Voluson® P8 / Voluson® P6/ Basic Service Manual • Voluson® P8 systems • Voluson® P6 systems Document Number: 5450839-100 Revision: 5 0459
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Important Precautions WARNING WARNING (EN) (EN) AVERTISSEMENT (FR) WARNUNG (DE) • THIS SERVICE MANUAL IS AVAILABLE IN ENGLISH ONLY. • IF A CUSTOMER’S SERVICE PROVIDER REQUIRES A LANGUAGE OTHER THAN ENGLISH, IT IS THE CUSTOMER’S RESPONSIBILITY TO PROVIDE TRANSLATION SERVICES. • DO NOT ATTEMPT TO SERVICE THE EQUIPMENT UNLESS THIS SERVICE MANUAL HAS BEEN CONSULTED AND IS UNDERSTOOD.
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 AVISO (ES) ATENÇÃO (PT) AVVERTENZA (IT) HOIATUS (ET) ii VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL • ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS. • SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA UN IDIOMA QUE NO SEA EL INGLÉS, LA TRADUCCIÓN ES RESPONSABILIDAD DEL CLIENTE. • NO SE DEBERÁ DAR SERVICIO TÉCNICO AL EQUIPO, SIN HABER CONSULTADO Y COMPRENDIDO ESTE MANUAL DE SERVICIO.
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL VAROITUS (FI) ΠΡΟΕΙ∆ΟΠΟΙΗΣΗ (EL) FIGYELMEZTETÉS (HU) VIÐVÖRUN (IS) • ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU. • EF ÞJÓNUSTUAÐILI VIÐSKIPTAMANNS ÞARFNAST ANNARS TUNGUMÁLS EN ENSKU, ER ÞAÐ Á ÁBYRGÐ VIÐSKIPTAMANNS AÐ ÚTVEGA ÞÝÐINGU. • REYNIÐ EKKI AÐ ÞJÓNUSTA TÆKIÐ NEMA EFTIR AÐ HAFA SKOÐAÐ OG SKILIÐ ÞESSA ÞJÓNUSTUHANDBÓK.
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL VÝSTRAHA (CS) ADVARSEL (DA) WAARSCHUWING (NL) • DENNE SERVICEMANUAL FINDES KUN PÅ ENGELSK. • HVIS EN KUNDES TEKNIKER HAR BRUG FOR ET ANDET SPROG END ENGELSK, ER DET KUNDENS ANSVAR AT SØRGE FOR OVERSÆTTELSE. • FORSØG IKKE AT SERVICERE UDSTYRET MEDMINDRE DENNE SERVICEMANUAL ER BLEVET LÆST OG FORSTÅET.
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL ĮSPĖJIMAS (LT) ADVARSEL (NO) • DENNE SERVICEHÅNDBOKEN FINNES BARE PÅ ENGELSK. • HVIS KUNDENS SERVICELEVERANDØR TRENGER ET ANNET SPRÅK, ER DET KUNDENS ANSVAR Å SØRGE FOR OVERSETTELSE. • IKKE FORSØK Å REPARERE UTSTYRET UTEN AT DENNE SERVICEHÅNDBOKEN ER LEST OG FORSTÅTT.
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL ОСТОРОЖНО! (RU) UPOZORNENIE (SK) VARNING (SV) • DEN HÄR SERVICEHANDBOKEN FINNS BARA TILLGÄNGLIG PÅ ENGELSKA. • OM EN KUNDS SERVICETEKNIKER HAR BEHOV AV ETT ANNAT SPRÅK ÄN ENGELSKA ANSVARAR KUNDEN FÖR ATT TILLHANDAHÅLLA ÖVERSÄTTNINGSTJÄNSTER. • FÖRSÖK INTE UTFÖRA SERVICE PÅ UTRUSTNINGEN OM DU INTE HAR LÄST OCH FÖRSTÅR DEN HÄR SERVICEHANDBOKEN.
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL (JA) (ZH-CN) (KO) - vii
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DAMAGE IN TRANSPORTATION All packages should be closely examined at time of delivery. If damage is apparent write “Damage In Shipment” on ALL copies of the freight or express bill BEFORE delivery is accepted or “signed for” by a GE representative or hospital receiving agent.
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL LEGAL NOTES The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of Voluson® P8 / Voluson® P6. Voluson® P8 / Voluson® P6 may revise this publication from time to time without written notice. TRADEMARKS All products and their name brands are trademarks of their respective holders.
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Revision History Revision Date Reason for change 1 August 2012 Initial release 2 September 2012 Updated Part numbers for Models 3 September 2012 Updated the rating plates and removed tilt indicator 4 September 2012 Upated isolation transformer information 5 January 2013 Updated Software version List of Effected Pages Pages Title Revision 5 Pages Revision Chapter 3 - Setup Instructions pag
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL CHAPTER 1 Introduction Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Purpose of Chapter 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Purpose of Service Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Typical Users of the Basic Service Manual . . . . . . . . . . . . . .
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL CHAPTER 2 Site Preparation Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1 Purpose of Chapter 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1 General Console Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 2 Environmental Requirements . . . .
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL CHAPTER 3 Setup Instructions Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1 The Purpose of Chapter 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1 Set Up Reminders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Average Installation Time . . . . . .
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Scanner Shutdown . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 27 Transducer Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 29 Printer Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 30 Installing Digital Black & White Printer Sony UP-D897 . . . . . . . . . .
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Stand-alone Voluson® P8 / Voluson® P6 . . . . . . . . . . . . . . . . . . . . . . Voluson® P8 / Voluson® P6 + PC (with 4D View Software) within a “Sneaker Net” . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Connection between Voluson® P8 / Voluson® P6 and DICOM Server Wireless Network Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL CHAPTER 4 Functional Checks Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1 Purpose of Chapter 4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1 Required Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1 General Procedure. . . . .
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Heart Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Calculations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Worksheet (Report) Pages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Probe/Connectors Usage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Connecting a probe . . .
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL CHAPTER 5 Components and Functions (Theory) Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1 Purpose of Chapter 5 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1 General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL SonoNT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Description of Hardware Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CW - Continuous Wave Doppler . . . . . . . . . . . . . . . . . . . . . . . . . . . . . DVR Recorder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ECG Module . . . . . . . . . .
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Power Distribution Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 40 Power Up Sequence Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 41 Power Up Sequence Description . . . . . . . . . . . . . . . . . . . . . . 5 - 41 Power Down Sequence Description . . . . . . . . . . . . . . . . . . . . 5 - 42 Mechanical Concept and Overview . . . . . . . . . . . . . .
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Monitor Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 54 Load Application Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 54 Boot Screen Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL CHAPTER 6 Service Adjustments Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1 Purpose of Chapter 6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1 Regulatory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL CHAPTER 7 Diagnostics/Troubleshooting Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1 Purpose of Chapter Voluson® P8 / Voluson® P6 . . . . . . . . . . . . . . . . . . . . 7 - 1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1 Collect Vital System Information. . . . .
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Storing SonoView images to Voluson® P8 / Voluson® P6 . . . . . . . . 7 - 29 Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL CHAPTER 8 Replacement Procedures Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 1 Purpose of Chapter 8 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 1 Returning/Shipping System, Probes and Repair Parts . . . . . . . . . . . . . . . . 8 - 3 System Software - Installation Instructions. . . . . . .
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Manpower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 23 Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 23 8-9-3 Removal Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 23 Installation Procedure . . . . . . . . . . . . . . . . . . . . . . . .
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Replacement of SIDE COVER L ASSY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Manpower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 52 Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 52 Removal Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 52 Installation Procedure . . . . . . . . . . . . . . . . . . . . .
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Replacement of the SOM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Manpower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 77 Removal Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 77 Installation Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 78 Replacement of the DVD . . . . . . . . . . . . . . . . . . . . . . . . . .
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Replacement of FAN with Cable ASSY II. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Manpower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL CHAPTER 9 Renewal Parts Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 1 Purpose of Chapter 9 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 1 List of Abbreviations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 2 Parts List Groups . . . .
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL CHAPTER 10 Care & Maintenance Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 1 Periodic Maintenance Inspections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 1 Purpose of Chapter 10 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 1 Why do Maintenance . . . . . . . . . . . . .
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Meter Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 20 Enclosure Leakage Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 21 Definition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 21 Generic Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Chapter 1 Introduction Section 1-1 Overview 1-1-1 Purpose of Chapter 1 This chapter describes important issues related to safely servicing the Voluson® P8 / Voluson® P6 and/or Voluson® P8 / Voluson® P6 ultrasound system. The service provider must read and understand all the information presented in this manual before installing or servicing a unit.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 1-1-3 1-1-4 Typical Users of the Basic Service Manual • GE Service Personnel (installation, maintenance, etc.).
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 1-1-5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Purpose of Operator Manual(s) The Operator Manual(s) should be fully read and understood before operating the Voluson® P8 / Voluson® P6 and also kept near the unit for quick reference. Section 1-2 Important Conventions 1-2-1 Conventions Used in this Manual Model Designations This manual covers the Voluson® P8 / Voluson® P6 ultrasound units listed on page 1-2.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 1-2-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Standard Hazard Icons Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could cause harm.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 1-2-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Product Icons The following table describes the purpose and location of safety labels and other important information provided on the equipment.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 Table 1-5 LABEL/SYMBOL DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Product Icons Voluson® P8 / Voluson® P6 PURPOSE/MEANING "Mains OFF" Indicates the power off position of the mains power switch. "OFF/Standby" Indicates the power off/standby position of the power switch.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 1-3 Safety Considerations 1-3-1 Introduction The following safety precautions must be observed during all phases of operation, service and repair of this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this manual, violates safety standards of design, manufacture and intended use of the equipment.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 1-3-4 1-3-4-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Electrical Safety Safe Practices To minimize shock hazard, the equipment chassis must be connected to an electrical ground. The system is equipped with a three-conductor AC power cable. This must be plugged into an approved electrical outlet with safety ground.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 1-3-5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Auxiliary Devices Safety (cont’d) Everybody who connects additional equipment to the signal input portion or signal output portion configures a medical system, and is therefore responsible that the system complies with the requirements of the system standard IEC 60601-1-1. Special care has to be taken, if the device is connected to computer network (e.g.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 1-3-6 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Labels Locations The Voluson® P8 / Voluson® P6 comes equipped with product labels and icons. These labels and icons represent pertinent information regarding the operation of the unit.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 1-3-6-1 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Main Label . The Main Label is located on the rear of the Voluson® P8 / Voluson® P6 system. 1 2 3 4 5 6 10 7 8 9 1 2 3 4 5 6 10 7 8 9 Figure 1-2 Main Label (located on left rear of Voluson® P8 / Voluson® P6) 1 Console Part No.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 1-3-6-2 DRAFT (JANUARY 17, 2013) UL Label The UL label is located on the rear of the system.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 1-3-6-3 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Multi Caution Label The Multi caution label is located on the rear of the Voluson P8/Voluson P6 system 1 2 3 4 5 Figure 1-4 Multi-Caution Label 1 Symbol “Follow instructions for use” 2 Symbol “Do not use the following devices near this system” 3 Symbol “Pushing prohibited’ 4 Rx Only mark 5 Caution Chapter 1 - Introduction 1-13
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 1-3-7 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Dangerous Procedure Warnings Warnings, such as the examples below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed. DANGER !! DANGER: DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND ADJUSTING.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 1-3-9 NOTE: DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Returning/Shipping System, Probes and Repair Parts (cont’d) The US Department of Transportation (DOT) has ruled that “items what were saturated and/or dripping with human blood that are now caked with dried blood; or which were used or intended for use in patient care” are “regulated medical waste” for transportation purpose and must be transported as a hazardous mater
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Section 1-5 Customer Assistance 1-5-1 Contact Information If this equipment does not work as indicated in this service manual or in the Basic User Manual, or if you require additional assistance, please contact the local distributor or appropriate support resource, as listed below.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL This page was intentionally left blank.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 1-18 DRAFT (JANUARY 17, 2013) Section 1-5 - Customer Assistance VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Chapter 2 Site Preparation Section 2-1 Overview 2-1-1 Purpose of Chapter 2 This chapter provides the information required to plan and prepare for the installation of a Voluson® P8 / Voluson® P6 ultrasound unit. Included are descriptions of the facility and electrical needs to be met by the purchaser.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Section 2-2 General Console Requirements 2-2-1 Environmental Requirements Table 2-2 Environmental Requirements Operating Temperature Operating Humidity 18 to 30oC 30 to 80% rH non-condensing (64 to 86oF) Heat Dissipation 1200BTU/h without on-board peripherals Storage Temperature -10 to 50 oC (14 to 122oF) Storage Humidity 10~ 90% rH noncondensing CAUTION If the system ha
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 2-2-2-1 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Voluson® P8 / Voluson® P6 Power Requirements Table 2-3 Electrical Specifications for Voluson® P8 / Voluson® P6 Voltage Tolerances Current Frequency 100 - 120 VAC ±10% 9.0 ... 7.0 A 50, 60 Hz (±2%) 220 - 240 VAC ±10% 4.0 ... 3.7 A 50, 60 Hz (±2%) Power Consumption nominal 900 VA including all on-board peripherals.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 2-2-3 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) EMI Limitations Ultrasound systems are susceptible to Electromagnetic Interference (EMI) from radio frequencies, magnetic fields, and transients in the air or wiring. Ultrasound machines also generate EMI. The Voluson® P8 / Voluson® P6 complies with limits as stated on the EMC label. However, there is no guarantee that interference will not occur in a particular installation.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 2-2-4 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Probe Environmental Requirements Operation: Ambient temperature 18° to 30° C Storage: -10° to 50° C NOTE: Temperature in degrees C. Conversion to degrees F = °C (9/5) + 32). NOTICE SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF -10 TO + 50 degrees C.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 2-3 Facility Needs 2-3-1 Purchaser Responsibilities The work and materials needed to prepare the site is the responsibility of the purchaser. Delay, confusion, and waste of manpower can be avoided by completing pre installation work before delivery. Use the Pre-installation checklist (provided in Table 2-6) to verify that all needed steps have been taken.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 2-3-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Mandatory Site Requirements The following are mandatory site requirements. Additional (optional) recommendations, as well as a recommended ultrasound room layout, are provided in Section 2-3-3 "Site Recommendations" .
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 2-3-3-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Recommended Ultrasound Room Layout Figure 2-1 below shows a floor plan illustrating the recommended layout of the Ultrasound Room and depicting the minimal room layout requirements. Dedicated Power Outlets VOLUSON P6/P8 Hospital Network Cabinet for Software and Manuals (optional) 35.5 IN. (92 CM) Figure 2-1 Recommended Floor Plan 4.3m x 5.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 2-3-4 2-3-4-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Networking Setup Requirements Stand-alone Unit (without Network Connection) None 2-3-4-2 Unit Connected to Hospital’s Network Supported networks: 2-3-4-3 • Ethernet • Wireless LAN Purpose of the DICOM Network Function DICOM (Digital Imaging and Communications in Medicine) services provide the operator with clinically useful features for moving images and pati
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) This page was intentionally left blank.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Chapter 3 Setup Instructions Section 3-1 Overview 3-1-1 The Purpose of Chapter 3 This chapter contains information needed to setup the Voluson® P8 / Voluson® P6. Included are procedures to receive, unpack and configure the equipment. A worksheet is provided (see: page 3-73 to page 3-74) to help ensure that all the required information is available, prior to setup the system.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 3-2-2 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Installation Warnings 1.) Since the Voluson® P8 / Voluson® P6 weighs approximately 70 kg (154.3 lbs.) without peripherals, two people are required to unpack it. 2.) There are no operator serviceable components. To prevent shock, do not remove any covers or panels. Should problems or malfunctions occur, unplug the power cord.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-2-2-4 WARNING WARNING !! WARNING: 3-2-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Brake Pedal Operation REMEMBER: If the front wheel brakes are engaged for transportation, release brake pedals (brakes on front wheels under the foot rest) to disengage the lock. Safety Reminders DANGER WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE AC GROUND LINE (I.E.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Section 3-3 Receiving and Unpacking the Equipment CAUTION Please read this section carefully before unpacking the Voluson® P8 / Voluson® P6 and its (optional) peripherals. !! CAUTION: The Voluson® P8 / Voluson® P6 ultrasound systemand accessories are shipped from the factory in a single durable shipping carton which is mounted on a raised wooden platform base.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 Section 3-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Receiving and Unpacking the Equipment (cont’d) The envelope with delivery address, packing list and invoice is located on the front panel of the crate. Check whether delivery is complete (according to packing list) and check visual damage! Figure 3-4 envelope at front panel of the crate Table 3-5 Step 1. Unpacking Procedure Task Loosen the screws.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 Table 3-5 DRAFT (JANUARY 17, 2013) Unpacking Procedure Step 3-6 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Task 2. Remove the top cover and top plate. 3. Remove the three Plastic Joints from the outer sleeve.. 4.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 Table 3-5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Unpacking Procedure Step Task 5. Cut the tape to unfold the package and remove the antistatic vinyl cover 6. Disengage the brakes, grasp the system at the handle, then slowly move unit down the ramp. Rear panel is used for ramp Note: Packing crate and material should be stored for future use.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 3-4 Preparing for Set Up 3-4-1 Verify Customer Order 1.) After unpacking the equipment, it is important to verify that all items ordered by the customer have been received. Compare all items listed on the packing slip (delivery note) with those received.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL this interference. For this reason, it is imperative that all covers and hardware are installed and secured before the unit is put into operation. Ensure that the system is protected from electromagnetic interference (EMI), as follows: • Operate the system at least 15 feet away from equipment that emits strong electromagnetic radiation.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 3-5 Connection of Auxiliary Devices NOTE: Normally auxiliary devices and peripherals come pre-installed with the system. Table 3-7 below outlines hardware installation procedures described in the sub-sections.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-5-1 NOTE: DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Connecting the LCD Monitor The LCD monitor comes pre-installed with the system.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 3-5-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Connecting the Black & White Printer 1.) Power OFF/Shutdown the system as described in: Section 3-6-3 on page 3-27. 2.) Connect the Black & White printer according to correct connection scheme as described in Figure 3-6 on page 3-13 NOTE: The Black & White printer should be connected to the USB2 port of the Voluson® P8 / Voluson® P6´s PC-part. 3.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-5-2-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Connection Scheme: B&W Printer Figure 3-6 B&W Printer connection Chapter 3 - Setup Instructions 3-13
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 3-5-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Connecting the Color Printer 1.) Power OFF/Shutdown the system as described in: Section 3-6-3 on page 3-27. 2.) Connect the Color printer according to correct connection scheme as described in Figure 3-7 on page 3-15. NOTE: The Color printer should be connected to the USB1 port of the Voluson® P8 / Voluson® P6´s PC-part. 3.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-5-3-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Connection Scheme: Color Printer Figure 3-7 Color Printer connection Chapter 3 - Setup Instructions 3-15
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 3-5-4 3-5-4-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Connecting the DeskJet Color Printer Connection via Bluetooth Adapter The DeskJet Color Printer can be connected to an external, non-isolated power source.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-5-4-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Connection via Bluetooth Adapter (cont’d) CAUTION Please observe that the complete Bluetooth Printer Assembly has to be located outside of the patient environment (according to IEC 60601-1 / UL 60601-1). !! CAUTION: CAUTION The printer being used may not be a medical device.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 3-5-5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Connecting the Wireless Network Adapter The Wireless Network Adapter can be connected to any accessible USB port of the Voluson® P8 / Voluson® P6. NOTE: Connection of the WLAN Adapter is always the same. Figure 3-9 Connection Scheme - Wireless (USB) Network Adapter 1.) Turn ON the power of the system and wait till the system has booted. 2.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-5-6 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Connecting the Footswitch The footswitch should be directly connected to any accessible USB-port on the Voluson® P8 / Voluson® P6 (e.g., on rear of the system). NOTE: Connection of the Footswitch is always the same.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 3-5-7 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Connecting the USB Flash Memory Stick NOTICE Before connecting an USB device, please read General Remarks and Hints when using external USB- Devices on page 3-21. !! NOTICE: The USB Flash Memory Stick may be connected to an accessible USB port of the Voluson® P8 / Voluson® P6 system (e.g., on back of control console).
