Specifications

Page 71 of 152
The control box CB6 has been specially developed for use together with LA27/LA40 actuator in
the care and rehab industry.
The control box is designed to be mounted on the actuator LA27/LA40 as with the CB9 and
LA31 system.
The CB6 control box has a LED power ON indicator, detachable mains cable and strain relief for
all cables.
The control box CB6 communicates with the LA27/LA40 actuator by means of the built-in end
stop signal switches in the actuator. Due to the signal switches the power to the motor will be
cut off in the control box and not in the actuator.
Recommendations
Be aware of the handset configuration (e.g. CH1, 2, 3 should be CH1, 3, 4).
Same cables variants as for CB9 CARELINE and CB12.
Connecting the system:
Do not connect the mains cable until all actuators have been connected to the control box.
Start by connecting the handset to the control box. The connection at the control box is marked with “HB”.
Connect the different actuators to the different channels on the control box. Each channel is marked with a number (e.g. “1”, “2”, “3”....).
Check that all plugs are well connected and firm pushed into the connection plug.
Due to the fact that LINAK control boxes are designed for a high IP degree, a firm force can be required.
CONNECT the mains and turn on the power!
Finally, connect the battery (BA18) with special T-cable or normal battery cable depending on the specified system.
The actuators can now be operated by pushing a button on the handset. Use only one button at a time.
If the control box is equipped with special software, an initialising process might be necessary. This process is described in the software specification.
Attention should be paid to the following:
Control boxes must only be connected to the mains voltage specified on the label.
The control box must be connected in such a way that the cables are not trapped, exposed to tension or sharp objects, when the application is
moved in different directions.
CB6 system diagram
2. CB6 (MEDLINE
®
CARELINE
®
)
Usage:
Duty cycle: 2/18; 2 minutes continuous use followed by 18 minutes not in use
CB6 is approved according to EN60601-1 / UL60601-1
The CB6 can only be combined with LA27/LA40 and HB30 / HB70 / HL70 / HB80 and HL80.