LiveOne Model: LM1P Clinical Non-invasive Blood Pressure Monitor USER MANUAL FOR HEALTH CARE PROVIDERS VERSION 3.
TABLE OF CONTENTS Your LiveOne Package Includes: ...................................................................................................................................................... 3 LiveOne Box.............................................................................................................................................................................. 3 Companion Device Box .....................................................................................................
YOUR LIVEONE PACKAGE INCLUDES: LIVEONE BOX LiveOne Wristband Charging Cable User Manual Pocket Instructions Card COMPANION DEVICE BOX Surface Tablet Charger User Manual Quick setup instructions Please read this user manual thoroughly before using the wristband and keep it for future reference. ABOUT LIVEONE Thank you for choosing LiveOne, LiveMetric’s innovative wrist-worn blood pressure monitor, for use in a clinical setting.
The device does not report any diagnosis but provides numerical values. It is the physician's responsibility to make proper judgments based on these numbers. LiveOne is also intended for use by a physician or other qualified healthcare provider for the calculation of heart rate and heart rate variability based on waveform analysis. INDICATION FOR USE The LiveOne is intended to be used in patients who have a need for a noninvasive blood pressure and hemodynamic monitor.
Do not use a mobile phone or other devices that emit electromagnetic fields, near the device; it may result in incorrect operation of the device Do not use the device with other medical electrical equipment simultaneously Do not use the device where high frequency surgical equipment, magnetic resonance imaging (MRI), computerized tomography (CT) scanner or X-ray machine is operating Do not use the device near electromagnetic fields emission equipment such as cellular phones, microwave ovens or televisions D
Home Screen (LiveMetric's logo) Terms of Use Screen Profile Setting Screen Pairing Screen Placement screen Monitoring Configuration Screen (in ABPM mode) Signals Screen ABPM Measurements Screen Single Measurement Mode Screen SETUP LIVEMETRIC Take your new LiveOne out of the box and carefully read the user manual.
Click next on Home Screen Pairing Screen will open Double tap the black casing on the wristband, a blue LED light should appear If you don’t see the blue light, double tap more firmly Click “pair” on the screen A message will confirm successful pairing and the LED light on the black casing should be blinking blue If pairing has not been successful – click the troubleshooting button for possible corrective actions Follow the suggested option, or consult LiveMetric customer service Placement Screen will open
The screen contains value boxes (systolic, diastolic, heart rate) and a plot area presenting the active sensor's waveform line To get the most updated values, click Refresh periodically Please instruct the patient to breath calmly and avoid any movements after clicking refresh, as the message "Measurement in Progress" is presented under the plot area When measurement process is finalized, the message "Measurement Done" In case no values are presented and an error message appears, consider repositioning the
The strap part with the holes on the bottom side of the wrist To readjust please consider moving the wristband: Inward or outwards Toward the palm or away from palm Please consider the following safety measures: Do not place wristband on injured skin Do not place wristband on clothing Place the wristband directly on intact skin IMPORTANT NOTICE FOR PATIENT INSTRUCTION: Research suggests that the following values can be used as a guide to high blood pressure for measurements taken at home: Systolic Blood Pre
15. Blue-Green LED light is not blinking (following 14) only blue LED light is on 16. Blue Green LED is not blinking and Blue LED light is not on (following 14) 17. Blue Green LED is not blinking and Blue LED light is not on (following 16); even after double tapping 18. Cannot place the wristband 19. The wristband is inflicting pain on my wrist 20. The wristband keeps moving on my wrist 21. Not sure the placement is good 22. The LED is stopped blinking 23.
34. Did not get any value; 35. The waveform line looks noisy 36. I got a pop up message that the connection cannot be restored 37. Did not get any values 38. Did not get any values Wristband is not blinking Single Measurement Screen ABPM Measurement Screen Please click Done to restart the session. The connection could not be restored Consider removing electromagnetic interference while monitoring Please click Done to restart the session.
The LiveOne LiveMetric wristband is made of a flexible, durable elastomer material similar to that used in many sports watches, Silicon 60 shore A (Shin-Etsu Silicon CHN) and includes a stainless steel black anodized buckle. The black casing is made of black anodized aluminum. The upper antenna cover is made of PC. The bottom support is made of ABS. WIRELESS TECHNOLOGY LiveMetric contains a Bluetooth 4.0 radio receiver and LiveOne. BATTERY LiveMetric contains a rechargeable 1 cell lithium-polymer battery.