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-5-9 WARNING WARNING !! WARNING: 3-5-9-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL General Remarks and Hints when using external USB-Devices Do not connect or disconnect any external USB-devices to or from the system while scanning a patient! The appearing dialogs could distract you from the scan! External USB-Devices - Connection When an external USB-storage device (such as a USB-memory stick or an external hard disk)
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL To stop the external device, select it and then click the STOP DEVICE button. Figure 3-12 Device can now be safely removed Confirm the ‘Stop Device‘ dialog with OK the “Connect USB and Network Drives” window.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 3-6 Completing the Set Up 3-6-1 Connecting the Unit to a Power Source The connection of the Voluson® P8 / Voluson® P6 ultrasound unit to a power source should be performed by a qualified person who has completed basic Voluson® P8 / Voluson® P6 System User Training.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 3.) Switch ON the Circuit at the rear of the system.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-6-2-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Back End Processor Boot Up NOTICE When AC power is applied to the scanner, the ON/OFF button on the control panel illuminates amber, indicating that the System (including the Back-end Processor) is in Standby mode. !! NOTICE: 4.) Hold down the On/Off button (see: Figure 3-14) on the control panel for ~3 seconds.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 3-6-2-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL During a normal boot, you may observe A.) Power is distributed to Peripherals, Operator Panel (control panel), Monitor, Front-End and Back-End Processor. B.) The Back-End Processor and rest of the scanner starts with the sequence listed in following steps: 1.) First of all, the BIOS version is shown on the monitor. 2.) Afterwards the “Boot Screen” is displayed.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-6-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Power Off / Shutdown NOTICE After turning off a system, wait at least 10 seconds before turning it on again. The system may not be able to boot if power is recycled too quickly. !! NOTICE: 3-6-3-1 Scanner Shutdown 1.) If not already in read mode, freeze the image. 2.) Press the ON/OFF button (see: Figure 3-14) on the control panel. Following dialog appears.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 NOTE: WARNING WARNING !! WARNING: DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL The mains outlet of the system for peripheral auxiliary equipment are commonly switched with the ON/OFF button. So the auxiliary equipment need not to be switched ON/OFF separately. Disconnection of the Main Power Cable is necessary! For Example: When repairing the system. 5.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-6-4 NOTE: DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Transducer Connection When the probe is connected, it is automatically activated. Once connected, the probes can be selected for different applications. Connect a transducer to one of the three rightmost transducer receptacle as follows: 1.) Inspect the probe and probe socket to verify that it is free of debris. 2.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 3-7 Printer Installation NOTE: For Connection schemes refer to Section 3-5 "Connection of Auxiliary Devices" on page 3-10.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-7-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Installing the Bluetooth Deskjet Printer NOTICE Before installing a Bluetooth color printer: • the Bluetooth Adapter and Software (see: Section 3-7-3-1 on page 3-31) MUST be installed first, • the Bluetooth Software has to be configured (see: Section 3-7-3-2 on page 3-33) next and • the Bluetooth Printer Adapter has to be connected (see: Section 3-7-3-3 on page 3-36), then
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 8.) Confirm the following dialog window with NEXT. The "License Agreement" window appears: Figure 3-21 Welcome to Install... and License Agreement 9.) Select "I accept the terms in the license agreement" and then click on NEXT. 10.)Confirm the following dialog with NEXT. The "Ready to Install the Program" window appears. Figure 3-22 Ready to Install the Program 11.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL away. If the “Bluetooth device not found” dialog box obstructs the view, simply move it away. Figure 3-23 “Bluetooth device not found” - DO NOT click any button 13.)Connect the Bluetooth Adapter (Belkin or DBT-122) to USB port E (see: Figure 3-8 on page 3-16) on the rear side. 14.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 4.) Press CTRL + ESC to open the Windows Start menu and select "My Bluetooth Places". Figure 3-25 select “My Bluetooth Places” and click on Next 5.) In the "Initial Bluetooth Configuration Wizard" dialog click on NEXT. 6.) In the following window, enter an unique computer name (e.g., D10013 = Serial number of the system) and then click NEXT.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 NOTE: DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Display of this window might be different. It depends on installed Bluetooth Adapter. deselect ALL services in this window move the scroll bar downwards to view all services Figure 3-27 Bluetooth Service Selection 9.) In the "Initial Bluetooth Configuration Wizard" window click on SKIP. Figure 3-28 Click on Skip and then Finish configuration 10.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 3-7-3-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Prepare the Printer and connect the Bluetooth Printer Adapter NOTICE Preparations as well as Printer Adapter connection depend on the used Bluetooth Line Printer. • in case of HP Officejet H470, proceed with Section 3-7-3-3-1 on page 3-36 !! NOTICE: 3-7-3-3-1 HP Officejet H470 Printer 1.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 2.) Select "View or modify configuration". Add a Bluetooth Device View or modify configuration Figure 3-30 My Bluetooth Places “View or modify configuration” 3.) In the LOCAL SERVICE page select "Printer" and click on PROPERTITES. 4.) Remove check mark at "Secure Connection" and then click OK. Figure 3-31 "Local Services" page / remove check mark at "Secure Connection" 5.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 6.) Remove check mark at "Secure Connection" and then click OK. Figure 3-32 "Client Applications" page / remove checkmark at "Secure Connection" 7.) Click OK again. 8.) In the displayed window, double-click "Add a Bluetooth Device" (see Figure 3-30 on page 3-37). 9.) In the "Bluetooth Setup" window (see: Figure 3-33 below) click on NEXT. 10.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL b.) If it looks like the right image, enter the passkey (0000) and then click on NEXT. a.) b.) Figure 3-34 Bluetooth Security Setup 12.)Finally, select "Hardcopy Cable Replacement" and then click FINISH.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 13.)If the "Add Printer Wizard" window appears, select the HAVE DISK button. [Browse] for Printer Driver on F:\ (where F: stands for the system´s DVD drive) Figure 3-36 Have Disk... / Browse for Printer Driver 14.)Use the BROWSE button to search the printer driver on F:\ (= DVD drive) and then click OPEN. NOTE: HP Officejet H470: browse for F:\HP470\Driver\autorun_bt.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-7-3-5-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL HP Officejet H470 - Printer Properties After driver installation, the “HP Officejet H470 series BT Properties” page appears. 1.) In the ADVANCED page click the PRINTING DEFAULTS... button. 2.) Select the appropriate paper size and confirm with APPLY and then OK. Figure 3-37 ADVANCED page / select paper size 3.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 3-7-4 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Printer Installation manually 1.) Press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen. 3.) On the left side of the screen select CONNECTIVITY and then click on the PERIPHERALS tab.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-7-4 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Printer Installation manually (cont’d) 7.) Select the corresponding Printer Port (e.g., Figure 3-41 = USB001) and click NEXT. Figure 3-41 Select Printer Port 8.) In the following window select the HAVE DISK button. Figure 3-42 Have Disk... 9.) Use the BROWSE button to search the Printer Driver path (R:\system\periph\xxxx).
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 3-7-4 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Printer Installation manually (cont’d) Figure 3-43 Select Printer Driver path (R:\system\periph\....) 10.)Click OPEN, select the “xxx.inf” file and click OPEN again. 11.)Verify the selected Printer Driver path and confirm with OK. Figure 3-44 verify selected Printer Driver path 12.)Select the manufacturer and model of your printer and confirm with the NEXT button.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-7-4 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Printer Installation manually (cont’d) 13.)Assign a name, decide if printer should be used as default printer and confirm with NEXT. Figure 3-45 Assign name NOTE: If the “Printer Sharing” window appears, select ‘Do not share this printer‘ and confirm with NEXT. 14.)The “Complete the Add Printer Wizard” window appears on the screen.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 3-7-5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Adjustment of Printer Settings 1.) After system restart, press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen. 3.) On the left side of the screen select CONNECTIVITY and then click on the PERIPHERALS tab. 4.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-7-5-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL UP-D897 - Printer Settings 1.) Call up the ‘Printer Preferences’; operation see: Section 3-7-5 "Adjustment of Printer Settings" . 2.) Select the LAYOUT page (see: Figure 3-48 below) and select: - Paper: 960x1280 - Orientation: Portrait - Interpolation Method: Bilinear Figure 3-48 Layout + Density Adjust page 3.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 3-7-5-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL P95D - Printer Settings 1.) Call up the ‘Printing Preferences’; operation see: Section 3-7-5 "Adjustment of Printer Settings" . 2.) Select the STANDARD page (see: Figure 3-49 below) and select: - Paper Size: 1280x960 - Orientation: Portrait 3.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-7-5-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL P95D - Printer Settings (cont’d) 4.) Select the IMAGE ADJUST page (see: Figure 3-50 below) and select: - Simple : brightness: 3, Contrast : -7 Figure 3-50 Image Adjust page 5.) For saving the adjusted printer settings click APPLY and then OK. Finally close the ‘Printers’ -window with the close button and exit System Setup with SAVE&EXIT. 6.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 3-7-5-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL UP-D25MD - Printer Settings 1.) Call up the ‘Printer Preferences’; operation see: Section 3-7-5 "Adjustment of Printer Settings" . NOTICE Settings for Paper Size MUST match with the used Paper (large/small) and also the right color ink !! NOTICE: cartridge has to be used. Otherwise you will get an error message at printing. 2.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-7-5-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL UP-D25MD - Printer Settings (cont’d) 3.) Select the GRAPHICS page. From the “Color Adjust” pop-up menu select: a.) Color Balance: Cyan = 0; Magenta = 0; Yellow = 0 b.) Gray Balance: X =50, Y=50 Figure 3-52 Graphics page (Color Balance + Gray Balance) c.) Color Correction: set Printer Hardware Color Correction d.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 3-7-5-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL UP-D25MD - Printer Settings (cont’d) e.) Adjust Value : Select nomal. Figure 3-54 Graphics page (Adjustment by color) 4.) For saving the adjusted printer settings click APPLY and then OK. Finally close the ‘Printers’-window with the close button and exit System Setup with SAVE&EXIT. 5.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-7-6 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Remote Control Selection To assign an auxiliary device (e.g., printer) to the remote keys P1, P2, and/or P3/DVR 1.) Press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen. 3.) On the left side of the screen select CONNECTIVITY and then click the BUTTON CONFIGURATION tab.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 3-8 System Configuration 3-8-1 System Setup Modifications of system parameters and settings are supported by 4 major groups. Each major group contains diverse dialog pages and sub windows. • General: Date, Time, Clinic Name, (EUM) Language, Screen saver, etc. • User Settings: to save User programs, 3D/4D programs, Auto Text, Doppler 2D Refresh, etc.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL right trackball key (left mouse button): sets, fixates markers and activates pages/buttons etc.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 3-8-1-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL How to enter Date and Time 1.) Invoke System Setup as described in Section 3-8-1-1 on page 3-54. 2.) On the left side of the screen select GENERAL and then click on the GENERAL tab.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-8-1-5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL How to activate Screen Lock 1.) Invoke System Setup as described in Section 3-8-1-1 on page 3-54. 2.) On the left side of the screen select GENERAL and then click on the GENERAL tab. 3.) Check mark “Screen Lock”. a.) If no password previously entered, the following dialog appears. Figure 3-57 Screenlock Password b.) Enter “New Password”.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL PERIPHERALS tab. 3.) Select desired function of the Footswitch Left, Middle and Right. Figure 3-58 Click SAVE&EXIT to save Settings and exit System Setup. Footswitch 3-8-1-7 How to change the Keyboard Layout see: Chapter 6 - Modification of Keyboard Layout, on page 6-3. 3-8-1-8 How to configure Service Platform see: Chapter 7 - To configure Service Platform, on page 7-15.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-8-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Measure Setup Modifications of system parameters are supported by diverse dialog pages and windows on the measure setup desktop: NOTICE Parameters and possible adjustments mostly depend on the selected Application! !! NOTICE: • Measure & Calc - shows all settings, which are used for generic measurements as well as calculations in different applications • Application Par
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 3-8-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL On-Board Optional Peripherals Mains outlet: AC outlet for on-board peripherals Output voltage : 100~250V.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-8-4 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) External I/O Connectors 3 1 4 2 Figure 3-59 External I/O Connectors - on Rear of System Table 3-9 Connector Name External I/O Connector - Description Item Table Number Description DVI Out 1 Table 3-11 Connector for external Monitor NETWORK 2 Table 3-12 DICOM input/output, twisted pair RJ-45 10/100 megabit/s USB 3 Table 3-13 USB-2.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 Table 3-12 3-62 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Network Connector, RJ45 Modular 8 Pin Pin No Output Signal Description 1 ETHER TD Ethernet RD+ 2 ETHER TD Ethernet RD- 3 ETHER RD Ethernet TD+ 6 ETHER RD Ethernet TD- Others NC No connection Section 3-8 - System Configuration
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-8-4-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL External I/O Pin Outs (cont’d) Table 3-13 USB Connectors Pin No Output Signal Description 1 VCC USB Power Supply 2 - Data USB Data (-) 3 + Data USB Data (+) 4 GND USB Power Ground Section 3-9 Available Probes See Chapter 9 - Probes, on page 9-15, for part numbers to be used when ordering new or replacement service probes.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 3-11 Connectivity Setup The Voluson® P8 / Voluson® P6 can be connected to various connectivity devices. The following sections describe how to connect the system to a remote archive/work station or a DICOM service, using a TCP/IP connection.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-11-1-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Dataflow Examples The local database is used for patient archiving. Images are stored to internal hard drive. The local database is used for patient archiving. Afterwards images are stored to a DVD/CD or external USB device, etc. PC with 4D View A remote database is used for patient archiving. Images are also stored to a remote archive.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 3-11-1-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Stand-alone Voluson® P8 / Voluson® P6 If digital images or 3D/4D data sets are stored, they should be saved in the Archive (Image Management System software). For Image Management functionality refer to the Basic User Manual of the Voluson® P8 / Voluson® P6.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-11-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Wireless Network Configuration NOTICE To configure the Voluson® P8 / Voluson® P6 ultrasound unit to work with WLAN, the hospital’s !! NOTICE: network administrator has to provide the required information. 1.) Connect the Wireless Network adapter as described in Section 3-5-5 on page 3-18. 2.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 3-11-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Wireless Network Configuration (cont’d) 5.) Select the country where you will use the wireless USB adapter from the “pull-down menu” and then click AGREE.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-11-2 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Wireless Network Configuration (cont’d) After a few seconds waiting, following information appear in the status bar at the bottom of the page.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Section 3-12 Network IP Address Configuration NOTE: Following Information must be provided by customer or hospital engineer before you can start: A Station name, AE Title, IP address and Port Number for the Voluson® P8 / Voluson® P6. The IP addresses for the default gateway and other routers at the site for ROUTING INFORMATION. Only if necessary (e.g. for Internet access). 1.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 3-12-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Map Network Drive 1.) Press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen. 3.) On the left side of the screen select CONNECTIVITY and then click the DRIVES tab. [Map Network Drive] button Figure 3-68 System Setup - Connectivity - DRIVES page 4.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 3-12-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Map Network Drive (cont’d) 7.) Select the CONNECT button to establish the connection to the remote machine. If successful, the DISCONNECT button becomes active. NOTICE If there is an error during the connection, a warning message appears inside the dialog. In this case, please verify the data in the dialog.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 3-13 Connectivity Setup Worksheet Site System Information Comments: Floor: Site: Dept: Room: Voluson® P8 / Voluson® Type: REV: CONTACT INFORMATION Name Title E-Mail Address Phone Remote Archive Setup TCP/IP Settings Name - AE Title: System IP Settings IP Address: Name - AE Title: Subnet Mask: IP Address: Default Gateway: Subnet Mask: Server Name: Defau
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 Section 3-13 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Connectivity Setup Worksheet (cont’d) Voluson® P8 / Voluson® P6 Local Port Host Name AE Title ROUTING INFORMATION IP Address . . . Net Mask . . . GATEWAY IP Addresses Destination IP Addresses Default . . . ROUTER1 . . . . . . ROUTER2 . . . . . . ROUTER3 . . . . . .
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 3-14 Paperwork NOTE: 3-14-1 NOTE: During and after installation, the documentation (i.e. User Manual, Installation Manual,...) for the peripheral units must be kept as part of the original system documentation. This will ensure that all relevant safety and user information is available during the operation and service of the complete system.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) This page was intentionally left blank.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Chapter 4 Functional Checks Section 4-1 Overview 4-1-1 Purpose of Chapter 4 This chapter provides procedures for quickly checking major functions of the Voluson® P8 / Voluson® P6 scanner diagnostics by using the built-in service software, and power supply adjustments.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 4-3 General Procedure CAUTION !! CAUTION: NOTICE SYSTEM REQUIRES ALL COVERS Operate this unit only when all board covers and frame panels are securely in place. The covers are required for safe operation, good system performance and cooling purposes.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 4-3-1-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Scanner Power On (cont’d) 5.) Hold down the On/Off button (see: Figure 4-2) on the control panel for ~3 seconds. NOTE: The power for on-board peripheral auxiliary equipments are commonly switched with the ON/OFF button. The power switch of any attached printer(s) needs to be in ON position before starting the system.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 4-3-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Power Off / Shutdown NOTICE After turning off a system, wait at least 10 seconds before turning it on again. The system may not be able to boot if power is recycled too quickly. !! NOTICE: 4-3-2-1 Scanner Shutdown 1.) If not already in read mode, freeze the image. 2.) Press the ON/OFF button (see: Figure 4-2) on the control panel. Following dialog appears.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 6.) Press on the brakes to block the front caster wheels. 7.) Disconnect probes. CAUTION DO NOT disconnect a probe while running (Live Scan “Write” mode)! A software error may occur. In this case switch the unit OFF (perform a reset).
