Adjustment instructions for the JUMBO application. Tool For the adjustment you must use a trimming screwdriver, which can be purchased from LINAK A/S. It is also possible to use other types of trimming screwdrivers for the adjustment. Ordinary screwdrivers cannot be used, as they will damage the potentiometer slot. When you receive the JUMBO from LINAK A/S it is adjusted to min. current cut-off. 1. Connect the JUMBO control box to the actuator. 2. Load the actuator with the required load. 3.
4. CBJ-Care (MEDLINE® CARELINE®) The control box CBJ Care is part of the JUMBO system. JUMBO is a modular system combining an actuator, control box (CBJ Care), battery (BAJ1, BAJL), wall charger (CHJ2), control box prepared for external charger by use of wall-plug charger CH01, and a hand control in a flexible solution, specially developed for patient lifts. The complete system contains a series of features which meet the patient’s need for a safe and comfortable lift.
Functionality – JUMBO Care with display Below you find information about what to read-out on the display version of JUMBO Care. Basically the functionality for the display version is the same as the diode version, but more information can be read out on the display. Driving information As long as a hand control button function is activated driving information will be shown on the display. Either lifting arm up, lifting arm down, legs in or legs out or tilt of sling.
Overload Channel 1 only: When overload occurs (according to the pre- defined current cut off limit) the overload symbol will be shown on the display. The ‘MAX’ text will blink 3 times and the overload symbol will be shown for 10 seconds in total. Service information read-out Basic service information can be read out on the display. To get the service information on the display please press the lifting arm up button (only ½ second press).
. CBJ-Home (MEDLINE® CARELINE®) The CBJ-Home is a specially developed solution for patient lifts. The complete system consists of a control box and a battery enclosed in a single elegant module. The system is approved according to medical safety standards and contains a series of features ensuring a safe comfortable lift, e.g. the CBJ-Home is equipped with a soft-start function, electrical emergency lowering, emergency stop etc. Usage: • CBJ Home with internal charger: Nominal current draw max.
. COBO (MEDLINE® CARELINE®) The COBO is an interface box specially developed for use together with the JUMBO battery pack (BAJ1/BAJ2 and BAJL Li-Ion) and the CU20 control unit. It is also possible to connect other 24V lead acid customer batteries or fixed power supply. Safety: The COBO has a monitoring circuit for the FET transistor. If the FET is damaged the CU20 will go into fatal error mode. In this case the COBO is defective and must be replaced.
Recommendations • Choose CU200XXXXX2XXXX if positioning/memory function is to be used. • It is recommended that the COBO is serviced according to the relevant national norms for the applications in which it is used, however all electrical parts must be checked at least once a year. • The COBO should be cleaned regularly, in order to maintain good hygiene. It is not allowed to use chemicals to clean the box. • Only use COBO together with CU20.
9. MBJ1/2/3 (MEDLINE® CARELINE®) Depending on of what your JUMBO system consists you need to use one of the following three mounting brackets. IP protection is only valid when the JUMBO system is mounted vertically. All three brackets include matching screws (IPX1, IPXX and IPX5 are delivered with stainless screws). The mounting screws for the control box, charger must be tightened with a maximum torque of 1 Nm. MBJ1 For use together with CBJ1 or CBJ2 or CBJC, CHJ2 and BAJ1 or BAJ2. I. e.
8. Information on specific accessories If the actuator is to be equipped with accessories, these must be specified when ordering the actuator from LINAK. There are the following possibilities: 1) TR6/TR7 External transformer If the TR6 or TR7 fixed cable connection becomes damaged the transformer must be replaced. 1. BA16 Lead acid (MEDLINE® CARELINE®) The battery box BA16 is developed for use together with the LINAK CA and CO control box series to support power backup.
Mounting instructions: BA16 must be mounted with M4 screws due to the battery weight. Make sure the surface touching the BA16 mounting surface is flat and use all 4 screws. The diameter of the screw cap must be maximum 8mm. Drawing No.: 0273044-SALES Drawing No.: 0273044-SALES SCALE : 2:1 d1 M4 d2 max 8 k 4 SCALE : 2:1 According to ISO 2009 Cable Cable Mini-fit (4 pole) with angle to Mini-fit (4 pole) straight Cable lock 0273044 For cable details see chapter 4.1.
