S430 Connections 6: Mains through 1 Female connection top entry, male connection side entry 3: Motor supply 4: Mains through 1 Female connection top entry, male connection side entry 6: Mains through 2 Male connection top entry, female connection side entry 3: Motor supply 4: Mains through 2 Male connection top entry, female connection side entry 6: Mains through 1 Female connection top entry, male connection side entry 3: Motor supply 4: Mains through 1 Female connection top entry, male connection si
S540 Connections 3: Motor supply 4: Mains through 2 Male connection top entry, female connection side entry 6: Mains through 2 Male connection top entry, female connection side entry Mounting of a product on the side of an LC3 It is possible to mount a product on all 4 sides of the LC3 at the same time. However, it requires that the holes for mounting are placed with different distances from the top and/or bottom plate. Otherwise the mounting screws will collide.
When mounting more than one LC3 you need to consider the fixation: The reason why it is important only to fix one column, is that the columns will not move exactly in parallel – even if you have positioning such as hall. If you do have a trend/anti-trend function in your application, you need to mount one or more of the LC3’s with a slider. If more than one column is fixed it can lead to dangerous situations. Having sliders prevents the column from bending as illustrated below.
Input/output specifications: dual Hall positioning Dual Hall digital (F3) with power switch Item Pin configuration Specification Comment Pin 1 GND Pin 2 VCC Pin 3 M+ Pin 4 HALL A Pin 5 HALL B Pin 6 M- VCC 4-15V Current Maximum 15 mA @10 kΩ and 1 nF load. See diagram. HALL A/B TState is minimum 5ms in all states (11,10,00,01) at a minimum mechanical load. Tested with the above specified load. Duty cycle Hall A 30-70% Duty cycle Hall B 30-70% Low level
Potentiometer Potentiometer cables: Columns with Potentiometer feedback option require specific actuator cables, both for termination through top-plate (option Txxx) and through side-entry (option Sxxx). Table 1, connection to the motor and to the potentiometer and their colour.
Recommendations: • • • • • • • Please follow the important LC3 mounting guidelines. LC3 is for use in push or pull applications, cable outlet from smallest profile (top) or biggest profile (bottom). See top and bottom plate dimensions. When washing according to IPX6 parameters please notice that the large profile of the LC3 column must be placed upwards (bottom up). We recommend making a functional test of the application with all accessories connected before putting it into operation.
5. Information on specific control boxes Please be aware if the control box is not visible after mounting, all information regarding limitation of use shall be marked on the end product. Output voltage On control boxes connected to the mains the voltage of the actuator output is dependent on load, and the no-load voltage can reach 50 V. Control boxes connected to a battery can reach a voltage of 30 V during charging and no load.
1. CA30/CA40 (MEDLINE® CARELINE®) The control boxes CA30 and CA40 are developed as part of a new control box platform for the care and rehab industry. The control boxes can be mounted separately on the application by means of unique slide-on brackets, but also on several actuator models using a specially designed actuator bracket. Usage: • Duty cycle: 10 % - 2/18 min. on/off continuos use.
Mounting instructions (Example CA30-CA40 with LA40) When mounting the control box onto the actuator (1) Simply slide on the bracket until you hear a clear click (2) Slide on the control box until you hear a click and the box is mounted (3) It is recommended that the CA30/CA40 is mounted in a position that allows water to escape. Recommended torque: 0.6 Nm +/-0.
Mounting of cables and cable lock: The control boxes CA30/CA40 have a uniquely designed cable lid. The lid also works as an integrated cable lock when closed. 1) Mount cable plugs in control box 2) Close lid until lock snaps into place (see arrows) To allow free access to the cables, the lid has a rest position when completely opened. It is possible to remove the lid by lifting it a few degrees and pulling it away from the housing under tight mounting conditions.
2. CA63 (MEDLINE® CARELINE® ) The LINAK control box CA63 with analogue input offers a consolidated range of unprecedented features – all utilising standardised technology, interfaces and compatibility. The CA63 for LINAK actuators is intended for the control of, for example, Home Care Bed.
Mounting bracket (frame flat) - article No. 1015W1001: Mounting bracket (frame flat) w/M4 nuts - article No. 1015W9009: Drawing No.: 1015W4001 Drawing No.: 1015W4009 CA63 - mounted on frame: Release hook Very small built-in dimension: Height: min. 80 mm (lid can be removed) Length: + 20 mm (locking system) Service lid Integrated hinge Removable if needed Easy-mount bracket 3-way cable exit Drawing No.