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 4-3-3-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Monitor Display Figure 4-6 Monitor Display Tour Table 4-2 4-6 Monitor Display Features 1. Logo 15. Measurement results 2. Patient Name (Last-, First-, Middle Name) 16. Status bar area 3. Patient ID-number ; GA (Gestational Age) 17. P buttons 4. Probe / Application 18. Run/Freeze/Cine 5. Depth / Frame rate 19. Trackball 6. Mechanical Index 20. - 7.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 4-4 Functional Checks For a basic functional check of the system’s different modes, following pages will familiarize you with image optimization for: • 2D Mode (B Mode), see: Section 4-4-1 on page 4-8 • Additional (optional) Operating Modes, see: Section on page 4-11 - XTD-View • M Mode, see: Section 4-4-2 on page 4-12 • Spectral Doppler Modes, see: Section 4-4-3 on
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 4-4-1 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) 2D Mode Checks Figure 4-7 2D Main and 2D Sub Menu Table 4-3 2D Mode Functions Step 1 Task 2D Mode Gain Rotate the 2D MODE key to adjust the sensitivity (brightness) of the entire image. Transmit Power & Transmit Power: Optimizes image quality and allows user to reduce beam intensity.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 Table 4-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 2D Mode Functions Step Task Expected Results Focus Depth To select the depth position of the actual focus zone(s). Arrows at the left edge of the 2D-Image mark the active focal zone(s) by their depth position. 3 Depth Adjusts the penetration depth range of the ultrasound image for the region of interest.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 Table 4-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 2D Mode Functions Step Task CINE Mode Expected Results While scanning a certain number of frames (2D images of the last examination sequence), this will be stored in the cine memory automatically.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 Table 4-3 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) 2D Mode Functions Step Task Expected Results TINT MAP A Tint map determines the Chroma value (color tone and saturation) of an echo in relationship to its amplitude. 23 For further details refer to the Voluson® P8 / Voluson® P6 Basic User Manual, Chapter 6, 2D Mode.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 4-4-2 NOTE: VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) M Mode Checks The AMM menu is only displayed if the Anatomical M-Mode option is installed. Figure 4-9 M Main and M Sub Menu Table 4-4 M Mode Functions Step 4-12 Task Expected Results 1 Cursor Position Adjust the M Cursor position with the TRACKBALL in the 2D Single image.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 Table 4-4 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) M Mode Functions Step Task Expected Results 12 ENHANCE Due to this function a finer, sharper impression of the image is produced. 13 DYNAMIC CONTROL. Dynamic Range enhances a part of the grayscale to make it easier to display pathology. 14 DISPLAY FORMAT For selection of different rations of display format.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 4-4-3 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Spectral Doppler Mode Checks NOTE: Different menus are displayed depending on which Spectral Doppler Mode (PW or CW) is selected. NOTE: The Continuous Wave Doppler Mode is an Option. The CW key is only illuminated if the option is installed and the selected probe is capable for the Continuous Wave Doppler Mode.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 Step DRAFT (JANUARY 17, 2013) Task VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Expected Results 9 Display Format Doppler spectrum is displayed below the B mode image or on its right side. 10 INVERT To invert the Doppler spectrum display in relation to the direction of the flow. 11 ANGLE The angle cursor can be turned in both directions without stop.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 4-4-4 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Color Doppler Mode Checks NOTE: Different menus are displayed depending on which Color Doppler Mode (CFM, PD, HD-FLow) is selected. NOTE: Pressing the CF key on the control panel activates the color mode. Selecting C on the alphanumeric keyboard, activates the CFM.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 Table 4-6 Step DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Color Doppler Mode (CFM, PD, HD-Flow, TD) Functions Task Expected Results 9 THRESHOLD After FREEZE you can adjust the Color Threshold. It eliminates small color noise or motion artifact signals in the color image. (small number cuts off less signals than a higher setting) 10 DISPL. MODE To select the CFM- Display Mode (V; V-T; V-Pow; Pow-T; or T).
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 4-4-5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Volume Mode Checks NOTICE 4D Real Time 4D, VCI, VCI Omni View are Options. If options are not part of the system configuration, checks can be omitted. !! NOTICE: NOTE: Different menus are displayed depending on which Menu and which Volume Mode is selected (4D Real Time or 3D Static) is selected.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 4-4-5-1 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Pre-Volume Mode Functions 1. Select the desired Volume Mode: The Volume Mode function is switched on, the “3D Pre” respectively “4D Pre” menu appears on the screen (write mode) 2. select the desired Visualization Mode 3. Start the Volume Acquisition with the right trackball key. Figure 4-13 Pre-Volume Mode menus (e.g.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 Table 4-7 Step DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Pre-Volume Mode Functions Task Expected Results • 4D REAL TIME Real Time 4D - continuous volume acquisition and parallel calculation of 3D rendered images • VCI A Volume Contrast Imaging - improves the contrast resolution and the signal / noise ratio and therefore facilitates finding of diffuse lesions in organs • SINLGE VIEW Single View (any plane) - improves
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 4-4-5-2 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Functions after the Acquisition Figure 4-14 Sectional Planes and Image Rendering Table 4-8 Functions after the 3D Acquisition Step Task Expected Results PW MODE rotary control: Rotation about X-axis and Y-axis of the reference image. 1 M MODE rotary control: Rotation about the Z-axis of the reference image. CF MODE rotary control: CF-Gain and CF-Mode.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 Table 4-8 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Functions after the 3D Acquisition Step 4-22 DRAFT (JANUARY 17, 2013) Task Expected Results 6 SRI / SRI 3D Speckle Reduction Imaging (SRI) to reduce speckle which interferes with the sectional planes or rendered 3D image. 7 ZOOM The 3D image as well as the sectional planes can be varied by their aspect ratio. 8 MIX To adjust the mix ratio between two calculated modes. 9 TH.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 4-4-5-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 3D/4D Sub Menu Figure 4-15 3D/4D Sub Menu Table 4-9 Step Sub Menus Task Expected Results 1 PROBE ORIENTATION Adjust and activate the display of direction markers at border of the image. 2 GRAY/TINT MAPS 3 RENDER VIEW DIRECTION Depending on individual requirements a “harder” or “softer” image can be obtained. To select the desired Render View Direction.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 4-4-6 4-4-6-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Using Cine Activating Cine Press FREEZE, then roll the TRACKBALL to display the images of the stored sequence one by one. 4-4-6-2 Cine-Split Function (Multiple Format) After FREEZE of a sequence in 2D Mode two or four different images of the sequence can be displayed simultaneously in Dual respectively Quad Display Mode.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 4-4-6-8 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Activating Cine Calc 1.) After 3D Volume or Real Time 4D acquisition select the CINE CALC key on the. 2.) Choose desired Cine Calc display mode. 3.) Select Step Size and Reference image. 4.) Select the START IMAGE and the END IMAGE of the sequence with the rotary controls. 5.) Select the CALCULATE CINE SEQUENCE key to start the calculation. 6.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 4-4-7 NOTE: DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Generic Measurements Different menus are displayed depending on: • the currently selected Application and Display Mode, • the selected “Study”, • and the settings in the Measure Setup - MEASURE & CALC page. For further details refer to the Voluson® P8 / Voluson® P6 Basic User Manual, Chapter 13.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 NOTE: 4-4-7-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL To exit Generic measurements, select the EXIT key on the Voluson® P8 / Voluson® P6, press the MEASUREkey or the EXIT key on the control panel. Circumference/Area Measurements 1.) Press the MEASURE key and then select GENERIC AREA on the Voluson® P8 / Voluson® P6. 2.) Select the corresponding item on the Voluson® P8 / Voluson® P6. An active cursor displays. 3.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 NOTE: 4-4-7-3-2 NOTE: 4-4-7-3-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL To erase the results, select the CLEAR key on the control panel. Measurements in Spectral Doppler Mode The Spectral Doppler image is displayed based on time (X-axis) and velocity (Y-axis). Auto Trace 1.) Press the MEASURE key. 2.) Select AUTO TRACE on the Voluson® P8 / Voluson® P6.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 4-4-8 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Calculations The Voluson® P8 / Voluson® P6 system supports calculation packages and application-oriented patient Worksheets (Reports) for following applications: • Abdomen Calculations • Small Parts Calculations • Obstetric Calculations • Cardiology Calculations • Urology Calculations • Vascular Calculations • Gynecology Calculations • Pediatric Calculations
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 4-4-8-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Worksheet (Report) Pages Press the REPORT key on the control panel to view the “application dependent” patient worksheet pages that contain the results of calculations and measurements. Any stored patient worksheet can be edited, printed, transferred, saved in the Archive or sent to DICOM server.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 4-4-9 4-4-9-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Probe/Connectors Usage Connecting a probe 1.) Place the probe's carrying case on a stable surface and open the case. 2.) Carefully remove the probe and unwrap the probe cable. 3.) DO NOT allow the probe head to hang free. Impact to the probe head could result in irreparable damage. 4.) Align the connector with the probe port and carefully push into place.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 4-4-10 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Patient Archive (Image Management) The Voluson® P8 / Voluson® P6 provides an Patient/Image Management System that allows fast and extremely easy patient, exam and image management.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 4-4-11 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Erasing DVD/CD 1.) Press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen. 3.) On the left side of the screen select CONNECTIVITY and then click the DRIVES tab. Figure 4-17 System Setup - Connectivity - DRIVES page 4.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 4-5 Backup and Restore Database, Preset Configurations and Images CAUTION It is recommended to Backup the Full System Configuration (Section 4-5-3 on page 4-39) and the Image Archive (Section 4-5-6 on page 4-44) once a week. !! CAUTION: 1.) Press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen. 3.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 4-5-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Save Image Settings Only The Image Settings contains: • Application Settings • User Programs • Auto Text • 3D/4D Programs • Scan Assistant 1.) Insert a DVD/CD+R/RW into the drive or connect an external USB device. 2.) Press the F2 key on the control panel. 3.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen. 4.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 4-5-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Load Image Settings Only CAUTION The loading procedure overwrites the existing image settings on the local hard drive. Make sure to insert the correct System DVD. Additionally you can load the image settings from D:\usersettings\FactoryDefault\Tetris !! CAUTION: 4-5-2-1 Preparations 1.) Insert a DVD/CD+R/RW into the drive or connect an external USB device. 2.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 4-5-2-2 NOTE: DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Load “Complete Backup” Following procedure should be used, if the user prefers factory default settings, which are adapted for the installed Application Software version. 1.) Perform Section 4-5-2-1 "Preparations" on page 4-36. 2.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 4-5-2-3 NOTE: DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Load only parts of the “Complete Backup” Following procedure should be used, if the user has saved own User Programs, 3D/4D Programs or AutoText entries, but new presets have to be added to match the installed Application Software version (e.g., User Programs and 3D/4D Programs for new probe, etc.). 1.) Perform Preparations on page 4-36. 2.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 4-5-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Save Full System Configuration (Full Backup) A backup of the Full System Configuration always contains the following data NOTE: • User Settings (databases and files containing User Programs, 2D/3D/4D Programs, Auto Text entries, gray curves and complete System settings such as language, time/date format, etc.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 4-5-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Save Full System Configuration (Full Backup) (cont’d) When the “Full Backup” is saved on a network drive (to map a network drive see: Section 3-12-1 "Map Network Drive" on page 3-71), it may be desirable to move the data (e.g., for backup or maintenance). The backups reside in sub folders of the main “fullbackup” -folder found at the root of the drive.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 4-5-4 WARNING WARNING !! WARNING: DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Load Full System Configuration (Full Backup) It is recommended to backup data before an upgrade; see: Section 4-5-3 on page 4-39.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 4-5-4 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Load Full System Configuration (Full Backup) (cont’d) 1.) Press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen. 3.) On the left side of the screen select BACKUP and then click the SYSTEM CONFIGURATION tab. 4.) Click the LOAD button of the “Full System Configuration” group.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 4-5-4 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Load Full System Configuration (Full Backup) (cont’d) 9.) Click the NEXT button and then select YES to start, or NO to cancel the restore procedure.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 4-5-6 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Archiving Images CAUTION It is recommended to Save Image Archive (Section 4-5-6-1 on page 4-44) and Backup the Full System Configuration (Section 4-5-3 on page 4-39) once a week.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 4-5-6-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Save Image Archive (cont’d) 5.) Choose archiving span time from the pop-up menu. NOTICE If for example “All images older than” 1 Day is chosen (see Figure 4-33 on page 4-44), images of the !! NOTICE: current day will not be archived! However, if you click the [Advanced] button you can put this right. 6.) If desired, check mark “Remove Local Images after Backup”. 7.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 4-5-6-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Load Image Archive 1.) Press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen. 3.) On the left side of the screen select BACKUP and then click the IMAGE ARCHIVE tab. 4.) Click the LOAD button (see: Figure 4-32 on page 4-44).
