CNR-CPO-8500T Patient Observer MODEL: CNR-CPO-8500T Patient Observer PC with Intel® RealSense™ Depth Camera D415, Intel® Core™ i5-8500T CPU, HDMI, GbE, Mic-in, Line-out, Four USB 3.1 Gen 1 and RoHS User Manual Page i Rev. 1.
CNR-CPO-8500T Patient Observer Revision Date Version Changes June 19, 2019 1.00 Initial release July 30, 2019 1.01 Modified FCC warning content August 21, 2019 1.02 Added stand mounting instruction (Section 3.5) Added System Troubleshooting (Section 3.8) Removed Red LED (Section 1.7, 1.8) August 27, 2019 Page ii 1.
CNR-CPO-8500T Patient Observer Copyright COPYRIGHT NOTICE The information in this document is subject to change without prior notice in order to improve reliability, design and function and does not represent a commitment on the part of the manufacturer. In no event will the manufacturer be liable for direct, indirect, special, incidental, or consequential damages arising out of the use or inability to use the product or documentation, even if advised of the possibility of such damages.
CNR-CPO-8500T Patient Observer Manual Conventions WARNING Warnings appear where overlooked details may cause damage to the equipment or result in personal injury. Warnings should be taken seriously. CAUTION Cautionary messages should be heeded to help reduce the chance of losing data or damaging the product. NOTE These messages inform the reader of essential but non-critical information.
CNR-CPO-8500T Patient Observer Table of Contents 1 INTRODUCTION........................................................................................................... 1 1.1 OVERVIEW ................................................................................................................ 2 1.2 USABILITY ................................................................................................................. 2 1.2.1 Intended Use .............................................................
CNR-CPO-8500T Patient Observer 3.8 SYSTEM TROUBLESHOOTING ................................................................................... 23 4 BIOS ............................................................................................................................. 24 4.1 INTRODUCTION........................................................................................................ 25 4.1.1 Starting Setup ......................................................................................
CNR-CPO-8500T Patient Observer B.1.3 Classification................................................................................................. 63 B.2 MAINTENANCE AND CLEANING PRECAUTIONS....................................................... 63 B.2.1 Maintenance and Cleaning......................................................................... 63 B.2.2 Cleaning Tools .............................................................................................. 64 C WATCHDOG TIMER...................
CNR-CPO-8500T Patient Observer List of Figures Figure 1-1: CNR-CPO-8500T Patient Observer PC ................................................................... 2 Figure 1-2: Front Panel ............................................................................................................. 4 Figure 1-3: I/O Panel .................................................................................................................. 5 Figure 1-4: Rear Panel ..................................................
CNR-CPO-8500T Patient Observer List of Tables Table 1-1: Technical Specifications .......................................................................................... 9 Table 2-1: Packing List ............................................................................................................ 13 Table 3-1: RJ-45 Connector Pinouts ....................................................................................... 17 Table 3-2: USB 3.1 Gen 1 Port Pinouts ....................................
CNR-CPO-8500T Patient Observer BIOS Menus BIOS Menu 1: Main .................................................................................................................. 27 BIOS Menu 2: Advanced ......................................................................................................... 28 BIOS Menu 3: CPU Configuration ........................................................................................... 29 BIOS Menu 4: ACPI Configuration ............................................
CNR-CPO-8500T Patient Observer Chapter 1 1 Introduction Page 1
CNR-CPO-8500T Patient Observer 1.1 Overview Figure 1-1: CNR-CPO-8500T Patient Observer PC The CNR-CPO-8500T is a patient observer PC with Intel® RealSense™ Depth Camera D415. It equips with the Intel® Core™ i5-8500T processor, and is preinstalled with one 4 GB DDR4 SO-DIMM and 128 GB M.2 SSD. Other peripherals, including one GbE port, four USB 3.1 Gen 1 port, one line-out jack, one mic-in jack and one HDMI port, ensure simplified connectivity to a variety of external peripheral devices.
