Bidirectional Wireless Communications Module (TCON) HTC-702 Instruction Manual ● Before using this device, read this “Instruction Manual” thoroughly. ● After reading, keep this manual near the device for future reference.
CAUTION Federal Law restricts this device to sale by or on the order of a physician. CAUTION: • The company and product names used in this manual are trademarks or registered trademarks. • If this manual has pages missing or out of order, contact Fukuda Denshi for replacement. • Only physician or persons instructed by physicians are allowed to use the equipment. • The information contained in this document is subject to change without notice due to improvement in the equipment.
Thank you for purchasing this product. Before using this product, please read through this “Instruction Manual” to make sure the product is used correctly and safely. Even if you are familiar with our devices, there are certain handling procedures that are unique to this device. Therefore, please make sure you read this manual thoroughly before use. Safety Precautions The safety precautions shown in this manual contain important details on the safe use of this product, and must be obeyed.
Precautions for Safety and Hazard Prevention This section contains general information on how to handle this device for safety use. Precautions unique to this device are detailed throughout the manual. CAUTION 1. Users should have a thorough knowledge of the operation before using this device. 2. Pay attention to the following when installing and storing the device. Do not install or store in an area where the equipment will be subject to splashing water.
CAUTION 6. If the device is damaged and in need of repair, users should not attempt to service the device themselves. Label the unit “OUT OF ORDER”, and contact Fukuda Denshi representative. 7. Do not disassemble or remodel the device. 8. Maintenance Check Make sure to periodically check the device and accessories. Before reusing a device that has been left unused for a while, make sure that the device works normally and safely.
Precautions for Use of Bidirectional Wireless Communications (TCON) CAUTION ■To operate the device correctly, read the following precautions carefully. The medical institution (hereinafter referred to as “Institution”) must execute investigation required to prevent interference including types of radio waves, frequencies, and antenna power if wireless equipment is already installed and being used in the facility.
CAUTION ■Precautions for Use The Bidirectional Wireless Communications Module (TCON) uses radio waves to transmit data. Therefore, the necessary precautions need to be taken for the characteristics and difficulties of using the device that emits radio waves. The TCON user should fully understand these precautions beforehand, and use the TCON device safely. Furthermore, situations in which interference may occur are outlined below.
FCC Radiation Exposure Statement WARNING This transmitter must not be co-located or operated in conjunction with any other antenna or transmitter. This equipment complies with FCC radiation exposure limits set forth for an controlled environment and meets the FCC radio frequency (RF) Exposure Guidelines in Supplement C to OET65. This equipment has very low levels of RF energy that it deemed to comply without maximum permissive exposure evaluation (MPE).
Contents Safety Precautions...................................................................................... i Precautions for Safety and Hazard Prevention....................................... ii Precautions for Use of Bidirectional Wireless Communications (TCON) ....................................................................................................... iv FCC Radiation Exposure Statement ....................................................... vi Contents..................................
Blank Page viii
1. Overview This device is a Bidirectional Wireless Communications Module (HTC-702) used for wirelessly communicating vital signs data monitored by the DS-7000 series patient monitor. This device can be used in combination with a wireless central monitor such as the DS-7600 to construct a patient monitoring system with a medical telemetry and the Bidirectional Wireless Communications (TCON) system. For details on compatibility with other patient monitors, contact Fukuda Denshi representative.
1.
2. Names of Parts and Their Functions ■HTC-702 (1) Antenna (3) Rating Label (2) Connecting Plug (1) Antenna Used to send and receive radio waves. (2) Connecting Plug Used to connect the HTC-702 with the patient monitor. Connect to the “COM1 to 3 port” or “F-LINK” connector on the patient monitor.
2.
3. Installation to the Patient Monitor CAUTION Before starting the installation work, switch the patient monitor main power switch OFF, and remove the power plug from the socket. ■Example Installation of HTC-702 to Bedside Monitor, DS-7100 1 On the rear side of the monitor unit (DS-7100 series), fix the OA-471 (Optional TOCN Installation Bracket for Patient Monitor) in place in 2 locations using the screws (small), which are included in the OA-471 accessories.