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 4-6 Software Configuration Checks 4-6-1 System Setup Press the F2 key on the control panel and then select SYSTEM SETUP. On the left side of the screen select the desired major group.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 4-6-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Measure Setup Press the F2 key on the control panel and then select MEASURE SETUP. The Measure Setup desktop offers three different pages to check. NOTICE Parameters and possible adjustments depend on the selected Application.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Section 4-7 Peripheral Checks Check that peripherals work as described below: Table 4-14 Peripheral Checks Step 4-7-1 Task to do Expected Result(s) 1 Press the FREEZE key. Stop image acquisition. 2 Press the remote key (P1, P2, P3 ) which is assigned to the BW printer. The image displayed on the screen is printed on the Black & White printer.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 1.) Remove the Monitor rear cover Monitor Rear cover Figure 4-37 2.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 Table 4-16 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Part list of Speaker plastics GEPN Description Qty 5433007 COVER LEFT SIDE, SPEAKER, PLASTIC 1 5433008 COVER RIGHT SIDE, SPEAKER, PLASTIC 1 2327596 SCREW WASHER HEAD M3 X 6 2 5433120 COVER, CAP 2 3.) Remove the two rubber caps and two screws beneath it Remove 2 screws 2327596 Remove 2 rubber caps 5433120 Figure 4-39 Rubber caps and screws 4.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) 5.) Remove keyboard base fasteners (5459100, OPIO CAPTIVE SCREW TETRIS) Figure 4-41 Keyboard base fasteners 6.) Remove 4 Probe holders from the keyboard assembly Table 4-17 Part list of Probe holders GEPN Description Qty 5433118 Probe Holder 2 1 5433119 Probe Holder 3 1 5442477 Vertical Probe holder 2 Figure 4-42 Probe holders 7.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Pull it downwards Figure 4-43 Keyboard assembly 8.) Remove upper Keyboard cover Upper keyboard cover Remove 4 screws 2159625, SCREW W/SP PH M4X8 Figure 4-44 Keyboard cover 9.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Cable tie 2160955 Figure 4-45 Cable tie 2160955 10.)Pull the LCD Cable through the LCD arm and Monitor holder LCD arm LCD cable Monitor holder Figure 4-46 LCD arm, cable and Monitor holder 11.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Keyboard lower plastic Keyboard assembly USB cables OPIO cable Power cable Figure 4-47 OPIO, Power and USB cables 12.)Disconnect Speaker cables Speaker cables Figure 4-48 Speaker cables 13.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) 14.)Disconnect the Power Cable and USB Cables from Keyboard upper cover USB cables Power cable Figure 4-50 USB and power cables 15.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) 16.)Hold the cast base firmly and lift it up as shown in the images below Hold the cast base firmly and lift it Lift the base and pull the LCD cable, OPIO cable and Speaker cables through the keyboard base hole Figure 4-52 Lift the base and pull the cables View after the Keyboard base removal Figure 4-53 Keyboard base is removed 17.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Assemble Extender as shown and pull all the cables through the Extender Extender Fasten 4 screws, 4 spring washers and 4 flat washers Figure 4-54 Extender 18.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 19.)Assemble the keyboard base Fasten 4 screws, 4 spring washers and 4 flat washers Figure 4-56 Assemble the keyboard base 20.)Re-Assemble LCD Cable 21.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 22.)Route the LCD cable as shown below. Insert the connector in the slot and fasten the screws. Figure 4-58 LCD cable 23.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) 24.)Connect the ground cable Ground cables, fasten 2 screws Figure 4-60 Ground cables 25.)Connect the Power cable and USB Cables Power cable USB cables Figure 4-61 Power and USB cables 26.)Reassemble the keyboard upper cover 27.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Fasten 2 screws and assemble the keyboard upper cover Figure 4-63 Keyboard upper cover Connect OPIO, Power and USB cables 28.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 29.)Reassemble 4 probe holders Figure 4-65 4 probe holders Figure 4-66 Fasten 9 screws 30.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Insert 2 rubber caps Fasten the screw Assemble LCD clamp and fasten a screw Route the LCD cable To avoid pinching of the LCD cable, make sure the cable is routed as shown above.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 4-7-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Brakes and Direction (Swivel) Locks Check the brakes and swivel locks function as described below. Table 4-19 Step Brakes and Direction (Swivel) Lock Task 1 Expected Result(s) Swivel lock released.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 4-8 Site Log Table 4-20 Date 4-66 Voluson® P8 / Voluson® P6 - Site Log (Paper Documentation)Voluson® P8 / Voluson® P6 Service person Problem Section 4-8 - Site Log Comments
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Chapter 5 Components and Functions (Theory) Section 5-1 Overview 5-1-1 Purpose of Chapter 5 This chapter explains Voluson® P8 / Voluson® P6’s system concepts, component arrangement, and subsystem function. It also describes the Power Distribution scheme and probes.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Section 5-2 General Information Voluson® P8 / Voluson® P6 is a digital beamforming curved-, linear- and phased array ultrasound imaging system.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 Section 5-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL General Information (cont’d) Among other significant features of the Voluson® P8 / Voluson® P6 are the following: - Integrated FrontEnd (uses advanced FGPA technologies) - Open connectivity using USB ports - Bluetooth - wireless connectivity - High performance 15” LCD monitor - Low profile, backlit Keyboard Voluson® P8 / Voluson® P6 has a digital beam formin
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 Section 5-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL General Information (cont’d) Figure 5-2 Basic Block diagram of Voluson® P8 / Voluson® P6 5-4 Section 5-2 - General Information
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 Section 5-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL General Information (cont’d) The Voluson® P8 / Voluson® P6 used digital beamforming technology which provides high resolution and high penetration performance. It is a general purpose, mobile, software controlled diagnostic ultrasound scanner. Its function is to acquire ultrasound data and to display the data of different modes.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 5-2-1 5-2-1-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Description of Voluson® P8 / Voluson® P6 Operating Modes B-Mode or 2D-Mode B-Mode or 2D-mode is a two-dimensional image of the amplitude of the echo signal. It is used for location and measurement of anatomical structures and for spatial orientation during operation of other modes.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 5-2-1-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Color Doppler Modes Color Doppler is used to detect motion presented as a two-dimensional display.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 5-2-1-4 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Pulsed (PW) Doppler PW Doppler processing is one of two spectral Doppler modalities. In spectral Doppler, blood flow is presented as a scrolling display, with flow velocity on the Y-axis and time on the X-axis. The presence of spectral broadening indicates turbulent flow, while the absence of spectral broadening indicates laminar flow.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 5-2-1-8 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 3D Rendering The 3D (volume) rendering is a calculation process to visualize certain 3D-structures of a scanned volume by means of a 2D-image. The gray value for each pixel of the 2D-image is calculated from the voxels along the corresponding projection path (analyzing beam) through the volume.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 5-2-2 5-10 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Block diagram Voluson® P8 / Voluson® P6 Figure 5-4 Voluson® P8 / Voluson® P6 - Block diagram Section 5-2 - General Information
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 5-2-2-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Front End Block diagram For further details : see: "Front End Processor - Block diagram" on page 5-20 see: "PIR - Blockdiagram" on page 5-21 see: "Beamformer + Frontend" on page 5-22 see: "RFS-Board - Block diagram" on page 5-23 5-2-2-2 Back End Block diagram For further details : see: "BackEnd Processor - Block diagram" on page 5-25 5-2-2-3 OPIO Block diagram For furth
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 5-2-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Data Flow Control Description This section describes the functions of the boards vs. different operation modes. 5-2-3-1 • BF64 - Beamformer Motherboard • RFS - Mid-Processor, System Control and DMA Controlle B-Mode 1.) RFS The RFS contains the Clock-Management and PRF-Generator.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL b.) XBEAM CRI (CrossBeam - Compound Resolution Imaging): Does not need any special functions of RFS. Image is composed of more than one different-direction-steered images. PC-calculated. 5-2-3-2 M-Mode 1.) RFS see: Section 5-2-3-1 "B-Mode" on page 5-12 a.) DMA section B-mode-Data is written via SP0 into SDRAM Fifo Buffer memory.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 5-2-3-5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL CFM-Mode (Color Flow Mode) 1.) RFS 5-2-3-6 - PRF-generator; see: Section 5-2-3-1 "B-Mode" on page 5-12 - After DC-cancelling the signal is mixed with RX-Frequency and brought to LF-Spectrum, where the LowPassFilter cuts HF. Mixer and Magnitude-Calculator arrange Complex Demodulation.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 5-2-4 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Description of Software Options To activate the software options: 1.) Press the F2 button on the OPIO. 2.) On the left side of the screen select SYSTEM SETUP to activate the setup desktop screen. 3.) On the left side of the screen select ADMINISTRATION and then click the OPTIONS tab.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 5-2-4-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 3D/4D Advance 5-2-4-1-1 Real Time 4D Real Time 4D mode is obtained through continuous volume acquisition and parallel calculation of 3D rendered images. In Real Time 4D mode the volume acquisition box is at the same time the render box. All information in the volume box is used for the render process. In Real Time 4D mode a “frame rate” of up 40 volumes/second is possible.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL tissue scatters, image speckle characteristics such as brightness, density or size have no apparent value in determining tissue structure or related properties. The elimination of or significant reduction in speckle improves the quality or diagnostic potential of the image.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 5-2-5 Description of Hardware Options Table 5-3 Hardware Options HW-Options 5-2-5-1 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Description 1 CW-Doppler 5-2-5-1 CW - Continuous Wave Doppler 2 DVR Recorder 5-2-5-2 DVR Recorder 3 ECG Module 5-2-5-3 ECG Module 4 WLAN Network Adapter 5-2-5-4 Wireless Network Adapter (WLAN - Wireless Local Area Network) CW - Continuous Wave Doppler CW Doppler mode provides real ti
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 - DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Printer Drivers LINUX partition: (not visible in Windows) - Linux operating system for rescue functionality Figure 5-5 Data Location - distribution of partitions Chapter 5 - Components and Functions (Theory) 5-19
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 5-3 FrontEnd Processor Figure 5-6 Front End Processor - Block diagram Table 5-4 below outlines the Voluson® P8 / Voluson® P6 Front End components described in the subsections.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 5-3-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL PIR - Probe Interface board Switches the Probe Connectors (3 RS -Connectors) and recognizes Probes • 3 RS-Probe Connectors • Probe Select Relays • Probe Recognition For Block diagrams refer to: - Figure 5-7: PIR - Blockdiagram Figure 5-7 PIR - Blockdiagram Chapter 5 - Components and Functions (Theory) 5-21
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 5-3-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL BF64 - Beam former board BF64 board has 64 TX/RX channels, respectively. beamformer board is a complete 64 channel ultrasound Front-end that includes the following features and functions:.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 5-3-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL RFS - Radio Frequency interface and System controller Figure 5-9 RFS-Board - Block diagram 5-3-3-1 RFS board - FPGA A.) Beamformer interface • LVDS-interface • Raw ultrasound data is transferred from the Beamformer to the Interface FPGA. • Configuration data (Probe Setup data, Beam Setup data) is transferred from the Interface FPGA to the Beamformer.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL • Stores this data in the Line Memory buffer (2x SDRAM) • Transfers buffered data by DMA-transfer via PCI and PCI-Express to the PC - Non-TimeMotion data (B, C): DMA0 - TimeMotion data (M, PW): DMA1. C.) PIR FPGA programming interface • This FPGA is the interface between the scanner connectors and the RFS-board. D.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 5-4 BackEnd Processor Figure 5-10 BackEnd Processor - Block diagram Chapter 5 - Components and Functions (Theory) 5-25
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 Table 5-5 DRAFT (JANUARY 17, 2013) Voluson® P8 / Voluson® P6 - RFS components Sub-section 5-4-1 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Description Page Number 5-4-1 Hard Disk Drive 5-26 5-4-2 SOM (System On Module) 5-26 5-4-3 DVR - Digital Video Recorder (option) 5-27 Hard Disk Drive The Hard Disk is the main storage device of the Voluson® P8 / Voluson® P6.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 - DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Control DVD drive (SATA), User Interface (USB), DVR (USB) Figure 5-11 SOM (System On Module) 5-4-3 DVR - Digital Video Recorder (option) For further details see: Section 5-9-1-1 "DVD Recorder" on page 5-37.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) Section 5-5 Internal I/O Figure 5-12 Voluson® P8 / Voluson® P6: Internal I/O 5-28 Section 5-5 - Internal I/O VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Section 5-6 OPIO (User Interface) 5-6-1 OPIO Block Diagram A/N Key USB Cable 8 wire (USB (4) + LED backlight) USB HUB * Stanby 3.3V Down Stream(2) Stanby 26Pin Flat Cable Power LED(4) Function S/W (9) HOST LED Control * USB-2 Communication * USB-1 Communication SUB Power Cable Assembly 6P Wire * +12V Power USB HUB * Stanby 3.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 5-6-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL OPIO layout Figure 5-14 Voluson® P8 / Voluson® P6 - OPIO Layout 5-6-3 5-30 Button Description Section 5-6 - OPIO (User Interface)
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) Chapter 5 - Components and Functions (Theory) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 5-31
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 5-32 DRAFT (JANUARY 17, 2013) Section 5-6 - OPIO (User Interface) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Function of the Trackball at Diverse Dialog Pages In general operations at diverse dialog pages and windows on the system desktop (e.g., Patient Data Entry, Usage of EUM, System Setup, Measurement Setup, etc.) are executed by trackball and the trackball keys (mouse emulation).
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) .
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Section 5-7 Monitor right > increase < left decrease On screen Display (OSD) OSD Menu key Figure 5-15 Monitor Adjustment buttons Menu key [OSD](Press the OSD (middle) key of the monitor controls over 10 sec.): This button will enable the On Screen Display to modify the settings and control the special Keyboard Keys features.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 5-8 External I/O 1 2 3 4 Figure 5-16 External I/O Connectors - on Rear of System (GES) Item Connector Name 1 USB Description USB-2.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 5-9 Peripherals 5-9-1 5-9-1-1 Recording Tools DVD Recorder The DVR recorder is implemmented by installing the DVR Module on RFS incorporating with DVD Drive. 5-9-2 5-9-2-1 Printers Black & White Digital Printer The B&W Digital Printer receives image data via the USB port.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Section 5-10 Power Distribution 5-10-1 ACFE - Primary Power Module Circuit breaker Equipotential connection Mains voltage indicator (shows input voltage level) pull-out protection Mains Power cable Figure 5-18 Primary Power Module 5-38 Section 5-10 - Power Distribution
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 5-10-1-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL CPS Block Diagram Figure 5-19 CPS Block Diagram Chapter 5 - Components and Functions (Theory) 5-39
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 5-10-1-2 5-40 DRAFT (JANUARY 17, 2013) Power Distribution Block Diagram Figure 5-20 CPS Block Diagram Section 5-10 - Power Distribution VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 5-10-1-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Power Up Sequence Description The Power Up Sequence can be divided in the following steps: • 5-10-1-3-1 Switch Circuit Breaker in CPS to ON position • Press the ON button on the Operator Panel • RFS power-up Power Up Sequence Description Figure 5-21 Power Up Sequence 1.) RFS, powered by 5Vstb, detects contact of Power-On button event. 2.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 5-10-1-3-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Power Down Sequence Description Figure 5-22 Power Down Sequence 1.) On/Off Switch is pressed on OPIO. 2.) ON/OFF signal is sent from On/Off Switch via OPIO cable to RFS 3.) “Shut down” dialog comes up and “Shut down” signal is sent. 4.) PC “Shut down” sequence is started in RFS. 5.) At the end of this sequence RFS sends PS_ON Signal to CPS. 6.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 NOTE: DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL All DC-supply voltages for built in peripherals are generated in DCDC module not inside ACFE. NOTICE The systems mains supply input voltage is a free voltage not need to configure for different AC power!! NOTICE: in.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 5-10-2 5-10-2-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL ACDC - primary DC Power Supply Function of ACDC With the AC input voltage from the ACFE, the ACDC module generates 12Vdc 5-10-3 DCDC - main DC Power Supply With the 12Vdc from ACDC, the DCDC module generates all DC power sorces: 5-10-4 • Front end voltages • Standby voltage • Tx voltages • DVD supply • 4D module supply • Fan supply 4D Module In addit
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Section 5-11 Service Platform 5-11-1 Introduction The Service Platform will increase service productivity and reduce training and service costs. 5-11-2 Access / Security The Service Platform has different access and security user levels. Each user is only granted access to the tools that are authorized for their use.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL For more detailed information and description refer to Section 5-12 on page 5-48. 5-11-2-2 Local(Extended) Access 1.) If not already in read mode, FREEZE the image. 2.) Connect the USB “standard GE” service dongle to the USB connector(s) on the rear of the system. As soon as the dongle is detected the “Extended Access” login window pops up.. 3.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 5-11-2-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Remote Access NOTICE Remote access is ONLY possible if the service platform is properly configured (either by the user or a GE technician at site). Operation see: Section 7-5-4 "CSD: Configuration" on page 7-15. !! NOTICE: This offers GE technicians the possibility to view the entire customer´s desktop and operation system.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 5-12 Common Service Desktop (CSD) 5-12-1 Internationalization The user interface provided by the service platform is designed for GE personnel and as such is in English only. There is no multi-lingual capability built into the Service Interface. There are different possibilities to access the Common Service Desktop and its available features: 1.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 5-12-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Error Logs When the Error Logs page is selected, different log viewing options are available. Log Viewer is displayed in a separate window. 5-12-3 Diagnostics Figure 5-26 Common Service Desktop - Error Logs The Diagnostic page uses a web-controlled user interface to provide access to common service components and perform diagnostics.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 5-12-4 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Image Quality In the Image Quality page, you can verify image quality. Figure 5-28 Common Service Desktop - Image Quality NOTE: 5-12-5 This page is not populated in this version. Configuration In the Configuration page, you can view and modify different device informations and configurations. in the “InsiteExC Agent Configuration” option field.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 5-12-6 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Utilities The Utilities page contains a variety of Windows utility tools to indicate the status of the system, in addition to various other tools. Figure 5-30 Common Service Desktop - Utilities 5-12-7 Replacement In the Replacement page, Hardware parts and their versions are displayed.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 5-13 Service Page 5-13-1 Introduction The Service Page contains specific software/hardware test modules, system setup, update, etc. for Voluson systems only. 5-13-2 Access / Security The service page has different access and security user levels. Each user is only granted access to the tools that are authorized for their use. 5-13-3 Service Login 1.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 5-13-3-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Update 5-13-3-2-1 FMI from DVD By means of the FMI FROM DVD button, the Systems C:\ image is partly or completely updated. The System Software parts to be upgraded depend on contents of the used System DVD. PRECONDITION: NOTE: 5-13-3-2-2 The first “Boot Device” in BIOS has to be Hard Disk Drive. During “FMI from DVD” the used system configuration (incl.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 5-13-3-8 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Keyboard Layout To change the keyboard layout to different languages. Operation see: Section 6-4 "Modification of Keyboard Layout" on page 6-3. NOTE: 5-13-3-9 Reconfigure the layout of the keyboard is only useful by changing the concerned keys also.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 5-14 Boot Screen Functions 5-14-1 Overview Following LINUX supported functions are available as soon as the “Boot Screen” appears: Voluson GE-Service Rollback Memtest Figure 5-36 Boot screen NOTICE After 3 sec. without pressing any key, the system will boot-up in windows (= Voluson item). !! NOTICE: 5-14-1-1 Voluson The System will boot-up in windows.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 5-14-1-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Rollback This function offers the possibility to simply restore the previously used system configuration (rollback), which was stored on R:\ during “FMI from DVD”. 1.) Turn system OFF and then back ON. 2.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 5-15 LINUX - Voluson Maintenance System 5-15-1 Overview The Voluson® P8 / Voluson® P6 systems contains the “Voluson Maintenance system” with improved service functionality (e.g., restore software/recover partitions) + possibility to perform extended PC Memtest (see: 5-15-8 "Memory Check in LINUX" on page 5-77).
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 5-15-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Before Recovering or Initializing the Hard disk CAUTION If the system is still responsive (system starts up without problems and the application software is running as usual, try to print out or write down System Setup and Measure Setup settings; because they could be lost.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL - Click the GENERAL tab and print out or write down the settings of this page. - Click the USER SETTINGS and print out or write down the settings of this page. - Click the PATIENT INFO DISPLAY tab and print out or write down the settings of this page. 8.) Press the F2 key on the control panel. 9.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 5-15-4 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Voluson Maintenance System Enter Linux as described in Section 5-15-2 "Access/Security" on page 5-57.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 5-15-5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Media Menu NOTICE The “Media Menu” is only accessible if the System DVD (see: Table 9-13) is inserted into the system’s DVD/CD+R/RW Drive 1.) If system functionality allows it, collect vital system information as described in 5-15-3 "Before Recovering or Initializing the Hard disk" on page 5-58. 2.) Insert the System DVD into the DVD/CD+R/RW Drive. 3.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Operation: 1.) Select [Restore] in the displayed “Media Menu” (see: Figure 5-39 above). NOTE: If the message: “Recovery system (Linux) needs to be updated ...” appears, confirm with YES. NOTE: If the currently installed software has to be upgraded to a newer version, and the system DVD being used is SW 13.0.0 or higher, a new software specific “Permanent key” is required. 2.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 4.) In the displayed menus select [Exit] and then [Exit/Reboot]. Figure 5-42 Exit -> Exit/Reboot 5.) Remove the System DVD from the DVD/CD+R/RW Drive drive. NOTICE The system boots with windows. An automated process was developed to check disk, assign an unique computer name and matching drive letters. The system will reboot at least 2 times! 6.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 8.) If following message appears, skip the “BT warning” with CONTINUE. Figure 5-44 Skip the BT warning with CONTINUE NOTE: Options will be recovered at a later step using “Full Backup” process. 9.) Customize the system as described in Section 5-15-7 on page 5-75. 5-15-5-3 Initialize This function performs operations which result in complete data loss of all files on C:\ and D:\
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL NOTICE However, if system functionality allows it: • save the patient images and corresponding patient data via System Setup - Backup - IMAGE ARCHIVE, as described in Section 4-5-6 on page 4-44 • gather full system configuration by saving a “Full Backup” via System Setup - Backup - SYSTEM CONFIGURATION, as described in Section 4-5-3 on page 4-39 Backed-up data can be restored
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 5.) Use “Arrow” keys [left] or [right] on the alphanumeric keyboard to select the [Boot] tab. Figure 5-45 Boot Menu 6.) Using the keys PG UP [up] and PG DN [down] on the alphanumeric keyboard select 1st Boot Device and then press ENTER.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 7.) Select Primary Master and then press ENTER. again.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 8.) Select 3rd Boot Device and then press ENTER. 9.) Select Third Master and then press ENTER. again. Figure 5-47 3rd Boot Device 10.)Press the F10 key to Save and Exit. 11.)Confirm the “Save configuration changes and exit setup” message with OK and ENTER. 12.)Proceed with SectionSection 5-15-5-4-2 on page 5-68.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 2.) In the displayed menus select [Media] -> [NewDisk] and then press ENTER. Figure 5-48 Media -> NewDisk NOTE: If the currently installed software has to be upgraded to a newer version, and the system DVD being used is SW 13.0.0 or higher, a new software specific “Permanent key” is required. 3.) The following warning message will be displayed.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Please wait until the final window (Figure 5-50 below) appears and then press ENTER. Figure 5-50 New Harddisk setup OK 6.) Remove the System DVD from the DVD/CD+R/RW Drive. In the displayed menus select [Exit] -> [Exit/Reboot]. Figure 5-51 Exit -> Exit/Reboot 7.) During startup, press the DEL key immediately, to enter the BIOS Setup. 8.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 10.)Wait until all processes are finished and the final “DOS” window (Figure 5-52 below) appear. Figure 5-52 press the “A” key to start UlSApp 11.)Press A key to activate the Ultrasound Application Software [UIS]. 12.)If following message appears, skip the “BT warning” with CONTINUE. Figure 5-53 Skip the BT warning with CONTINUE 13.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 5-15-6 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Advanced Functions Menu 1.) If system functionality allows it, collect vital system information as described in Section 5-15-7 "After Recovering or Initializing the Hard disk" on page 5-75. 2.) Enter Linux as described in Section 5-15-2 "Access/Security" on page 5-57. 3.) In the “Voluson Maintenance System” main menu (Figure 5-38) select [Advanced]. 4.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 5-15-6-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL SSHD By selecting this item, a SSH-daemon is started on the system. The [SSHD] function is reserved for extended troubleshooting. Not intended for use in the field. 5-15-6-4 Restore This function takes the image stored on the Rescue (R:) partition to replace the existing image of the “C” partition. NOTE: Data on the D: [User] partition (e.g.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 5-15-6-9 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL USBTree The USB Tree displays connected devices and their respective hub. This enables easy control, whether all devices are correctly wired. Figure 5-55 USB Tree (example) 5-15-6-10 SmartHDDtest This item starts the SMART (Self Monitoring, Analysis and Reporting Technology) long test.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 5-15-7 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL After Recovering or Initializing the Hard disk NOTICE In case that “Full Backup” is NOT available, contacting the OLC will be necessary to get “Permanent key”. Before calling OLC, please prepare the following data: • SNO of Voluson® P8 / Voluson® P6 • Needed option 1.) If not currently done, restart the system (with new HDD) and boot up completely. 2.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 12.)Press F2, select the MEASURE SETUP item and check and match settings of all the pages. 13.)Click the CLOSE button to store settings and exit the Measure Setup desktop screen. NOTICE Save any presets, configurations and settings to HDD and/or DVD, etc. to have it available for next time. Operation see: Section 4-5-3 "Save Full System Configuration (Full Backup)" on page 4-39" .
