Users Manual Part 6

IBA | 43-12 |
Clinical User’s Guide
Volume 1 - Treatment Session
|Part IX
- adaPT
deliver
in Physics Mode Using adaPTdeliver in Physics Mode
|
Remote positioning controls: in Physics mode, the Manual mode of the
remote positioning controls is available by default. That is not the case in Clinical
mode, where the Manual mode of the remote positioning controls is only
available upon explicit request.
Figure 43-10. Equipment Preparation Screen
Physics Mode
Verifying the Position of Devices
Note: If you want to use another gantry angle than the one from the prescription,
you should edit the related field from the B
EAM SELECTION SCREEN (see Figure 43-9)
before entering the Preparation Screen.
The Set Range is done with the gantry angle defined in the B
EAM SELECTION SCREEN,
not the current gantry angle.
Most of the differences in terms of PMS between Clinical mode and Physics mode
are visible on the PMS I
NTERFACE PANEL.
WARNING Be aware that in Physics Mode, you are not blocked if you choose to
irradiate with a gantry angle different from the prescription; no warning
message will appear.

Summary of content (112 pages)