used for the Rad-87 Pulse CO-Oximeter parameters to enable measurement of Carboxyhemoglobin (SpCO), Methemoglobin (SpMet) and Total Hemoglobin (SpHb). Rainbow sensors will only measure SpO2 and pulse rate on devices without Masimo Rainbow SET Technology. RAINBOW REUSABLE SENSORS SpO2, SpCO, SpMet, SpHb and pulse rate accuracy for the Rainbow sensors is specified in the following table.
sensors & patient cables 8 Masimo SpO2 Sensors The Rad-87 may use standard Masimo LNOP, LNOPv and LNCS SpO2 sensors, when used with Red PC, Red LNC or Rainbow RC cables respectively. Select the appropriate patient cable to attach the LNOP or LNCS sensor to the device. RED DIRECT CONNECT SENSORS Masimo Red sensors can be used with the Rad-87 to enable measurement of SpO2 and pulse rate only. Red sensors will only function with oximeter devices equipped with Masimo Rainbow SET technology.
8 sensors & patient cables TM LNOPv ADHESIVE SENSORS LNOPv adhesive sensors must be used in conjunction with Red PC cables SENSOR Weight Range LNOPv In 3 - 20 kg Saturation Accuracy No Motion Motion ± 2% ± 3% Pulse Rate Accuracy No Motion Motion ± 3 bpm ± 5 bpm Low Perfusion Accuracy Saturation Pulse Rate ± 2% ± 3 bpm LNOPv Ne < 3 kg ± 3% ± 3% ± 3 bpm ± 5 bpm ± 3% ± 3 bpm LNOPv Ad > 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm ® LNOP SPECIALTY SENSORS LNOP specialty sensors must be
sensors & patient cables 8 ® LNCS SPECIALTY SENSORS LNCS specialty sensors must be used in conjunction with Red LNC cables or Rainbow RPC cables.
service / maintenance 9 Introduction This chapter covers how to test the operation, properly clean and how to obtain service for the Rad-87 Pulse CO-Oximeter. Under normal operation, no internal adjustment or recalibration is required. Safety tests and internal adjustments should be done by qualified personnel only. Safety checks should be performed at regular intervals or in accordance with local and governmental regulations.
9 service / maintenance Battery Service WARNING: THE BATTERY SHOULD BE INSTALLED AND/ OR REMOVED FROM THE RAD-87 BY QUALIFIED PERSONNEL ONLY. Performance verification To test the performance of the Rad-87 after repairs or during routine maintenance, follow the procedure outlined in this section. If the Rad-87 fails any of the described tests, discontinue its use and correct the problem before returning the device back to the user.
service / maintenance 9 LED BRIGHTNESS: 1. With the monitor turned on, press the Brightness Button once to enter the LED Brightness menu. The display will show the default setting Level 2. 2. Continue pressing the Brightness Button to scroll through the settings. 3. Press the Enter Button to accept the desired setting and exit to the home display screen. TESTING THE RAD-87 WITH MASIMO SET TESTER (OPTIONAL): 1. Turn the Rad-87 off and then on again. 2.
9 service / maintenance Service and repair REPAIR POLICY Masimo or an authorized Service Department must perform warranty repair and service. Do not use malfunctioning equipment. Have the device repaired. WARNING: AN OPERATOR MAY ONLY PERFORM MAINTENANCE PROCEDURES SPECIFICALLY DESCRIBED IN THIS MANUAL. REFER SERVICING TO QUALIFIED SERVICE PERSONNEL TRAINED IN THE REPAIR OF THIS EQUIPMENT.
service / maintenance 9 Sales & End-User License Agreement THIS DOCUMENT IS A LEGAL AGREEMENT BETWEEN YOU (“PURCHASER”) AND MASIMO CORPORATION (“MASIMO”) FOR THE PURCHASE OF THIS PRODUCT (“PRODUCT”) AND A LICENSE IN THE INCLUDED OR EMBEDDED SOFTWARE (“SOFTWARE”). EXCEPT AS OTHERWISE EXPRESSLY AGREED IN A SEPARATE CONTRACT FOR THE ACQUISITION OF THIS PRODUCT, THE FOLLOWING TERMS ARE THE ENTIRE AGREEMENT BETWEEN THE PARTIES REGARDING YOUR PURCHASE OF THIS PRODUCT.
9 service / maintenance Exclusions continued THIS WARRANTY, TOGETHER WITH ANY OTHER EXPRESS WRITTEN WARRANTY THAT MAY BE ISSUED BY MASIMO IS THE SOLE AND EXCLUSIVE WARRANTY AS TO THE PRODUCT AND SOFTWARE. THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY ORAL OR IMPLIED WARRANTIES, INCLUDING WITHOUT LIMITATION ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
service / maintenance 9 Restrictions 1. Copyright Restrictions: The Software and the accompanying written materials are copyrighted. Unauthorized copying of the Software, including Software that has been modified, merged, or included with other software, or the written materials is expressly forbidden. Purchaser may be held legally responsible for any copyright infringement that is caused or incurred by Purchaser’s failure to abide by the terms of this Agreement.
Section 10 Part Numbers part numbers 10 Part Numbers PART NUMBER 32750 32751 32752 32753 32754 32755 32756 32757 32758 32759 DESCRIPTION Rad-87 Operator's Manual (SpHb), French Rad-87 Operator's Manual (SpHb), German Rad-87 Operator's Manual (SpHb), Italian Rad-87 Operator's Manual (SpHb), Spanish Rad-87 Operator's Manual (SpHb), Japanese Rad-87 Operator's Manual (SpHb), Dutch Rad-87 Operator's Manual (SpHb), Portuguese Rad-87 Operator's Manual (SpHb), Danish Rad-87 Operator's Manual (SpHb), Swedish Ra
31374/4525E-0608 www.masimo.com Instruments and sensors containing Masimo Rainbow SET technology are identified with the Masimo Rainbow SET logo.