DESPERADOES L-5160 Operation Manual @ Before using the L-5160 you must first thoroughly read this manual. @ Remember to keep this operation manual in an easily accessible place near the unit for future reference.
Important Information Only a physician or a person under the guidance of a physician can use this product. +The information contained in this document is subject to change without notice due to improvement in the equipment. Federal law restricts this device to sale by or on the order of a physician. Users are advised to periodically contact the FCC or specified frequency coordinator and determine if other or your transmitter frequencies that may cause interference.
TELEMETRY PRECAUTIONS For proper management of the telemetry installation, consult your Fukuoka Dens hi representative concerning the following: Plan the installation of your telemetry system taking into account your entire medical facility needs and plant requirements. ¢ Be sure the antenna system installed meets the facility and plant requirements. This radio frequency device is susceptible to interference from outside sources. Interference may prevent the monitoring of patients connected to this devices.
ii Thank you for purchasing this instrument from Fukuoka Dens hi. Familiarize yourself with the correct operation of this instrument prior to operation. PRECAUTIONS FOR SAFE OPERATION AND HAZARD PREVENTION Installation and Storage Install or store the instrument where it is free from moisture, splashed water, excessive dust, salt and direct sunlight. * Avoid excessive vibration and shock during operation or transport.
CLEANING AND DISINFECTION Electrodes and electrode cables For disinfection, the electrodes and electrode cables should be rubbed with a swab or cloth moistened with a formaldehyde solution, such as Cid ex 5%. Under no circumstances may the electrode cables be immersed in any cleaning fluid, nor may they be subjected to heat sterilization with water, steam, or air, or be subjected to ether sterilization. Transmitter The transmitter cabinet may be cleaned and disinfected in the following manner.
iv UNION Some pacemaker pulses are difficult or not possible to detect. This is dependent on the amplitude and width of the pacemaker pulse in addition te the type of pacemaker and lead type used (uni polar, bipolar, etc.). QRS detection allows for detection of low amplitude ECG. But if excessive artifact is present on the ECG waveform, the noise may be detected as QRS in error. Make sure each receiving telemetry channel corresponds to that of the transmitter worn by the patient.
CONTENTS . GENERAL DESCRIPTION PFE PS AAR DEB ET ABS NNT EEA ERGO GD 1 . CONTROLS AND INDICATORS cr tcrsvsorcvnscconsacnnns 2 , PREPARATION & OPERATION carthorse Rabat cane Ann Loading Battery Fee TMA TY NHR m TE EAE Yeas Een 4 2) Turn the power switch to "ON" TTY ANA LS NT Yuma 5 3) Attaching Electrodes SRM NN TAI YN AEN Erase xed a 6 4) Connecting lead wires TEI SE SE SER 9 5) Connect the lead wire set firmly to the transmitter 9 . SET THE GROUP CODE Tyree raa Tae era une urea 10 .
1. GENERAL DESCRIPTION The L-5160 is a radio telemetry transmitter designed for monitoring the ECG and Respiration. The transmitter will operate 5 continuous days from one "AA" size alkaline battery. On the top panel, battery and electrodes status displays are provided. The L-5160 is transmitting a digitized code that includes the transmitting channel number and group codes to prevent interference by other radio apparatus. Read the receiver and display unit operation manual before using this transmitter.
2. CONTROLS AND INDICATORS (Decreasing.
(Decreasing. Input Connector Connect the accessory patient cable. (@Poor Electrode Display This lamp will light for two minutes when the electrodes are making poor contact or the patient cables are broken or disconnected. Display When the power switch is turned on, the battery check lamp will light for about 10 seconds. If the battery is low, the lamp will not light after turning on the power switch. The battery should be replaced with a new one. Also, when the Event switch is pressed, this lamp will light.
3. PREPARATION & OPERATION Loading a battery The L-5160 uses one "AA" size alkaline cell (LR-6) for its power source. When installing a battery taking note of the polarity. To close the battery compartment, align the mark (yellow) on the cap and the mark (yellow) on the lower part of the transmitter, then gently push in and turn the cap clockwise to the mark (white) on the transmitter to align the dot-marks (white & yellow) with each other.