CNR-CPO-8500T Patient Observer Knowledge: Minimum - read and understand "westernized Arabic" numerals when written in Arial font can distinguish: every parts of body as described in user manual understand the potential risk hazard regarding to misjudgment of diagnostic images Language understanding: English Permissible impairments: mild reading vision impairment or vision corrected to log MAR 0,2 (6/10 or 20/32) one arm/hand system capable of guiding and holding device average degree of aging-related
CNR-CPO-8500T Patient Observer 1.3 Features The CNR-CPO-8500T features the following: Built-in Intel® RealSense™ Depth Camera D415 On-board Intel® Core™ i5-8500T processor 4 GB DDR4 SO-DIMM preinstalled 128 GB M.2 SSD preinstalled One GbE port Four USB 3.1 Gen 1 (5Gb/s) ports One HDMI 1.4 port 802.11 ac with Bluetooth v4.1 RoHS compliant 1.4 Front Panel The followings are found on the front panel.
CNR-CPO-8500T Patient Observer 1.5 I/O Panel The I/O panel is protected by a cover that can be opened by loosening the two captive screws. The followings are found on the I/O panel. 1 x 19 V DC input jack 1 x RJ-45 4 x USB 3.
CNR-CPO-8500T Patient Observer 1.6 Rear Panel The rear panel has the VESA 100 mounting screw holes for mounting (Figure 1-4).
CNR-CPO-8500T Patient Observer 1.7 Side Panels Both side panels contain the Wi-Fi antenna connectors and the LED indicators for indicating camera status (Figure 1-5). The LED indicator behavior is listed below. OFF: Camera is not active. Patient can be assured of privacy. GREEN: Camera is on/active. YELLOW: Camera is in privacy mode.
CNR-CPO-8500T Patient Observer 1.8 System Specifications The CNR-CPO-8500T technical specifications are listed in Table 1-1. Specifications Processor Chipset Memory CNR-CPO-8500T Intel® CoreTM i5-8500T processor (6-core, 2.1 GHz, 9MB cache, TDP=35 W) Intel® H310 2 x 260-pin 2666 MHz single-channel DDR4 SO-DIMM slot Preinstalled with one 4 GB (system max. 64 GB) Storage 128 GB M.2 SSD (up to 500 GB) Ethernet 1 x GbE by Intel® I219 PHY 802.11 ac with Bluetooth v4.
CNR-CPO-8500T Patient Observer Power Input 19 V DC input jack (4-pin DIN) 120 W medical power adapter (Darfon, H1120-B0) Power Adapter Input AC: 100 V ~ 240 V, 2.0-1.0A Output DC: 19 V, 6.31A Mounting VESA 100 Safety: 60601-1 EMC: 60601-1-2 EMI/Safety FCC R&TTE Temperature Humidity Operating: 30ºC (room temperature) Storage/Transportation: -20ºC ~ 60ºC Operating/Storage/Transportation: 10% ~ 90% (non-condensing) Pressure Operating/Storage/Transportation: 700hpa ~ 1060hpa Weight 2.
CNR-CPO-8500T Patient Observer 1.9 Dimensions The dimensions are shown below.
CNR-CPO-8500T Patient Observer Chapter 2 2 Unpacking Page 11
CNR-CPO-8500T Patient Observer 2.1 Anti-static Precautions WARNING! Static electricity can destroy certain electronics. Make sure to follow the ESD precautions to prevent damage to the product, and injury to the user. Make sure to adhere to the following guidelines: Wear an anti-static wristband: Wearing an anti-static wristband can prevent electrostatic discharge. Self-grounding: Touch a grounded conductor every few minutes to discharge any excess static buildup.
CNR-CPO-8500T Patient Observer 2.3 Packing List NOTE: If any of the components listed in the checklist below are missing, do not proceed with the installation. Contact the reseller or vendor the CNR-CPO-8500T was purchased from.
CNR-CPO-8500T Patient Observer Chapter 3 3 Installation Page 14
CNR-CPO-8500T Patient Observer 3.1 Anti-static Precautions WARNING: Failure to take ESD precautions during the maintenance of the CNR-CPO-8500T may result in permanent damage to the CNR-CPO-8500T and severe injury to the user. Electrostatic discharge (ESD) can cause serious damage to electronic components, including the CNR-CPO-8500T. Dry climates are especially susceptible to ESD.
