4) Unit: Units used for measurement. a. Use the button to move to ‘Unit’ and then press the button. b. Use the button to set the unit appropriate for the test environment. * Unit can be set to be cm or feet and inches. Unit setup c. Save changes by pressing the button. 5) Offset value: Setting an offset value allows height measurements to be adjusted automatically. a. Use the button to move to ‘Offset’ and then press the button. b.
. Optional Settings 1. After turning on the BSM170/170B, press and hold the button and the measurement unit simultaneously for at least 3 seconds to enter the Setup menu. button on the 2. In the Setup menu, you can calibrate the equipment or reset factory default settings. 1) Calibration: The equipment can be calibrated to ensure correct measurement values. a. Raise the measurement unit to eye level as illustrated below.
c. Use the button to move to ‘Yes’ and then press the button. Calibration will begin. d. Lower the measurement unit all the way down of the footplate and press the * The measurement unit must be lowered until it is resting against the footplate. e. The number ‘34.3’ appears on the LCD screen. 16 button.
f. Press the button to move to ‘Exit’ and then press the button. BSM170/170B is now ready for use. g. Erect the Test Jig provided at the logo on the footplate and lower the measurement unit so that it touches the top of the Test Jig as illustrated below. * The Test Jig must be positioned perpendicular to the footplate. * The Test Jig must be positioned correctly.
h. Check that the ̒ 88.9cm ̕ is shown on LCD screen of the measurement unit. * If ̒ 88.9cm ̕ is not shown on the LCD screen(error range: ±0.1cm), please contact InBody. 2) Reset: Reset all settings to factory settings. a. Use the button to move to ‘Reset’ and then press the b. Use the button to move to ‘Yes’ and then press the button. button. All stored settings will be reset. Reset E.
II. Height Measurement A. Precautionary Measurements ION NOTE 참고 DANGER 주의 Warning/Caution 참조 주의 • If the equipment is unstable or unbalanced when you step on the footplate, use the footplate adjustment screws to ensure that the equipment is level before taking measurements. • Take care when lowering the measurement unit to not catch your fingers. • Do not run around run around or play around the footplate. This may result in serious injury.
B. Measurement Instructions 1. Move the measurement unit so that a number is displayed on the screen. 2. Step on the footplate barefooted and lower the measurement unit so that the upper part of the measurement unit touches the examinee’s head. The height measurement is shown on the LCD screen, as illustrated below. * If the measurement value hold feature has been enabled in Setup, pressing the displayed on the LCD, locks the measurement for the set hold time.
C. Measurement Posture The examinee must maintain proper posture to have accurate measurement results. Tuck in the chin, stomach, and chest, and look forward. Allow your backside to gently touch the stand. Put the arms down naturally. Put both the feet together. Do not bend the knees. Keep the legs straight. Touch the end of the footplate with the heels.
III. Transportation and Storage A. Repacking Instructions 1. Prepare the BSM170/170B packaging box. 2. Place the paper pad on the bottom and put 2 small Polyethylene pads on both sides of the packaging box, as illustrated below. Polyethylene pads 3. Position the disassembled components as illustrated below.
IV. Frequently Asked Questions Even if no problems arise from the equipment, users may still have many questions, especially regarding clinical procedures. A few more of the common questions and answers are listed below. If you encounter problems while using the BSM170/170B or have questions concerning height measurement, please refer to the possible solutions below. If your problem cannot be resolved from the following table, please contact InBody. Question • My BSM170/170B does not turn on.
V. Others * The BSM170/170B is registered for electromagnetic compatibility in accordance with the Framework Act on Telecommunications and the Radio Waves Act. * The BSM170/170B is designed, manufactured, and inspected according to h the quality assurance system of InBody. InBody complies with the ISO9001 which are international quality management systems. A. Exterior and Functions The followings are names and functions of each part of the BSM170/170B.
2. Upper section ❶ Stand (upper): Upper part of the equipment Stand. ❷ Measurement tape: Used to adjust the footplate to be level for accurate measurements. ❸ Stand (lower): Lower part of the equipment Stand. ❶ ❷ ❸ 3. Lower section ❶ Stand insertion hole: The Stand is inserted here for assembly with the footplate. ❷ Footplate: The examinee steps on the footplate when having height measured. ❸ Leveling screws: Used to adjust the footplate to be level for accurate measurements.
B. Symbols Safety Symbol Warning/Caution Other Symbols Manufacturer 1177 SN SN European Conformity SN 1177 1177 EC Direct current 1177 1177 Serial number SN SN Authorized representative in the EUROPEAN COMMUNTY REP EC EC REP REP EC EC REP REP Disposal of old Electrical & Electronic Equipment (Application in the European Union and other European countries with separate collection system.) This symbol indicates that this product shall not be treated as household waste.
C. Specifications Intended use Hand-operated mechanical stadiometer Measurement item Height Heights measured 350 ~ 2100mm (1 ft. 1.78 in. ~ 6 ft. 10.67 in.) Height measurement error range ± 1mm (± 0.04 in.) Result display Customized LCD Height measurement Manual measurement Measurement unit Select between ‘cm’ and ‘inch’ Power DC 6V (4 AA-sized 1.5V batteries) External interface Bluetooth *Only for BSM170B Dimension 15.35 (W) × 19.70 (L) × 86.23(H): inch 390 (W) × 500.6 (L) × 2190.