2. BA18 (MEDLINE® CARELINE®) The BA18 is a cost-effective lead acid battery without integrated charger that can be used in combination with CO53 and with the long-established control box range, of which many are now legacy products.
3. BA19 Lead acid (MEDLINE® CARELINE®) The BA19 lead acid backup battery has been developed specifically for use with the new control boxes CA30/CA40 and CO61. It is a compact and cost-efficient battery with built-in charger and cable management.
Recommendations: • • • • • Do not exceed the storage temperature as it will shorten the product life and reduce performance. Allow the battery to settle to room temperature before use. Do not exceed the duty cycle 2/18 as it will shorten the life, reduce performance, and eventually activate overcurrent protection. BA19 is not intended for use in outdoor applications. If the battery is completely discharged, then recharge the battery before storage.
4. BA21 Li-Ion (MEDLINE® CARELINE®) The BA21 Li-Ion back-up battery pack has been specially developed for use with the new control boxes COxx and CAxx, e.g. CO61 and CA40, etc. It is a low weight battery with built-in charger and high performance and safety. Features and Options Weight: 0.
• Disconnect the mains cable to the application at the power outlet. • Release the control box off the application by pressing the tab on the mounting clip. • Open the lid to the control box by releasing the locking clips. • Open the lid to the battery box by releasing the locking clips. • Open the lid for access. • Remove the power cable from the control box by inserting a screwdriver into the locking clip marked.
• If the clips are not engaged fully, the tab will be sticking out as indicated. • Insert the battery connection cable into the battery port in the control box. • Reconnect any actuator cables or hand control cables to the correct ports. • Close the control box lid, ensuring that the locking clips fully engage. • Ensure that it is fully connected. • Reconnect the mains cable to the control box, ensuring that the locking clip engages. • Turn on or reconnect the mains outlet.
Safety feature BA21 Li-Ion contains several mechanisms to protect itself from being damaged due to excessive use. In case of overheating, the device will activate a thermal protection. No power output will be available until the temperature has returned to normal operating range. Overheating may occur by extensive use at high temperature or by exceeding the 1/19 duty cycle. BA21 safety LINAK Li-Ion batteries for medical use are designed and manufactured to be safe through the product life.
5. CS16 (TECHLINE®) The CS16 electronic limit switch is connected between the LINAK® actuator and a non-LINAK power supply, where it cuts out the current to the actuator in end position of if an obstacle is encountered. The PCB contains a variable current limit setting and is available in different versions, depending on the actuator with which it is to be used.
7. Massage Motor Medical (MEDLINE® CARELINE®) The massage motor can be added to all kinds of couches and tables, chairs or beds for treatment and examination. It enables comfort, relaxation and tension release for patients and clients. The massage motors are directly connected to the actuator port at the control box – no extra wiring at the application, simple and easy mounting.
8. MJB2 (MEDLINE® CARELINE®) The MJB2 is a compact 2-port repeater designed for use together with analogue or OpenBus™ control boxes. It is optimised for use in systems where 1 extra port is needed for easy connection of a hand control, a foot switch or an accessory like the UBL. It is easy to integrate in a wide range of healthcare applications such as hospital beds, surgery tables, and treatment chairs.
9. MJB5 Plus Port Repeater (MEDLINE® CARELINE®) The modular junction box MJB5 Plus is designed for use together with OpenBus™ control boxes. The MJB5 Plus makes it possible to connect multiple hand controls and attendant controls. It can even be used for charging or to connect the Under Bed Light and 3rd party products. MJB5 Plus Port Repeater, version 000: The MJB5 Plus version 000 is used where there is a need for more ports than available in the control box.
MJB5 Plus mounting without bracket MJB5 Plus can be mounted directly on the application from the MJB5 Plus side or via the top. It is recommended to use screw type DIN912 M4 for mounting without the bracket. When mounting, ensure that a screw torque limit of 1.0 Nm is not exceeded.
Unlocking of cable locking mechanism If a cable needs to be replaced or added, we recommend to open the cable locking mechanism as described below and demonstrated in our video on LINAK.com - MJB5 Plus section brochures and manuals. CLICK Video guide available on www.linak.com CLICK 1. Insert a pin or screwdriver with a diameter of 3.5 to 4 mm. 2. Pull downwards until a click sound is heard. Repeat this for the other locking hole. 3. The cable mechanism is now unlocked.