CNR-CPO-8500T Patient Observer Professional installation instruction 1. Installation personal This product is designed for specific application and needs to be installed by a qualified personal who has RF and related rule knowledge. The general user shall not attempt to install or change the setting. 2. Installation location The product shall be installed at a location where the radiating antenna can be kept 20cm from nearby person in normal operation condition to meet regulatory RF exposure requirement. 3.
CNR-CPO-8500T Patient Observer Anti-static Discharge: If a user open the bottom panel of the CNR-CPO-8500T, to configure the jumpers or plug in added peripheral devices, ground themselves first and wear an anti-static wristband. 3.3 I/O Interface Connectors This section provides an overview of the I/O interface connectors of the CNR-CPO-8500T. 3.3.1 RJ-45 Connector The CNR-CPO-8500T has one RJ-45 connector. The RJ-45 connector enables connection to WAN.
CNR-CPO-8500T Patient Observer Pin Description Pin Description Table 3-2: USB 3.1 Gen 1 Port Pinouts Figure 3-2: USB 3.1 Gen 1 Port 3.3.3 VGA Connector The CNR-CPO-8500T has a HDMI 1.4 connector on the I/O panel. The HDMI connector can connect to an HDMI device. The pinouts for the HDMI connector are listed in Table 3-3.
CNR-CPO-8500T Patient Observer 3.4 Installing the Power Adapter CAUTION: The Darfon, H1120-B0 power adapter came with the CNR-CPO-8500T is a forming part of the medical device. There is a space in the bottom part of the CNR-CPO-8500T for installing the provided power adapter. Be sure to install the power adapter before start using the patient observer. Step 1: Loose the two captive screws shown below, and open the cover.
CNR-CPO-8500T Patient Observer Step 3: Orient the power adapter to a proper position. Take the provided power cord and pass it through the hole to connect to the power adapter. Place the bracket onto the power adapter and install the two previously-removed screws to secure it. Figure 3-6: Secure the Power Adapter Step 4: Organize the power adapter cable, and place it in the space between the power adapter and the bottom panel. Connect the power adapter to the power jack on the I/O panel.
CNR-CPO-8500T Patient Observer 3.5 Mounting the System CAUTION: Use suitable mounting apparatus to avoid risk of injury. WARNING! Two or more people should mount the CNR-CPO-8500T. Dropping or bumping the system during installation can cause serious or irreparable damage to the system. The CNR-CPO-8500T is VESA (Video Electronics Standards Association) compliant, please follow the steps below. Step 1: Locate the screw holes on the rear of the CNR-CPO-8500T. This is where the bracket will be attached.
CNR-CPO-8500T Patient Observer Step 2: Align the screw holes with the mounting bracket of the stand. Step 3: To secure the CNR-CPO-8500T to the bracket, insert the retention screws (M4*10) came with the stand into the screw holes and tighten them.S te p 0 : 3.6 Powering On the System WARNING: To avoid risk of electric shock, this equipment must only be connected to supply mains with protective earth. To power on the system, follow the steps below: Step 1: Follow the procedure described in Section 3.
CNR-CPO-8500T Patient Observer 3.7 Resetting the System The reset button enables the user to reboot the system when the system is turned on. The reset button location is shown in Figure 3-10. Press the reset button to reboot the system. Figure 3-10: Reset Button 3.
CNR-CPO-8500T Patient Observer Chapter 4 4 BIOS Page 24
CNR-CPO-8500T Patient Observer 4.1 Introduction The BIOS is programmed onto the BIOS chip. The BIOS setup program allows changes to certain system settings. This chapter outlines the options that can be changed. NOTE: Some of the BIOS options may vary throughout the life cycle of the product and are subject to change without prior notice. 4.1.1 Starting Setup The UEFI BIOS is activated when the computer is turned on. The setup program can be activated in one of two ways.
CNR-CPO-8500T Patient Observer Key Function Page Up key Move to the previous page Page Dn key Move to the next page Main Menu – Quit and not save changes into CMOS Esc key Status Page Setup Menu and Option Page Setup Menu -- Exit current page and return to Main Menu F1 General help, only for Status Page Setup Menu and Option Page Setup Menu F2 Load previous values F3 Load optimized defaults F4 Save changes and Exit BIOS Table 4-1: BIOS Navigation Keys 4.1.
