Da Vinci IS5000 System User Manual 555500-01 Rev. A
Appendix E System Specifications
Contents
E.1 Design Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-1
E.2 Power Requirements Main Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-1
E.3 Power Specifications Main Components . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-4
E.4 Power Specifications Tower Components . . . . . . . . . . . . . . . . . . . . . . . . . . .E-5
E.5 Power Specifications Integrated Equipment . . . . . . . . . . . . . . . . . . . . . . . . . .E-5
E.6 Laser Specifications Robot . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-6
E.7 Physical Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-6
E.8 Environmental Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-7
E.9 Crate Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-7
E.10 Electromagnetic Compatibility. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .E-7
E.11 Wireless Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-16
E.1 Design Life
The system has a design life of 5 years. However, even after 5 years, the standard maintenance
contract provides preventive maintenance, inspection, and re-qualification by Intuitive personnel
that extends the service life of the system until the next scheduled service, which occurs at least
every 2 years.
E.2 Power Requirements Main Components
The system has three main components requiring electrical power: the Console, the Robot, and the
Tower. The integrated VIO dV electrosurgical unit also requires electrical power. To ensure optimum
performance, make sure each component of the system (including the VIO dV) is connected to a
dedicated, noise-free, and well-grounded AC power outlet.
CAUTION: To avoid overloading circuits, all three components Tower, Robot, and Console
must operate on separate, dedicated power circuits. Do not connect ancillary devices such
as insufflators or energy devices on the same circuit as the system components, particularly
the Tower because it has large power requirements. Ancillary devices must be connected to
wall outlets on separate circuits from all system components.
A label on the rear of the Tower reinforces this caution. Its text appears below the label image.
Figure E.1 Tower power caution label
“Vision Cart requires dedicated circuit. Do not connect auxiliary equipment to the circuit
powering the Vision Cart.”