Operation of DR Detector
73
Exposure time for Aero Sync mode of the
AeroDR 3 Detector
IMPORTANT
If the conditions are not met, X-rays may not be detected or
streaks may occur in the image.
The minimum exposure time applies to the time when X-
ray is exposed onto the center of the DR Detector. If X-ray
cannot be exposed onto the center, contact Konica Minolta
technical representatives.
The cycle time will be extended by setting the longer maxi-
mum exposure time. It is recommended that the optimal
maximum exposure time be set.
Set the maximum exposure time according to the
sensitivity button setting of the image processing
controller. The available values for the maximum
exposure time are as follows.
Maximum
exposure time
Extra
Large
Large Med
0.8 seconds
Can be
set
Can be
set
Can be
set
1.8 seconds
Can be
set
Cannot be
set
Cannot be
set
4.0 seconds
Can be
set
Cannot be
set
Cannot be
set
When using Aero Sync mode, set the minimum
exposure time as follows:
DR Detector
Minimum
exposure time
AeroDR 3 Detector 1 millisecond
HINT
When using ImagePilot as the image processing controller,
it is not possible to change the maximum exposure time.
Identify the DR Detector that users can
use in Aero Sync mode
IMPORTANT
The AeroDR 1012HQ, AeroDR 2 Detector, and AeroDR 3
Detector can use the Aero Sync mode.
Users can not use Aero Sync Mode in the DR Detector if it
is AeroDR 1417S.
Users cannot use Aero Sync mode in the DR Detector if the
rst four digits of its SerialNumber are shown in the follow-
ing table in the AeroDR 1417HQ or AeroDR 1717HQ.
DR Detector
SerialNumber that can-
not be used
AeroDR 1417HQ A45Y
AeroDR 1717HQ A54T
Confirm that if the following identification is in the
DR Detector that users can use in Aero Sync mode.
AeroDR P-11
1417HQ
()
Identication
DR Detector Identication
AeroDR 1417HQ 1417HQ
AeroDR 1717HQ 1717HQ
Exposure with the Aero Sync mode
To conrm X-ray detection, make sure the following
conditions are met.
For the AeroDR 1417HQ, AeroDR 1717HQ and
AeroDR 1012HQ:
Perform exposure by following the survey results
of the installation. For details, contact Konica Mi-
nolta technical representatives.
For the AeroDR 2 Detector:
Perform X-ray exposure at a range of 10 cm
square or more at the center of the DR Detector.
For the AeroDR 3 Detector:
Perform X-ray exposure at a range of 10 cm
square or more at the center of the DR Detector.
You can select the maximum exposure time from
0.8 seconds or shorter, 1.8 seconds or shorter and
4.0 seconds or shorter in the Extra Large mode.