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 5-15-8 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Memory Check in LINUX 1.) Turn system OFF and then back ON. 2.) As soon as the “Boot Screen” appears (see: Figure 5-36 on page 5-55), press the PG DN [↓ Arrow down] key on the keyboard until the Memtest item is highlighted, then press ENTER. NOTICE After 3 sec. without pressing any key, the system will boot-up in windows.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 5-78 DRAFT (JANUARY 17, 2013) Section 5-15 - LINUX - Voluson Maintenance System VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of Chapter 6 This chapter describes how to test and adjust the mechanical capabilities of a scanner that may be out of specification. Although some tests may be optional they should only be performed by qualified personnel.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 6-3 LCD Monitor Adjustment The Voluson® P8 / Voluson® P6 has a free adjustable LCD monitor in relation to the user interface. To adjust the brightness: 1. Contrast Indicator 2. Brightness Indicator 3. Dim Brightness Indicator 4. Light Button 5. Adjustment (-) Button 6. Toggle Button 7.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 6-4 Modification of Keyboard Layout NOTE: Configuring the layout of the keyboard is only useful by changing the concerned keys also. see: Section 8-38 "Replacement of AN KEY TOP" on page 8-83 6-4-1 Setup the Voluson® P8 / Voluson® P6 Keyboard Language Layout 1.) Press the F2 key on the control panel. 2.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) 6.) Select default input language from the drop down menu.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Chapter 7 Diagnostics/Troubleshooting Section 7-1 Overview 7-1-1 Purpose of Chapter VOLUSON® P8 / VOLUSON® P6 This section describes how to setup and run the tools and software that help maintain image quality and system operation. Basic host, system, and board level diagnostics are run whenever power is applied. Some Service Tools may be run at the application level.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 7-2 Collect Vital System Information 1.) Press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen. 3.) On the left side of the screen select ADMINISTRATION and then click the SYSTEM INFO tab.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 Section 7-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Collect Vital System Information (cont’d) Move the scroll bars downwards to view additional information about installed software/hardware (e.g., Console information). Figure 7-1 System Setup - Administration - SYSTEM INFO page (e.g.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 7-2-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Request for Service (RFS) NOTICE Service Connectivity has to be checked out once before you can request for service. !! NOTICE: i.e., Service platform has to be configured properly; see: CSD: Configuration on page 7-15. There are two possibilities to contact GE: • by means of the GE “Remote Status Icon” that is displayed on the bottom of the screen.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 7-2-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Request for Service (RFS) (cont’d) A request confirmation screen is displayed: write down and keep the Reference Number Figure 7-3 Request submitted 5.) Write down and keep the Reference Number for follow up procedures, then click OK. NOTE: If the service platform is not configured an Error message is displayed.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 7-2-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Shortcuts List Press the CTRL + H key simultaneous to display the shortcuts list and a description of what they do. Figure 7-5 Shortcut list (Ctrl + H) 7-6 Shortcut Description Ctrl + C copy text Alt + D export system state (Full Backup and Dump files) Ctrl + H shortcuts list (see: Figure 7-5 above) Alt + P stores screenshot on D:\export Ctrl + V paste, insert t
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 7-3 Check Points Voltages 7-3-1 7-3-1-1 How to check power Common Power Supply (CPS) Turn on mains switch and check green LED inside Power. LED Figure 7-6 Check green LED inside CPS NOTE: If LED is on, CPS has full function (all voltages are within the specified range).
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 7-4 Screen Captures and Logs There may be times when the customer or field engineer will want to capture a presentation on the screen so it may be recovered by the OnLine Center. This is accomplished by saving the image(s): A.) to Archive and export them (as jpg, bmp or tiff) to DVD/CD+R/RW or external USB drive. B.) as jpg and bmp to D:\export by pressing the ALT + P key on
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 7-4-1-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL How to use a Capture Filter Use filters to limit the captured data. Filter by IP and port number or by IP only. by IP and port number 1.) Click on CAPTURE FILTER (see: Figure 7-9 on page 7-9). 2.) Select NEW. 3.) Edit “Filter name”. For example: DICOM port 104 4.) Enter the “Filter string” with the following syntax: port xxx and host yy.yy.yy.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 7-4-1-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL How to use a Screen Filter In addition to the “pre” capture filter, use the Filter tool in the screen to filter what is displayed. Filter the DICOM packets, since they are the most probable for the troubleshooting. 1.) Type dicom in the “Filter” field and click on APPLY. The display should filter all DICOM packets, filtering out image data.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 7-4-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Export Log’s and System Data There are two possibilities to export system data (and log´s): 1.) by pressing the ALT + D key to save a snapshot and “Full Backup” of the System state; see: Section 7-4-2-1 on page 7-11 2.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 7-4-2-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Export Log´s and System Data (via Service Page) 1.) Press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen. 3.) If not already done, connect the USB “standard GE” service dongle to the USB connector(s) on the rear of the system. As soon as the dongle is detected the login window pops up. 4.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 7-5 Remote Access to Service Platform 7-5-1 General This offers GE technicians the possibility to view the entire customer´s desktop and operation system. Using VCO (Virtual Console Observation) a service technician or the OnLine Center can access and modify all PC settings and programs or run diagnostics on the customer´s ultrasound scanner.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 7-5-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Common Service Desktop (CSD) As soon as the Common Service Desktop (CSD) is started, the Service [Home] Page appears. Figure 7-14 Common Service Desktop - Home NOTE: 7-14 As described in Chapter 5 - Common Service Desktop (CSD), on page 5-48, the service platform uses a web-based user interface (UI) to provide access to common service components.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 7-5-4 7-5-4-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL CSD: Configuration To configure Service Platform NOTICE If a Proxy server is available, the information MUST be properly entered, otherwise remote control !! NOTICE: does not work. There is no possibility to detect proxy server information automatically. 1.) Ask the hospital's network Administrator for the Proxy Server Address and the Proxy Server Port. 2.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL NOTE: 1.) It requires all bold fields should be filled out. The mandatory are : 1.) Device Name (Not Change) 2.) CRM No 3.) Continent 4.) Country 5.) City 6.) State(Prov) 7.) Enterprise Server 8.) Service Center 9.) File watcher 10.)Proxy and Proxy Authentication (If it requires) 2.) Device name is pre-populated. Do not change this Value. 3.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL [Start] Auto Tester Figure 7-16 Service Tools window 7.) Activate the “Auto Tester” program by clicking START.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 Section 7-6 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL How to use the Auto Tester program (cont’d) The following message box appears. Figure 7-17 Message Box 8.) Click OK. 9.) Press the ALT GR key on the alphanumeric keyboard. 10.)Activate the “Auto Tester” program by clicking the “Record” icon on the displayed screen. “Record” icon Figure 7-18 Start Auto Tester 11.)Start scanning.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 Section 7-6 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL How to use the Auto Tester program (cont’d) The following screen will appear. “DVD/CD Burn” icon “Disc” icon Figure 7-19 Auto Tester Finished 12.)Insert an empty DVD/CD+R/RW in the Drive and select the “DVD/CD Burn” icon. Or hit “Disc icon” to save to export. 13.)Enter a Filename. Figure 7-20 Enter a Filename 14.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 7-7 Troubleshooting Trees, Instructions and Tech Tips Table 7-1 below outlines Voluson® P8 / Voluson® P6 troubleshooting trees and instructions shown in the sub-sections.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 7-7-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL System does not boot up System doesn’t boot Is Standby Switch illuminated? No Verify that - Power Cord is connected to Mains - Power Switch is turned on (on the backside of the system) Yes Is Standby Switch orange? Yes No Press Standby Switch till it changes to green System starts booting Is there any picture on the monitor ? No Check cables between Console and Monit
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 7-7-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Noise disturbs the Image Noise is shown in image. Is there any electrical device that affect the Scanner? NO YES Keep the scanner away from those electrical devices. use another wallsocket Check probes.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 7-7-3 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Trackball - Impaired Sensitivity The trackball has low sensitivity! NO Is it difficult to roll the trackball? Contact GE service representative to replace the trackball module. YES Set up the appropriate trackball response time (via [Trackball Speed] button in the System Setup - User Settings page). Still low sensitive... Remove the trackball and clean it from dust.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 7-7-4 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Printer Malfunction Figure 7-25 Printer Malfunction 7-24 Section 7-7 - Troubleshooting Trees, Instructions and Tech Tips
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 7-7-5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Monitor Troubleshooting Fault symptom Check these items Check the main panel cable is properly connected to Monitor. No image/No signal Check the main panel cable is properly connected to RFS. Check no pins of the main panel cable are bent. Video test pattern are not clear, bright parallel or square Replace the monitor.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 7-7-6 NOTE: DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DVD/CD+R/RW Drive Test Only DVD+R/RW and CD+(R)W medias are available. 1.) Insert an empty DVD/CD+R/RW into the Drive. 2.) Enter “Patient Archive” by pressing the PATIENT key on the control panel. Archive choose Source select Exam (with images) Figure 7-26 Patient Archive - ARCHIVE 3.) On the left side of the screen select ARCHIVE as shown in Figure 7-26 ABOVE. 4.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 7-7-6 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DVD/CD+R/RW Drive Test (cont’d) 6.) Export images of the selected exam to DVD/CD+R/RW: a.) Click the EXPORT button. b.) If not already selected, choose “DVD/CD” drive from pop-up menu. c.) Enter “File name”. d.) Select any Voluson Format (*.4dv) from the pop-up menu. e.) Click the SAVE button. step b.) choose “Source” step c.) step e.) step d.) step a.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 7-7-7 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Network Troubleshooting 7-7-7-1 No Connection to the Network at All Gray = Cable disconnected or no network signal on a connected cable. Green = Cable connected to a network. Does not imply proper network settings. 1.) Check that the network cable between the Voluson® P8 / Voluson® P6 system and the wall network is connected and well seated in both ends.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 7-7-8 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Tech Tips Use Table 7-2 below to access Voluson® P8 / Voluson® P6 (troubleshooting) Technical Tips listed. Each entry in the table will hyper-link to the issue, cause and solution.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 7-8 Error Messages NOTICE If the problem (error message) still exists after performing the described actions, call technical support.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Error Messages Actions BC Ensemble Out Of Range press ok and save this user-setting once again BC lines_per_sequenz < 1 Action: restart the system restart the system BC_Dynamic Out Of Range press ok and save this user-setting once again BC_Gain Out Of Range press ok and save this user-setting once again BC_Lines: BC_LineDensity out of limit restart the system BCMC_Balan
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) Error Messages VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Actions cannot read file restart the system Cannot import volume data to 3D dll Action: Load volume files from other storage medium. Call technical support Load volume files from other storage medium. Can't add chosen files to the virtual folders! restart the system Can't create hardware configuration index file in D:\\SERVICE Action: restart the system restar
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Error Messages Actions CW quant TxFreq out of range restart the system CW_BaseLinePos Out Of Range Press ok and save this user-setting once again. CW_Gain Out Of Range Press ok and save this user-setting once again. CW-hardware doesn't support pencil probe + CWHardware not available, restart the system CW-TxFrequency == 0 restart the system D:\\SERVICE\\ConfIndexFile is
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) Error Messages VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Actions Enhance Out Of Range press ok and save this user-setting once again Error in apertureShape, lateral start pos of the aperture incorrect! Action: Please contact technical support restart the system Error in CreateCineImage try to store again, restart the system Error in execution due to: %s, please call technical support.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) Error Messages VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Actions File Volume size not consistent. Action: Load volume files from other storage medium. load volume files from other storage medium FilterBuffer " Memory allocation failed".
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) Error Messages VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Actions iPWCW_CloseProcessing in RT_PWorCWFilterBlock::execute() returned error code xx !! Action: restart the system, If after restart not ok then call technical support restart the system iPWCW_InitializeProcessing in RT_PWorCWFilterBlock::execute() returned error code xx !! Action: restart the system, If after restart not ok then call technical support restart
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Error Messages Actions No CD Writer found check the connection and the Power cable - plug the cable off and on and try again.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) Error Messages VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Actions Probe-ID valid, but Flexible-EEPROM-Datamodel not implemeted yet on left probe connector (A). Probe-ID valid, but Flexible-EEPROM-Datamodel not implemeted yet on middle probe connector (B). Probe-ID valid, but Flexible-EEPROM-Datamodel not implemeted yet on right probe connector (C).
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) Error Messages VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Actions RepresentationManager::vCreate: Dimension unknown, arguments of vSetReplayParams() incorrect!! restart the system RepresentationManager::vCreate: Nr of Units = xx, wrong arguments in Constructor and/or vSetReplayParams() used!! Parameters => Mode[xx] BufferSize[xx] unitSize[xx] m_iSizeDim0[xx] m_iNrDim0[xx] m_iNrDim1[xx] m_iNrDim2[xx] restart the system
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) Error Messages Actions RT_MColorFilterBlock::Memory allocation failed.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Error Messages Actions RT_RFProcessingBlockTissue::vCheckRFDataSizeAndUpdateBuffers: Memory allocation failed. Action: restart the system, If after restart not ok then call technical support restart the system RT3D.cpp rot phased: unexpected O_AxisAngle ? restart the system RT-TissueFilter DLL call failed. Action: restart system restart the system SamplePRF too big.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Error Messages Actions The System detected a problem with your harddisk. Please do not reboot or shut down the machine! Please contact your service department as soon as possible.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Error Messages Actions Unhandled Probe-EEPROM data type on middle probe connector (B). Connect another probe Unhandled Probe-EEPROM data type on right probe connector (C). Connect another probe Unhandled Probe-EEPROM data type on upper-left probe connector (CW).
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 7-44 DRAFT (JANUARY 17, 2013) Section 7-8 - Error Messages VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter contains replacement procedures for different modules and their subsystems. NOTICE The Manpower, time and Tools indicated in the Sub-sections include all requirements from Preparations to Installation Procedures.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 Table 8-1 DRAFT (JANUARY 17, 2013) Chapter 8 Contents Section 8-2 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Description Page Number 8-15 Replacement of the AIR FILTER 8-37 8-16 Replacement of the Probe Holder(KIT) 8-39 8-17 Replacement of SIDE COVER L ASSY 8-40 8-18 Replacement of SIDE COVER R ASSY 8-42 8-19 Replacement of the Trackball 8-44 8-20 Replacement of the TGC BOARD 8-47 8-21 Replacement of AN KEYBOARD ASSY 8-49 8
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 8-1-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Returning/Shipping System, Probes and Repair Parts When returning or shipping the Voluson® P8 / Voluson® P6 system in the original packaging: NOTE: • system must be lowered to its minimum height with monitor flapped down (see: Figure on page 311) • the Control Console has to be centered and locked in “unextended” position For Control Console Positioning refer to 5-11-3
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-2 System Software - Installation Instructions 8-2-1 Manpower One person ~ 1 hour (depends on contents of System DVD, peripherals, etc.) 8-2-2 Tools System DVD - Base image software part number for Voluson P6/P8 - 5438266 1.) Insert the CD/DVD 2.) The CD/DVD is located as shown below.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Select Boot option Figure 8-2 Boot option 4.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 5.) Select New Disk Select New Disk Figure 8-4 New Disk 6.) Figure below shows the Installing screen Figure 8-5 Installing 7.) After installation, remove the DVD from the system. 8.) Reconfigure the 1st boot device to hard disk and restart the system. 9.) On second reboot, system will ask to confirm activation of Software stating - “Do you want to activate: Type A”.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Select A Figure 8-6 Activate Software 10.)After UISApp is running, check the installed version by following these steps: A.) Press F2 (Utility) hard key on the console B.) Select "System Setup", and select "Administration". C.) Select "System Info" tab. D.) Check the version tag from the System Info Software panel as shown in the example figure below. Figure 8-7 Version Info E.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-3 System Software - Installation/Upgrade Procedure 8-3-1 Introduction NOTICE To update the Systems C:\ image via the FMI FROM DVD button in the System Setup SERVICE page, see: Section 8-3-5 "System Software - Installation Procedure (FMI from DVD)" on page 8-12. The Software parts to be upgraded (e.g.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 8-3-4 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Preparations (cont’d) Table 8-1 Mode 2D Mode Quality Receiver Frequency M Mode Volume Mode Triplex Mode System Function Task Expected Result Connect an abdominal probe, press the PROBE key, select the “Abdomen” - Application and start the “Default” -Program. Record a 2D image of the Liver.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL which Software/Hardware version is installed in the unit.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 8-3-4 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Preparations (cont’d) 4.) Select the OPTIONS tab to see which Options (and Application Packages) are installed.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 8-3-5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL System Software - Installation Procedure (FMI from DVD) The system software installation procedure starts with saving and recording the settings present on the system (silent “Rollback”). Then the new software is written to the hard disk using the System DVD. Application Settings are automatically updated, to match with new Software version.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 8-3-5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL System Software - Installation Procedure (FMI from DVD) (cont’d) 9.) Click the FMI FROM DVD button (see: Figure 8-10 on page 8-12) for updating the System Software. 10.)Verify that only the DVD drive is connected to the system, then click OK (Figure 8-11 below). Figure 8-11 Verify that USB devices are disconnected, then click OK 11.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 8-3-5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL System Software - Installation Procedure (FMI from DVD) (cont’d) 12.)Check disk is performed automatically - restart. NOTICE Do NOT skip checkdisk. !! NOTICE: Figure 8-15 Check disk is performed automatically 13.)3 dots (one after the other) appear on the screen (see: Figure 8-16 below).