CNR-CPO-8500T Patient Observer 4.2 Main The Main BIOS menu (BIOS Menu 1) appears when the BIOS Setup program is entered. The Main menu gives an overview of the basic system information. Aptio Setup Utility – Copyright (C) 2019 American Megatrends, Inc. Advanced Chipset Security Boot Save & Exit Main BIOS Information BIOS Vendor Core Version Compliancy Project Version Build Date and Time American Megatrends 5.13 UEFI 2.6; PI 1.4 B531AT04.ROM 04/03/2019 15:25:27 iWDD Vendor iWDD Version iEi B531ET15.
CNR-CPO-8500T Patient Observer System Date [xx/xx/xx] Use the System Date option to set the system date. Manually enter the day, month and year. System Time [xx:xx:xx] Use the System Time option to set the system time. Manually enter the hours, minutes and seconds. 4.3 Advanced Use the Advanced menu (BIOS Menu 2) to configure the CPU and peripheral devices through the following sub-menus: WARNING: Setting the wrong values in the sections below may cause the system to malfunction.
CNR-CPO-8500T Patient Observer 4.3.1 CPU Configuration Use the CPU Configuration menu (BIOS Menu 3) to view detailed CPU specifications or enable the Intel Virtualization Technology. Aptio Setup Utility – Copyright (C) 2019 American Megatrends, Inc. Advanced CPU Configuration Type Intel(R) Core(TM) i5-8500T CPU @ 2.
CNR-CPO-8500T Patient Observer All 1 Enable one core in the processor package. 2 Enable two cores in the processor package. 3 Enable three cores in the processor package. 4 Enable four cores in the processor package. 5 Enable five cores in the processor package. DEFAULT Enable all cores in the processor package.
CNR-CPO-8500T Patient Observer 4.3.2 ACPI Settings The ACPI Settings menu (BIOS Menu 4) configures the Advanced Configuration and Power Interface (ACPI) options. Aptio Setup Utility – Copyright (C) 2019 American Megatrends, Inc. Advanced ACPI Settings ACPI Sleep State [S3 (Suspend to RAM)] Select the highest ACPI sleep state the system will enter when the SUSPEND button is pressed. ---------------------: Select Screen ↑ ↓: Select Item Enter: Select +/-: Change Opt.
CNR-CPO-8500T Patient Observer 4.3.3 iWDD H/W Monitor The iWDD H/W Monitor menu (BIOS Menu 5) contains the fan configuration submenu, and displays operating temperature, fan speeds and system voltages. Aptio Setup Utility – Copyright (C) 2019 American Megatrends, Inc. Advanced PC Health Status Smart Fan Mode Select CPU temperature : +44ºC CPU_FAN1 Speed : N/A --------------------: Select Screen CPU_CORE : +0.844 V ↑ ↓: Select Item +5V : +5.131 V Enter: Select +DDR : +1.196 V +/-: Change Opt. +3.
CNR-CPO-8500T Patient Observer Aptio Setup Utility – Copyright (C) 2019 American Megatrends, Inc. Advanced Smart Fan Mode Configuration Smart Fan Mode Select CPU_FAN1 Smart Fan Control Auto mode fan start temperature Auto mode fan off temperature Auto mode fan start PWM Auto mode fan slope PWM [Auto Mode] 40 30 30 2 --------------------: Select Screen ↑ ↓: Select Item Enter: Select +/-: Change Opt. F1: General Help F2: Previous Values F3: Optimized Defaults F4: Save & Exit ESC: Exit Version 2.20.
CNR-CPO-8500T Patient Observer Auto mode fan slope PWM Use the + or – key to change the Auto mode fan slope PWM value. Enter a decimal number between 1 and 8. 4.3.4 RTC Wake Settings The RTC Wake Settings menu (BIOS Menu 7) enables the system to wake at the specified time. Aptio Setup Utility – Copyright (C) 2019 American Megatrends, Inc. Advanced Wake system with Fixed Time [Disabled] Enable or disable System wake on alarm event. When enabled, System will wake on the date::hr::min::sec specified.