Turn the power switch to "ON" Turn the power switch, which is located at the bottom of the transmitter, on the battery cap, to the "ON" position. The battery life will be approximately 5 days of continuous operation. Do not use in high humidity or in areas of high oxygen concentration.
Attaching the electrodes Wipe the skin at the electrode site with alcohol. Then detach the electrode from the electrode mount and attach it to the electrode site. Color coding of the electrode positions are shown below. Adhesive tape should be used to tape the electrode wire to the skin to prevent the lead wire from moving on the electrode. With the L-5160 impedance measurements are used for respiration monitoring.
For two lead ECG and respiration monitoring, use the CM-85C for AHA color code or CM-86C for IEC color code patient cable (four electrodes).
For dual channel ECG and respiration monitoring, use the CM-85D for AHA color code or CM-86D for IEC color code patient cable (five electrodes).
Connect the lead wires to the electrode. Connect the tip of lead wire to the center of the electrode and gently swing it right and left as shown below. Connect the patient cable firmly to the ECG/Resp. input connector of the transmitter. When disconnecting the patient cable, do not disconnect the lead wire set by pulling on the wires. SON . Confirm the direction of the keyed plug to match the transmitter’s guide key on the connector.
10 4. SET THE GROUP CODE The L-5160 transmits a digitized code which includes the transmitting channel number and group code to prevent interference from other radio apparatus or a neighboring hospital's transmitter. There are 64 group codes. Zero is set for factory adjustment.
5. SELECT THE CORRESPONDING RECEIVER CHANNEL Select the receiver channel at the patient monitor to correspond to the telemetry transmitter. The channel number will be displayed on the screen display. If the receiver channel and transmitter channel do not match, the monitor will display a caution and the unique waveform shown below. This function will prevent telemetry channel interference from other transmitters or external sources.
12 6. CLEANING AND DISINFECTION Clean the transmitter and patient cable with gauze or sanitary cotton dampened with alcohol or inert soap. Be sure not to get cleaning liquid into the patient lead connector and battery compartment. Do not use cleaner containing organic solution, thinners, toluene, benzene. Do not autoclave or heat the unit and patient cable above 60°C.
7. MAINTENANCE AND INSPECTION Stems in this section include routine daily and periodic checks of the equipment to ensure it is operating properly. It is recommended that to maintain the safety and reliability of functions and performance of the unit, the daily and periodical checks given in this section be followed. Do not open the unit or attempt service. Refer service to Fukuoka Dens hi.
Daily Check List No. Date Checker Installation Place Unit L-5160 S/Ne. Purchase Date Items Details of the Check Criteria Judgment Visually check for any No abnormality. [1 OK damage, cracks, chink, chips 0 NG Appearance and peeled nameplate on the outer enclosure. Battery Visually check the connecting | No spring transformed, [0 OK Compartmentalizing on the inside of the deformed or rusted. 0 NG men device and battery cap. Turn ON/OFF power to verify | With battery in, the Bart.
8. SPECIFICATIONS Parameters ECG input impedance ECG max. input range ECG freq. response ECG time constant Resp. measurement Resp. max. input range Resp. meas. electrodes Resp. freq. response Defibrillator protection Status information Transmission freq.
9. ACCESSORIES HM Standard Accessories + Patient cable : One CM-85B or CM-86B, three electrodes type or CM-85C or CM-86C, four electrodes type. The type of the cable depends on the destination. Pouch : Two AB-95, disposable pouch Security Strap : One OB-25 belt kit Battery : One LR-6, 1.
HM Optional Accessories *CM-85B or CM-86B, Patient cable for single ECG lead and respiration. *CM-85C or CM-86C, Patient cable for two ECG lead I, I and respiration. *CM-85D or CM-86D, Patient cable for dual ECG for any two channel differential ECG and Respiration, five electrode type.