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 8-3-5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL System Software - Installation Procedure (FMI from DVD) (cont’d) 14.)Booting auto..... Figure 8-17 Boot screen - auto 15.)Updating will take some time...... Figure 8-18 Please wait.... 16.)Please wait until all processes are finished (100 percent completed). 17.)The system is rebooting into windows (Boot screen - Voluson).
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 8-3-5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL System Software - Installation Procedure (FMI from DVD) (cont’d) Figure 8-19 automatic processes are running 18.)Please wait until all processes are finished. Finally the 2D screen is displayed on the monitor. NOTE: If the BT warning dialog (Figure 8-20 below) appears, enter the proper “Permanent key” and then click SUBMIT.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-4 Software and Functional Checks after Installation/Upgrade Procedure 1.) Press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen. 3.) On the left side of the screen select ADMINISTRATION and then click the OPTION tab. 4.) Verify the correct settings of the OPTIONS page; see: Figure 8-9 on page 8-11.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-5 Image Settings Only - Loading Procedure 8-5-1 Introduction The Image Settings contain: 8-5-2 • Application Settings • User Programs • Auto Text • 3D/4D Programs • Scan Assistant • Measure configuration Loading Procedure see: Section 4-5-2 "Load Image Settings Only" on page 4-36 Section 8-6 Full Backup (Full System Configuration) - Loading Procedure 8-6-1
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-8 Replacement or Activation of Options Following Software options are available: Table 8-2 Software options Options VP6 VP8 CFM/M-CFM1 Option Standard Basic 3D/4D2 Option Option Advanced 3D/4D3 Not Available Option DICOM Option Option XTD (Extended View) Not Available Option AMM Not Available Option Sono NT4 Not Available Option 1 - CFM/MCFM opt
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 8-8-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Operation for activating Options 1.) Press the F2 key on the control panel. 2.) In the “Utilities” menu select SYSTEM SETUP to invoke the setup desktop on the screen. 3.) On the left side of the screen select ADMINISTRATION and then click the OPTIONS tab (see: Figure 8-9 on page 8-11.) 8-8-1-1 Operation for installing a “Demo Key” or a “Permanent Key”: 1.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Section 8-9 Replacement of Monitor 8-9-1 Manpower 2 people, 10 minutes 8-9-2 Tools Philips screwdriver, Allen Wrench 8-9-3 Remove Procedure 1.) Open the LCD LID COVER. Refer to the figure below. LCD Cover Figure 8-22 Removing the LCD COVER 2.) Loosen the cable screw. Refer to the figure below.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL WARNING WARNING The monitor arm can spring up and cause serious damage or personal injury while separating the monitor. Hold the monitor arm while separating the monitor. 8-9-4 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-8: Replacement of Monitor - Functional Tests.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-10 Replacement of the MONITOR STAND ARM 8-10-1 Manpower Two persons, 20 minutes 8-10-2 Tools Philips screwdriver, Nipper, Allen Wrench 8-10-3 8-9-3 Removal Procedure 1.) Disassemble the LCD MONITOR. Refer to the section 8-8 Replacement Procedure of MONITOR. 2.) Remove the LCD ARM assembly. Refer to the figure below. 3.) 4.) 8-10-4 Installation Procedure 1.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-11 Replacement of OPIO 8-11-1 Manpower One person, 15 minutes 8-11-2 Tools Philips screwdriver, Nipper, Allen Wrench 8-11-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 8-11-4 Removal Procedure 1.) Remove 2 screws from the OPIO COVER. Refer to the figure below. Figure 8-24 Removing the RUBBER CAPS 2.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 3.) Unscrew 9 screws located in the bottom of the OPIO and lift up the MAIN OPIO. Refer to the figure below. Figure 8-26 Seperating the MAIN OPIO 4.) Lift up the MAIN OPIO and disconnect the cables from the OPIO HUB PWA. Refer to the figure below.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) 5.) Remove the ribbon cable. Ribbon cable 8-11-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-10: Replacement of OPIO - Functional Tests.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-12 Replacement of the SPEAKER 8-12-1 Manpower One person, 30 minutes 8-12-2 Tools Philips screwdriver 8-12-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 8-12-4 Removal Procedure 1.) Remove the MAIN OPIO and SUB OPIO. Refer to the Section 8-11 on page 8-24 OPIO. 2.) Unscrew each 2 screws of SPEAKER ASSY L, R.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 3.) Disconnect the 2 SPEAKER CONNECTORS. Refer to the figure below. Figure 8-30 Disconnecting SPEAKER CONNECTOR 8-12-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-11: Replacement of the SPEAKER - Functional Tests.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Section 8-13 Remove the Pedestal Bottom Cover PEDESTAL BOTTOM COVER PEDESTAL FRONT COVER Figure 8-31 Unscrewing 1 screw 3.) Unscrew 2 screws located in the Pedestal Front Cover and remove the part. Refer to the above figure. 4.) Unscrew 2 screws of POWER CORD FIX BRACKET. Then disconnect POWER CORD. Refer to the figure below.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) 5.) Remove the REAR COVER ASSY. Refer to the figure below.. Figure 8-33 Unscrewing 5 screws and Remove REAR COVER ASSY 6.) Unscrew 7 screws to remove the CABLE COVER BRACKET. Refer to the figure below.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 7.) Untie the CABLE CLAMP with finger. Refer to the figure below.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) 8.) Unscrew 2 screws to remove the MONITOR CABLE BRACKET. Then unscrew the screw of MAIN PANEL LONG CABLE and disconnect the CABLE. Refer to the figure below. MONITOR CABLE BRACKET Figure 8-36 Unscrewing 2 screws and remove the MONITOR CABLE BRACKET 9.) Disconnect other 2 cables. (OPIO PANEL CABLE, VS SPEAKER CABLE ASSY). Refer to the figure below.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-14 Replacement of the NEST 8-14-1 Manpower One person, 10 minutes 8-14-2 Tools Philips screwdriver, Allen Wrench 8-14-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 8-14-4 Removal Procedure 1.) Follow the Replacement procedure of section 8-11-4 (step 13 ~ 19). 2.) Disconnect 2 connectors. (DVD SATA CABLE, DVD POWER CABLE).
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 Table 8-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Cable Connections with FRU information # Cables FRU 1 LCD Monitor cable 5444914 3 OPIO cable 6020400 Peripheral Cable Assembly (ODD SATA Shield Cable) (Dual Printer PWR cable ASSY) 4 (BW Color USB cable ASSY) 5445881 (ODD PWR cable ASSY) (VS Speaker Cable ASSY) 5 Main Power cable see Table 9-16 Power Cord DVD POWER CABLE DVD SATA CABLE Figure 8-39 Disconn
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 3.) Set the 3 RS KNOBS to unlock position to separate from the SIDE COVER R. and unscrew one screw from the each 3 RS KNOB. Refer to the figure below. Figure 8-40 Removing the 3 RS KNOBS 4.) Remove the FRONT COVER ASSY and unscrew 2 screws. Refer to the figure below.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) 5.) Unscrew 2 hexagonal socket screws. Refer to the figure below. Figure 8-42 Unscrewing 2 hexagonal socket screws 6.) Unscrew 7 screws to open the rear bracket. 7.) Pull out the NEST using the 2 handles. Refer to the figure below.. Figure 8-43 Pulling out the NEST 8-14-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-13: Replacement of the NEST - Functional Tests.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-15 Replacement of the AIR FILTER 8-15-1 Manpower One person, 1 minutes 8-15-2 Tools None 8-15-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 8-15-4 Removal Procedure 1.) Pull out the FRONT COVER ASSY. Refer to the figure below. Figure 8-44 Removing the FRONT COVER ASSY 2.) Pull out the AIR FILTER. Refer to the figure below.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 8-15-5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-14: Replacement of the AIR FILTER - Functional Tests.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-16 Replacement of the Probe Holder(KIT) 8-16-1 Manpower One person, 1 minutes 8-16-2 Tools None 8-16-3 Removal Procedure 1.) Pull out the PROBE HOLDER to be replaced. Refer to the figure below. Figure 8-46 Removing the PROBE HOLDER 8-16-4 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Section 8-17 Replacement of SIDE COVER L ASSY 8-17-1 Manpower One person, 5 minutes 8-17-2 Tools Philips Screwdriver 8-17-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 8-17-4 Removal Procedure 1.) Pull out the MID CABINET ASSY. Refer to the figure below. MID CABINET ASSY Figure 8-47 Removing the MID CABINET ASSY 2.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 8-17-5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-16: Replacement of SIDE COVER L ASSY - Functional Tests.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Section 8-18 Replacement of SIDE COVER R ASSY 8-18-1 Manpower One person, 5 minutes 8-18-2 Tools Philips Screwdriver 8-18-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 8-18-4 Removal Procedure 1.) Pull out the MID CABINET ASSY. Refer to the figure below. MID CABINET ASSY Figure 8-49 Removing the MID CABINET ASSY 2.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 3.) Unscrew one screw from the each 3 RS knobs. Refer to the figure below. Figure 8-51 Removing the 3 RS KNOBS 8-18-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-17: Replacement of SIDE COVER R ASSY - Functional Tests.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-19 Replacement of the Trackball 8-19-1 Manpower One person, 10 minutes 8-19-2 Tools Philips Screwdriver 8-19-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 8-19-4 Removal Procedure 1.) Remove 4 PROBE HOLDERS from the OPIO.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 2.) Unscrew 9 screws located in the bottom of the OPIO and lift up the MAIN OPIO. Refer to the figure below. Figure 8-53 Unscrewing 9 screws and lifting up the MAIN OPIO 3.) Lift up the MAIN OPIO and disconnect 3 cables from the OPIO HUB PWA. Refer to the figure below.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 4.) Disconnect the TRACKBALL CABLE from the OPIO HUB PWA. Refer to the figure below. Figure 8-55 Disconnecting the TRACKBALL CABLE 5.) Unscrew 2 screws and then pull out the TRACKBALL from the MAIN OPIO. Refer to the figure below.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-20 Replacement of the TGC BOARD 8-20-1 Manpower One person, 10 minutes 8-20-2 Tools Philips Screwdriver 8-20-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 8-20-4 Removal Procedure 1.) Follow the removal procedure of Section 8-19-4 (step 1~3) 2.) Remove 8 TGC KNOBS. Refer to the figure below.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 3.) Disconnect TGC BOARD CABLE. Refer to the figure below. Figure 8-58 Disconnecting TGC BOARD CABLE 4.) Unscrew 4 screws. then pull out the TGC BOARD. Refer to the figure below. Figure 8-59 Unscrewing 4 screws and pulling out the TGC BOARD 8-20-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-21 Replacement of AN KEYBOARD ASSY 8-21-1 Manpower One person, 15 minutes 8-21-2 Tools Philips Screwdriver 8-21-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 8-21-4 Removal Procedure 1.) Follow the removal procedure of Section 8-19-4 (step 1~3) 2.) Disconnect the 2 HUB TO MAIN OPIO CABLES. Refer to the figure below.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 8-21-5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-20: Replacement of AN KEYBOARD ASSY - Functional Tests.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-22 Replacement of the Caster Cap 8-22-1 Manpower One person, 2 minutes 8-22-2 Tools None 8-22-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. NOTE: 8-22-4 Lock Caster before before removing it. Removal Procedure 1.) Remove the Side Tray by removing the 2 screws below. Refer to the figure below.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-23 Replacement of the CASTER 8-23-1 Manpower One person, 10 minutes 8-23-2 Tools Monkey Wrench 8-23-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. NOTE: 8-23-4 Lock casters before removing them. Removal Procedure 1.) Remove the SIDE TRAY and CASTER CAP as explained above. 2.) Unfasten the NUT. And pull out the CASTER.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Figure 8-64 Unfasten the nut 8-23-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-22: Replacement of the CASTER - Functional Tests.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-24 Replacement of the SIDE TRAY 8-24-1 Manpower One person, 2 minutes 8-24-2 Tools Philips screwdriver 8-24-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 8-24-4 Removal Procedure 1.) Unscrew 2 screws and remove the SIDE TRAY. Refer to the figure below.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-25 Replacement of the REAR HANDLE 8-25-1 Manpower One person, 5 minutes 8-25-2 Tools Philips screwdriver 8-25-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 8-25-4 Removal Procedure 2.) Remove the Pedestal Bottom Cover. 3.) Unscrew 2 screws located in the Pedestal Front Cover and remove the part. Refer to the figure below.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 4.) Unscrew 2 hexagonal Socket screws to remove the REAR HANDLE. Refer to the figure below. Figure 8-67 Unscrewing 2 screws and remove REAR HANDLE 8-25-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-26: Replacement of the REAR HANDLE - Functional Tests.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-26 Replacement of the OP COVER L,R 8-26-1 Manpower One person, 5 minutes 8-26-2 Tools Philips screwdriver 8-26-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 8-26-4 Removal Procedure 1.) Remove 2 RUBBER CAPS from the OPIO COVER. Refer to the figure below. Figure 8-68 Removing the RUBBER CAP 2.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 8-26-5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-27: Replacement of the OP COVER L,R - Functional Tests.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Section 8-27 Replacement of the CPS 8-27-1 Manpower One person, 30 minutes 8-27-2 Tools Philips screwdriver, Allen Wrench 8-27-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 2.) Remove the rear cover as described in Section 8-25 on page 8-55, except for pulling out the NEST. 8-27-4 Removal Procedure CPS Figure 8-70 CPS 1.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 8-27-5 DRAFT (JANUARY 17, 2013) Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-28: Replacement of the CPS - Functional Tests.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-28 Replacement of the RFS 8-28-1 Manpower One person, 30 minutes 8-28-2 Tools Philips screwdriver, Allen Wrench 8-28-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 2.) Remove the rear cover as described in Section 8-25 on page 8-55, except for pulling out the NEST. 8-28-4 Removal Procedure RFS Figure 8-72 RFS 1.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 8-28-5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perfom 8-28-6 Serial Number - Reprogamming Procedure to reprogram the serial number.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Figure 8-75 Hardware ID on label RFS-board 2.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 3.) Program the System Serial number and re-enter the Permanent Key To program the System SNO, click the OPTIONS tab: Field to enter the “Serial Key” Field to enter Permanent key Figure 8-76 System Setup - Administration - Options page a.) Enter the "Serial key" (alphanumeric letters based on system serial number and RFS Hardware ID) in the field "Serial Number" (e.g.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-29 Replacement of the SOM 8-29-1 Manpower One person, 30 minutes 8-29-2 Tools Philips screwdriver, Allen Wrench 8-29-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 2.) Remove the rear cover as described in Section 8-25 on page 8-55, except for pulling out the NEST. 3.) Remove the RFS board as described in 8-28-3 on page 8-61.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 CAUTION DRAFT (JANUARY 17, 2013) When replacing with new SOM, Push the SOM in both side of RFS ASSEMBLY without putting weight on the board. !! CAUTION: Figure 8-78 Pushing the SOM 8-29-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-30: Replacement of the SOM - Functional Tests..
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-30 Replacement of the HDD 8-30-1 Manpower One person, 30 minutes 8-30-2 Tools Philips screwdriver, Allen Wrench 8-30-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 2.) Remove the rear cover as described in Section 8-25 on page 8-55, except for pulling out the NEST. 3.) Remove the RFS board as described in 8-28-3 on page 8-61.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 8-30-5 DRAFT (JANUARY 17, 2013) Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Set up the new hard disk as described in 5-15-5-4-2 on page 5-68. 3.) Perform: FRU8-31: Replacement of the HDD - Functional Tests..
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-31 Replacement of the DVR 8-31-1 Manpower One person, 30 minutes 8-31-2 Tools Philips screwdriver, Allen Wrench 8-31-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 2.) Remove the rear cover as described in Section 8-25 on page 8-55, except for pulling out the NEST. 3.) Remove the RFS board as described in 8-28-3 on page 8-61.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) 2.) Remove the dummy board from the RFS. Figure 8-82 Installing DVR 8-31-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-32: Replacement of the DVR - Functional Tests..
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Section 8-32 Replacement of the BF64 Manpower One person, 30 minutes 8-32-1 Tools Philips screwdriver, Allen Wrench 8-32-2 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 2.) Remove the rear cover as described in Section 8-13 on page 8-29, step 5 and step 6. 8-32-3 Removal Procedure BF/PIR Figure 8-83 BF/PIR 1.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) 2.) Unscrew 6 screws. 3.) Remove BF64 from the EMI Shield cover.. EMI Shield cover Figure 8-85 Unscrewing 6 screws 8-32-4 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-33: Replacement of the BF64 - Functional Tests..
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Section 8-33 Replacement of the PIR 8-33-1 Manpower One person, 30 minutes 8-33-2 Tools Philips screwdriver, Allen Wrench, Hirose ejector 8-33-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 2.) Remove the rear cover as described in Section 8-25 on page 8-55, except for pulling out the NEST. 3.) Remove 3 RS Knobs.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) 2.) Unscrew 6 screws. 3.) Remove PIR from the EMI Shield cover. Figure 8-88 Removing the bracket Figure 8-89 Bottom of PIR 8-33-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-34: Replacement of the PIR - Functional Tests..
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-34 Replacement of the CSI 8-34-1 Manpower One person, 30 minutes 8-34-2 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 8-34-3 Removal Procedure 1.) Unscrew 4 screws to remove the LED Indicator bracket. Figure 8-90 Opening the rear bracket 2.) Unscrew 4 screws to remove CSI board and Separate the LED connector and FAN connector.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 8-34-4 DRAFT (JANUARY 17, 2013) Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-35: Replacement of the CSI - Functional Tests..