CNR-CPO-8500T Patient Observer selected: Wake up every day Wake up date Wake up hour Wake up minute Wake up second After setting the alarm, the computer turns itself on from a suspend state when the alarm goes off. 4.3.5 USB Configuration Use the USB Configuration menu (BIOS Menu 8) to read USB configuration information and configure the USB settings. Aptio Setup Utility – Copyright (C) 2019 American Megatrends, Inc. Advanced USB Configuration Enables Legacy USB support.
CNR-CPO-8500T Patient Observer Legacy USB Support [Enabled] Use the Legacy USB Support BIOS option to enable USB mouse and USB keyboard support. Normally if this option is not enabled, any attached USB mouse or USB keyboard does not become available until a USB compatible operating system is fully booted with all USB drivers loaded. When this option is enabled, any attached USB mouse or USB keyboard can control the system even when there is no USB driver loaded onto the system.
CNR-CPO-8500T Patient Observer 4.3.7 IEI Feature Use the IEI Feature menu (BIOS Menu 10) to configure One Key Recovery function. Aptio Setup Utility – Copyright (C) 2019 American Megatrends, Inc. Advanced iEi Feature Auto Recovery Function [Disabled] Auto Recovery Function Reboot and recover system automatically within 10 min, when OS crashes. Please install Auto Recovery API service before enabling this function. --------------------: Select Screen ↑ ↓: Select Item Enter: Select +/-: Change Opt.
CNR-CPO-8500T Patient Observer 4.4 Chipset Use the Chipset menu (BIOS Menu 11) to access the PCH IO and System Agent (SA) configuration menus. WARNING! Setting the wrong values for the Chipset BIOS selections in the Chipset BIOS menu may cause the system to malfunction. Aptio Setup Utility – Copyright (C) 2019 American Megatrends, Inc.
CNR-CPO-8500T Patient Observer 4.4.1 System Agent (SA) Configuration Use the System Agent (SA) Configuration menu (BIOS Menu 12) to configure the System Agent (SA) parameters. Aptio Setup Utility – Copyright (C) 2019 American Megatrends, Inc. Chipset System Agent (SA) Configuration VT-d VT-d VT-d capability Supported [Disabled] > Memory Configuration > Graphics Configuration > PEG Port Configuration --------------------: Select Screen ↑ ↓: Select Item Enter: Select +/-: Change Opt.
CNR-CPO-8500T Patient Observer 4.4.1.1 Memory Configuration Use the Memory Configuration submenu (BIOS Menu 13) to view memory information. Aptio Setup Utility – Copyright (C) 2019 American Megatrends, Inc. Chipset Memory Configuration DIMM1 Size Number of Ranks Manufacturer DIMM2 Populated & Enabled 4096 MB (DDR4) 1 UnKnown Not Populated/Disabled --------------------: Select Screen ↑ ↓: Select Item Enter: Select +/-: Change Opt.
CNR-CPO-8500T Patient Observer 4.4.1.2 Graphics Configuration Use the Graphics Configuration (BIOS Menu 14) menu to configure the video device connected to the system. Aptio Setup Utility – Copyright (C) 2019 American Megatrends, Inc. Chipset Graphics Configuration Internal Graphics DVMT Pre-Allocated DVMT Total Gfx Mem [Enabled] [32M] [MAX] Keep IGFX enabled based on the setup options. If user want to support Dual Display by Internal Graphics & External Graphics.
CNR-CPO-8500T Patient Observer then only be used as graphics memory, and is no longer available to applications or the operating system. Configuration options are listed below: 32M 64M Default DVMT Total Gfx Mem [MAX] Use the DVMT Total Gfx Mem option to select DVMT5.0 total graphic memory size used by the internal graphic device.
CNR-CPO-8500T Patient Observer 4.4.1.3 PEG Port Configuration Aptio Setup Utility – Copyright (C) 2019 American Megatrends, Inc. Chipset PEG Port Configuration Enable or Disable the Root Port PEG 0:1:0 Enable Root Port Max Link Speed PEG 0:1:1 Enable Root Port Max Link Speed Not Present [Enabled] [Auto] Not Present [Enabled] [Auto] > PEG Port Feature Configuration ---------------------: Select Screen ↑ ↓: Select Item Enter: Select +/-: Change Opt.