Multi-flexible mounting bracket for MJB5 Plus - article no. 1015W1010-A Drawing no.: 0835012 Mount the bracket on the side or the back of the MJB5 Plus. If it is necessary to dismount the MJB5 Plus from the application, we recommend to use the bracket for mounting. Use special screws type WN1423 K60x16, ordering no. 0002645. The screws must be mounted with a torque of maximum 1,0 Nm. The screw head will then be flush with the bracket.
The MJB5 Plus bracket is for mounting on the CO and CB OpenBus control boxes. Mounting examples MJB5 Plus with bracket mounted on the side. MJB5 Plus mounted on the side of the control box. MJB5 Plus with bracket mounted on the back. MJB5 Plus mounted on the side of the control box.
10. MJB5 Plus SMPS (MEDLINE® CARELINE®) The modular junction box MJB5 Plus is designed for use together with OpenBus™ control boxes. The MJB5 Plus makes it possible to connect multiple hand controls and attendant controls. It can even be used for charging or to connect the Under Bed Light and 3rd party products. MJB5 Plus with Switch Mode Power Supply (SMPS) 5V SMPS, version 502-010. The SMPS is to be used where there is a need for power supply near the bed. For example, to charge electronical devices.
MJB5 Plus mounting without bracket MJB5 Plus can be mounted directly on the application from the MJB5 Plus side or via the top. It is recommended to use screw type DIN912 M4 for mounting without the bracket. When mounting, ensure that a screw torque limit of 1.0 Nm is not exceeded.
Unlocking of cable locking mechanism If a cable needs to be replaced or added, we recommend to open the cable locking mechanism as described below and demonstrated in our video on LINAK.com - MJB5 Plus section brochures and manuals. CLICK Video guide available on www.linak.com CLICK 1. Insert a pin or screwdriver with a diameter of 3.5 to 4 mm. 2. Pull downwards until a click sound is heard. Repeat this for the other locking hole. 3. The cable mechanism is now unlocked.
Multi-flexible mounting bracket for MJB5 Plus - article no. 1015W1010-A Drawing no.: 0835012 Mount the bracket on the side or the back of the MJB5 Plus. If it is necessary to dismount the MJB5 Plus from the application, we recommend to use the bracket for mounting. Use special screws type WN1423 K60x16, ordering no. 0002645. The screws must be mounted with a torque of maximum 1,0 Nm. The screw head will then be flush with the bracket.
The MJB5 Plus bracket is for mounting on the CO and CB OpenBus control boxes. Mounting examples MJB5 Plus with bracket mounted on the side. MJB5 Plus mounted on the side of the control box. MJB5 Plus with bracket mounted on the back. MJB5 Plus mounted on the side of the control box.
11. MJB5 Plus Gateway (MEDLINE® CARELINE®) The modular junction box MJB5 Plus is designed for use together with OpenBus™ control boxes. The MJB5 Plus makes it possible to connect multiple hand controls and attendant controls. It can even be used for charging or to connect the Under Bed Light and 3rd party products.
Relay 1 Relay 1 is NO when connected to mains and NC when no mains, this means the relay is “active”, when connected to mains (closed loop). The closed loop principal is to ensure that a notification is sent if power is missing on the bed. When mains is disconnected (power is missing), the relay will go from NO to NC which will automatically result in a notification (status indicator) on the OpenBus. Relay 1 can be activated via the patient control (Key1/Key4) or the attendant control (Key2 / Key3).
Example of closed loop: 3rd party site LINAK site In this state no notification Example of open loop: 3rd party site LINAK site In this state no notification Modular plug cable, narrow/wide alignment grooves: Modular plug with wide alignment groove. To be used with LINAK products. Can be connected to all ports in the MJB5 Plus, both ports with narrow and wide alignment grooves. Page 205 of 271 Modular plug with narrow alignment groove. To be used with 3rd party products.
MJB5 Plus mounting without bracket MJB5 Plus can be mounted directly on the application from the MJB5 Plus side or via the top. It is recommended to use screw type DIN912 M4 for mounting without the bracket. When mounting, ensure that a screw torque limit of 1.0 Nm is not exceeded.