Mounting instructions (Example CA63 with LA40) When mounting the control box onto the actuator (1) Simply slide on the bracket until you hear a clear click (2) Slide on the control box until you hear a click and the box is mounted (3) It is recommended that the CA63 is mounted in a position that allows water to escape. Recommended torque: 0.6Nm +/-0.
Mounting of cables and cable lock: CA63 has a uniquely designed cable lid. The lid also works as an integrated cable lock when closed. 1) Mount cable plugs in control box 2) Close lid until lock snaps into place (see arrows) To allow free access to the cables, the lid has a rest position when completely opened. It is possible to remove the lid by lifting it a few degrees and pulling it away from the housing under tight mounting conditions.
Recommendations • If there is a risk that the system is overloaded and therefore shuts down thermally, LINAK recommends to use quick release actuators. These will allow functions to be lowered manually in case of a CA63 malfunction due to misuse/abuse. • If the customer has other essential performance than “no unintended movement”, he must consider this in his own risk analysis. LINAK disclaims any liability.
3. CB6 (MEDLINE® CARELINE®) The control box CB6 has been specially developed for use together with LA27/LA40 actuator in the care and rehab industry. The control box is designed to be mounted on the actuator LA27/LA40 as with the CB9 and LA31 system. The CB6 control box has a LED power ON indicator, detachable mains cable and strain relief for all cables. The control box CB6 communicates with the LA27/LA40 actuator by means of the built-in end stop signal switches in the actuator.
4. CB6P2 (MEDLINE® CARELINE®) The CB6P2 platform is introduced to obtain a powerful and optimised solution to customers looking for existing analogue input systems. It is based on OpenBus™ technology, but to meet existing analogue systems it has an analogue input and therefore OpenBus™ accessories cannot be connected. Combination Overview CB6P2 is meant for use with: LA27 std. motor with Hall (cable type ‘A’) LA27 std. motor (cable type ‘B’) LA31 std. / fast motor with / without Hall LA34 std.
6. CB8A (MEDLINE® CARELINE® TECHLINE®) The CB8A is a battery powered control box operating up to 3 actuators individually. One of these channels can be used either as an external emergency stop device or for battery charging. Simple design and high quality construction make the CB8A an ideal control box choice for mains-free operation of beds, chairs, tables and many other mobile applications. Usage: • Duty cycle: • Ambient temperatures: • Approvals: Max. 5% or 3 min. in use followed by 57 min.
9. CB9 CARELINE and CB9 CARELINE Basic (MEDLINE® CARELINE®) The CB9 CARELINE and CB9 CARELINE Basic have been developed for use together with LA31/ LA31R, LA34/LA34R* in the Care & Rehab industry. CB9 and LA31 can be fully integrated which saves mounting and wiring or be installed separately. Exchangeable mains cables, Electronic Overload Protection (EOP), EAS, earth connection (Class 1) and exchangeable mains fuse makes CB9 a good choice for the simple hospital and care beds. Usage: • Duty cycle: Max.
10. CB20 (MEDLINE® CARELINE®) CB20 is a platform which offers a unique safety concept, logging off service data and it is possible to connect a variety of accessories to the control box. The CB20 consists of 3 modules : CP20 = Control Power CU20 = Control Unit BA20 = Battery. For recommendations on maintenance and storage of the BA20 battery packs, please refer to chapter 5. Usage: • Compatible with specific versions of LA23, LA31, LA34, LA40, LA43, LA44 and BL1 • Duty cycle: 10 % ˜ max. 2 min.
12. CO53 (MEDLINE® CARELINE® ) The LINAK® control box CO53 is a new member of the OpenBus™ range. The control box offers a 190 Watt power supply, a small and compact design and the ability to handle multiple hand controls or other OpenBus accessories for applications with a need for more advanced features. Features: • Duty cycle: 10% - 2/18 min.
LED indicator CO53 is equipped with a green LED for indication of mains power connection and an orange LED for indication of hand or foot control activation. When the CO53 is connected to mains, the LED is green and when connected to battery, the LED is off. Connected to MAINS LED colour Indication of operation Green On mains, not activated by hand or foot control. The system is working ok and is ready for normal operation. Green and orange On mains, not activated by hand or foot control.
Screw bracket spacer - article No. 1015W1001: Drawing No.: 1054W4001 CO53 - mounted on frame: Release hook Very small built-in dimension: Height: min.