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-35 Replacement of the CBP 8-35-1 Manpower One person, 30 minutes 8-35-2 Tools Philips screwdriver, Allen Wrench 8-35-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 2.) Remove the rear cover as described in Section 8-14 on page 8-33. 8-35-4 Removal Procedure 1.) Remove the CSI. Refer to the replacement of CSI. 2.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) 3.) Unscrew 6 screws. Figure 8-93 Unscrewing 6 screws 8-35-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-36: Replacement of the CBP - Functional Tests.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-36 Replacement of the DVD 8-36-1 Manpower One person, 10 minutes 8-36-2 Tools Philips screwdriver 8-36-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 8-36-4 Removal Procedure 1.) Remove FRONT PLASTIC COVER. Refer to the below picture. Figure 8-94 Removing FRONT PLASTIC COVER 2.) Unplug two cables from the DVD.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) 3.) Unscrewing 2 screws to remove the DVD . Figure 8-96 Removing the DVD 8-36-5 Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-37: Replacement of the DVD - Functional Tests..
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-37 Replacement of the OPIO USB Sub PWA 8-37-1 Manpower One person, 30 minutes 8-37-2 Tools Philips screwdriver 8-37-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 2.) Remove the rear cover as described in Section 8-11 on page 8-24. 8-37-4 Removal Procedure 1.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 8-37-5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-39: Replacement of the OPIO USB Sub PWA - Functional Tests.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-38 Replacement of AN KEY TOP 8-38-1 Manpower One person, 30 minutes 8-38-2 Tools Flat head screw driver 8-38-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 8-38-4 Key Caps - Removal Procedure 1.) Remove “1” and “Enter” keys with Tool A. Figure 8-99 removing “1” key and “ Enter” key 2.) Confirm that two screw holes exist.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) Section 8-39 Replacement of FULL AN KEY 8-39-1 Manpower One person, 30 minutes 8-39-2 Tools Philips screwdriver 8-39-3 Removal Procedure 1.) Remove 9 screws.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 2.) Lift up the OPIO Assy. And then, disconnect 3 cables from OPIO ASSY. Figure 8-101 Disconnecting 3 cables from OPIO ASSY 3.) Separate the cable that connects with HUB Board. Figure 8-102 Disconnecting all cables from HUB Board 4.) Remove 12 screws, and separate HUB Board..
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 8-39-4 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Installation Procedure 1.) Install the new parts in the reverse order of removal. 2.) Perform: FRU8-41 Replacement of FULL AN KEY - Functional Tests.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-40 Replacement of the Caps for Hardkeys NOTICE Please observe that replacement procedure depend on key caps that have to be replaced! • if just the circle key caps have to be replaced. • if trackball buttons or mode key slices have to be replaced too, please contact your local distributor or GE service representative.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-41 Replacing optional Peripherals NOTICE Normally auxiliary devices and peripherals come pre-installed with the Voluson® P8 / Voluson® P6 system. !! NOTICE: 8-41-1 Manpower One person, 70 minutes 8-41-2 Tools Standard Field Service Tool Kit NOTICE READ and UNDERSTAND these instructions thoroughly before proceeding with the installation.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 EMI Gaskets (5448934) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) EMI Gaskets (5447303) EMI Gaskets (5457475) Figure 8-105 B/W Printer assembly Chapter 8 - Replacement Procedures 8-89
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) 5.) Disassemble front printer cover assembly 6.) Remove dummy printer bracket assembly Figure 8-106 Disassemble printer cover 7.) Disassemble power cord bracket 8.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) 9.) Unscrew 7 fasteners and disassemble rear EMI cover. 10.)Unscrew 2 fasteners and disassemble printer EMI cover 1 and 2. Unscrew 7 fasteners 2 fasteners and printer EMI covers 1 and 2 Rear side view after removing the EMI covers 1 and 2 Figure 8-108 Disassemble printer EMI covers 11.)Disassemble front cover.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 12.)Assemble Printer assembly. 13.)Push the printer assembly till it makes contact with sheet metal chassis. Printer Assembly Push the printer assembly Front cover Figure 8-110 Assemble printer 14.)Carefully insert the printer base plate in to the chassis slot. 15.)Assemble the printer base plate with chassis using 2 screws (2159634-screw BH M4X10). 16.)Assemble front cover.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) 17.)Insert the printer USB cable and printer power cable (5410088 - printer cable set). 18.)Insert the printer power cable in to the nest box. 19.)Insert the printer USB cable in to the nest box.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) 20.)Check the printer USB cable and power cable to passing through in EMI cover slot. Then fasten 2 screws and assemble printer EMI cover 1 and 2. 21.)Route the B/W printer USB and power cable through the left side cable clamp. Left side clamp EMI cover slot Figure 8-114 Route the printer cables 22.)Fasten the 7 screws and fix the rear EMI cover. 23.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 8-41-3-2 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Color Printer Assembly 1.) Assemble 2 EMI gaskets (5448934 - EMI gasket, side, b-w printer bkt-3) on printer base plate. 2.) Assemble 2 EMI gaskets (5448934 - EMI gasket, side, b-w printer bkt-3) on printer base plate. 3.) Assemble 2 EMI gaskets (5457475 - EMI gasket, side, b-w printer bkt-1) on printer base plate. 4.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 5.) Remove the front printer cover assembly. Figure 8-117 Front cover 6.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) 7.) Unscrew 7 fasteners and disassemble rear EMI cover. 8.) Unscrew 2 fasteners and disassemble printer EMI cover 1 and 2. Unscrew 7 fasteners 2 fasteners and printer EMI covers 1 and 2 Rear side view after removing the EMI covers 1 and 2 Figure 8-119 Disassemble printer EMI covers 9.) Disassemble front cover. 10.)Remove left side printer bracket. 11.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) 13.)Assemble the printer assembly. 14.)Assemble the printer base plate with chassis using 2 screws (2159634 - screw BH M4X10). 2 screws (2159634 -screw BH M4X10) Figure 8-121 Assemble printer base plate 15.)Insert the printer USB cable and printer power cable (5410088 - printer cable set). 16.)Insert the printer power cable in to the nest box.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Printer Power cable in Nest box Printer USB cable Printer Power cable Figure 8-123 Printer cables 17.)Insert the printer USB cable in to the nest box. 18.)Route the color printer USB and power cable through the left side cable clamp.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) 19.)Fasten the 7 screws and fix the rear EMI cover. 20.)Assemble the rear cover, power cord and power cord bracket. Fasten the 7 screws Assemble rear cover, power cord and power cord bracket Figure 8-125 Assemble the rear cover 21.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 22.)Assemble 4 numbers of Holding Catch, female with DVD front cover using 8 screws (2306562 - screw FH M3x6) 23.)Assemble 2 numbers of holding catch male with sheet metal chassis using 4 screws (2306562 - screw FH M3x6) 24.)Assemble DVD front cover plastic with sheet metal chassis. Holding Catch, Male Figure 8-127 Assemble Holding Catch, male and DVD front cover 25.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 8-41-4 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL HORIZONTAL TV PROBE HOLDER Installation 1.) Install HORIZONTAL TV PROBE HOLDER on the left side bottom of OPIO with 3 screws.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 8-41-5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL VERTICAL TV PROBE HOLDER Installation 1.) Install VERTICAL TV PROBE HOLDER BASE on the right side bottom of OPIO with 2 screws. Figure 8-131 Assemble TV PROBE HOLDER BASE with 2 screws (2159634, BH M4x10) 2.) Place the TV PROBE HOLDER RUBBER on the PROBE HOLDER BASE.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-42 Replacement of FAN with Cable ASSY II 8-42-1 Manpower One person, 30 minutes. 8-42-2 Tools Philips screwdriver, Allen Wrench 8-42-3 Preparations 1.) Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 2.) Remove the rear cover as described in Section 8-14 on page 8-33. 8-42-4 Removal Procedure 1.) Unscrew 4 screws and remove fan connector.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 8-43 Functional Checks after FRU replacement (Debrief Guidelines) 1.) Perform required Functional and/or Leakage Current tests based upon the FRU being replaced. 2.) Clarify the scripts to debrief a Service Dispatch using your pole's dispatch tool. Use this script in the Service Comments when debriefing a Service Dispatch.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Section 8-44 Replacement of CWD 8-44-1 Manpower One person, 30 minutes. 8-44-2 Tools Philips screwdriver, Hirose ejector 8-44-3 Preparations Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 8-44-4 Removal Procedure 1.) Unscrew 2 screws and remove POWER CORD FIX BRKT and power cord.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 2.) Unscrew 5 screws and remove REAR COVER. Figure 8-135 Rear cover 3.) Remove the REAR EMI COVER.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) 4.) Unscrew Monitor Cable fix screw with "-" Driver and Disconnect all cable connectors. Peripherals Power Cable Keyboard Cable Monitor Cable Fix Screw Peripherals USB Cable SATA Cable Monitor Cable Speaker Figure 8-137 Cable connectors 5.) Unscrew 9 screws and remove NEST EMI BRKT.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 6.) Pull out BF64/PIR Boards. Figure 8-139 BF64/PIR Boards 7.) Unscrew 6 screws and replace BF64(p/n:5396937) to BF64C Boards which have CWD Circuits. Unscrew 6 screws Figure 8-140 BF64C Board 8-44-5 Installation Procedure 1.) Install the new board assy in the reverse order of removal procedure. 2.) Perform: - Functional Tests .
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) Section 8-45 Replacement of ECG Kit assy 8-45-1 Manpower One person, 10 minutes. 8-45-2 Tools Philips screwdriver 8-45-3 Preparations Power Off/Shutdown the system as described in 4-3-2 on page 4-4. 8-45-4 Removal Procedure 1.) Remove 4 probe holders. 2.) Unscrew 9 numbers of fasteners below the keyboard base.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 3.) Open keyboard assembly. 4.) Remove ECG clamp. Figure 8-142 ECG Clamp 8-45-5 Installation Procedure 1.) Assemble ECG kit on keyboard base. 2.) Fasten ECG kit with Keyboard base using 2 screws (2159625-screw w/SP PH m4x8). Tie the USB cable with tie (2160955- cable-tie, DACT-100A) as shown.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) 3.) Connect the USB cable with Keyboard assembly as shown. 4.) Assemble Keyboard assembly. Figure 8-144 USB cable 5.) Fasten 9 numbers of fasteners below the Keyboard base. 6.) Assemble 4 probe holders.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) 7.) Attach ECG label. ECG label Figure 8-146 ECG label 8.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL FRU8-5: Full Backup (Full System Configuration) - Loading Procedure go to 4-5-4 Load Full System Configuration (Full Backup) FRU8-6: Image Archive - Loading Procedure go to 4-5-6-2 Load Image Archive FRU8-7: Replacement or Activation of Options Functional and/or Leakage Current Test See Section Table 8-1 Debrief Script Proprietary Service Manual, System Function Voluson® P8 /
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL FRU8-10: Replacement of OPIO Functional and/or Leakage Current Test See Section Debrief Script 4-3-1 Power On / Boot Up Proprietary Service Manual, 7-11-1 Console Test Voluson® P8 / Voluson® P6, Table 8-1 4-3-2 Revision Voluson® P8 / Voluson® P6, Section 8-11 Equipment passed all required tests and is ready for use.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL FRU8-14: Replacement of the AIR FILTER Functional and/or Leakage Current Test - N/A FRU8-15: Replacement of the Probe Holder(KIT) Functional and/or Leakage Current Test - N/A FRU8-16: Replacement of SIDE COVER L ASSY Functional and/or Leakage Current Test - N/A FRU8-17: Replacement of SIDE COVER R ASSY Functional and/or Leakage Current Test- N/A FRU8-18: Replacement of the Track
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL FRU8-20: Replacement of AN KEYBOARD ASSY Functional and/or Leakage Current Test See Section Debrief Script 4-3-1 Power On / Boot Up Proprietary Service Manual, 7-11-1 Console Test Voluson® P8 / Voluson® P6, Table 8-1 4-3-2 Revision Voluson® P8 / Voluson® P6, Section 8-21 Equipment passed all required tests and is ready for use.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL FRU8-26: Replacement of the REAR HANDLE Functional and/or Leakage Current Test See Section 4-3-1 Debrief Script Proprietary Service Manual, Power On / Boot Up Voluson® P8 / Voluson® P6, 4-3-2 Revision Voluson® P8 / Voluson® P6, Section 8-25 Equipment passed all required tests and is ready for use.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL FRU8-30: Replacement of the SOM Functional and/or Leakage Current Test See Section 4-3-1 Table 8-1 4-3-2 Debrief Script Power On / Boot Up Proprietary Service Manual, System Function Voluson® P8 / Voluson® P6, Power Off / Shutdown Revision Voluson® P8 / Voluson® P6, Section 8-29 Equipment passed all required tests and is ready for use.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL FRU8-34: Replacement of the PIR Functional and/or Leakage Current Test See Section 4-3-1 Table 8-1 4-3-2 Debrief Script Power On / Boot Up Proprietary Service Manual, System Function Voluson® P8 / Voluson® P6, Power Off / Shutdown Revision Voluson® P8 / Voluson® P6, Section 8-33 Equipment passed all required tests and is ready for use.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL FRU8-37: Replacement of the DVD Functional and/or Leakage Current Test See Section 4-3-1 Table 8-1 Debrief Script Power On / Boot Up Proprietary Service Manual, System Function Voluson® P8 / Voluson® P6, 7-7-6 DVD/CD+R/RW Drive Test 4-3-2 Power Off / Shutdown Revision Voluson® P8 / Voluson® P6, Section 8-36 Equipment passed all required tests and is ready for use.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL FRU8-41 Replacement of FULL AN KEY Functional and/or Leakage Current Test See Section Debrief Script 4-3-1 Power On / Boot Up Proprietary Service Manual, 7-11-1 Console Test Voluson® P8 / Voluson® P6, Table 8-1 4-3-2 Revision Voluson® P8 / Voluson® P6, Section 8-39 Equipment passed all required tests and is ready for use.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) Functional and/or Leakage Current Test See Section VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Debrief Script 4-3-1 Power On / Boot Up Proprietary Service Manual, 10-6-2 Functional Checks Voluson® P8 / Voluson® P6, 4-4-3 Spectral Doppler Mode Checks 4-3-2 Power Off / Shutdown Revision Voluson® P8 / Voluson® P6, Equipment passed all required tests and is ready for use. FRU8-46: Perform: - Functional Tests .
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 8-124 DRAFT (JANUARY 17, 2013) Section 8-45 - Replacement of ECG Kit assy VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Chapter 9 Renewal Parts Section 9-1 Overview 9-1-1 Purpose of Chapter 9 This chapter gives an overview of replacement parts available for the Voluson® P8 / Voluson® P6. .
