Luminex® IS Versions 2.1 and 2.2 Upgrade Instructions to xPONENT® 3.
© LUMINEX CORPORATION, 2001-2009. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into any language or computer language, in any form or by any means without prior express, written consent of: LUMINEX CORPORATION 12212 Technology Boulevard Austin, Texas 78727-6115 U.S.A. Voice: (512) 219-8020 Fax: (512) 219-5195 Luminex® IS Versions 2.1 and 2.2 Upgrade Instructions to xPONENT® 3.1 SP1 PN 89-00002-00-213 Rev.
Contents Upgrade Procedure .............................................................................1 Important Upgrade Information..........................................................2 Recommendations...............................................................................2 Training...............................................................................................2 System Requirements .........................................................................2 Kit Components ......
ii
Upgrade Procedure 1
Luminex IS Versions 2.1 and 2.2 Upgrade Instructions to xPONENT 3.1 SP1 - For In Vitro Diagnostic Use Important Upgrade Information This Luminex xPONENT 3.1 SP1 product conversion kit is intended for use on a single Luminex analyzer (Luminex 100 or Luminex 200™) purchased after October 1, 2000. If you install this kit on multiple analyzers, you accept the legal and financial risk associated with violating software licensing.
Luminex IS Versions 2.1 and 2.2 Upgrade Instructions to xPONENT 3.1 SP1 - For In Vitro Diagnostic Use • • Kit Components Note: For information about calibration and verification frequency and procedures, see Luminex xPONENT 3.1 Software Manual. Contact Luminex Technical Support with the analyzer serial number to confirm that it meets the requirements. A Luminex XYP purchased after May 30, 2001.
Luminex IS Versions 2.1 and 2.2 Upgrade Instructions to xPONENT 3.1 SP1 - For In Vitro Diagnostic Use These license packages are part of the upgrade kit only if you purchase them from Luminex: • 21 CFR Part 11 package • LIS package • Automation package • Security package • Extra seats package Upgrade Instructions You must run Luminex xPONENT 3.1 SP1 on a PC which meets the minimum specifications listed in "System Requirements" on page 2. To upgrade your PC, contact Luminex Technical Support.
Luminex IS Versions 2.1 and 2.2 Upgrade Instructions to xPONENT 3.1 SP1 - For In Vitro Diagnostic Use Back Up Luminex IS Version 2.1 or 2.2 Database Note: If you are installing a new PC, you do not need to back up the database. You cannot import data into xPONENT 3.1 SP1 from IS 2.1 or 2.2; however, you can store the data to import to another IS 2.1 or 2.2 system at a later time. To back up your Luminex IS Version 2.1 or 2.2 database before removing the Luminex IS Version 2.1 or 2.2 software: 1.
Luminex IS Versions 2.1 and 2.2 Upgrade Instructions to xPONENT 3.1 SP1 - For In Vitro Diagnostic Use 12. Click on MSDE, then click Change/Remove. Select Remove, and follow the prompts to remove it. 13. Restart your computer. Install Luminex xPONENT 3.1 SP1 Software Note: If you purchased a new PC from Luminex as part of the upgrade, you do not need to perform these steps. xPONENT 3.1 SP1 software is already Note: Steps 2, 3, and 4 may not occur in all installations. Luminex xPONENT 3.
Luminex IS Versions 2.1 and 2.2 Upgrade Instructions to xPONENT 3.1 SP1 - For In Vitro Diagnostic Use To install the xCONVERT Data Conversion Utility: 1. Click Install xPONENT file conversion utility. A File Download – Security Warning dialog box opens. 2. Click Run. An Internet Explorer – Security Warning dialog box opens. 3. Click Run. An xCONVERT Setup dialog box opens (.NET Framework 2.0). Note: Steps 3 and 4 may not occur in all installations. 4. Click Accept.
Luminex IS Versions 2.1 and 2.2 Upgrade Instructions to xPONENT 3.1 SP1 - For In Vitro Diagnostic Use 1 2 3 6 4 5 LxFirmware Update Tool Dialog Box 1 2 3 4 5 6 LXR Runtime Version Luminex Analyzer Firmware Version Movement Device Textbox Luminex XYP/SD Instrument Firmware Version Status of Upgrade Movement Device / SD Device Selection Button 6. Select the SD Device option button (item 6, above).
Luminex IS Versions 2.1 and 2.2 Upgrade Instructions to xPONENT 3.1 SP1 - For In Vitro Diagnostic Use 11. An entry in the Movement Device text box should read: Firmware is up-to-date. The firmware version should be 2.0.5. 12. Turn the power off to the SD. 13. Reconnect the serial cable to the XYP. Luminex 100/200 Analyzer Firmware Upgrade To upgrade the Luminex 100/200 analyzer firmware: 1. Turn on the power to the analyzer. 2. Close any open applications. 3.
Luminex IS Versions 2.1 and 2.2 Upgrade Instructions to xPONENT 3.1 SP1 - For In Vitro Diagnostic Use LX200 Sample Probe 5. The Download In Progress dialog box opens displaying the progress. Read the information presented on the dialog box. When the download is finished the Status text box displays Download Complete. Click OK. 6. Turn off the power to the analyzer. 7.
Luminex IS Versions 2.1 and 2.2 Upgrade Instructions to xPONENT 3.1 SP1 - For In Vitro Diagnostic Use (item 4 in "LxFirmware Update Tool Dialog Box" on page 8) displays the XYP firmware version and status. This process may take up to five minutes. The example above may not match the dialog box on your system. 5. If the second line indicates that the firmware is up-to-date, then click Close. No further action is required; continue with step 8. If an update version is available, continue with step 6. 6.
Luminex IS Versions 2.1 and 2.2 Upgrade Instructions to xPONENT 3.1 SP1 - For In Vitro Diagnostic Use Luminex Contact Information Contact Luminex Technical Support by telephone in the U.S. and Canada at 1-877-785-BEAD (2323), by telephone outside the US and Canada at +1 512-381-4397, or by email at support@luminexcorp.com. Additional information is available on the Luminex website. Search on the desired topic, navigate through menus, or review the website FAQ section. Enter http://www.luminexcorp.