CNR-CPO-8500T Patient Observer 4.4.1.3.1 PEG Port Feature Configuration Use the PEG Port Feature Configuration submenu (BIOS Menu 16) to configure the SA PCIe settings. Aptio Setup Utility – Copyright (C) 2019 American Megatrends, Inc. Chipset PEG Port Feature Configuration Detect Non-Compliance Device Detect Non-Compliance PCI Express Device in PEG [Disabled] --------------------: Select Screen ↑ ↓: Select Item Enter Select + - Change Opt.
CNR-CPO-8500T Patient Observer 4.4.2 PCH-IO Configuration Use the PCH-IO Configuration menu (BIOS Menu 17) to configure the PCH parameters. Aptio Setup Utility – Copyright (C) 2019 American Megatrends, Inc. Chipset PCH-IO Configuration Auto Power Button Status Restore AC Power Loss [Disable (ATX)] [Last State] > PCI Express Configuration > HD Audio Configuration Power Saving Function(ERP) USB Power SW1 [Disabled] [+5V DUAL] Select AC power state when power is re-applied after a power failure.
CNR-CPO-8500T Patient Observer Power Saving Function(ERP) [Disabled] Use the Power Saving Function(ERP) BIOS option to enable or disable the power saving function. Disabled Enabled Power saving function is disabled. DEFAULT Power saving function is enabled. It will reduce power consumption when the system is off. USB Power SW1 [+5V DUAL] Use the USB Power SW1 BIOS option to configure whether to provide power to the four external USB 3.
CNR-CPO-8500T Patient Observer 4.4.2.1 PCI Express Configuration Use the PCI Express Configuration submenu (BIOS Menu 18) to configure the PCIe Mini slot. Aptio Setup Utility – Copyright (C) 2019 American Megatrends, Inc. Chipset PCI Express Configuration PCI Express Root Port Settings. > M.2(A) Slot ---------------------: Select Screen ↑ ↓: Select Item Enter: Select +/-: Change Opt. F1: General Help F2: Previous Values F3: Optimized Defaults F4: Save & Exit ESC: Exit Version 2.20.1271.
CNR-CPO-8500T Patient Observer 4.4.2.1.1 M2 (A) Slot Aptio Setup Utility – Copyright (C) 2019 American Megatrends, Inc. Advanced M.2(A) Slot PCIe Speed Detect Non-Compliance Device [Enabled] [Auto] [Disabled] Select PCI Express port speed. --------------------: Select Screen ↑ ↓: Select Item Enter: Select +/-: Change Opt. F1: General Help F2: Previous Values F3: Optimized Defaults F4: Save & Exit ESC: Exit Version 2.20.1271. Copyright (C) 2019 American Megatrends, Inc.
CNR-CPO-8500T Patient Observer Disabled DEFAULT Disables to detect if a non-compliance PCI Express device is connected to the PCI Express slot. Enables to detect if a non-compliance PCI Enabled Express device is connected to the PCI Express slot. 4.4.2.2 HD Audio Configuration Use the HD Audio Configuration menu (BIOS Menu 20) to configure the PCH Azalia settings. Aptio Setup Utility – Copyright (C) 2019 American Megatrends, Inc.
CNR-CPO-8500T Patient Observer Enabled DEFAULT The onboard High Definition Audio controller is enabled. 4.5 Security Use the Security menu (BIOS Menu 21) to set system and user passwords. Aptio Setup Utility – Copyright (C) 2019 American Megatrends, Inc. Advanced Chipset Security Boot Save & Exit Main Password Description Set Administrator Password If ONLY the Administrator’s password is set, then this only limits access to Setup and is only asked for when entering Setup.
CNR-CPO-8500T Patient Observer 4.6 Boot Use the Boot menu (BIOS Menu 22) to configure system boot options. Aptio Setup Utility – Copyright (C) 2019 American Megatrends, Inc.
CNR-CPO-8500T Patient Observer Quiet Boot [Enabled] Use the Quiet Boot BIOS option to select the screen display when the system boots. Disabled Enabled Normal POST messages displayed DEFAULT OEM Logo displayed instead of POST messages UEFI Boot [Disabled] Use the UEFI Boot BIOS option to allow the system to boot from the UEFI devices. Enabled Disabled Enables to boot from the UEFI devices. DEFAULT Disables to boot from the UEFI devices.
