3G Gateway Instruction (H3G-800) H3 System Co.
< Revision Notes > No.
USA-Federal Communications Commission (FCC) This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of FCC Rules.
Welcome Thank you for purchasing 3G Gateway (H3G-800). The 3G gateway receive health data from devices such as blood pressure monitor, glucometer, and weight scale The 3G Gateway transmit them into a healthcare center through CDMA(verizon) mobile network Packaging List 0.
This document describes instructions regarding the H3 3G gateway (H3G-800, H3 System) and the related components. This document includes - How to install and configure the H3 3G gateway - How to use the H3 3G gateway - How to setup a server system for the H3 3G gateways 1. Specification 1. Dimension & weight - Size : 125 x 100 x33 mm - Weight : 160g 2. Communication technology - Bluetooth : class 2 - Range : 10m - Profile : SPP (Serial Port Profile) 3.
D. Final inspection i. Confirm the mode selection switch at “N.” If not, set the switch at “N.” ii. Prepare dry cells for backup batteries. iii. Close the bottom cover. 2) Installation at user’s home A. Place a gateway at a suitable place. B. Connect a power adapter. C. Insert dry cells if required. 8. Gateway configuration 1) Components A. A serial cable provided by H3 System B. Configuration software i. For admin: Serial numbers can be modified. ii. For user: Serial number change is not allowed.
v. Port: DataCollection Server port vi. Devices List: List of paired devices. Types and BT addresses. vii. Delete List: Delete all pairing info. viii. Clear Button: Clear window ix. Send Button: Send a new configuration to the gateway x. 3G Log: 3G AT-command logs 3) Steps A. Turn off the gateway. B. Connect a serial cable (provided by H3 System) to a gateway and a PC. C. Run the configuration software. D. Select the suitable COM port (You can check the COM port at the device manager.) and click “Open.” E.
B. “P1”: Pairing mode as a slave. Used for A&D weighing scale (UC-321PBT) and blood pressure monitor (UA-767PBT). C. “P2”: Used for GMA (BA-110) for Lifescan’s blood glucose meter (OneTouch Ultra, OneTouch Mini/Easy, OneTouch Ultra2), IEM blood pressure monitor (Stabil-OGraph), ForaCare’s thermometer (FORA IR20b), and Beijing Choice’s Pulse Oximeter(MD300C318). 2) Pairing process A. General process i. Turn off the gateway. ii. Set the mode selection switch at suitable mode. iii.
at “N” and turn on the gateway for normal operation. C. Pairing GMA (BA-100). i. Set the mode selection switch at “P2.” ii. Let the GMA be at pairing mode. iii. Turn on the gateway. iv. The gateway shows the number of currently paired devices with LEDs for 5 seconds. v. The “Readings” LED turns GREEN. vi. Wait for about a few minutes. vii. At success, the “Readings” LED blinks GREEN 5 times. viii. At failure, the “Readings” LED blinks RED 5 times. ix. The gateway goes to the step “iv.” x.
ii. iii. iv. v. vi. vii. viii. ix. x. xi. xii. Take a reading with the oximeter. When the Bluetooth icon of the oximeter flashes, remove the finger. Turn on the gateway. The gateway shows the number of currently paired devices with LEDs for 5 seconds. The “Readings” LED turns GREEN. Wait for about a few minutes. At success, the “Readings” LED blinks GREEN 5 times. At failure, the “Readings” LED blinks RED 5 times. The gateway goes to the step “v.” You can pair another device at P2 mode successively.
A. The gateway is supposed to connect to a server when an AC adapter is connected and disconnected. B. It sends “Power up” and “Power down” signals, respectively. C. Backup batteries are required. 5) Heart beat A. Server access is taken every six hours for indication “I’m alive.” B. If there is any reading unsent, the reading is sent to a server together. 11. Data collection software 1) Setup A. Default path : C:\Program Files\H3SYSTEM\DataCollection_GPRS\DataCollection_GPRS.exe B.
Ex) Data XML file name: H3_20101004194422_1111111111.xml Creation logic of xml file name: Datetime + gateway serial no Ex) Log file name: H3_20101004194422_156.log Creation logic of log file name: Datetime 12. XML file PAGE 13spo2="90" battery="10"/>
Item Common battery="3.
spo2="90" battery="10" Gateway information ManualShelf © 2013-2024