Troubleshooting Guide

FHR Gaps Troubleshooting Guide
SKIN PREPARATION GOOD
REPEAT SKIN PREP
REPEAT SKIN PREP
DO NOT BYPASS
2.4 "/6cm
1"/2-3cm
1"/2-3cm
FHR
MHR
UA
1
The patient is ambulating
POSSIBLE CAUSE ACTION & SOLUTION
1
HELP/TIP
Electrode ‘detached’ or makes
poor contact with skin
SKIN PREPARATION GOOD
REPEAT SKIN PREP
REPEAT SKIN PREP
DO NOT BYPASS
2.4 "/6cm
1"/2-3cm
1"/2-3cm
FHR
MHR
UA
2
Patient position/posture
SKIN PREPARATION GOOD
REPEAT SKIN PREP
REPEAT SKIN PREP
DO NOT BYPASS
2.4 "/6cm
1"/2-3cm
1"/2-3cm
FHR
MHR
UA
3
SKIN PREPARATION GOOD
REPEAT SKIN PREP
REPEAT SKIN PREP
DO NOT BYPASS
2 .4"/6cm
1"/2-3cm
1"/2-3cm
FHR
MHR
UA
4
Lost MHR and UA
as well as FHR
SKIN PREPARATION GOOD
REPEAT SKIN PREP
REPEAT SKIN PREP
DO NOT BYPASS
2.4"/6cm
1"/2-3cm
1"/2-3cm
FHR
MHR
UA
5
Bypassed red X on
electrode/skin checkscreen
l Return patient to bed
l Consider using a maternity belt to support pannus during
ambulation or upright position
a) Allow 1015 minutes of monitoring before starting ambulation.
b) The patient should not be encouraged to ambulate unless the FHR
trace is good and the signal indicator on the Novii Interface shows
3 green squares.
l Use a pillow behind back or head
to make patient more comfortable
l Return patient to a position where
Novii worked well
l Encourage patient to lie on left or right
l If patient on side, support abdomen with
a pillow/rolled blanket to re-position
abdomen so that the Patch is centred
over the uterus
l Check electrodes and move or stick back down
l Re-position Patch or electrode to avoid the skin problem
a) The Interface will alert user with a visual message, but only when electrode is fully detached.
b) Check electrodes after a shower, clinical procedure, ambulation or position change.
c) Electrode(s) should not be placed over a skin lesion, skin fold, umbilicus, stretch mark,
pronounced linea nigra.
d) If necessary use a strip of micropore tape to prevent electrode lifting or detachment.
l Check Interface for help message
l Interface is o
l Exfoliate skin under ‘bad’ electrode
l Restart the monitoring session if location of ‘bad’ electrode not known
SKIN PREPARATION GOOD
REPEAT SKIN PREP
REPEAT SKIN PREP
DO NOT BYPASS
2.4 "/6cm
1"/2-3cm
1"/2-3cm
FHR
MHR
UA
a) If Interface has lost power Bluetooth pairing with POD is lost.
Remove POD from Patch, place in charging bay and start new monitoring
episode when power is returned.
b) Interface does not have a battery back-up.
l Start-Screen is displayed
POD has switched o – remove POD from Patch and place in charging well.
Start new monitoring session with other POD.
a) Peel the X electrode back, remove excess gel from skin and inish with alcohol wipe.
Wait until skin is dry then exfoliate skin and reapply electrode.
b) May need to use micropore tape to hold in place.
Remove POD from Patch, place in charging bay and start new monitoring episode.
HELP/TIP
HELP/TIP
HELP/TIP
HELP/TIP
HELP/TIP
l Faulty ‘Y’ Connector - replace
Signal