User's Manual
Please note the important information below before reading this manual. NOTE 고 주의 NOTE 고 DANGER Warning 참조 주의 Failure to comply with safety warnings and regulations can cause serious injury or death. DANGER 주의 Caution 참조 주의 Failure to comply with safety cautions and regulations can cause injury or property damage. For any problems with BSM170/170B or any clinical questions, please contact us at the following: InBody Co., Ltd. [HEAD OFFICE] InBody Bldg.
User’s Manual Contents I. Stadiometer Installation A. Product Components 4 B. Operating Environment 5 C. Installation Instructions 5 D. Stadiometer Settings 12 E. Maintenance 18 II. Height Measurement A. Precautionary Measurements 19 B. Measurement Instructions 20 C. Measurement Posture 21 III. Transportation and Storage A. Repacking Instructions 22 B. Transportation and Storage Environment 22 IV. Frequently Asked Questions 23 V. Others A.
I. Stadiometer Installation A. Product Components The BSM170/170B contains of the following components. Make sure all of the following components are present. * Please inspect each component of the BSM170/170B for damage prior to installation.
B. Operating Environment ON Please make sure that the environment is adequate for the BSM170/170B installation. This equipment is designed for indoor use. If installing outdoors, the following requirements must be fulfilled. NOTE Temperatures range 10 ~ 40℃ (50 ~ 104℉) Relative humidity 30 ~ 75% RH Atmospheric pressure 70 ~ 106kPa DANGER 참고 주의 • Warning 참조 주의 Before installing BSM170/170B, check that the ceiling is at least 2.3 m high. The maximum height of this equipment is 2.19m.
3. Hold the “Stand (lower)” in place to prevent it from sliding out of the insertion hole of the footplate, tilt the equipment sideways towards the floor. 4. Screw each of the 4 footplate leveling screws into the holes located on the opposite side of the stand insertion hole of the footplate. 6 5. Tighten the Stand mounting screws by turning them clockwise with the Phillips screwdriver.
UTION 6. Once all four screws as fastened tightly, return the “Stand (lower)” and footplate to an upright position. 7. Insert the black end of “Stand (upper)” into the “Stand (lower)” as illustrated below. NOTE 참고 DANGER Caution 참조 주의 • 주의 Hold the “Stand (upper)” as vertically as possible so that it is not inserted in a slanting position. 8. After inserting the “Stand (upper)” into “Stand (lower)”, fasten the two stands together using the mounting screws and Phillips screwdriver.
9. Insert the measuring tape into the “Stand (upper)” with the ̒↓̓ facing down. Refer to the illustration below. After the measurement tape is completed inserted, remove the sticker attached to the end of the measurement tape to affix the tape to the stand. UTION NOTE 참고 8 DANGER 주의 Caution 참조 주의 • The measurement tape must be inserted with the arrow marked ‘↓’ facing down. • Correct measurements cannot be taken if the measurement tape is inserted reversed.
AUTION 10. Tilt the Stand to a comfortable position and insert the measuring unit onto the stand as illustrated below. NOTE 참고 DANGER 주의 • Caution 참조 주의 The LCD on the measurement unit and the measurement tape must be visible on the same side. 11. The BSM170/170B uses batteries as its power source. * When the measurement standby time expires, the equipment automatically turns off. ➊ BSM170 standby time: 10 seconds.
12. After placing the BSM170/170B on a flat surface, the footplate is not level, use the footplate adjustment screws to adjust until level. * Leveling the equipment is necessary for accurate height measurements. footplate leveling screws 13. When taking measurements with the BSM170/170B, the operator must move the measurement unit up and down manually. By default, the adjustment screw is tight to keep the measurement unit from sliding down too freely.
14. Turns on the BSM170/170B. 15. Move the measurement unit up or down until the ‘Move’ text disappears. The equipment will enter the measurement standby mode and the display will change as shown below.
D. Stadiometer Settings 1. Requried Settings 1. After turning on the BSM170/170B, press and hold the for at least 3 seconds to enter the Setup menu. button and the button simultaneously 2. In the Setup menu, set the measurement value hold time, beep, screen brightness, units, and offset value.
2) Beep: Enable or disable the beep alert. a. Use the button to move to ‘Beep’ and then press the b. Use the button to turn on or off the beep alert. button. * Beep can be set to on or off. Beep setup c. Save changes by pressing the button. 3) Screen brightness: Brightness of the LCD screen. a. Use the button to move to ‘Light’ and then press the button. b. Use the button to set the screen brightness that is appropriate for the test environment.