Unlocking of cable locking mechanism If a cable needs to be replaced or added, we recommend to open the cable locking mechanism as described below and demonstrated in our video on LINAK.com - MJB5 Plus section brochures and manuals. CLICK Video guide available on www.linak.com CLICK 1. Insert a pin or screwdriver with a diameter of 3.5 to 4 mm. 2. Pull downwards until a click sound is heard. Repeat this for the other locking hole. 3. The cable mechanism is now unlocked.
Multi-flexible mounting bracket for MJB5 Plus - article no. 1015W1010-A Drawing no.: 0835012 Mount the bracket on the side or the back of the MJB5 Plus. If it is necessary to dismount the MJB5 Plus from the application, we recommend to use the bracket for mounting. Use special screws type WN1423 K60x16, ordering no. 0002645. The screws must be mounted with a torque of maximum 1,0 Nm. The screw head will then be flush with the bracket.
The MJB5 Plus bracket is for mounting on the CO and CB OpenBus control boxes. Mounting examples MJB5 Plus with bracket mounted on the side. MJB5 Plus mounted on the side of the control box. MJB5 Plus with bracket mounted on the back. MJB5 Plus mounted on the side of the control box.
12. MJB5 Plus UBL (int) (MEDLINE® CARELINE®) The modular junction box MJB5 Plus is designed for use together with OpenBus™ control boxes. The MJB5 Plus makes it possible to connect multiple hand controls and attendant controls. It can even be used for charging or to connect the Under Bed Light and 3rd party products. MJB5 Plus with Under Bed Light (int.
MJB5 Plus mounting without bracket MJB5 Plus can be mounted directly on the application from the MJB5 Plus side or via the top. It is recommended to use screw type DIN912 M4 for mounting without the bracket. When mounting, ensure that a screw torque limit of 1.0 Nm is not exceeded.
Unlocking of cable locking mechanism If a cable needs to be replaced or added, we recommend to open the cable locking mechanism as described below and demonstrated in our video on LINAK.com - MJB5 Plus section brochures and manuals. CLICK Video guide available on www.linak.com CLICK 1. Insert a pin or screwdriver with a diameter of 3.5 to 4 mm. 2. Pull downwards until a click sound is heard. Repeat this for the other locking hole. 3. The cable mechanism is now unlocked.
Modular plug cable, narrow/wide alignment grooves: Modular plug with wide alignment groove. To be used with LINAK products. Can be connected to all ports in the MJB5 Plus, both ports with narrow and wide alignment grooves. Modular plug with narrow alignment groove. To be used with 3rd party products. Can only be connected to ports in the MJB5 Plus with narrow slit. This is to prevent 3rd party products to interfere with the OpenBus™ connections.
Multi-flexible mounting bracket for MJB5 Plus - article no. 1015W1010-A Drawing no.: 0835012 Mount the bracket on the side or the back of the MJB5 Plus. If it is necessary to dismount the MJB5 Plus from the application, we recommend to use the bracket for mounting. Use special screws type WN1423 K60x16, ordering no. 0002645. The screws must be mounted with a torque of maximum 1,0 Nm. The screw head will then be flush with the bracket.
The MJB5 Plus bracket is for mounting on the CO and CB OpenBus control boxes. Mounting examples MJB5 Plus with bracket mounted on the side. MJB5 Plus mounted on the side of the control box. MJB5 Plus with bracket mounted on the back. MJB5 Plus mounted on the side of the control box.
13. MJB5 Plus UBL (ext) (MEDLINE® CARELINE®) The modular junction box MJB5 Plus is designed for use together with OpenBus™ control boxes. The MJB5 Plus makes it possible to connect multiple hand controls and attendant controls. It can even be used for charging or to connect the Under Bed Light and 3rd party products. MJB5 Plus with Under Bed Light (ext.
MJB5 Plus mounting without bracket MJB5 Plus can be mounted directly on the application from the MJB5 Plus side or via the top. It is recommended to use screw type DIN912 M4 for mounting without the bracket. When mounting, ensure that a screw torque limit of 1.0 Nm is not exceeded.
Unlocking of cable locking mechanism If a cable needs to be replaced or added, we recommend to open the cable locking mechanism as described below and demonstrated in our video on LINAK.com - MJB5 Plus section brochures and manuals. CLICK Video guide available on www.linak.com CLICK 1. Insert a pin or screwdriver with a diameter of 3.5 to 4 mm. 2. Pull downwards until a click sound is heard. Repeat this for the other locking hole. 3. The cable mechanism is now unlocked.
