Thank you for purchasing InBody H20 or H20Ⓑ. H20Ⓑ has Bluetooth function but H20 doesn’t. Please read this manual carefully and use it with proper operation. Keep it handy for future reference. ✽This product is medical equipment.
ⓒ 1996-2012 Biospace Co., Ltd. All rights reserved. Notice This user’s manual cannot be copied, translated or transformed into different media without a prior written consent from Biospace Co., Ltd, according to the copyright laws. This user’s manual and product pertinent to this manual may have misprints or technical flaws, which can be modified without prior notice to the readers. Biospace is not liable for any damage caused by a failure to meet the requirements in this user’s manual. Biospace Co., Ltd.
How to use this manual This user’s manual explains the functions of the InBody H20. Follow the instructions below for effective use this manual. 1. Please read all the instructions in this manual thoroughly before operation. 2. Fully utilize the helpful materials, such as the diagrams and illustrations, to get a clear understanding of the product. 3. Before calling Biospace for assistance, please refer to “ Chapter 11. One-minute check before calling for service ”. 4.
Safety Information Make sure not to use this equipment with medical electrical devices, such as a pacemaker. NO USAR USARLO CON EQUIPO MEDICO ELECTRONICO ASI COMO CON MARCAPASOS. Keep this equipment away from liquid. MANTENER USARLO LEJOS DEL AGUA. 1. Do not use this equipment in combination with the following electronic medical devices.
. You may slip off the device if you step on it with wet feet. 6. The arm consists of a hand electrode. Do not force the arm in the wrong direction. The resulting damage may affect the functioning of the internal cable and circuit board. 7. Individuals with any kind of contagious disease or any kind of injury on the palm of their hand or sole of their foot should not come in contact with this product. 8.
1. Cross Contamination Individuals with any kind of contagious disease or infection must not use or come in contact with the platform or this product. Please be sure to clean the platform with appropriate disinfectant after each use. Never pour any liquid directly on the platform, as it may leak and cause internal damage. Use a soft cloth and appropriate ethyl alcohol to wipe off the platform. Do not wipe the platform with strong chemicals. 2.
8. This equipment should be serviced only by qualified personnel. Contact Biospace for any Customer Service issues. 9. Please use your hands to operate buttons. Using feet may lead to malfunction of buttons. 1. This equipment may cause the above mentioned medical devices to malfunction. This equipment may cause harmful interference to other devices in the vicinity, if not installed and used in accordance with the installation specifications.
Information 1. This product has been designed, manufactured, and inspected under the full quality assurance system of Biospace. Biospace fulfills European MDD(Medical Device Directive), and InBody H20 has acquired the CE conformity marking. 2. This equipment may cause the above mentioned medical devices to malfunction. 3. This equipment may cause harmful interference to other devices in the vicinity if not installed and used in accordance with the installation specifications.
Indicators & Safety Symbols Disposal of old Electrical & Electronic Equipment (Application in the European Union and other European countries with separate collection system) This symbol indicates that this product shall not be treated as household waste. Instead, it shall be handed over to the applicable collection point for the recycling of electrical and electronic 10 equipment.
Workplace Requirements □ Operation Environment Temperature range : 10 – 40 ℃ Relative humidity : 30 – 75 % Atmospheric pressure range : 70 – 106 kPa □ Storage Environment Temperature range : -20 – 70 ℃ Relative humidity : 10 – 95 % (No condensation) Atmospheric pressure range : 50 – 106 kPa □ Battery Ratting : AA Type Battery(1.
Introduction Body Composition consists of 4 major components: Water, Protein, Minerals and Fat. These four elements are the fundamental ingredients the body is comprised of, and it is important for them to be in balance. Body composition analysis is expected to quantify and measure these ingredients. Until recently, diagnosing obesity has focused on appearance, without considering a balanced body composition.
Notice Please read thoroughly and follow the instructions. 1. Please do not press with other object on the platform. 2. Please use your hands to operate buttons. Using feet may lead to malfunction of buttons. 3. Please be sure to avoid keeping or using InBody in wet area, i.e. bathroom. It may cause malfunction of the device. 4. You may slip off the device if you step on it with wet feet. 5. Please wrap InBody and keep it level while storing for a long period. 6.