3. Installation to the Patient Monitor 3 Attach the Connecting Plug of the HTC-702 cable to the Serial Connector on the right side of the monitor unit. Attach the Cord Clamp, which is included in the HTC-702 accessories, to the monitor unit, and pass the connecting cable through the clamp. Serial Connector Cord Clamp 4 Switch the DS-7100 power on, and set the TCON channel and ID in the TCON Setup screen. For details on the setting procedure, see the DS-7100 operation manual.
3. Installation to the Patient Monitor ■Example Installation of HTC-702 to Central Monitor, DS-7600 NOTE The recommended connection procedure for the HTC-702 and DS-7600 is to use the optional OA-702 and CJ-702 to install the HTC-702 in a high location on a wall or on the ceiling in the corridor. However, if the DS-7600 is to be moved during use, install it to the rear side of the DS-7600. In this case, the TCON communication range is smaller than when installed on the ceiling or wall.
3. Installation to the Patient Monitor 2 Attach the Connecting Plug of the HTC-702 cable to the Serial Connector on the left side of the DS-7600. Attach the Cord Camp, which is included in the HTC-702 accessories, to the DS7600, and pass the connecting cord through the clamp. Serial Connector Cord Clamp 3 This completes the installation. 4 Switch the DS-7600 power on, and set the TCON channel and ID in the TCON Setup screen. For details on the setting procedure, see the DS-7600 operation manual.
4. System Configuration Example TCON Bidirectional Wireless Communications (Generally within 60m) (Nurse Station) HTC-702 (TCON) Medical Telemetry Wireless (Generally within 10m) Ceiling antenna for telemetry system Note 1) In order to improve the communication status of TCON, Fukuda Denshi recommends using the optional extension cable to install the HTC-702 to a good location near the ceiling.
4. System Configuration Example NOTE If the TCON is installed in a location that exceeds the TCON communication range (within 60m) in the TCON system on the previous page, it is possible to use the system by installing a “TCON Repeater”. Certain installation conditions and limitations apply to the “TCON Repeater”. For details, contact Fukuda Denshi representative.
5. Cleaning, Disinfection, and Sterilization Clean, disinfect, and sterilize the module unit and connecting cords as shown below. ■Cleaning Wipe using gauze or absorbent cotton etc. that has been soaked in alcohol, or a weak acidic, weak alkaline, or neutral detergent and wrung. At this time, make sure the chemicals do not enter inside the connectors or main unit.
5.
6. Maintenance Check In order to use this equipment safely and to maintain the functions, performance and reliability of this equipment, the “Daily Check” and “Periodic Inspection” must be performed. Fukuda Denshi is not liable for accidents arising from lack of maintenance. CAUTION Do not open the main unit housing. Make sure that alcohol and other liquids do not enter inside the device. ■Daily Check Perform the following daily check according to the “Daily Check List” on the next page.
6. Maintenance Check Daily Check List No. Checked Date Checked by Installation Location Device Type HTC-702 Serial Number Date of Purchase Item Check Details Visually inspect for damage, cracks, breaks, chips, loose label, loose screws, installation External appearance condition or bending of the connecting connector and antenna. Connect to the monitor, operate under normal operating conditions, and check the Functions communication function and operation.
7. Specifications Communication Parameters (Depends on the patient monitor used) : Bedside monitor measurement values, alarm limits, NIBP control, etc. Communication Channel : Max.
7.
8. Accessories In order to satisfy product performance requirements, always use the accessories specified by Fukuda Denshi. ■Standard Accessories No.
8. Accessories ■Optional Accessories The following items are available as optional accessories. Purchase them as required separately. No.
39-4, Hongo, 3-chome, Bunkyo-ku, Tokyo, Japan Phone:+81-3-3815-2121 Fax:+81-3-3814-1222 Printed in Japan 20101008