Product Manual

Seite/page 1 von/of 3
Verfasser/author: Stan Schnüttgen
Datum/date: 2019-07-09
Instructions for host product manufacturers
1. General
This instruction is related to the following customized Read/Write RFID module.
RFID-Module: TS-HW42
FDD-ID: 2AS5K-TSHW42
IC: 25004-TSHW42A
2.2. List of applicable FCC and ISED rules
FCC 47 CFR Part 15c
FCC 47 CFR 2.1091
FCC 47 CFR 2.1093
FCC KDB 447498 D01
ISED RSS-210, Issue 9
ISED RSS-Gen, Issue 5
ISED RSS-102, Issue 5
2.3. Specific operational use conditions
The module is only approved for use by W.O.M. World Of Medicine GmbH in our own products and
not intended for sale to third parties. The RFID module is intended to be used together with the
dedicated antenna TS-AH20 only. Because the module will not be sold separately, all information
about operational use conditions are described in the IFU of the host system.
2.4. Limited module procedures
The module gets a limited approval and will only be used in dedicated host systems which are
approved separately regarding more applicable FCC/IC rules. The selling of the RFID module
separately is not intended.
2.5. Trace antenna designs
Trace antenna design is not applicable for this module.
2.6. RF exposure considerations
The RFID-Module is approved as device type portable in the exposure category general public with a
maximum permissible exposure at 0cm from a person’s body and as device type mobile in the
exposure category general public with a maximum permissible exposure at 20cm from a person’s
body. The host products must contain the previously mentioned information in their end-product
manuals.

Summary of content (3 pages)