2. CB6 (MEDLINE® CARELINE®) The control box CB6 has been specially developed for use together with LA27/LA40 actuator in the care and rehab industry. The control box is designed to be mounted on the actuator LA27/LA40 as with the CB9 and LA31 system. The CB6 control box has a LED power ON indicator, detachable mains cable and strain relief for all cables. The control box CB6 communicates with the LA27/LA40 actuator by means of the built-in end stop signal switches in the actuator.
3. CB6 OBMe (MEDLINE® CARELINE®) The CB6 OBMe is a powerful control box with switch mode power supply typically used for applications like hospital beds, couches/tables for treatment and examination and other medical applications. The universal input voltage makes the control box adaptable to the world-wide market irrespective of the input voltage. Approvals: CB6 OBMe is approved according to UL60601-1 and EN60601-1. Possible combinations. CB6 OBMe is meant for use with: LA27 std.
If the LED has turned yellow AND the HB has NOT been activated, it indicates that a failure has occurred (1st Failure). NOTE: Even though the LED illuminates yellow before the HB is activated it is possible to operate the CB6, however the 1st. Failure is somehow still in the CB and must be removed to prevent a further failure causing a hazardous failure to happen. • To meet the safety requirements the device must have a dual switch safety concept.
5. CB6S (MEDLINE® CARELINE®) CB6S is part of the LINAK OpenBus™ product range - that provides more flexible solutions no matter which actuator concept is preferred. Mounting of CB6S – article no. 9271233 Apart from LA27 and LA40, it is not possible to mount CB6S onto LINAK actuators. In such case you need a CB6 mounting bracket which can be welded or likewise to the application. This bracket is shown below: 28.25 22 17 6.8 13.15 17.8 80 35.8 40 6.8 11.5 12 14 85 128.
Recommendations NOTE - HOT PLUGGING Removing or adding any OpenBus cables is not allowed when the CB is powered by mains supply! If needed anyway, follow the below procedure: 1. Remove mains and wait 5 sec. 2. Mount or dismount the required cables If this procedure is NOT followed, it may result in a damaged OpenBus driver circuit. The risk of a damaged circuit increases if the accessory has a high start current (in rush current). OBF - hot-plugging protected from January 2016.
. CB7 (HOMELINE®) Compared to other LINAK control boxes the CB7 is very small and compact in design. The CB7 is designed to slide onto an LA31 actuator for easy fitting e.g. in a recliner application where “mounting” space is limited. The control box function is divided in two parts. The actual control box CB7, which slides onto the LA31 actuator and a separate external power supply transformer box TR6 or TR7, which can be wall mounted or placed on the floor moulding next to the mains.
8. CB8-T (MEDLINE® CARELINE® TECHLINE®) The CB8-T is developed for use with LINAK A/S’ actuators and handsets. The control box can operate up to 2 actuators individually. The simple compact design combined with high quality makes the control box ideal for use with beds, chairs, tables and many other applications. Important: Individual current cut-off: The current to each actuator is monitored and when this reaches a specified value, the current to that actuator is cut-off.
CB9 is equipped with a green LED for indication of mains power connected. • When th CB9 is connected to mains, the LED is green. • Connected only to battery, the LED is off. The cables are locked in position with a locking mechanism. The locking mechanism is shownhere (not in position) CB9 for mounting on application. Shows the two holes used for mounting CB9 mounted on actuator, LA31. Screw used for mounting CB9 can be mounted and fixed together with LA31 with one screw.
11. CB9 CARELINE Basic (MEDLINE® CARELINE®) The CARELINE® Basic CB9 has been developed for use together with LA31 and LA34 in the Care & Rehab industry.CB9 and LA31 can be fully integrated thus saving mounting and wiring or they can be installed separately. Exchangeable mains cables, Electronic Overload Protection (EOP) and earth connection (Class 1) mains fuse makes CB9 a good choice for simple hospital and care beds. The CARELINE® Basic CB9 series is only available as an analogue (Ax) type.
13. CB14 (MEDLINE® CARELINE® TECHLINE®) The CB14 with microprocessor is developed for systems with a need to run up to five actuators or two actuators / lifting columns in parallel and / or with memory function. The effective toroidal transformer and the many features such as battery backup, earth outlet, wet alarm makes the control box suitable for a variety of applications. CB14 with battery backup The CB14 with battery backup has an acoustic alarm, which sounds when the batteries are low, approx. 17-18V.
• If the battery voltage is at ‘Medium level’ a battery alarm beeps as long as a key is activated. (Medium level means that battery charging is necessary to maintain the best possible lifetime). • If the battery voltage is at ‘Low level’, the battery alarm function shuts down all movement immediately. The OpenBus is still active for approx. 15 seconds. (If trying to operate the system anyway, the battery could get deep drained or the actuator system could get damaged.