Mounting instructions (Example CO53 with LA40) When mounting the control box onto the actuator (1) Simply slide on the bracket until you hear a clear click (2) Slide on the control box until you hear a click and the box is mounted (3) It is recommended that the CO53 is mounted in a position that allows water to escape. Recommended torque: 0.6Nm +/-0.
Mounting of cables and cable lock: CO53 has a uniquely designed cable lid. The lid also works as an integrated cable lock when closed. 1) Mount cable plugs in control box 2) Close lid until lock snaps into place (see arrows) To allow free access to the cables, the lid has a rest position when completely opened. It is possible to remove the lid by lifting it a few degrees and pulling it away from the housing under tight mounting conditions.
Recommendations • To avoid unintended activation of actuators if hand control cables short-circuit, LINAK® recommends to use an OpenBus™ system (CO53). • If there is a risk that the system is overloaded and therefore shuts down thermally, LINAK recommends to use quick release actuators. These will allow functions to be lowered manually in case of a CO53 malfunction due to misuse/abuse.
13. CO61 (MEDLINE® CARELINE® ) The LINAK control box CO61 offers a consolidated range of unprecedented features – all utilising standardised technology, interfaces and compatibility. The CO61 for LINAK actuators is intended for the control of, for example, hospital bed movement.
Mounting bracket (frame flat) - article No. 1015W1001: Mounting bracket (frame flat) w/M4 nuts - article No. 1015W9009: Drawing No.: 1015W4001 Drawing No.: 1015W4009 CO61 - mounted on frame: Release hook Very small built-in dimension: Height: min.
Mounting instructions (Example CO61 with LA40) When mounting the control box onto the actuator (1) Simply slide on the bracket until you hear a clear click (2) Slide on the control box until you hear a click and the box is mounted (3) It is recommended that the CO61 is mounted in a position that allows water to escape. Recommended torque: 0.6 Nm +/-0.
Mounting of cables and cable lock: CO61 has a uniquely designed cable lid. The lid also works as an integrated cable lock when closed. 1) Mount cable plugs in control box 2) Close lid until lock snaps into place (see arrows) To allow free access to the cables, the lid has a rest position when completely opened. It is possible to remove the lid by lifting it a few degrees and pulling it away from the housing under tight mounting conditions.
Recommendations • To avoid unintended activation of actuators if hand control cables short-circuit, LINAK recommends to use an OpenBusTM system (CO61). • If there is a risk that the system is overloaded and therefore shuts down thermally, LINAK recommends to use quick release actuators. These will allow functions to be lowered manually in case of a CO61 malfunction due to misuse/abuse.
14. CO65 (MEDLINE® CARELINE® ) The LINAK control box CO65 offers a consolidated range of unprecedented features – all utilising standardised technology, interfaces and compatibility. Usage: • Operation temperature: • Storage temperature: • Relative humidity: • Atmospheric pressure: • Meters above sea level: • Approvals: +5 °C to +40 °C -10 °C to +50 °C 20% to 80% - non-condensing 700 to 1060 hPa (3000 m) Max. 2000 meters IEC60601-1 edition 3.
Mounting bracket (frame flat) - article No. 1015W1001: Mounting bracket (frame flat) w/M4 nuts - article No. 1015W9009: Drawing No.: 1015W4001 Drawing No.: 1015W4009 CO65 - mounted on frame: Release hook Very small built-in dimension: Height: min.
Mounting instructions (Example CO65 with LA40) When mounting the control box onto the actuator (1) Simply slide on the bracket until you hear a clear click (2) Slide on the control box until you hear a click and the box is mounted (3) It is recommended that the CO65 is mounted in a position that allows water to escape. Recommended torque: 0.6 Nm +/-0.
Mounting of cables and cable lock: CO65 has a uniquely designed cable lid. The lid also works as an integrated cable lock when closed. 1) Mount cable plugs in control box 2) Close lid until lock snaps into place (see arrows) To allow free access to the cables, the lid has a rest position when completely opened. It is possible to remove the lid by lifting it a few degrees and pulling it away from the housing under tight mounting conditions.
Recommendations • To avoid unintended activation of actuators if hand control cables short-circuit, LINAK recommends to use an OpenBus system (CO65). • If there is a risk that the system is overloaded and therefore shuts down thermally, LINAK recommends to use quick release actuators. These will allow functions to be lowered manually in case of a CO65 malfunction due to misuse/abuse.
15. CO71 (MEDLINE® CARELINE® ) The LINAK control box CO71 offers a consolidated range of unprecedented features – all utilising standardised technology, interfaces and compatibility. The CO71 for LINAK actuators is intended for the control of, for example, hospital bed movement.