CNR-CPO-8500T Patient Observer 4.7 Save & Exit Use the Save & Exit menu (BIOS Menu 23) to load default BIOS values, optimal failsafe values and to save configuration changes. Main Aptio Setup Utility – Copyright (C) 2019 American Megatrends, Inc. Advanced Chipset Security Boot Save & Exit Save Changes and Reset Discard Changes and Reset Reset the system after saving the changes.
CNR-CPO-8500T Patient Observer Appendix A A Regulatory Compliance Page 54
CNR-CPO-8500T Patient Observer DECLARATION OF CONFORMITY This equipment is in conformity with the following EU directives: EMC Directive (2004/108/EC, 2014/30/EU) Low-Voltage Directive (2006/95/EC, 2014/35/EU) RoHS II Directive (2011/65/EU, 2015/863/EU) Ecodesign Directive 2009/125/EC If the user modifies and/or install other devices in the equipment, the CE conformity declaration may no longer apply.
CNR-CPO-8500T Patient Observer Español [Spanish] IEI Integration Corp declara que el equipo cumple con los requisitos esenciales y cualesquiera otras disposiciones aplicables o exigibles de la Directiva 2014/53/EU.
CNR-CPO-8500T Patient Observer Româna [Romanian] IEI Integration Corp declară că acest echipament este in conformitate cu cerinţele esenţiale şi cu celelalte prevederi relevante ale Directivei 2014/53/EU. Slovensko [Slovenian] IEI Integration Corp izjavlja, da je ta opreme v skladu z bistvenimi zahtevami in ostalimi relevantnimi določili direktive 2014/53/EU. Slovensky [Slovak] IEI Integration Corp týmto vyhlasuje, že zariadenia spĺňa základné požiadavky a všetky príslušné ustanovenia Smernice 2014/53/EU.
CNR-CPO-8500T Patient Observer FCC WARNING This equipment complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: This device may not cause harmful interference, and This device must accept any interference received, including interference that may cause undesired operation. This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules.
CNR-CPO-8500T Patient Observer This transmitter must not be co-located or operating in conjunction with any other antenna or transmitter. This device meets all the other requirements specified in Part 15E, Section 15.407 of the FCC Rules. Radiation Exposure Statement: This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment. This equipment should be installed and operated with minimum distance 20cm between the radiator & your body.
CNR-CPO-8500T Patient Observer Appendix B B Safety Precautions Page 60
CNR-CPO-8500T Patient Observer WARNING: The precautions outlined in this chapter should be strictly followed. Failure to follow these precautions may result in permanent damage to the CNR-CPO-8500T. B.1 Safety Precautions Please follow the safety precautions outlined in the sections that follow: B.1.1 General Safety Precautions Please ensure the following safety precautions are adhered to at all times. To prevent the risk of electric shock, make sure power cord is unplugged from wall socket.
CNR-CPO-8500T Patient Observer Do not drop or insert any objects into the ventilation openings of the CNR-CPO-8500T. If considerable amounts of dust, water, or fluids enter the CNR-CPO-8500T, turn off the power supply immediately, unplug the power cord, and contact the CNR-CPO-8500T vendor. DO NOT: Drop the CNR-CPO-8500T against a hard surface. In a site where the ambient temperature exceeds the rated temperature B.1.
CNR-CPO-8500T Patient Observer B.1.3 Classification Power by Class I power adapter No Applied Part No protection against the ingress of water: IPX0 Mode of operation: Continuous Operation The equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide: Not AP or APG Category. B.2 Maintenance and Cleaning Precautions When maintaining or cleaning the CNR-CPO-8500T, please follow the guidelines below. B.2.
CNR-CPO-8500T Patient Observer B.2.2 Cleaning Tools Some components in the CNR-CPO-8500T may only be cleaned using a product specifically designed for the purpose. In such case, the product will be explicitly mentioned in the cleaning tips. Below is a list of items to use when cleaning the CNR-CPO-8500T. Cloth – Although paper towels or tissues can be used, a soft, clean piece of cloth is recommended when cleaning the CNR-CPO-8500T.