Contents How to use this manual Ⅰ Safety Information Ⅱ Indicators & Safety Symbols Ⅶ Workplace Requirements Ⅷ Introduction Ⅸ Notice Ⅹ 1. Product Package 1 2. Insert Battery 2 3. Connecting InBody H20Ⓑ and PC 3 4. Weight Measurement Mode 4 5. InBody Test Mode 5 6. Maintenance 10 7. Understanding of Results 11 8. Error messages 14 9.
1. Product Package When opening the box, check to make sure all the following items are included. Included Items ① InBody H20 unit ② Quick Installation Guide ③ Battery (AA 1.5 V 4 ea) ④ Bluetooth dongle To prevent physical shock, use Biospace’s packing material when shipping or transporting the equipment.
2. Insert Battery < Rear part> (1) Open the battery cover at the bottom of the device. (2) Insert the batteries (AA 1.5 V 4 ea) in right direction. (3) Close the battery cover until it clicks. Battery is low when is still on after you turn on InBody H20. Please replace the batteries when is blinking after you turned on InBody H20.
3. Connecting InBodyH20Ⓑ and PC (1) Connect Bluetooth dongle(Bluetooth dongle from product package) to PC. (2) Access http://www.myinbody.com and register. (3) Click “InBody PC Program Install Guide” and install the program. Ⓑ (4) Turn on H20 Ⓑ When the bluetooth is being connected to InBodyH20 , the logo will display on the LCD screen. Please use the Bluetooth dongle provided by Biospace only.
4. Weight Measurement Mode A. Body Composition Analyzing (1) Step on the foot board to measure your weight. (2) When the measurement is over there will be a beep and you will see your weight on the screen. (3) When the weight is not correct, press the foot plate to turn on H20 and wait for 5 seconds. Then you can re-measure your weight. B. Power Off (1) Weight measurement is done, step down from the foot plate and the power will automatically be turned off.
5. InBody Test Mode A. Tips Before Body composition analyzing (1) Make yourself as light as possible. Wear light clothes for measurement. (2) Please go to the bathroom before measurement. Urine and feces affect your weight and consequently the body composition analysis. (3) Please conduct the measurement before meals or at least two hours after meals. Food in the stomach before digestion affects body weight, causing possible errors in the result.
B. Body Composition Analyzing (1) Push the button in the center to turn on H20. (2) Rotate the button in the center to set your height. Wait for few seconds once it is set. (3) The height will appear on the LCD screen. (4) After entering the height hold the handle and step on the foot board to measure your weight. (5) When weight is set, the measurement position icon will blink while checking if the hand and feet electrodes are connected. (6) With measuring sound, InBody measurement will start.
C. How to Hold InBody Electrode Take the hand grip and hold it with the palms and thumbs touching the hand electrodes. Step on the foot electrodes at the platform. Make sure your hands and feet on both sides are not touching each other. Please refer to next page for the right measurement posture. The measurement begins automatically when your position is set. (1) Hand Electrodes Fingers must not leave the surface of the electrodes. Right hand and left hand must not touch each other.
D. Recommended measurement posture ➊ Please keep your eyes to the front. ➋ Grab the hand electrodes with bare hands and be careful not to let your hands touch each other. ➌ Thighs should not touch each other if possible. ✽Tip: For those whose thighs contact each other, it is recommended to wear shorts, which come down below the thighs for accurate measurement. ➍ No moisture or too much callus tissues on the palms or soles of feet may lower the accuracy of the measurement.
E. Power Off (1) InBody test is done, step down from the foot plate and the power will automatically be turned off after 60 seconds. Turn the dial to see the measurement results again in 60 seconds. (2) If you step down the foot plate while measuring, the device will not be turned off for 60 seconds. If you step on the foot plate within 60 seconds the measurement will restart.
6. Maintenance (1) After usage, wipe electrodes with a wet tissue (also known as wet naps, surface wipes, antibacterial tissues, etc.) Wet tissues can be purchased at any local store. (2) Handle the arms with care. Never apply excessive force near the hand electrode joint. The damage caused by misuse may affect the function of the internal cable and electric board. (3) Do not place anything on the platform or apply any pressure onto it when the InBody H20 is not in use.