Under Bed Light cable: 0964135: Under Bed Light cable, Length 2500 mm. Please use metal screws for mounting. Screw placement Modular plug cable, narrow/wide alignment grooves: Modular plug with wide alignment groove. To be used with LINAK products. Can be connected to all ports in the MJB5 Plus, both ports with narrow and wide alignment grooves. Modular plug with narrow alignment groove. To be used with 3rd party products. Can only be connected to ports in the MJB5 Plus with narrow slit.
Multi-flexible mounting bracket for MJB5 Plus - article no. 1015W1010-A Drawing no.: 0835012 Mount the bracket on the side or the back of the MJB5 Plus. If it is necessary to dismount the MJB5 Plus from the application, we recommend to use the bracket for mounting. Use special screws type WN1423 K60x16, ordering no. 0002645. The screws must be mounted with a torque of maximum 1,0 Nm. The screw head will then be flush with the bracket.
The MJB5 Plus bracket is for mounting on the CO and CB OpenBus control boxes. Mounting examples MJB5 Plus with bracket mounted on the side. MJB5 Plus mounted on the side of the control box. MJB5 Plus with bracket mounted on the back. MJB5 Plus mounted on the side of the control box.
14. MJB5 Plus AOC (MEDLINE® CARELINE®) The modular junction box MJB5 Plus is designed for use together with OpenBus™ control boxes. The MJB5 Plus makes it possible to connect multiple hand controls and attendant controls. It can even be used for charging or to connect the Under Bed Light and 3rd party products. MJB5 Plus – Analogue to Openbus Converter (AOC) MJB5 Plus versions 509 The AOC is an Analogue to OpenBus™ Converter.
AOC pin connections Please be aware of the cable and pin orientation: Pin 10 Pin 1 Port/pins Port used for Port 1 OpenBus connection Port 4 Port 5 Common ground Port 2 Port 3 Common positive Common positive OpenBusTM connection OR extra 3rd party connection 3rd party connections 3rd party connections GND GND GND Vperm* Vperm* Pin 1 Pin 2 Pin 3 Input 1 Input 9 Pin 4 Input 2 Input 10 Pin 5 Input 3 Input 11 Pin 6 Input 4 Input 12 Input 17 Input 5 Input 13 Input 14 Pin 7 Pin
Dimension Page 224 of 271
Standard configuration Table 1: Overview of pin activation and OpenBus codes When a pin is activated, different OpenBus codes from port 4 and port 5 will be sent. Table 1 is the standard configuration. Example: If Pin 3 is activated, it will send OpenBus code H1 from port 4 and H11 from port 5.
MJB5 Plus mounting without bracket MJB5 Plus can be mounted directly on the application from the MJB5 Plus side or via the top. It is recommended to use screw type DIN912 M4 for mounting without the bracket. When mounting, ensure that a screw torque limit of 1.0 Nm is not exceeded.
Unlocking of cable locking mechanism If a cable needs to be replaced or added, we recommend to open the cable locking mechanism as described below and demonstrated in our video on LINAK.com - MJB5 Plus section brochures and manuals. CLICK Video guide available on www.linak.com CLICK 1. Insert a pin or screwdriver with a diameter of 3.5 to 4 mm. 2. Pull downwards until a click sound is heard. Repeat this for the other locking hole. 3. The cable mechanism is now unlocked.
Modular plug cable, narrow/wide alignment grooves: Modular plug with wide alignment groove. To be used with LINAK products. Can be connected to all ports in the MJB5 Plus, both ports with narrow and wide alignment grooves. Modular plug with narrow alignment groove. To be used with 3rd party products. Can only be connected to ports in the MJB5 Plus with narrow slit. This is to prevent 3rd party products to interfere with the OpenBus™ connections.
Multi-flexible mounting bracket for MJB5 Plus - article no. 1015W1010-A Drawing no.: 0835012 Mount the bracket on the side or the back of the MJB5 Plus. If it is necessary to dismount the MJB5 Plus from the application, we recommend to use the bracket for mounting. Use special screws type WN1423 K60x16, ordering no. 0002645. The screws must be mounted with a torque of maximum 1,0 Nm. The screw head will then be flush with the bracket.