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 9-2 List of Abbreviations AC ADC A/N Key ASIC Assy BF64 BF128 CBP CPS CPU CSI CWDM DAC DC DSP DVD DVI ECGP EUM FRU 1 FRU 2 HDD Int I/O LCD PCI PWA RFS SOM USB CRU 9-2 Alternating Current Analog to Digital Converter Alpha Numeric Keys Application Specific Integrated Circuit Assembly 64ch Beamformer module 128ch Beamformer module Backplane interface module Common Power Suppl
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 9-3 Parts List Groups Table 9-2 Item Mechanical and user accessible parts Part Group Name Table Number Description 101-108 Plastics Covers Table 9-3 on page 9-4 Housing Covers 201-202 LCD Monitor Table 9-2 on page 9-5 Monitor and its covers 301-311 OPIO Table 9-5 on page 9-7 User interface parts & speaker 401-413 Nest Box Table 9-6 on page 9-9 501-510 M
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Section 9-4 Plastics Covers (Front/Sides/Rear) 104 103 102 101 106 107 107 107 105 Figure 9-1 Table 9-3 Item Part Name Plastic Covers Plastics Covers Part Number Description Qty CRU FRU 101 Side Cover R Assy 5432941 Plastics, RS probe connector Voluson® P8 / Voluson® P6 1 1 102 Side Cover L Assy 5432943 Plastic, w/ ‘cap’ for Voluson® P8 / Voluson® P6 1 1 103
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 9-5 LCD Monitor 201 Figure 9-2 LCD Monitor Table 9-4 Item 201 Part Name 15" LCD Monitor LCD Monitor Part Number 5439256 Description 15" LCD Monitor Chapter 9 - Renewal Parts Qty 1 CRU FRU 1 9-5
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Section 9-6 OPIO 302 305 303 301 304 306 309 307 308 303 310 9-6 Section 9-6 - OPIO
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Figure 9-3 OPIO Table 9-5 Item OPIO Part Name Part Number Description Qty CRU FRU 301 OPIO USB Sub PWA Assy 5461194 OPIO USB Sub PWA Assy 1 1 302 Alphanumeric Keyboard Assy 5459665 Full Alphanumeric keyboard 1 1 303 TGC Assy 5449578 TGC Assembly 1 1 304 OPIO Button Sets 5450806 OPIO BUTTON SETS 1 1 305 Encoder set 5437573 ENCODER SET 1 1 306 O
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 9-7 Nest Box Parts 401 403 402 405 404 406 407 408 409 412 410 411 413 Figure 9-4 Nest Box parts 9-8 Section 9-7 - Nest Box Parts
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 Table 9-6 Item Part Name DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Nest Box Part Number Description Qty CRU FRU 5427038 PIR Assy 1 1 401 PIR Assy 402 BF64 Assy 5396937-2 BF64 Assy 1 1 403 RFS Assy 5364098-2 RFS Assy 1 1 404 SOM w/ 2GB Memory 5324556-3 COM Express MODULE with Intel Core 2 Duo 2.53Ghz, DDR3 2G Memory, Heatsink 1 1 405 2.5 Inch 500 GB SATA HDD 5459671 2.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 9-8 Mechanical Parts Figure 9-5 Mechanical Parts 502 501 503 503 505 504 506 509 508 507 512 511 510 Table 9-7 Item 501 9-10 Part Name Total Lock Caster Mechanical Parts Part Number Description Qty 5400391-2 Castor Total Lock 2 POLARIS 3 Section 9-8 - Mechanical Parts CRU FRU 1
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 Table 9-7 Item Part Name DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Mechanical Parts Part Number Description Qty CRU FRU 5400392-2 Castor Dual Lock 2 POLARIS 1 1 502 Swivel Lock Caster 503 Caster Cap Set 5408087 Caster Cap Set 1 1 504 Rear Handle 5374864 Rear Handle 1 1 505 LCD Arm 5310100 LCD Arm Assy (without cables) 1 1 506 LCD Arm Plastic Covers 5310093 LCD Arm Plastic Covers 1 1 507
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Section 9-9 Options 602 603 601 604 Figure 9-6 Options Table 9-8 Item Options Part Name Part Number Description Qty CRU FRU 601 Vertical TV probe holder 5408092 Vertical TV probe holder 1 1 602 Horizontal TV probe holder 5408514 Horizontal TV probe holder 1 1 603 USB Footswitch, 3 button 5380960 3 button Foot Switch 1 604 DVR Assy 5388734 DVR Assy 1
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 9-10 Optional Peripherals and Accessories 9-10-1 Printers and options 701 701 703 702 705 704 706 Figure 9-7 Optional Peripherals and Accessories - Printers Table 9-9 Item 701 702 Part Name Tetris Software B&W Printer Digital (SONY, UP-D897) Optional Peripherals and Accessories Part Number Description Qty 5438266 Tetris Software - 1 H44492LB (5160406) B&W P
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) Section 9-11 System Manuals Table 9-10 System Manuals - Voluson® P8 / Voluson® P6 Part Name Part # VP8/VP6 5450839-100 Service Manual System User Manuals 9-14 Basic User Manual, English 5454053-100 Basic User Manual, German 5454053-108 Basic User Manual, Spanish 5454053-106 Basic User Manual, Portuguese 5454053-127 Basic User Manual, Italian 5454053-111 Basic User Manual, French 5454053-101 Basic User Manual, Ja
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Section 9-12 Probes Table 9-11 below outlines the replacement parts described in the sub-sections.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 9-12-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL 2D-Probes - Linear Array Transducer 803 Figure 9-9 2D Linear Array Transducer Table 9-13 2D linear array Transducers Item Part Name Part Number Description Qty CRU FRU 803 12L-RS 5154514 Small and lightweight linear-array transducer 4.0 - 10.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 9-12-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Real-Time 4D Volume Probes 806 805 804 Figure 9-10 Real-Time 4D Volume Probes Table 9-14 Real-Time 4D Volume Probes Item Part Name Part Number Description 804 RAB2-6-RS KTZ302893 2.0 - 8.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 9-12-4 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) 2D-Probes - Phased Array Transducers 807 Figure 9-11 2D Phased Array Probe Table 9-15 Item Part Name 807 3Sc-RS 2D-Probes - Phased Array Transducers Part Number 47237516 9-18 (H45041DL) Description Qty CRU FRU Broadband phased array transducer, 1.45-4.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 9-13 Power Cord 903 902 901 906 905 904 907 908 909 910 911 912 913 Figure 9-12 Power Cords Chapter 9 - Renewal Parts 9-19
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 Table 9-16 Item DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Power Cord Part Name Part Number Description Qty CRU FRU 901 Power Cord for India 5182611 Power Cord for India 1 1 1 902 Power Cord for Switzerland 5182235 Power Cord for Switzerland 1 1 1 903 Power Cord for UK/HK 5182816 Power Cord for UK/HK 1 1 1 904 Power Cord for ANZ 5182296 Power Cord for ANZ 1 1 1 905 Power Cord for Denmark
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Section 9-14 Biopsy Needle Guide 1001 1002 1003 1004 Figure 9-13 Biopsy Needle Guides Table 9-17 Biopsy Needle Guides Item Part Name Part Number Description Qty CRU FRU 1001 ***** (disposable) E8385MJ disposable Biopsy needle guide for probe E8C-RS needle diameter: < 1.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 Section 9-14 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Biopsy Needle Guide (cont’d) 1005 1006 1007 1008 1009 1010 Figure 9-14 Biopsy Needle Guides 9-22 Section 9-14 - Biopsy Needle Guide
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 Table 9-18 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Biopsy Needle Guides Item Part Name Part Number Description Qty CRU FRU 1005 PEC74 H48621Y reusable Biopsy needle guide for probe RAB4-8-RS, RAB2-5-RS needle diameter: < 1 mm, 1.4 mm, 2.2 mm - 1 N bracket H48681ML Non Sterile Single Angle Bracket needle guide for probe RAB2-6-RS needle diameter: > 0.6 mm - < 2.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 9-24 DRAFT (JANUARY 17, 2013) Section 9-14 - Biopsy Needle Guide VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Chapter 10 Care & Maintenance Section 10-1 Overview 10-1-1 Periodic Maintenance Inspections It has been determined by engineering that your Voluson® P8 / Voluson® P6 system does not have any high wear components that fail with use, therefore no Periodic Maintenance Inspections are mandatory.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL CAUTION DO NOT operate this unit unless all board covers and frame panels are securely in place, to ensure optimal system performance and cooling. When covers are removed, EMI may be present. !! CAUTION: Section 10-2 Why do Maintenance 10-2-1 Keeping Records It is good business practice that ultrasound facilities maintain records of quality checks and corrective maintenance.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 10-3-1 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) How often should care & maintenance tasks be performed? (cont’d) Table 10-2 Customer Care Schedule Item AC Mains Cable Service at Indicated Time D W M • Inspect AC Mains Cable Check if all cables are fixed well seated at Cables and Connectors the correct position and if there is no mechanical damage visible.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 Table 10-2 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Customer Care Schedule Item A Notes Endocavity Probe Leakage Current Checks • Also after corrective maintenance or as required by your facilities QA program. Measurement Accuracy Checks • Also after corrective maintenance or as required by your facilities QA program. • Also after corrective maintenance or as required by your facilities QA program.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Section 10-4 Tools Required Table 10-3 Overview of GE-1 Tool Kit Contents Tool ID Description Tool ID Description 9-45358 Pliers Retaining Ring 9-XL9971MM Xcelite-hex Blade 1.27mm 9-4078 Scribe 9-XL9972MM Xcelite-hex Blade 1.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 Table 10-3 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Overview of GE-1 Tool Kit Contents (Continued) Tool ID Description Tool ID Description 9-46696 Hex Keys 9-XL100X Wire Stripper/Cutter 5 inch - 100X 9-39829 Torpedo Level, Magnetic 9-XL87CG Pliers - very fine needle nose-87CG 9-38461 Hammer, Ball Peen, 4 oz 9-WEWDT-07 Weller-Soldering-Replacement Tip(1) 9-4280 Universal Joint 1/4 inch 9-WS175-E Wiss - S
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 10-5 Tools Required 10-5-1 10-5-1-1 Special Tools, Supplies and Equipment Specific Requirements for Care & Maintenance Table 10-6 Overview of Requirements for Care & Maintenance Tool Part Number Digital Volt Meter (DVM) Comments minimum 5% accuracy, 3.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 10-6 System Maintenance 10-6-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. Refer to the system user documentation whenever necessary. Table 10-7 10-8 System Checks Step Item Description 1 Ask & Listen 2 Paperwork Fill in the top of the Ultrasound Inspection Certificate (see: page 10-32). Note all probes and system options.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 10-6-2 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Functional Checks The functional checks take about 60 minutes to perform. Refer to the Voluson® P8 / Voluson® P6 Basic User Manual whenever necessary. 10-6-2-1 System Checks Table 10-8 NOTE: System Functional Checks Step Item Description 1 B Mode Verify basic B Mode (2D) operation. Check the basic system controls that affect this mode of operation.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 10-6-2-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted. Refer to Table 3-8, “Approved Peripherals,” on page 3-60 for a list of approved peripherals. Table 10-9 Step Item 1 B/W Printer Verify hardcopy output of the B/W video page printer. Clean heads and covers if necessary.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 10-6-5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Physical Inspection Table 10-12 Physical Checks Step Item Description 1 Labeling Verify that all system labeling is present and in readable condition. 2 Scratches & Dents Inspect the console for dents, scratches or cracks. 3 LCD Monitor Display Inspect the LCD Monitor Display for scratches and raster burns.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 10-6-7 10-6-7-1 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Probe Maintenance Probe Related Checks Table 10-13 Probe Related Checks 10-6-7-2 Step Item 1 Probes 2 Probe Holder Description Thoroughly check the system probe connectors and remove dust from inside the connector sockets if necessary. Visually check for bent, damaged or missing pins. Clean probe holders (they may need to be soaked to remove excess gel).
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 10-6-7-4 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Disinfection by means of the RIC-Holder Especially for Real-time 4D endocavity probes (RIC), it is necessary to take extreme care when transporting the system with the probe attached, or during the disinfection process. Inadequate handling may lead to dead elements, shocked head mechanics, etc.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 10-7 Using a Phantom (Optional) The use of a Phantom is not required during Preventive maintenance. Customer may use it as part of their Quality Assurance Program tests. Section 10-8 Electrical Safety Tests 10-8-1 Safety Test Overview The electrical safety tests in this section are based on and conform to IEC60601-1 standard including national deviations.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 10-8-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL GEHC Leakage Current Limits for Voluson® P8 / Voluson® P6 The following limits are summarized for IEC60601-1 standard including national deviations and IEC 62353 Medical Electrical Equipment - Recurrent test and test after repair of medical electrical equipment. Measurement limits per IEC60601-1 standard including national deviations.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Country Conditions 10-16 USA Others Close Protective earth and close neutral with reverse polarity 0.01 mA* 0.01 mA*** Open Protective earth and close neutral with normal polarity 0.05 mA* 0.05 mA*** Open Protective earth and close neutral with reverse polarity 0.05 mA* 0.05 mA*** Open neutral and close Protective earth with normal polarity 0.05 mA* 0.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Table 10-18 Type CF Patient Auxiliary Limits Country Conditions USA Others Close Protective earth and close neutral with normal polarity 0.01 mA* 0.01 mA*** Close Protective earth and close neutral with reverse polarity 0.01 mA* 0.01 mA*** Open Protective earth and close neutral with normal polarity 0.05 mA* 0.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 NOTE: DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL * Measurement limits per UL 60601-1 Medical Electrical Equipment Safety Standards Table IV ** Measurement limits per UL 60601-1 Medical Electrical Equipment Safety Standards, Table 19.5DV.1 *** Measurement limits per IEC 60601-1 Ed 2.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 10-8-3 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Outlet Test - Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated. Any problems found should be reported to the hospital immediately and the receptacle should not be used.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 10-8-4 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Grounding Continuity CAUTION Electric Shock Hazard! The patient or operator MUST NOT come into contact with the equipment during this test !! CAUTION: Measure the resistance from the third pin of the attachment plug to the exposed protectively-earthed metal parts of the case. The ground wire resistance should be less than 0.2 ohms.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 10-8-5 10-8-5-1 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Enclosure Leakage Current Test Definition This test measures the current that would flow in a grounded person who touched accessible nonprotectively-earthed parts of equipment if the ground wire were to break. The test verifies the isolation of the power line from the chassis.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL ii)Others Hot : Brown Neutral : Blue Ground : Green-yellow Electrical leakage testing may be accomplished with any calibrated Electrical Safety Analyzer tool compliant with AAMI/ESI 1993 or IEC 60601 or AS/NZS 3551. When using a test instrument, its power plug may be inserted into the wall outlet and the equipment under test is plugged into the receptacle on the panel of the meter.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 10-8-5-3 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Data Sheet for Enclosure Leakage Current The test passes when all readings measure less than the value shown in Table 10-14. Record all data on the Ultrasound Inspection Certificate.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 10-8-6-2 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Generic Procedure The testing meter is connected from accessible protectively-earthed parts of the equipment to ground. Measurements should be made with power line polarity normal and reversed, and with the neutral open and closed. Record the highest reading of current.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Table 10-24 Typical Data Sheet for Earth Leakage Current Tester Neutral Tester Polarity Switch Measured Earth Leakage Current Enter Name of tested peripheral here 10-8-7 10-8-7-1 Closed Normal Closed Reversed Open Normal Open Reversed Type CF Patient Leakage Current Test - ECG Leads Definition This test measures the current which would flow to ground from any of the i
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Neutral : White Ground : Green or Green-yellow ii) Others Hot : Brown Neutral : Blue Ground : Green-yellow Table 10-25 Typical Data Sheet for Type CF Patient Leakage Current Test - ECG Leads 10-8-8 10-8-8-1 Tester Neutral Tester Polarity Switch Ground Switch Closed Normal Closed Closed Normal Open Closed Reversed Closed Closed Reversed Open Open Normal Closed Ope
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 10-8-8-2 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Generic Procedure on Leakage Current Measurements should be made with the ground open and closed, with power line polarity normal and reversed, and with neutral open and closed. For each combination, the probe must be active to find the worst case condition.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) Table 10-27 Typical Data Sheet for Type BF Patient Auxiliary Current Test 10-8-10 Tester Neutral Tester Polarity Switch Ground Switch Closed Normal Closed Closed Normal Open Closed Reversed Closed Closed Reversed Open Open Normal Closed Open Reversed Closed Measured Leakage Current Type BF Patient Leakage current - Mains to Probe Reference the procedure in th
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 10-8-12 10-8-12-1 DRAFT (JANUARY 17, 2013) Tester Neutral Tester Polarity Switch Ground Switch Closed Reversed Closed VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Measured Leakage Current Total Patient leakage current : Type CF probe to ground(Only for Europe) Definition This test measures the current that would flow to ground from all ECG leads through a patient who is being scanned and becomes grounded by touching some other grounded surface
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 10-8-13 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Total Patient leakage current : Mains to Type CF probe (Only for Europe) Reference the procedure in the IEC 60601-1 Ed 3. Measure leakage current flow from mains to all ECG leads connected together. DANGER Electric Shock Hazard. Line voltage is applied to the ECG leads during this test.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Section 10-9 When There's Too Much Leakage Current... 10-9-1 Earth and/or Enclosure Fails Follow the below checklist in case of earth and/or enclosure leakage current failure. NOTE: 10-9-2 • Confirm the ground of the power cord and plug for continuity. • Ensure the ground is not broken, frayed or intermittent. If that, replace any defective part. • Tighten all grounds.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL ULTRASOUND INSPECTION CERTIFICATE Customer Name: System ID: Dispatch Number / Date Performed: Warranty/Contract/HBS System Type Model Number: Serial Number: Manufacture Date: Probe 1: Frequency: Scan Format*: Model Number: Serial Number: Probe 2: Frequency: Scan Format*: Model Number: Serial Number: Probe 3: Frequency: Scan Format*: Model Number: Serial Number:
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL DRAFT (JANUARY 17, 2013) ELECTRICAL SAFETY Outlet and Ground Continuity test Outlet/Ground Continuity Max Value Allowed Value Measured Outlet (correct ground &wiring config.) N/A N/A System Ground Continuity 0.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Type CF Patient Leakage Current Test - ECG Leads ECG lead 1 Tester Neutral Tester Polarity Switch Ground Switch Max Value Allowed Closed Normal Closed 0.01mA Closed Normal Open 0.05mA Closed Reversed Closed 0.01mA Closed Reversed Open 0.05mA Open Normal Closed 0.05mA Open Reversed Closed 0.
GE HEALTHCAREDRAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Type BF Patient Leakage Current Test - Probes (This test should be done for all probe) Tester Neutral Tester Polarity Switch Ground Switch Max Value Allowed Closed Normal Closed 0.1mA Closed Normal Open 0.5mA Closed Reversed Closed 0.1mA Closed Reversed Open 0.5mA Open Normal Closed 0.5mA Open Reversed Closed 0.
GE HEALTHCARERAFT DIRECTION 5450839-100, REVISION 5 DRAFT (JANUARY 17, 2013) VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Total Patient Leakage Current - Type CF probe to Ground Tester Neutral Tester Polarity Switch Ground Switch Max Value Allowed Closed Normal Closed 0.05mA Closed Normal Open 0.1mA Closed Reversed Closed 0.05mA Closed Reversed Open 0.1mA Open Normal Closed 0.1mA Open Reversed Closed 0.
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 Numerics 2D Mode Checks Functional Checks, 4-8 A,B ACDC - primary DC Power Supply Theory, 5-44 ACFE - Primary Power Module Theory, 5-38 Acoustic Noise Output, 2-5 Activation of Options, 8-19 Additional Operating Modes Functional Checks, 4-11 Adjustment LCD Monitor, 6-2 Printer Settings, 3-46 Administration of Full Backup data, 4-40 Air Filter Replacement Procedure, 8-37 AN Keyboard assy Replacement Procedure, 8-49 Archiving Images, 4-44 Auxiliary Devices Con
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL Electromagnetic Compatibility, 1-15 EMI Protection, 3-8 Erasing DVD/CD, 4-33 Error Messages, 7-30 External I/O Theory, 5-36 External USB-Devices Disconnection, 3-21 7-13 How to use the Auto Tester, 7-16 Human Safety, 1-7 I F FMI from DVD Installation Procedure, 8-12 Theory, 5-53 FrontEnd BF128/BF64 - Beamformer Board, 5-22 PIR - Probe Interface board, 5-21 RFS - Radio Frequ.
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 Network IP Address Configuration, 3-70 O OPIO Sub PWA Replacement Procedure, 8-81 Options Activation, 8-19 Replacement Procedure, 8-19 P Patient Archive (Image Management) Functional Checks, 4-32 Peripherals Disk Drive Module (GEM), 5-37 Replacement Procedure, 8-88 Physical Dimensions, 2-5 PIR Replacement Procedure, 8-73 Power Distribution, 5-38 4D Module, 5-44 ACDC - primary DC Power Supply, 5-44 ACFE - Primary Power Supply, 5-38 DCDC - main DC Power Supply
GE HEALTHCARE DIRECTION 5450839-100, REVISION 5 VOLUSON® P8 / VOLUSON® P6 BASIC SERVICE MANUAL How to change Language, 3-56 How to change the EUM Language, 3-56 How to configure Service Platform, 3-58 How to enter Date and Time, 3-56 How to enter Hospital Name, 3-56 System Maintenance, 10-8 System Manufacturer, 1-16 System Software Installation/Upgrade, 8-4, , 8-8 System Software (FMI from DVD) Installation, 8-12 T Tech Tip Storing Sonoview images to VE8, 7-29 TGC Board Replacement Procedure, 8-47 Theory
GE HEALTHCARE DIRECTION 5450839-100, REVISION 4 BASICSERVICE MANUAL - 1-1
GE HEALTHCARE DIRECTION 5450839-100, REVISION 4 BASICSERVICE MANUAL © Voluson® P8 / Voluson® P6 GE Healthcare 65-1, Sangdaewon-dong, Jungwon-gu, Seongnam-si, Gyeonggi-do 462-120 Korea www.gehealthcare.