15. CB20 (MEDLINE® CARELINE®) CB20 is a platform which offers a unique safety concept, logging off service data and it is possible to connect a variety of accessories to the control box. The CB20 consists of 3 modules : CP20 = Control Power CU20 = Control Unit BA20 = Battery. For recommendations on maintenance and storage of the BA20 battery packs, please refer to chapter 5. Usage: • Compatible with specific versions of LA23, LA27C, LA31, LA34, LA40, LA43, LA44 and BL1 • Duty cycle: 10% ˜ max. 2 min.
17. CO41 (MEDLINE® CARELINE® ) The LINAK control box CO41 offers a consolidated range of unprecedented features – all utilising standardised technology, interfaces and compatibility. The CO41 for LINAK actuators is intended for the control of, for example, hospital bed movement.
CO41 - mounted on frame: Release hook Very small built-in dimension: Height: min. 80 mm (lid can be removed) Length: + 20 mm (locking system) Service lid Integrated hinge Removable if needed Easy-mount bracket 3-way cable exit Mounting bracket (frame flat) - article No. 1015W1001: Mounting bracket (frame flat) w/M4 nuts - article No. 1015W1009: Page 84 of 152 ...
Mounting instructions (Example CO41 with LA40) When mounting the control box onto the actuator (1) Simply slide on the bracket until you hear a clear click (2) Slide on the control box until you hear a click and the box is mounted (3) It is recommended that the CO41 is mounted in a position that allows water to escape. Recommended torque: 0.6 Nm +/-0.
Mounting of cables and cable lock: CO41 has a uniquely designed cable lid. The lid also works as an integrated cable lock when closed. 1) Mount cable plugs in control box 2) Close lid until lock snaps into place (see arrows) To allow free access to the cables, the lid has a rest position when completely opened. It is possible to remove the lid by lifting it a few degrees and pulling it away from the housing under tight mounting conditions. See illustrations: 1 Cable management: 2 Page 86 of 152 ...
Recommendations • To avoid that hand control cables short-circuit, LINAK recommends to use an OpenBus system (CO41). • When the system is overloaded, LINAK recommends to use quick release actuators in the application. These will allow functions to be lowered manually in case of a CO41 malfunction due to misuse/abuse. • If the customer has other essential performance than “no unintended movement”, he must consider this in his own risk analysis. LINAK disclaims any liability.
. CO61 (MEDLINE® CARELINE® ) The LINAK control box CO61 offers a consolidated range of unprecedented features – all utilising standardised technology, interfaces and compatibility. The CO61 for LINAK actuators is intended for the control of, for example, hospital bed movement.
CO61 - mounted on frame: Release hook Very small built-in dimension: Height: min. 80 mm (lid can be removed) Length: + 20 mm (locking system) Service lid Integrated hinge Removable if needed Easy-mount bracket 3-way cable exit Mounting bracket (frame flat) - article No. 1015W1001: Mounting bracket (frame flat) w/M4 nuts - article No. 1015W1009: Page 89 of 152 ...
Mounting instructions (Example CO61 with LA40) When mounting the control box onto the actuator (1) Simply slide on the bracket until you hear a clear click (2) Slide on the control box until you hear a click and the box is mounted (3) It is recommended that the CO61 is mounted in a position that allows water to escape. Recommended torque: 0.6 Nm +/-0.
Mounting of cables and cable lock: CO61 has a uniquely designed cable lid. The lid also works as an integrated cable lock when closed. 1) Mount cable plugs in control box 2) Close lid until lock snaps into place (see arrows) To allow free access to the cables, the lid has a rest position when completely opened. It is possible to remove the lid by lifting it a few degrees and pulling it away from the housing under tight mounting conditions. See illustrations: 1 Cable management: 2 Page 91 of 152 ...
Recommendations • To avoid that hand control cables short-circuit, LINAK recommends to use an OpenBus system (CO61). • When the system is overloaded, LINAK recommends to use quick release actuators in the application. These will allow functions to be lowered manually in case of a CO61 malfunction due to misuse/abuse. • If the customer has other essential performance than “no unintended movement”, he must consider this in his own risk analysis. LINAK disclaims any liability.
19. CO71 (MEDLINE® CARELINE® ) The LINAK control box CO71 offers a consolidated range of unprecedented features – all utilising standardised technology, interfaces and compatibility. The CO71 for LINAK actuators is intended for the control of, for example, hospital bed movement.