The MJB5 Plus bracket is for mounting on the CO and CB OpenBus control boxes. Mounting examples MJB5 Plus with bracket mounted on the side. MJB5 Plus mounted on the side of the control box. MJB5 Plus with bracket mounted on the back. MJB5 Plus mounted on the side of the control box.
15. Simulator tool (MEDLINE® CARELINE® ) The Simulator Tool is a software that can be used to simulate hand control functions on OpenBus™ and analogue actuator systems. With the Simulator Tool, sequences of actuator movements can be programmed and repeated in order to test actuator systems. USB to OpenBus™ gateway: The gateway acts as an interface between the Simulator Tool Software and the OpenBus control box. Together with the Simulator Tool Software, it can be used for test and demo purposes only.
16. SLS (MEDLINE® CARELINE® TECHLINE®) There are 2 types of SLS switches, an analogue and an OpenBus™ type. The analogue SLS can be used as safety feature to cut off the current to the actuator. The SLS is available with 1 or 2 switches (activated by the same button). It can be placed to prevent an unintentional positioning of the various frame segments in relation to each other or simply as an external limit switch to protect the frame against the full thrust of the actuator in end position.
The ultrasonic welded SLS is fully compatible with the SLS fitted with screws. All outer dimensions as well as the activation point are unchanged. Cable length Safety Limit Switch Cable length Signal Limit Switch 1 17. Under Bed Light (MEDLINE® CARELINE®) The Under Bed Light (UBL) can be mounted under the bed to provide a discrete guiding light when the patient leaves the bed during night.
18. Under Bed Light 2 (MEDLINE® CARELINE®) Usage: • Operating temperature: • Storage temperature: • Relative humidity: • Atmospheric pressure: • Height above sea level: • Approvals: The Under Bed Light (UBL2) provides a powerful light with a good distribution. The UBL2 is to be used for beds within hospitals, nursing homes and in homecare.
19. QLCI2 (MEDLINE® CARELINE®) The Quad Load Cell Interface 2 (QLCI2) for the hospital and care segment is a scale system accessory with weighing capabilities and Out of Bed functionality. The housing, makes the QLCI2 easy to mount by unique slide-on brackets and has an IPX6 Washable DURATM ingress protection. The QLCI2 supports the LINAK OpenBusTM system offering a high level of customisation. LINAK A/S delivers the OpenBus™ system only and is not responsible for any products (i.e.
Mounted on frame: Service lid Integrated hinge Removable if needed Release hook 3-way cable exit Very small built-in dimension: Height: Min. 90 mm (lid can be removed) Length: + 20 mm (locking system) Easy-mount bracket Drawing No.: 1013W4008 Mounting bracket (frame flat) Article No. 1015W1001: Mounting bracket (frame flat) w/M4 nuts Article No. 1015W9009: Drawing No.: 1015W1009 Drawing No.: 1015W4001 It is recommended to mount the QLCI2 in a position that allows water to escape.
Recommendations It is recommended to mount the load cells on the bottom frame to ensure a stable system. Shielded load cell cables will be damaged if exposed to sharp bends. Therefore, if bended, cables should have a minimum bending radius of 60 mm. It is not allowed to bend load cell cables repeatedly, so mount cables on non-moving parts, like the bed frame. Load cell cables should not exceed a length of 2700 mm. In Europe weight systems are subject to important legal restrictions.
9. Information on specific LIFT products 1. CAL40 (MEDLINE® CARELINE®) The control box CAL40 is part of the LIFT40 product series specially developed for patient lifts. LIFT40 is a complete system consisting of the control box CAL40 or CAL40+, a battery BAL40 and an external charger CHL40 in a flexible solution. Combined with one or more actuators and a hand control you have a complete system for modern patient lifts. Usage: • With internal charger: Nominal current draw max.
General functionality – LIFT40 Remove battery: 1) Lift handle upwards to release lock 2) Grab handle, pull out and slightly to the right 3) Lift off the battery - carry in handle Grab handle Lift off 2 3 Open up 1 Mounting battery: 1) With open handle - position battery base over the guide track and lower it in place 2) Steer in upper part of battery 3) Press down handle to lock battery in place Steer top in 2 Position 1 Page 239 of 271 Lock handle 3
Cable mounting and cable cover: LIFT40 control boxes have a uniquely designed cable cover which also works as an integrated cable cover when closed.