7. Understanding of Results A. Result Screen During measurement, the InBody H20 displays information of an examinee’s body composition on the LCD. The results are shown while you are on the foot plate. ⓛ Weight (kg) Ideal weight is based on examinee’s height. Using the classical BMI method, the InBody 370 identifies the standard BMI as 22kg/m2 for males, 21.5kg/m2 for western females and 21kg/m2 for asian females.
② Percentage of Body Fat (%) Percentage of Body Fat indicates the percentage of body fat to body weight. The ideal percentage of body fat is 15% for men and 23% for women, while the standard range of Body Fat Mass for men is 10-20% of the weight and 18-28% of the weight for women. In the case of children under the age of 18, children’s standard is used. Ref. Samuel J. Fomon, et al. (1982): Body Composition of reference children from birth to age 10 years.
④ BMI (Body Mass Index, kg/㎡) BMI is determined by using only weight and height and diagnoses superficial obesity. The standard values are 22kg/m2 for male and 21.5kg/m2 for western female and 21kg/m2 for asian female. Formula) BMI = weight (kg) ÷ height2 (㎡) Determination 1) WHO Standard BMI(kg/㎡) Classification Diagnosis <18.5 Underweight Under Infectious disease, malnutrition related disease 18.5~24.9 Normal Standard Least risk at most disease 25.0~29.9 Overweight 30.0~34.9 Obese1 35.
8. Error Messages InBody H20 helps users respond to possible malfunctions that may occur by displaying error messages described below. This is the list of typical error messages and adequate responses. A. Error 1. Cause 1 Bad body posture due to poor contacts at hand or foot electrodes. Actuion1 See “Chapter 6 - C. How to Hold InBody Electrode” and correct the body posture. B. Error 2. Cause 1 Fundamental problem of the device suspected. Actuion1 Please contact Customer Support . (info@inbody.
9. One-Minute Check Before Calling for Service This section lays out the order of steps you have to take in case of malfunction, with the assumption that you have some basic knowledge about how to operate the equipment. If you still have the problem after taking the following steps, contact the local distributor or Biospace. A. I cannot turn on the device even though I press InBody or Weight button. Cause 1 The batteries might have been fully discharged. Action 1 Replace the batteries.
C. Measurement data is not transferred. (When the Bluetooth is connected, the data will be automatically transferred after the measurement.) Cause 1 Poor connection of the Bluetooth dongle. Its connection can be checked by the Bluetooth icon. Action 1 Please check if there is a Bluetooth icon on the LCD and reconnect the Bluetooth dongle. Action 2 Please find the Bluetooth icon from Control Panel and double click, then below screen will appear.
Appendix More about the InBody H20 A. How does BIA work? The Bioelectrical Impedance Analysis (BIA) method is based on the fact that the human body consists of conductors and non-conductors. Generally, 50~70% of the human body consists of water which functions as a conductor, whereas body fat functions as a non-conductor. The classic BIA method measures the impedance of the whole body on the assumption that the human body can be considered a cylinder for the application of this model.
B. Core Technology The following are key features that make the InBody H20 extremely convenient, timely, and accurate. (1) Tetrapolar 8-Point Tactile Electrode It was a complex and inconvenient procedure to attach and detach the electrodes to a specific spot every time. Trained technicians were needed for each measurement. The InBody H20 uses tactile electrodes to avoid the possibility of errors and inaccuracies.
C. Classifications Type of protection against electric shock: Internally Powered Equipment Degree of protection against electric shock: Type B Applied Parts Degree of protection against the ingress of water: Ordinary equipment: IPX0 Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
D.
E. Customer Service Information Corporate agents’ addresses are listed below. Contact us for assistance or more information. Biospace Co., Ltd. 518-10, Dogok 2-Dong, Gangnam-gu, Seoul 135-784, Korea TEL : 82-2-501-3939 FAX : 82-2-578-2716 Homepage://www.inbody.com E-mail : info@inbody.com Biospace Japan Inc. [JAPAN] Second Floor Ayabe Bldg., 2-17-3 Sotokanda, Chiyoda-ku, Tokyo JAPAN TEL : 81-03-5298-7667 FAX : 81-03-5298-7668,7669 Homepage : http://www.biospace.co.jp E-mail : biospace@biospace.co.