Safety Read this information before start using your device. About this Publication While this information is presented in good faith and believed to be accurate, Honeywell disclaims the implied warranties of merchantability and fitness for a particular purpose and makes no express warranties except as may be stated in its written agreement with and for its customers. In no event is Honeywell liable to anyone for any indirect, special, or consequential damages.
equipment. This device complies with Part 15 of the FCC Rules. Operation to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. Industry Canada Statement This device contains licence-exempt transmitter(s)/receiver(s) that comply with Innovation, Science and Economic Development Canada’s licence-exempt RSS(s).
Cabling At a very minimum, all cables should include a braided shield. Ideal results have been obtained with twisted pair cabling which has a foil shield surrounding each pair plus foil and 90% braid shielding around the bundle. In addition, ensure local electrical code requirements are met. The following cable parameters must be considered: Braid Foil Twisted Pair Stranded Pair Must have a minimum 90% coverage When used with braid, provides 100% coverage. Do not use foil alone. It has a tendency to break.
Introduction Learn what you need to know about the Honeywell Vertex® SF Continuous Monitor System before operating. System Overview The Honeywell Analytics Vertex SF™ System continuously monitors for toxic gases.
System Components The following photos illustrate the Vertex SF System views, ports, connections and controls.
Analyzer — Front view 1. Analyzer status LED 2.
Chemcassette Detection System The Chemcassette® Analyzer module is a self- contained, microprocessor controlled analyzer that occupies one slot in a Vertex tier. Sample lines and the vacuum source are connected to the Chemcassette® via a single 10-tube connector. The system powers up in the same state as when powered down. Data is stored in the module’s memory until the data acquisition computer retrieves it. The Vertex Analyzer modules use the Honeywell Analytics’ Chemcassette® optical detection system.
When monitoring a location, the system detects and measures a specific gas or a family of gases in the sample. The microprocessor in the analyzer module interprets the data and responds appropriately. In the Closed Loop Optics (CLO) detection system, a reference detector monitors and controls the intensity of the LED.
The microprocessor in the Chemcassette analyzer module interprets the stain. It then calculates and stores a precise concentration level in the module’s memory. Gas concentrations are reported in parts-per-million (ppm), parts-per- billion (ppb) or milligrams-per-cubic-meter (mg/ m3).
Install an Analyzer 1. 2. 3. 4. Reach into the slot and position cables out of the way. Place analyzer on slides. Push in until locks engage. Slide Analyzer out until fully extended. Reach behind the analyzer; connect and lock the tubing harness. NOTE When reconnecting the analyzer, connect the Analyzer Communications cable to the analyzer first. 5. Connect the Analyzer Communications, 24V Power Supply, Multifunction Connector. 6.
Specifications Overall system dimension Size 76"X24"X35" Weight - Full loaded condition 1,000lbs (454kg) Weight - Empty rack ≤730lbs (331kg) Analyzer Size 22" x 13" x 6" Weight ≤ 27lbs(12kg) Power Requirements Operating Voltage DC 24V / 2.5A Power Consumption Normal condition 1125 Watt Maximum condition <1400 Watt Outputs Visual 15.