CNR-CPO-8500T Patient Observer Appendix C C Watchdog Timer Page 65
CNR-CPO-8500T Patient Observer NOTE: The following discussion applies to DOS. Contact IEI support or visit the IEI website for drivers for other operating systems. The Watchdog Timer is a hardware-based timer that attempts to restart the system when it stops working. The system may stop working because of external EMI or software bugs. The Watchdog Timer ensures that standalone systems like ATMs will automatically attempt to restart in the case of system problems.
CNR-CPO-8500T Patient Observer NOTE: The Watchdog Timer is activated through software. The software application that activates the Watchdog Timer must also deactivate it when closed. If the Watchdog Timer is not deactivated, the system will automatically restart after the Timer has finished its countdown.
CNR-CPO-8500T Patient Observer Appendix D D Hazardous Materials Disclosure Page 68
CNR-CPO-8500T Patient Observer D.1 Hazardous Materials Disclosure Table for IPB Products Certified as RoHS Compliant Under 2011/65/EU Without Mercury The details provided in this appendix are to ensure that the product is compliant with the Peoples Republic of China (China) RoHS standards. The table below acknowledges the presences of small quantities of certain materials in the product, and is applicable to China RoHS only.
CNR-CPO-8500T Patient Observer Part Name Toxic or Hazardous Substances and Elements Lead Mercury Cadmium Hexavalent Polybrominated Polybrominated (Pb) (Hg) (Cd) Chromium Biphenyls Diphenyl (CR(VI)) (PBB) Ethers (PBDE) Housing O O O O O O Display O O O O O O Printed Circuit O O O O O O O O O O O O O O O O O O Fan Assembly O O O O O O Power Supply O O O O O O O O O O O O Board Metal Fasteners Cable Assembly Assemblies Battery O: This toxic
CNR-CPO-8500T Patient Observer 此附件旨在确保本产品符合中国 RoHS 标准。以下表格标示此产品中某有毒物质的含量符 合中国 RoHS 标准规定的限量要求。 本产品上会附有”环境友好使用期限”的标签,此期限是估算这些物质”不会有泄漏或突变”的 年限。本产品可能包含有较短的环境友好使用期限的可替换元件,像是电池或灯管,这些元 件将会单独标示出来。 部件名称 有毒有害物质或元素 铅 汞 镉 六价铬 多溴联苯 多溴二苯 (Pb) (Hg) (Cd) (CR(VI)) (PBB) 醚 (PBDE) 壳体 O O O O O O 显示 O O O O O O 印刷电路板 O O O O O O 金属螺帽 O O O O O O 电缆组装 O O O O O O 风扇组装 O O O O O O 电力供应组装 O O O O O O 电池 O O O O O O O: 表示该有毒有害物质在该部件所有物质材料中的含量均在 SJ/T 11363-2006 (
CNR-CPO-8500T Patient Observer Appendix E E EMC Table Page 72
CNR-CPO-8500T Patient Observer Guidance and manufacturer’s declaration – electromagnetic emissions The model CNR-CPO-8500T is intended for use in the electromagnetic environment specified below. The customer or the user of the model CNR-CPO-8500T should assure that it is used in such an environment. Emissions test Compliance Electromagnetic environment – guidance RF emissions Group 1 The model CNR-CPO-8500T uses RF energy only for its internal function.
CNR-CPO-8500T Patient Observer Recommended separation distances between portable and mobile RF communications equipment and the model CNR-CPO-8500T The model CNR-CPO-8500T is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
CNR-CPO-8500T Patient Observer Guidance and manufacturer’s declaration – electromagnetic immunity The model CNR-CPO-8500T is intended for use in the electromagnetic environment specified below. The customer or the user of the model CNR-CPO-8500T should assure that it is used in such an environment.
CNR-CPO-8500T Patient Observer Power frequency 30 A/m 30 A/m Power frequency (50/60 Hz) magnetic fields should be magnetic field at levels characteristic of a typical location in a typical commercial or hospital environment. IEC 61000-4-8 NOTE: UT is the a.c. mains voltage prior to application of the test level. Guidance and manufacturer’s declaration – electromagnetic immunity The model CNR-CPO-8500T is intended for use in the electromagnetic environment specified below.
CNR-CPO-8500T Patient Observer Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range. b Interference may occur in the vicinity of equipment marked with the following symbol: NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.