CO71 - mounted on frame: Release hook Very small built-in dimension: Height: min. 80 mm (lid can be removed) Length: + 20 mm (locking system) Service lid Integrated hinge Removable if needed Easy-mount bracket 3-way cable exit Mounting bracket (frame flat) - article No. 1015W1001: Mounting bracket (frame flat) w/M4 nuts - article No. 1015W1009: Page 94 of 152 ...
Mounting instructions (Example CO71 with LA40) When mounting the control box onto the actuator (1) Simply slide on the bracket until you hear a clear click (2) Slide on the control box until you hear a click and the box is mounted (3) It is recommended that the CO71 is mounted in a position that allows water to escape. Recommended torque: 0.6 Nm +/-0.
Mounting of cables and cable lock: CO71 has a uniquely designed cable lid. The lid also works as an integrated cable lock when closed. 1) Mount cable plugs in control box 2) Close lid until lock snaps into place (see arrows) To allow free access to the cables, the lid has a rest position when completely opened. It is possible to remove the lid by lifting it a few degrees and pulling it away from the housing under tight mounting conditions. See illustrations: 1 Cable management: 2 Page 96 of 152 ...
Recommendations • To avoid that hand control cables short-circuit, LINAK recommends to use an OpenBusTM system (CO71). • When the system is overloaded, LINAK recommends to use quick release actuators in the application. These will allow functions to be lowered manually in case of a CO71 malfunction due to misuse/abuse. • If the customer has other essential performance than “no unintended movement”, he must consider this in his own risk analysis. LINAK disclaims any liability.
6. Information on specific controls 1. ACC (MEDLINE® CARELINE®) The ACC (Attendant Control Compact) is fitted to advanced hospital and patient care beds for use where patient positioning must be carefully controlled by medical staff. 2. ACK (MEDLINE® CARELINE®) With the OpenBusTM system it is possible to use keypads as attendant control or hand controls integrated in the bed side rails.
4. ACM (MEDLINE® CARELINE®) The ACM (Mini Attendant Control) box is fitted to hospital and care beds for use where patient positioning must be carefully controlled by the medical staff. The compact design and simple operation makes it easy for the nursing staff to retain direct control over critical positioning functions whilst giving the patient a limited degree of adjustment. The ACM must be mounted correctly on a flat surface to ensure IP degree. A short circuit in the cable can cause movement.
7. ACP (MEDLINE® CARELINE®) The ACP (Attendant Control Panel) controller is fitted to advanced hospital and patient care beds for use where patient positioning must be carefully controlled by medical staff. The ACP allows nursing staff to retain direct control over critical positioning functions while giving the patient a limited degree of adjustment. It is recommended to clean all surfaces, also surfaces covered by lids and the inside of the lids.
The recommended screw type is self-tapping BN84229 50 x 12. A grounding cable (article no.: 1009W7004) will be supplied together with the ACT and must be connected to the OpenBusTM system and the other end to the bed frame to decouple electrical fields. Ferrite core : If upward cable exit is used, a ferrite core needs to be added as close as possible to the ACT.
Technical recommendations: • • • • • • • Always use the cable locking mechanism and an O ring. Prepare a system/bit overview as the unit may conflict with other OpenBusTM accessories such as HB, ACP etc. Avoid large metal parts in the vicinity of the RFID reader (approx. 50 mm). When designing the application, be aware of the ACT position in the application to avoid impact to the front glass. Allow for a power consumption of max. 200 mA on both V-Bus and Vpermanent when designing a system.
10. FPP (MEDLINE® CARELINE®) The FPP is for use with a variety of different bed types and is therefore compatible with control boxes that use an OpenBus™ interface i.e. CB6S, CB16 and CB20. Usage: • Ambient temperature: • Approvals: • Current consumption: +5° to +40° LINAK and UL International DEMKO A/S are currently in the process of certifying FPP re. UL60601 and EN60601-1. 15 mA (Standard versions with 5 no.
As illustrated from the pictures below the panel itself can be moved and angled in a number of positions. The arm can also be bent to move it closer to or move it further away from the user. The lock function Between the arm and the panel there is a lock/unlock function, (a hose type connection). It enables the user to turn the panel into a preferred position. Locking of the Panel: Turn the panel to a preferred position. With one hand on the panel turn the hose clockwise with the other hand.
11. FS (MEDLINE® CARELINE® TECHLINE®) The Foot Switch is a modular system, developed for use together with some of LINAK control boxes. The LINAK Foot Switch is designed for control of physiotherapeutic beds, hospital beds, dentist chairs, gynaecologist chairs, computer workstations, and working desks etc. It can also be used as a “stand alone” item for industrial applications. Footswitch Consist of: FS (a pedal unit) and FSE (electronics unit), which can activate one or more actuators.