User Manual

Installation Instructions for the
HIH-4000 Humidity Sensors
ISSUE 3
50009357
Sensing and Control
WARNING
PERSONAL INJURY
DO NOT USE these products as safety or emergency
stop devices or in any other application where failure of
the product could result in personal injury.
The sensor’s housing does not provide electrical safety
isolation. Mount the sensor in a location where it cannot
be touched, or where a surrounding barrier prevents
human contact.
Failure to comply with these instructions could result in
death or serious injury.
CAUTION
IMPROPER HANDLING
Do not remove the sensor from its original protective
packaging until it is ready to be installed.
Do not touch the sensor surface. Use latex finger cots.
Handle the sensor by its package edges or leads.
Do not allow objects to enter the cavity of the sensor
element.
Failure to comply with these instructions may result in
product damage.
NOTICE
Under condensing conditions where enough liquid water
forms on the sensor to create a parasitic leakage path, the
HIH-4000 Series Humidity Sensor produces an erroneous
reading of 0% humidity. If this erroneous reading is
assumed to be correct by your control function, excess
humidity is likely to be introduced into the system. Once
the liquid water evaporates from the sensor and the
environment returns to a non-condensing state, the device
returns to normal functionality.
Shade the sensor from direct light. Intense direct light can
flood junctions in the CMOS (Complementary Metal Oxide
Semiconductor) device and drive the output signal to the
minimum. This does not harm the sensor or affect
calibration. Proper operation resumes shortly after the
direct light is removed. Ambient scattered light normally
does not affect performance.
At the end of its working life, dispose the sensor in
accordance with Directive 2002/96/EC (WEEE).
TABLE 1. RECOMMENDED PCB MOUNTING SOCKETS
Catalog Listing Mill-Max Socket Number
HIH-4000-001
HIH-4000-003
HIH-4000-005
310-41-132-41-001 or similar
HIH-4000-002
HIH-4000-004
851-43-032-10-001 or similar
SOLDERING/ASSEMBLY
CAUTION
IMPROPER CLEANING
Insert and solder the sensor after the PCB cleaning
process.
Clean sensor with isopropyl alcohol after soldering.
Failure to comply with these instructions may result in
product damage.
CAUTION
IMPROPER SENSOR POSITIONING
Position the sensor so that its face is exposed to the
atmosphere being monitored.
Failure to comply with these instructions may result in
product damage
Hand soldering is acceptable. If wave soldering is required,
use a no-clean flux. Limit the contact of the flux to the leads
only.
Recommended PC board wave soldering temperature is
250
o
C to 260
o
C [482
o
F to 500
o
F].
MOISTURE SEALING THE LEADS
If, in the presence of intermittent moisture or other
contaminants, there is the possibility of galvanic paths between
the leads, moisture seal the leads and exposed pads.

Summary of content (4 pages)